PMP120 Series

MEDICAL
PMP120 Series
9 6 - 1 2 0 W AT T S - A C / D C S I N G L E O U T P U T
FEATURES
•
Low safety ground leakage current
•
Class I and Class II models are certified
to medical and ITE safety standards
•
Wide input range 90 to 264 VAC
•
Optional output connectors
•
100% burn-in
•
Over voltage protection
•
Over current protection
•
Compliant with CEC and Energy Star Efficiency level V
•
RoHS compliant
SPECIFICATIONS
INPUT
OPERATING
Voltage range
90-264 VAC
Frequency
Inrush current
47-63 Hz
60 A @ 115 VAC or 120 A @ 230 VAC, at
25°C cold start
1.60 A (rms) for 115 VAC
0.80 A (rms) for 230 VAC
180 µA max. @ 264 VAC, 63 Hz
Input current
Earth leakage current
Efficiency
Switching frequency
150,000 hours minimum at full load at
25°C ambient, calculated per MIL-HDBK217F
Average active 87% min. (except 85%
min. for PMP150-12 and PMP150-13)
50-110 KHz
ENVIRONMENT
Touch current
100 µA max. @ 264 VAC, 63 Hz
Operating temperature:
Power factor
0.98 Typical at 115 VAC
Storage temperature:
-40°C to +85°C
Input – Output: 4000VAC
Input – Case:1500VAC
Output – Case: 500VAC
Relative humidity:
Derating:
5% to 95% non-condensing
Derate from 100% at +40°C linearly to
50% at +60°C
Output voltage
See table
Safety
Output current
See table
UL 60601-1, CSA C22.2 No. 601.1, TÜV EN
60601-1, UL 60950-1, CSA C22.2 No.
60950-1, TÜV EN 60950-1
Ripple and noise
1% peak to peak maximum at full load
C-Tick
AS/NZS CISPR11 Group1, Class A
Line regulation
±0.5% maximum at full load
EMC standards
Load regulation
See table
Minimum load current
0A all models
(IEC60601-1-2) EN55011 / EN55022 and
FCC, VCCI: Class B conducted. Class B
radiated
Transient response
Maximum excursion of 4% or better on
all models, recovering to 1% of final
value within 500 us after a 25% step
load change
Immunity standards
EN61000-3-2, EN61000-3-3, EN61000-42, EN61000-4-3, EN61000-4-4, EN610004-5, EN61000-4-6, EN61000-4-8,
EN61000-4-11
Overvoltage protection
Provided and set at 112-140% of its
nominal output voltage
Protected to short circuit conditions
Isolation
OUTPUT
Overcurrent protection
382
MTBF
0°C to +60°C
STANDARDS AND APPROVALS
Temperature coefficient
±0.04% /°C maximum
Hold-up time
15 ms minimum at 115 VAC
MECHANICAL
Dimensions
160x76x43.7mm
Weight
780g
Doc. No. PBC9 Rev.C 28-4-11
PMP120 Series
OUTPUT VOLTAGE/CURRENT RATING CHART
MODEL(1)
CLASS I
PMP120-12
PMP120-13
PMP120-13-1
PMP120-13-2
PMP120-13-3
PMP120-14
PMP120-16
PMP120-17
PMP120-18
OUTPUT
CLASS II
V1
PMP120F-12
PMP120F-13
PMP120F-13-1
PMP120F-13-2
PMP120F-13-3
PMP120F-14
PMP120F-16
PMP120F-17
PMP120F-18
12 V
15 V
18 V
19 V
20 V
24 V
30 V
36 V
48 V
MIN.
CURRENT
MAX.
CURRENT
0A
0A
0A
0A
0A
0A
0A
0A
0A
8.00 A
7.00 A
6.67 A
6.32 A
6.00 A
5.00 A
4.00 A
3.34 A
2.50 A
TOL.
RIPPLE &
NOISE (2)
MAX POWER
AVERAGE ACTIVE
EFFICIENCY (TYPICAL)
@ 115 / 230 VAC
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
120 mV
150 mV
180 mV
190 mV
200 mV
240 mV
300 mV
360 mV
480 mV
96 W
105 W
120W
120W
120W
120W
120W
120W
120w
86 / 86 %
86 / 86 %
87 / 86 %
87 / 86 %
87 / 86 %
88 / 88 %
89 / 88 %
89 / 88 %
88 / 88 %
MEDICAL
9 6 - 1 2 0 W AT T S - A C / D C S I N G L E O U T P U T
NOTES:
1. Class I models are equipped with IEC320/C14 inlet, and class II models with IEC320/C18 inlet.
2. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges,
and with a 10 µF tantalum capacitor in parallel with a 0.1 µF ceramic capacitor across the output.
OUTPUT POWER DERATING CURVE
MECHANICAL SPECIFICATIONS
NOTES:
1.
2.
3.
4.
5.
Dimensions shown in inches [mm]
Tolerance 0.02 [0.5] maximum
Weight: 780 grams (1.716 lbs.) approx.
Output connector is 4 pin plug without lock, mating with Kycon P/N KPJX-4S-S socket or equivalent.
Refer to Powerbox for “OPTIONAL OUTPUT CONNECTORS”. Add the suffix assigned for a selected connector to a wanted model number, e.g.
PMP120-14-B1, for ordering.
PIN CHART
MODEL
PMP120-12
PMP120-13
PMP120-13-1
PMP120-13-2
PMP120-13-3
PMP120-14
PMP120-16
PMP120-17
PMP120-18
PIN
PMP120F-12
PMP120F-13
PMP120F-13-1
PMP120F-13-2
PMP120F-13-3
PMP120F-14
PMP120F-16
PMP120F-17
PMP120F-18
Doc. No. PBC9 Rev.C 28-4-11
1
2
3
4
V1 Return
+V1
V1 Return
+V1
Powering Progress – www.powerbox.com.au
383