MEDICAL PMP120 Series 9 6 - 1 2 0 W AT T S - A C / D C S I N G L E O U T P U T FEATURES • Low safety ground leakage current • Class I and Class II models are certified to medical and ITE safety standards • Wide input range 90 to 264 VAC • Optional output connectors • 100% burn-in • Over voltage protection • Over current protection • Compliant with CEC and Energy Star Efficiency level V • RoHS compliant SPECIFICATIONS INPUT OPERATING Voltage range 90-264 VAC Frequency Inrush current 47-63 Hz 60 A @ 115 VAC or 120 A @ 230 VAC, at 25°C cold start 1.60 A (rms) for 115 VAC 0.80 A (rms) for 230 VAC 180 µA max. @ 264 VAC, 63 Hz Input current Earth leakage current Efficiency Switching frequency 150,000 hours minimum at full load at 25°C ambient, calculated per MIL-HDBK217F Average active 87% min. (except 85% min. for PMP150-12 and PMP150-13) 50-110 KHz ENVIRONMENT Touch current 100 µA max. @ 264 VAC, 63 Hz Operating temperature: Power factor 0.98 Typical at 115 VAC Storage temperature: -40°C to +85°C Input – Output: 4000VAC Input – Case:1500VAC Output – Case: 500VAC Relative humidity: Derating: 5% to 95% non-condensing Derate from 100% at +40°C linearly to 50% at +60°C Output voltage See table Safety Output current See table UL 60601-1, CSA C22.2 No. 601.1, TÜV EN 60601-1, UL 60950-1, CSA C22.2 No. 60950-1, TÜV EN 60950-1 Ripple and noise 1% peak to peak maximum at full load C-Tick AS/NZS CISPR11 Group1, Class A Line regulation ±0.5% maximum at full load EMC standards Load regulation See table Minimum load current 0A all models (IEC60601-1-2) EN55011 / EN55022 and FCC, VCCI: Class B conducted. Class B radiated Transient response Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change Immunity standards EN61000-3-2, EN61000-3-3, EN61000-42, EN61000-4-3, EN61000-4-4, EN610004-5, EN61000-4-6, EN61000-4-8, EN61000-4-11 Overvoltage protection Provided and set at 112-140% of its nominal output voltage Protected to short circuit conditions Isolation OUTPUT Overcurrent protection 382 MTBF 0°C to +60°C STANDARDS AND APPROVALS Temperature coefficient ±0.04% /°C maximum Hold-up time 15 ms minimum at 115 VAC MECHANICAL Dimensions 160x76x43.7mm Weight 780g Doc. No. PBC9 Rev.C 28-4-11 PMP120 Series OUTPUT VOLTAGE/CURRENT RATING CHART MODEL(1) CLASS I PMP120-12 PMP120-13 PMP120-13-1 PMP120-13-2 PMP120-13-3 PMP120-14 PMP120-16 PMP120-17 PMP120-18 OUTPUT CLASS II V1 PMP120F-12 PMP120F-13 PMP120F-13-1 PMP120F-13-2 PMP120F-13-3 PMP120F-14 PMP120F-16 PMP120F-17 PMP120F-18 12 V 15 V 18 V 19 V 20 V 24 V 30 V 36 V 48 V MIN. CURRENT MAX. CURRENT 0A 0A 0A 0A 0A 0A 0A 0A 0A 8.00 A 7.00 A 6.67 A 6.32 A 6.00 A 5.00 A 4.00 A 3.34 A 2.50 A TOL. RIPPLE & NOISE (2) MAX POWER AVERAGE ACTIVE EFFICIENCY (TYPICAL) @ 115 / 230 VAC ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% 120 mV 150 mV 180 mV 190 mV 200 mV 240 mV 300 mV 360 mV 480 mV 96 W 105 W 120W 120W 120W 120W 120W 120W 120w 86 / 86 % 86 / 86 % 87 / 86 % 87 / 86 % 87 / 86 % 88 / 88 % 89 / 88 % 89 / 88 % 88 / 88 % MEDICAL 9 6 - 1 2 0 W AT T S - A C / D C S I N G L E O U T P U T NOTES: 1. Class I models are equipped with IEC320/C14 inlet, and class II models with IEC320/C18 inlet. 2. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 µF tantalum capacitor in parallel with a 0.1 µF ceramic capacitor across the output. OUTPUT POWER DERATING CURVE MECHANICAL SPECIFICATIONS NOTES: 1. 2. 3. 4. 5. Dimensions shown in inches [mm] Tolerance 0.02 [0.5] maximum Weight: 780 grams (1.716 lbs.) approx. Output connector is 4 pin plug without lock, mating with Kycon P/N KPJX-4S-S socket or equivalent. Refer to Powerbox for “OPTIONAL OUTPUT CONNECTORS”. Add the suffix assigned for a selected connector to a wanted model number, e.g. PMP120-14-B1, for ordering. PIN CHART MODEL PMP120-12 PMP120-13 PMP120-13-1 PMP120-13-2 PMP120-13-3 PMP120-14 PMP120-16 PMP120-17 PMP120-18 PIN PMP120F-12 PMP120F-13 PMP120F-13-1 PMP120F-13-2 PMP120F-13-3 PMP120F-14 PMP120F-16 PMP120F-17 PMP120F-18 Doc. No. PBC9 Rev.C 28-4-11 1 2 3 4 V1 Return +V1 V1 Return +V1 Powering Progress – www.powerbox.com.au 383