PM100 - PROTEK POWER

100 WATT MEDICAL & ITE POWER SUPPLIES
DESCRIPTION
PM100 SERIES
The PM100 series of compact, open PCB constructed, AC-DC
(LVD)
switching power supplies are capable of delivering 100 watts of
RoHS
continuous output power at convection cooling. They are suited for
medical, information technology and industrial applications, but not for
life-supporting medical equipment. Approval to both EN60601-1 and
EN60950-1 safety standards improves design-in time and reduces
end equipment compliance costs.
FEATURES
z
Medical and ITE approvals
z
Compact size 2” × 4” × 1.26”
z
High power density 10 W/cubic inch
z
100 W output with convection cooling up to +50℃
z
Low earth leakage current
z
EN55011 /55022 class B emissions
UL 60601-1, CSA C22.2 No. 601.1
z
RoHS compliant
File No. E178020
SAFETY STANDARD APPROVALS
INPUT SPECIFICATIONS
Input voltage:
TÜV EN 60601-1
90-132 /180-264 VAC (Universal mains
supply operation)
Input frequency:
47-63 Hz
Input current:
1.9 A (rms) for 100-120 VAC
UL 60950-1, CSA C22.2 No. 60950-1
1.1 A (rms) for 200-240 VAC
Earth leakage current:
TÜV EN 60950-1
150 μA max. @ 264 VAC, 63 Hz
GENERAL SPECIFICATIONS
OUTPUT SPECIFICATIONS
Output voltage/current:
See rating chart.
Total output power:
100 watts maximum
Ripple and noise:
150 mV peak to peak on 5.0 V model,
1% peak to peak on other models
Overvoltage protection:
100 KHz (typical)
Efficiency:
88-90% @ 230 VAC full load
Hold-up time:
12 ms minimum at 110 VAC
Line regulation:
±0.2% maximum at full load
Inrush current:
40 A @ 115 VAC or 80 A @ 230 VAC, at 25℃
Withstand voltage:
4000 VAC from input to output,
cold start
Provided on output; set at 110-140% of its
nominal output voltage
Overcurrent protection:
Switching frequency:
1500 VAC from input to ground,
All outputs protected to short circuit
500 VAC from output to ground
conditions
Temperature coefficient:
All outputs ±0.04% /℃ maximum
Transient response:
Maximum excursion of 4% or better on all
MTBF:
270,000 hours at full load at 25℃ ambient
temperature, calculated per MIL-HDBK-217F
EMC Performance
models, recovering to 1% of final value
EN55011 /EN55022:
FCC:
Class B conducted, class B radiated
Class B conducted, class B radiated
VCCI:
Class B conducted, class B radiated
EN61000-3-2:
Harmonic distortion, class A
EN61000-3-3:
Line flicker
ENVIRONMENTAL SPECIFICATIONS
EN61000-4-2:
ESD, ±8 KV air and ±6 KV contact
Operating temperature:
-10℃ to +70℃
EN61000-4-3:
Radiated immunity, 3 V/m
Storage temperature:
-40℃ to +85℃
EN61000-4-4:
Fast transient/burst, ±2 KV
Relative humidity:
5% to 95% non-condensing
EN61000-4-5:
Surge, ±1 KV diff., ±2 KV com.
Derating:
Derate from 100% at +50℃ linearly to
EN61000-4-6:
Conducted immunity, 3 Vrms
50% at +70℃
EN61000-4-8:
Magnetic field immunity, 3 A/m
Convection
EN61000-4-11:
Voltage dip immunity, 30% reduction for 500
ms (criteria A @ 230 VAC, criteria B @ 100
VAC), 60% reduction for 100 ms (criteria B),
>95% reduction for 10 ms (Criteria A)
within 500 us after a 25% step load
change
Cooling:
9
UNIVERSAL INPUT
PM100 MEDICAL & ITE SERIES
OUTPUT VOLTAGE/CURRENT RATING CHART
Output
(1)
Model
PM100-10A
PM100-12A
PM100-13A
PM100-13-1A
PM100-14A
PM100-15A
PM100-17A
PM100-18A
NOTES: 1.
2.
V1
5V
12 V
15 V
18 V
24 V
28 V
36 V
48 V
Min. Current
0A
0A
0A
0A
0A
0A
0A
0A
Max. Current
20.00 A
8.34 A
6.70 A
5.56 A
4.20 A
3.58 A
2.78 A
2.10 A
Tol.
±2 %
±2 %
±2 %
±2 %
±2 %
±2 %
±2 %
±2 %
Ripple &
(2)
Noise
150 mV
120 mV
150 mV
180 mV
240 mV
280 mV
360 mV
480 mV
Max. Power
100 W
100 W
100 W
100 W
100 W
100 W
100 W
100 W
Average Active
Efficiency (typical)
@ 115/230 Vac
85/88%
86/89%
86/89%
86/89%
87/90%
87/90%
87/89%
87/89%
Safety approvals are for PCB form only. To order models with metallic L-bracket or box, change suffix “A” to “B” for
L-bracket form, to “C” for enclosed form (see Outline Drawing of Cased Internal Switchers), e.g. PM100-14C.
Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line
voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across
the output.
MECHANICAL SPECIFICATIONS
OUTPUT POWER DERATING CURVE
NOTES:
1. Dimensions shown in inches [mm]
2. Tolerance 0.02 [0.5] maximum
3. Connector P1: Molex header 09-65-2038 or equivalent, mating with Molex
housing 09-50-1031 or equivalent.
4. Connector P2: Molex header 09-65-2068 or equivalent, mating with Molex
housing 09-50-1061 or equivalent.
5. To ensure compliance with level B emissions, connect the three “*
“ marked mounting holes with metallic standoffs to chassis.
6. Weight: 190 grams (0.44 lbs.) approx.
PIN CHART
PIN
1
2
3
4
5
6
V1 Return
V1 Return
V1 Return
+V1
+V1
+V1
MODEL
PM100-10A
PM100-13-1A
PM100-17A
PM100-12A
PM100-14A
PM100-18A
PM100-13A
PM100-15A
10