PMP60 Series

PMP60 Series
MEDICAL
3 0 - 6 0 W AT T S - A C / D C S I N G L E & M U LT I P L E O U T P U T
FEATURES
•
•
•
•
•
•
•
•
Single, dual or triple outputs
Optional output connectors
Optional on/off switch
Wide input range
Input surge current protection
Over voltage protection
Over current protection
Energy Star Efficiency level IV
RoHS compliant
SPECIFICATIONS
INPUT
OPERATING
MTBF
Voltage range
85-264 VAC
Frequency
47-63 Hz
50 A @ 115 VAC or 100 A @ 230 VAC, at
25°C cold start
1.22 A (rms) for 100 VAC
0.68 A (rms) for 240 VAC
200 µA max. @ 264 VAC, 63 Hz
Inrush current
Input current
Earth leakage current
Withstand voltage
Input – Output: 4000VAC
Input – Case:1500VAC
Output – Case: 500VAC
OUTPUT
Efficiency
Switching frequency
150,000 hours at full load at 25°C
ambient, calculated per MIL-HDBK-217F
85% minimum on single output models,
68-74% minimum on the others
40 KHz-130 KHz
ENVIRONMENT
Operating temperature:
0°C to +60°C
Storage temperature:
-40°C to +85°C
Relative humidity:
Derating:
5% to 95% non-condensing
Derate from 100% at +40°C linearly to
50% at +60°C
Output voltage
See table
Output current
See table
Safety
Ripple and noise
66 mVp-p maximum on 3.3 V output,
100 mVp-p maximum on 5 V output and
1% maximum on other voltage outputs
(12 V, 15 V …, 48 V etc.)
UL 60601-1, CSA C22.2 No. 601.1, TÜV EN
60601-1, UL 60950-1, CSA C22.2 No.
60950-1(except PMP60-14 by UL) TÜV EN
60950-1
C-Tick
AS/NZS CISPR11 Group1, Class A
Line regulation
±0.5% maximum at full load
EMC standards
Load regulation
See table
Transient response
Maximum excursion of 4% or better on
all models, recovering to 1% of final
value within 500 us after a 25% step
load change
IEC60601-1-2, EN55011 / EN55022 and
FCC, VCCI: Class B conducted . class B
radiated
Immunity standards
Overvoltage protection
Temperature coefficient
Provided on output #1 only, set at
112-140% of its nominal output voltage
All outputs protected to short circuit
conditions
All outputs ±0.04% /℃ maximum
EN61000-3-2, EN61000-3-3, EN61000-42, EN61000-4-3, EN61000-4-4, EN610004-5, EN61000-4-6, EN61000-4-8,
EN61000-4-11
Hold-up time
10 ms minimum at 110 VAC
Overcurrent protection
Doc. No. PBC9 Rev.C 28-4-11
STANDARDS AND APPROVALS
MECHANICAL
Dimensions
147x75x46mm
Weight
600g
Powering Progress – www.powerbox.com.au
377
MEDICAL
PMP60 Series
3 0 - 6 0 W AT T S - A C / D C S I N G L E & M U LT I P L E O U T P U T
SELECTION TABLE
OUTPUT #1
MODEL
OUPUT #2
MIN.
MAX.
CURRENT CURRENT
V1
TOL.
PMP60-12
PMP60-13
PMP60-13-1
PMP60-14
PMP60-16
PMP60-17
PMP60-18
11-13 V
13-17 V
17-21 V
21-27 V
27-33 V
33-39 V
46-50 V
0A
0A
0A
0A
0A
0A
0A
5.46 A
4.62 A
3.53 A 2.86
A
2.23 A 1.82
A
1.31 A
±5%
±5%
±5%
±5%
±3%
±3%
±3%
PMP60-23
PMP60-30
PMP60-31
PMP60-32
PMP60-36
+5.0 V
+3.3 V
+5.0 V
+5.0 V
+5.0 V
1A
1A
1A
1A
1A
5.0 A
6.0 A
5.0 A
5.0 A
5.0 A
±5%
±5%
±5%
±5%
±5%
V2
MIN.
MAX.
CURRENT CURRENT
OUPUT #3
TOL.
V3
MIN.
MAX.
CURRENT CURRENT
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(NA)
+12 V
+5 V
+12 V
+15 V
+24 V
0.5A
0.5 A
0.5 A
0.4 A
0.3 A
TOL.
MAX
OUTPUT
POWER
(N/A)
(N/A)
(N/A
(N/A)
(N/A)
(N/A)
(N/A)
3.0 A
3.0 A
3.0 A
2.3 A
1.5 A
±5%
±5%
±5%
±5%
±5%
60 W
60 W
60 W
60 W
60 W
60 W
60 W
(N/A)
+12 V
-12 V
-15 V
+12 V
0.1 A
0.1 A
0.1 A
0.1 A
0.7 A
0.7 A
0.7 A
0.7 A
±10%
±10%
±10%
±10%
40 W
30 W
40 W
40 W
40 W
NOTES:
1. The output voltages of a multiple output model may go outside of the stated tolerance when an output load current is out of stated limits. All models
may be operated at no-load without damage.
2. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load
ranges, and with a 10 µF tantalum capacitor in parallel with a 0.1 µF ceramic capacitor across the output.
OUTPUT POWER DERATING CURVE
MECHANICAL SPECIFICATIONS
PIN CHART
MODEL
PIN
PMP60-12 PMP60-14
PMP60-13 PMP60-16
PMP60-13-1
PMP60-23
PMP60-30
PMP60-31
378
PMP60-32
PMP60-17
PMP60-18
PMP60-36
1
2
3
4
5
V1 Return
V1 Return
+V1
V1 Return
+V1
Common Return
Common Return
V1
N.C.
V2
Common Return
Common Return
V1
V3
V2
Doc. No. PBC9 Rev.C 28-4-11