RPS-75 series

RPS-75 series
75W Single Output Medical Type
MODEL:RPS-75-36
OUTPUT FUNCTION TEST
NO
1
TEST ITEM
SPECIFICATION
TEST CONDITION
RIPPLE & NOISE
V1: 300
I/P: 230VAC
mVp-p (Max )
RESULT
VERDICT
V1:
18
mVp-p (Max )
O/P:FULL LOAD
Ta:25℃
2
OUTPUT VOLTAGE
CH1: 34.2V~ 39.6V
ADJUST RANGE
P
I/P: 230 VAC
33
V~
41.4
V/ 230 VAC
I/P: 115 VAC
33
V~
41.4
V/ 115 VAC
V1:
0.05
%~
-0.05
P
O/P:MIN LOAD
Ta:25℃
3
OUTPUT VOLTAGE
V1: 1 %~ -1 % (Max)
I/P: 100VAC / 264 VAC
TOLERANCE
4
LINE REGULATION
%
O/P:FULL/ MIN LOAD
Ta:25℃
V1: 0.5 %~ -0.5 % (Max)
I/P: 100VAC ~ 264 VAC
P
V1:
0.02
%~
-0.02
%
O/P:FULL LOAD
Ta:25℃
5
LOAD REGULATION
V1: 1 %~ -1 % (Max)
I/P: 230 VAC
P
V1:
0.04
%~
-0.04
%
O/P:FULL ~MIN LOAD
Ta:25℃
6
SET UP TIME
P
230VAC:
500
ms (Max)
I/P: 230 VAC
230VAC/
343
ms
115 VAC:
500
ms (Max)
I/P: 115 VAC
115VAC/
324
ms
O/P:FULL LOAD
Ta:25℃
7
RISE TIME
230VAC:
115VAC:
30
30
ms (Max)
I/P: 230 VAC
230VAC/
11
ms
ms (Max)
I/P: 115 VAC
115VAC/
11
ms
O/P:FULL LOAD
Ta:25℃
8
HOLD UP TIME
230VAC:
80
ms (TYP)
I/P: 230 VAC
230VAC/
115
ms
115VAC:
20
ms (TYP)
I/P: 115 VAC
115VAC/
25.9
ms
<5
%
O/P:FULL LOAD
Ta:25℃
9
OVER/UNDERSHOOT TEST
I/P: 230 VAC
< +5%
TEST:
O/P:FULL LOAD
Ta:25℃
10
DYNAMIC LOAD
V1:
3600
mVp-p
I/P: 230 VAC
P
P
P
P
209
mVp-p
O/P:FULL /Min LOAD
P
90%DUTY/1KHZ
Ta:25℃
Model: RPS-75-36
Test Report
1 / 5
RPS-75 series
75W Single Output Medical Type
INPUT FUNCTION TEST
NO
1
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
INPUT VOLTAGE RANGE
90VAC~264 VAC
I/P:TESTING
47 V~264V
VERDICT
O/P:FULL LOAD
Ta:25℃
I/P:
TEST: OK
LOW-LINE-3V= 87 V
P
HIGH-LINE+15%=300 V
O/P:FULL/MIN LOAD
ON: 30 Sec . OFF: 30 Sec 10MIN
( AC POWER ON/OFF NO DAMAGE )
2
3
INPUT FREQUENCY RANGE
EFFICIENCY
47HZ ~63 HZ
I/P: 90VAC ~ 264 VAC
NO DAMAGE OSC
O/P:FULL~MIN LOAD
Ta:25℃
86 % (TYP)
I/P: 230 VAC
TEST: OK
P
86.9 %
O/P:FULL LOAD
Ta:25℃
4
INPUT CURRENT
P
230V/
1
A (TYP)
I/P: 230 VAC
I=
0.83
A/ 230 VAC
115V/
1.5
A (TYP)
I/P: 115 VAC
I=
1.34
A/ 115 VAC
P
O/P:FULL LOAD
Ta:25℃
5
INRUSH CURRENT
230V/
50
A (TYP)
I/P: 230 VAC
I=
37
A/ 230 VAC
115V/
25
A(TYP)
I/P: 115 VAC
I=
17
A/ 115 VAC
COLD START
6
LEAKAGE CURRENT
<
<
150
100
P
O/P:FULL LOAD
Ta:25℃
uA/ 264 VAC
I/P: 264 VAC
L-FG:
119
uA
for
N-FG:
119
uA
leakage current
O/P:Min LOAD
Ta:25℃
uA/264 VAC
I/P: 264 VAC
L-V+ :
84
uA
for
O/P:Min LOAD
Ta:25℃
L-V-:
84
uA
N-V+:
84
uA
N-V-:
84
uA
earth
touch
leakage current
P
P
PROTECTION FUNCTION TEST
NO
1
2
3
TEST ITEM
SPECIFICATION
TEST CONDITION
OVER LOAD PROTECTION
140%~ 180 %
I/P: 230 VAC
162 %/ 230 VAC
I/P: 115 VAC
157 %/ 115 VAC
