PMP60 Series MEDICAL 3 0 - 6 0 W AT T S - A C / D C S I N G L E & M U LT I P L E O U T P U T FEATURES • • • • • • • • Single, dual or triple outputs Optional output connectors Optional on/off switch Wide input range Input surge current protection Over voltage protection Over current protection Energy Star Efficiency level IV RoHS compliant SPECIFICATIONS INPUT OPERATING MTBF Voltage range 85-264 VAC Frequency 47-63 Hz 50 A @ 115 VAC or 100 A @ 230 VAC, at 25°C cold start 1.22 A (rms) for 100 VAC 0.68 A (rms) for 240 VAC 200 µA max. @ 264 VAC, 63 Hz Inrush current Input current Earth leakage current Withstand voltage Input – Output: 4000VAC Input – Case:1500VAC Output – Case: 500VAC OUTPUT Efficiency Switching frequency 150,000 hours at full load at 25°C ambient, calculated per MIL-HDBK-217F 85% minimum on single output models, 68-74% minimum on the others 40 KHz-130 KHz ENVIRONMENT Operating temperature: 0°C to +60°C Storage temperature: -40°C to +85°C Relative humidity: Derating: 5% to 95% non-condensing Derate from 100% at +40°C linearly to 50% at +60°C Output voltage See table Output current See table Safety Ripple and noise 66 mVp-p maximum on 3.3 V output, 100 mVp-p maximum on 5 V output and 1% maximum on other voltage outputs (12 V, 15 V …, 48 V etc.) UL 60601-1, CSA C22.2 No. 601.1, TÜV EN 60601-1, UL 60950-1, CSA C22.2 No. 60950-1(except PMP60-14 by UL) TÜV EN 60950-1 C-Tick AS/NZS CISPR11 Group1, Class A Line regulation ±0.5% maximum at full load EMC standards Load regulation See table Transient response Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change IEC60601-1-2, EN55011 / EN55022 and FCC, VCCI: Class B conducted . class B radiated Immunity standards Overvoltage protection Temperature coefficient Provided on output #1 only, set at 112-140% of its nominal output voltage All outputs protected to short circuit conditions All outputs ±0.04% /℃ maximum EN61000-3-2, EN61000-3-3, EN61000-42, EN61000-4-3, EN61000-4-4, EN610004-5, EN61000-4-6, EN61000-4-8, EN61000-4-11 Hold-up time 10 ms minimum at 110 VAC Overcurrent protection Doc. No. PBC9 Rev.C 28-4-11 STANDARDS AND APPROVALS MECHANICAL Dimensions 147x75x46mm Weight 600g Powering Progress – www.powerbox.com.au 377 MEDICAL PMP60 Series 3 0 - 6 0 W AT T S - A C / D C S I N G L E & M U LT I P L E O U T P U T SELECTION TABLE OUTPUT #1 MODEL OUPUT #2 MIN. MAX. CURRENT CURRENT V1 TOL. PMP60-12 PMP60-13 PMP60-13-1 PMP60-14 PMP60-16 PMP60-17 PMP60-18 11-13 V 13-17 V 17-21 V 21-27 V 27-33 V 33-39 V 46-50 V 0A 0A 0A 0A 0A 0A 0A 5.46 A 4.62 A 3.53 A 2.86 A 2.23 A 1.82 A 1.31 A ±5% ±5% ±5% ±5% ±3% ±3% ±3% PMP60-23 PMP60-30 PMP60-31 PMP60-32 PMP60-36 +5.0 V +3.3 V +5.0 V +5.0 V +5.0 V 1A 1A 1A 1A 1A 5.0 A 6.0 A 5.0 A 5.0 A 5.0 A ±5% ±5% ±5% ±5% ±5% V2 MIN. MAX. CURRENT CURRENT OUPUT #3 TOL. V3 MIN. MAX. CURRENT CURRENT (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) (NA) +12 V +5 V +12 V +15 V +24 V 0.5A 0.5 A 0.5 A 0.4 A 0.3 A TOL. MAX OUTPUT POWER (N/A) (N/A) (N/A (N/A) (N/A) (N/A) (N/A) 3.0 A 3.0 A 3.0 A 2.3 A 1.5 A ±5% ±5% ±5% ±5% ±5% 60 W 60 W 60 W 60 W 60 W 60 W 60 W (N/A) +12 V -12 V -15 V +12 V 0.1 A 0.1 A 0.1 A 0.1 A 0.7 A 0.7 A 0.7 A 0.7 A ±10% ±10% ±10% ±10% 40 W 30 W 40 W 40 W 40 W NOTES: 1. The output voltages of a multiple output model may go outside of the stated tolerance when an output load current is out of stated limits. All models may be operated at no-load without damage. 2. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 µF tantalum capacitor in parallel with a 0.1 µF ceramic capacitor across the output. OUTPUT POWER DERATING CURVE MECHANICAL SPECIFICATIONS PIN CHART MODEL PIN PMP60-12 PMP60-14 PMP60-13 PMP60-16 PMP60-13-1 PMP60-23 PMP60-30 PMP60-31 378 PMP60-32 PMP60-17 PMP60-18 PMP60-36 1 2 3 4 5 V1 Return V1 Return +V1 V1 Return +V1 Common Return Common Return V1 N.C. V2 Common Return Common Return V1 V3 V2 Doc. No. PBC9 Rev.C 28-4-11