GSM06U05-P1J

 Test Report : GSM06x05 AC‐DC Green Medical Adaptor  DESIGN VERIFY TEST
Output Function Test
Input Function Test
Protection Function Test
 SAFETY TEST
Safety Test
 RELIABILITY TEST
Environment Test
Other test
GSM06
series
RESULT
VERDICT
 DESIGN VERIFY TEST
OUTPUT FUNCTION TEST
NO
1
TEST ITEM
RIPPLE & NOISE
SPECICATION
50mVp-p (Max)
TEST CONDITION
I/P:230VAC
O/P:FULL LOAD
Ta:25℃
2
VOLTAGE TOLERANCE
-5% ~ +5% (Max)
LINE REGULATION
-0.5% ~ +0.5% (Max)
LOAD REGULATION
-5% ~ +5% (Max)
SET UP TIME
1000mS
RISE TIME
50mS
HOLD UP TIME
P
-2.42% ~ +2.66%
P
648mS
P
47.8mS
P
13.6mS
P
I/P:230VAC
O/P:FULL LOAD
Ta:25℃
7
-0.31% ~ +0.11%
I/P:230VAC
O/P:FULL LOAD
Ta:25℃
6
P
I/P:230VAC
O/P:FULL ~MIN LOAD
Ta:25℃
5
-2.80% ~ +2.66%
I/P:90VAC ~264VAC
O/P:FULL LOAD
Ta:25℃
4
P
I/P:90VAC~264VAC
O/P:FULL~MIN. LOAD
Ta:25℃
3
30mVp-p
12mS (Min)
I/P:115VAC
O/P:FULL LOAD
Ta:25℃
INPUT FUNCTION TEST
NO
1
TEST ITEM
VOLTAGE RANGE
SPECICATION
90VAC ~ 264VAC
TEST CONDITION
O/P:FULL LOAD
Ta:25℃
2
3
4
5
FREQUENCY RANGE
EFFICIENCY
AVERAGE EFFICIENCY
AC CURRENT
P
TEST: OK
P
O/P:FULL LOAD
Ta:25℃
71.62%
P
I/P:115/230VAC
O/P:25%、50%、75%、100% LOAD
70.68% (115VAC)
Ta:25℃
71.55% (230VAC)
I/P: 100VAC ~ 240VAC
NO DAMAGE OSC
O/P:FULL~MIN LOAD
Ta:25℃
68% (Typ.)
I/P:230VAC
0.18A (Max)
NO LOAD POWER
CONSUMPTION
0.3W
P
I/P: 100VAC
O/P:FULL LOAD
Ta:25℃
6
VERDICT
68V ~ 264V
50HZ - 60HZ (Typ)
69.6% ( LEVEL Ⅴ)
RESULT
I/P:TESTING
0.155A
P
0.2W
P
I/P:230VAC
O/P: NO LOAD
Ta:25℃
Model: GSM06x05 Test Report
2 / 4
GSM06
7
8
INRUSH CURRENT
LEAKAGE CURRENT
< 30A
I/P:230VAC
COLD START
O/P:FULL LOAD
Ta:25℃
< 0.05mA
I/P:264VAC
O/P:Min LOAD
Ta:25℃
20.1A
series
P
L-FG: 0.02mA
N-FG: 0.02mA
P
PROTECTION FUNCTION TEST
NO
1
2
3
TEST ITEM
OVER LOAD PROTECTION
OVER VOLTAGE PROTECTION
SHORT PROTECTION
SPECICATION
>105%
110%~140%
TEST CONDITION
RESULT
I/P:230VAC
1.44A (120%)
O/P:TESTING
Ta:25℃
HICCUP MODE
RESET : AUTO RECOVER
VERDICT
P
I/P:230VAC
O/P:MIN LOAD
Ta:25℃
Clamp by ZENER diode
SHORT OUTPUT
I/P:264VAC
NO DAMAGE
1 HOUR NO DAMAGE
O/P:FULL LOAD
Ta:25℃
HICCUP MODE
RESET AUTO RECOVER
P
P
 SAFETY TEST
SAFETY TEST
NO
1
2
TEST ITEM
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
SPECICATION
I/P-O/P:5656 VDC/min
I/P-O/P:500VDC>50MΩ
TEST CONDITION
RESULT
I/P-O/P:4242 VDC/min
Ta:25℃
I/P-O/P: 0.03uA
I/P-O/P:500 VDC
Ta:25℃
I/P-O/P>100MΩ
NO DAMAGE
NO DAMAGE
VERDICT
P
P
Model: GSM06x05 Test Report
3 / 4
GSM06
series
RESULT
VERDICT
 RELIABILITY TEST
ENVIRONMENT TEST
NO
1
TEST ITEM
TEMPERATURE RISE TEST
SPECICATION
TEST CONDITION
1. ROOM AMBIENT BURN-IN:4HRS
I/P:230VAC O/P:100% LOAD Ta=25℃
2. ROOM AMBIENT BURN-IN:4HRS
I/P:115VAC O/P:100% LOAD Ta=25℃
3. ROOM AMBIENT BURN-IN:40HRS
I/P:230VAC O/P:100% LOAD Ta=40℃
NO
Position
P/N
1
2
3
1
BD1
58.9℃
65.8℃
72.3℃
2
I/P C2
LT 2B06
Ltec 6.8u/400V TY105℃ 8*16
61.1℃
66.3℃
75.1℃
OB2536AP
72.7℃
80.3℃
88.1℃
90.1℃
3
2
LOW TEMPERATURE TURN ON
U1
4
T1
EE-15
76.1℃
79.1℃
5
O/P D2
88.6℃
90.6℃ 102.8℃
6
O/P C5
SB340
N.C.C 680u/10V KY 105℃ 8*16
72.3℃
74.1℃
87.0℃
7
C7
FuhYin 10u/50V HFR 105℃ 5*11.5
53.4℃
56.2℃
-----
TURN ON AFTER 2 HOURS
TEST
I/P : 230VAC
O/P : 100﹪LOAD
Ta= -0℃
TEST:OK
P
P
OTHER
NO
1
TEST ITEM
CAPACITOR
LIFE CYCLE
SPECICATION
I/P:115 VAC
2
MTBF
TEST CONDITION
RESULT
SUPPOSE C5 IS THE MOST CRITICAL COMPONENT
I/P:230 VAC
O/P:100% LOAD Ta=25℃ LIFE TIME= 27878HRS
O/P:100% LOAD
P
Ta=25℃ LIFE TIME= 51089HRS
MIL-KDBK-217F NOTICES 2 PARTS COUNT
TOTAL FAILURE RATE:1.127
P
M.T.B.F:886925HRS
TEST RESULT
PASS
VERDICT
TESTER
APPROVAL
PETER
VINCENT
Model: GSM06x05 Test Report
4 / 4