O/P:TESTING
Ta:25℃
Hiccup Mode
OVER VOLTAGE PROTECTION
SHORT PROTECTION
CH1: 41.4 V~ 48.6 V
RESULT
I/P: 230 VAC
45 V/ 230 VAC
I/P: 115 VAC
45 V/ 115 VAC
O/P:MIN LOAD
Ta:25℃
Shunt down Re- power ON
SHORT EVERY OUTPUT
I/P: 264 VAC
NO DAMAGE
1 HOUR NO DAMAGE
O/P:FULL LOAD
Ta:25℃
Hiccup Mode
Model: RPS-75-36
VERDICT
P
P
P
Test Report
2 / 5
RPS-75 series
75W Single Output Medical Type
ONTROL FUNCTION TEST
NO
1
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
No load power consumption
<0.75W
I/P: 240 VAC
0.63 W/240VAC
VERDICT
P
O/P:NO LOAD
2
Remote sense
>0.3V
I/P: 230 VAC
>0.3 V
O/P:FULL LOAD
Ta:25℃
P
ENVIRONMENT TEST
NO
1
TEST ITEM
SPECIFICATION
TEMPERATURE RISE TEST
MODEL: RPS-75-24 WITH FAN
TEST CONDITION
RESULT
VERDICT
1. ROOM AMBIENT BURN-IN:1HRS
I/P: 230VAC O/P: FULL LOAD Ta= 31 ℃
2. HIGH AMBIENT BURN-IN:2.5 HRS
I/P: 230VAC O/P: FULL LOAD Ta= 60.7 ℃
2
OVER LOAD BURN-IN TEST
NO
Position
1
2
LF2
BD1
3
ROOM AMBIENT HIGH AMBIENT
Ta= 31℃
Ta= 60.7℃
47.0℃
75.9℃
P/N
TR-719
45.4℃
74.7℃
C6
US4K80 4A/800V SHI
100U/400V NCC 105℃
36.6℃
65.7℃
4
C36
100U/50V NCC 105℃ KY
39.4℃
68.4℃
5
T1 COIL
TF-1535 JSI
63.1℃
89.6℃
STP9NK70ZFP 7.5A/700V
48.2℃
79.6℃
6
Q1
7
ZD1
P6KE220A LT
50.2℃
77.8℃
8
D1
HER306 3A/600V REC
30.5℃
76.6℃
9
D100
FMX-12S 10A/200V SAN
51.5℃
78.3℃
10
C105
330U/35V NCC 105℃ KY
39.0℃
66.5℃
11
R6
330Ω/3W R/MO
56.6℃
86.7℃
12
R3
68KΩ/2W R/MO
40.9℃
68.0℃
13
U1
NCP1230D165R2 ON
38.4℃
66.8℃
NO DAMAGE
I/P: 230 VAC
O/P: 159 ﹪LOAD
1 HOUR ( MIN )
P
TEST:OK
P
Ta:25℃
3
LOW TEMPERATURE
TURN ON AFTER
2 HOUR
TURN ON TEST
I/P: 230 VAC
O/P:100 ﹪LOAD
TEST:OK
P
Ta= -20℃
4
AFTER 12 HOURS
I/P: 272 VAC
HIGH TEMPERATURE
IN CHAMBER ON
CONTRO 60℃
O/P:FULL LOAD
Ta= 60℃
NO DAMAGE
+ 0.03 %(0~50℃)
I/P: 230 VAC
HIGH VOLTAGE
TURN ON TEST
5
TEMPERATURE
VIBRATION TEST
HUMIDITY=
P
95 %R.H
+ 0.01 %(0~50℃)
P
O/P:FULL LOAD
COEFFICIENT
6
TEST:OK
HIGH HUMIDITY
1 Carton & 1 Set
TEST:OK
(1) Waveform: Sine Wave
(2) Frequency:10~500Hz
(3) Sweep Time:10min/sweep cycle
P
(4) Acceleration:2G
(5) Test Time:1 hour in each axis (X.Y.Z)
(6) Ta:25℃
Model: RPS-75-36
Test Report
3 / 5
RPS-75 series
75W Single Output Medical Type
SAFETY TEST
NO
1
2
3
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
WITHSTAND VOLTAGE
I/P-O/P: 4 KVAC/min
I/P-O/P: 4.2 KVAC/min
I/P-O/P:
2.18
mA
I/P-FG: 1.5 KVAC/min
I/P-FG: 1.8 KVAC/min
I/P-FG:
1.32
mA
O/P-FG: 1.5 KVAC/min
O/P-FG: 1.8 KVAC/min
Ta:25℃
O/P-FG:
1.46
mA
I/P-O/P:500VDC>100MΩ
I/P-O/P: 500
VDC
I/P-O/P:
25
GΩ
I/P-FG: 500VDC>100MΩ
I/P-FG:
500
VDC
O/P-FG:500VDC>100MΩ
O/P-FG: 500
Ta:25℃
VDC
I/P-FG:
30
GΩ
