GSM120B series

120W AC-DC Single Output Medical Adaptor
GSM120B
series
EN 60601-1
IEC 60601-1
8
Features
Applications
Universal AC input / Full range
2 pole AC inlet IEC320-C8
Built-in active PFC function, PF>0.93
High efficiency up to 91.5%
Low leakage current <100 A
Protections: Short circuit / Overload / Over voltage / Over temperature
Fully enclosed plastic case
Medical safety approved (2 x MOPP between primary to secondary)
Class Ⅱ power (without earth pin)
LED indicator for power on
No load power consumption<0.15W
ErP step2 compliant (level V)
Meet EISA 2007 (Energy Independence and Security Act)
100% full load burn-in test
Suitable for BF application with appropriate system consideration
3 years warranty
Mobile clinical workstation
Oral irrigator
Portable hemodialysis machine
Breath Machine
Medical computer monitor
Description
GSM120B is a highly reliable, 120W single-output green medical adaptor series. This product is equipped
with a 2-pin (no FG) standard IEC320-C8 power plug, adopting the input range from 80VAC to 264VAC.
The entire series supplies different output voltages between 12VDC and 48VDC that can satisfy the demands
for various kinds of medical electrical devices. The circuitry design meets the international medical standards
(2*MOPP), having an ultra low leakage current (<100 A), fitting the medical devices in direct electrical
contact with the patients.
With the efficiency up to 91.5% and the extremely low no-load power consumption below 0.15W, the design
of GSM120B observes the latest energy regulation (Level V); the supreme feature allows the adaptor to
save the energy when it is either under the operating mode or the standby mode. The entire series utilizes
the 94V-0 flame retardant plastic case, providing the double insulation that effectively prevents electrical
shock. GSM120B is approved with the international medical safety certificates.
Model Encoding
GSM120 B 12 - R7B
DC plug type
Output voltage
IEC320-C8 AC inlet
Output wattage
Series name
File Name:GSM120B-SPEC 2015-02-03
120W AC-DC Single Output Medical Adaptor
GSM120B
series
SPECIFICATION
ORDER NO.
SAFETY MODEL NO.
DC VOLTAGE
OUTPUT
48V
20V
24V
7A
6A
5A
2.5A
CURRENT RANGE
0 ~ 8.5A
0 ~ 7A
0 ~ 6A
0 ~ 5A
0 ~ 2.5A
RATED POWER (max.)
102W
120W
105W
120W
120W
RIPPLE & NOISE (max.) Note.3 120mVp-p
150mVp-p
180mVp-p
200mVp-p
240mVp-p
VOLTAGE TOLERANCE Note.4 ±5.0%
LINE REGULATION
Note.5 ±1.0%
±5.0%
±1.0%
±5.0%
±5.0%
±1.0%
±4.0%
±3.0%
±1.0%
±3.0%
±2.5%
±1.0%
±2.5%
90%
91.5%
±5.0%
Note.6 1500ms, 30ms / 230VAC
2000ms, 30ms / 115VAC at full load
HOLD UP TIME (Typ.)
VOLTAGE RANGE
20ms / 230VAC
20ms / 115VAC at full load
80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE
47 ~ 63Hz
POWER FACTOR (Typ.)
PF>0.93 / 230VAC
EFFICIENCY (Typ.)
88%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.4A / 115VAC
OVER VOLTAGE
PF>0.97 / 115VAC at full load
89%
35A / 115VAC
70A / 230VAC
Touch current < 100 A/264VAC
105 ~ 160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
105 ~ 135% rated output voltage
Protection type : Shut down o/p voltage, re-power on to recover
Shut down o/p voltage, re-power on to recover
WORKING TEMP.
WORKING HUMIDITY
-30 ~ +60℃ (Refer to "Derating Curve")
20% ~ 90% RH non-condensing
SAFETY STANDARDS
SAFETY & ISOLATION LEVEL
WITHSTAND VOLTAGE
EMC
(Note. 7)
ISOLATION RESISTANCE
89%
0.7A / 230VAC
OVER TEMPERATURE
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
NOTE
GSM120B48-R7B
GSM120B48
15V
OVERLOAD
CONNECTOR
GSM120B24-R7B
GSM120B24
Note.2 12V
LEAKAGE CURRENT(max.)
OTHERS
GSM120B20-R7B
GSM120B20
8.5A
SETUP, RISE TIME
PROTECTION
GSM120B15-R7B
GSM120B15
RATED CURRENT
LOAD REGULATION
INPUT
GSM120B12-R7B
GSM120B12
-40 ~ +85℃, 10 ~ 95% RH
±0.03% / ℃ (0 ~ 40℃)
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1 / ES60601-1-11, TUV EN60601-1 / EN60601-1-11 approved
Primary-Secondary: 2xMOPP
I/P-O/P: 4KVAC
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
EMC IMMUNITY
Compliance to EN55011(CISPR11) class B, EN61000-3-2,3, FCC PART 15 class B
MTBF
DIMENSION
372K hrs min. MIL-HDBK-217F(25℃)
167*67*35mm (L*W*H)
0.6Kg; 20pcs/13,0Kg/0.9CUFT
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3 medical level, criteria A
PACKING
PLUG
See page 2 ; Other type available by customer requested
CABLE
See page 2 ; Other type available by customer requested
1. All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.
2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1uf & 47uf capacitor.
4. Tolerance: includes set up tolerance, line regulation, load regulation.
5. Line regulation is measured from low line to high line at rated load.
6. Length of set up time is measured at first cold start. Turning ON/
7. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
File Name:GSM120B-SPEC 2015-02-03
GSM120B
120W AC-DC Single Output Medical Adaptor
Derating Curve
series
Static Characteristics
100
100
90
80
60
70
LOAD (%)
LOAD (%)
50
20
-30
0
10
20
30
40
50
60
70
(HORIZONTAL)
60
50
40
80
90
AMBIENT TEMPERATURE (℃)
100
110
120
140
160
180
200
240
264
INPUT VOLTAGE (VAC) 60Hz
Mechanical Specification
35
220
Case No.
Unit:mm
UL2464 16AWGx4C 1000±50mm for 12V
UL2464 18AWGx4C 1200±50mm for 15V ~ 48V
167
30 3mm
67
23
14
POWER LED
Plug Assignment
Standard plug (all models)
R7B
PIN NO. OUTPUT
2
1
3
4
1,4
+V
2,3
-V
Installation Manual
Please refer to : http://www.meanwell.com/webnet/search/InstallationSearch.html
File Name:GSM120B-SPEC 2015-02-03