O/P-FG:
30
GΩ
ISOLATION RESISTANCE
APPROVAL
VERDICT
P
NO DAMAGE
P
NO DAMAGE
TUV: Certificate NO : TA 50096088
P
UL: File NO :E227340
E.M.C TEST
NO
1
2
3
4
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
HARMONIC
EN61000-3-2
I/P: 230 VAC/50HZ
PASS
CLASS A
O/P:FULL LOAD
Ta:25℃
EN55022
I/P: 230 VAC (50HZ)
PASS
CLASS B
O/P:FULL/50% LOAD
Ta:25℃
Test by certified Lab
CONDUCTION
RADIATION
E.S.D
5
E.F.T
EN55022
I/P: 230 VAC (50HZ)
PASS
O/P:FULL LOAD
Ta:25℃
Test by certified Lab
EN61000-4-2
MEDICAL
I/P: 230 VAC/50HZ
CRITERIA A
EN61000-4-4
MEDICAL
SURGE
P
P
O/P:FULL LOAD
Ta:25℃
I/P: 230 VAC/50HZ
P
CRITERIA A
O/P:FULL LOAD
Ta:25℃
INPUT: 2KV
6
P
CLASS B
AIR:8KV / Contact:6KV
VERDICT
IEC61000-4-5
MEDICAL
I/P: 230 VAC/50HZ
P
CRITERIA A
O/P:FULL LOAD
Ta:25℃
L-N :1KV
P
L,N-PE:2KV
7
Test by certified Lab & Test Report Prepare
M.T.B.F & LIFE CYCLE CALCULATION
NO
1
TEST ITEM
SPECIFICATION
CAPACITOR
RPS-75-24 : SUPPOSE C105 IS THE MOST CRITICAL COMPONENT
I/P: 230VAC O/P:FULL LOAD Ta= 25℃ LIFE TIME= 1101254 HRS
LIFE CYCLE
I/P: 230VAC
2
3
TEST CONDITION
RESULT
VERDICT
P
O/P:FULL LOAD Ta= 60℃ LIFE TIME= 113398 HRS
MTBF
MIL-HDBK-217F NOTICES2 PARTS COUNT
DMTBF/Accelerated Life Test
TOTAL FAILURE RATE: 446.8K HRS
Demonstration Mean Time Between Failure : Above 30,000 hours @ TA 50℃
P
P
Model: RPS-75-36
Test Report
4 / 5
RPS-75 series
75W Single Output Medical Type
COMPONENT STRESS TEST
NO
1
TEST ITEM
SPECIFICATION
TEST CONDITION
RESULT
Power Transistor
Q1 Rated
I/P:High-Line +3V = 267 V
(1)
558
V
VERDICT
( D to S) or (C to E) Peak Voltage
STP9NK70Z : 700V 7.5A
O/P: (1)Full Load Turn on
(2)
664
V
P
(2) Output Short
Ta:25℃
2
Diode Peak Voltage
D100 Rated
I/P:High-Line +3V = 267 V
(1)
396
V
SF10LC40 : 400V 10A
O/P: (1)Full Load Turn on
(2)
400
V
(1)
530
V
P
(2)Output Short
Ta:25℃
3
Clamp Diode Peak Voltage
D1 Rated
I/P:High-Line +3V = 267 V
HER306 : 600V 3A
O/P: (1) Dynamic Load
90%Duty/1KHz
P
Ta:25℃
4
Input Capacitor Voltage
C5 Rated
:100u / 400V/ 105℃
I/P:High-Line +3V = 267 V
(1)
378
V
O/P: (1)Full Load Turn on /Off
(2)
384
V
(2) Min load Turn on /Off
(3)
384
V
(3)Full Load /Min load Change
Ta:25℃
5
Control IC Voltage Test
P
U1 Rated
I/P:High-Line +3V = 267 V
(1)
1230D165 : 18 V
O/P: (1)Full Load Turn on /Off
11.38
V
(2) Min load Turn on /Off
(2)
9.93
V
(3)
11.38
V
P
(3)Full Load /Min load Change
Ta:25℃
DATE
2006/9/4
2006/11/14
SAMPLE
TEST RESULT
TESTER
APPROVAL
RD SAMPLE
PASS
VINCENT TSENG
MAX LIN
PRODUCT SAMPLE
W0610A12
PASS
VINCENT TSENG
MAX LIN
2003/12/12
Model: RPS-75-36
A50-F023
Test Report
5 / 5