PMP60 - PROTEK POWER

30-60 WATT MEDICAL & ITE POWER SUPPLIES
DESCRIPTION
PMP60 SERIES
(LVD)
The PMP60 series of AC/DC switching power supplies are for
30-60 watts of continuous output power. They are enclosed in
a 94V-1 rated polyphenylene-oxide case with an IEC320/C14
inlet to mate with interchangeable cord for world-wide use. All
models meet EN55011, EN55022 and FCC class B emission
limits, and are designed for medical and ITE applications, not
for life supporting equipment.
RoHS
FEATURES
z
z
z
z
z
z
z
z
z
z
12 standard desktop models
Single, dual or triple outputs
Optional output connectors
Optional on/off switch
100% burn-in
W i d e i n p u t r a n g e 8 5 - 2 6 4 VA C
Input surge current protection
Overvoltage protection
Overcurrent protection
Single output models compliant with CEC and
E n e r g y St a r E f f i c i e n c y l e v e l I V r e q u i r e m e n t s
* No load power consumption less than 0.5 W
* Average active efficiency ≧85%
z Compliant with RoHS requirement
SAFETY STANDARD APPROVALS
UL 60601-1, CSA C22.2 No. 601.1
File No. E178020
TÜV EN 60601-1
UL 60950-1, CSA C22.2 No. 60950-1
(except PMP60-14 by UL)
TÜV EN 60950-1
GENERAL SPECIFICATIONS
INPUT SPECIFICATIONS
Input voltage:
85-264 VAC
Input frequency:
47-63 Hz
Input current:
1.22 A (rms) for 100 VAC
200 μA max. @ 264 VAC, 63 Hz
(Touch current)
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
Transient response:
85% minimum on single output models,
Hold-up time:
10 ms minimum at 110 VAC
Line regulation:
±0.5% maximum at full load
Inrush current:
50 A @ 115 VAC or 100 A @ 230 VAC, at
Withstand voltage:
OUTPUT SPECIFICATIONS
Output voltage /current:
Maximum output power:
Ripple and noise:
40 KHz-130 KHz
Efficiency:
68-74% minimum on the others
0.68 A (rms) for 240 VAC
Earth leakage current:
Switching frequency:
See rating chart.
See rating chart.
66 mVp-p maximum on 3.3 V output,
100 mVp-p maximum on 5 V output
and 1% maximum on other voltage
outputs (12 V, 15 V …, 48 V etc.)
Provided on output #1 only, set at
112-140% of its nominal output voltage
All outputs protected to short circuit
conditions
All outputs ±0.04% /℃ maximum
Maximum excursion of 4% or better on
all models, recovering to 1% of final
value within 500 us after a 25% step
load change
MTBF:
25℃ cold start
4000 VAC from input to output,
1500 VAC from input to ground,
500 VAC from output to ground
150,000 hours minimum at full load at 25℃
ambient, calculated per MIL-HDBK-217F
EMC Performance (IEC60601-1-2)
EN55011 / EN55022:
Class B conducted, class B radiated
FCC:
Class B conducted, class B radiated
VCCI:
Class B conducted, class B radiated
EN61000-3-2:
Harmonic distortion, class A and D
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8 KV air and ±6 KV contact
EN61000-4-3:
Radiated immunity, 3 V/m for 80-2500 MHz
EN61000-4-4:
Fast transient/burst, ±2 KV
EN61000-4-5:
Surge, ±1 KV diff., ±2 KV com
EN61000-4-6:
Conducted immunity, 3 Vrms
EN61000-4-8:
Magnetic field immunity, 3 A/m
EN61000-4-11:
Voltage dip immunity, 30% reduction for 500
ENVIRONMENTAL SPECIFICATIONS
ms, 60% reduction for 100 ms and >95%
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
reduction for 10 ms
0℃ to +60℃
-40℃ to +85℃
5% to 95% non-condensing
Derate from 100% at +40℃ linearly to
50% at +60℃
47
UNIVERSAL INPUT
PMP60 MEDICAL & ITE SERIES
OUTPUT VOLTAGE/CURRENT RATING CHART
Model
PMP60-12
PMP60-13
PMP60-13-1
PMP60-14
PMP60-16
PMP60-17
PMP60-18
PMP60-23
PMP60-30
PMP60-31
PMP60-32
PMP60-36
V1
11-13
13-17
17-21
21-27
27-33
33-39
46-50
+5.0
+3.3
+5.0
+5.0
+5.0
V
V
V
V
V
V
V
V
V
V
V
V
Output
Min.
current
0A
0A
0A
0A
0A
0A
0A
1A
1A
1A
1A
1A
#1
Max.
current
5.46 A
4.62 A
3.53 A
2.86 A
2.23 A
1.82 A
1.31 A
5.0 A
6.0 A
5.0 A
5.0 A
5.0 A
Tol.
±5%
±5%
±5%
±5%
±3%
±3%
±3%
±5%
±5%
±5%
±5%
±5%
V2
+12 V
+5 V
+12 V
+15 V
+24 V
Output #2
Min.
Max.
current current
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
0.5 A
3.0 A
0.5 A
3.0 A
0.5 A
3.0 A
0.4 A
2.3 A
0.3 A
1.5 A
Tol.
V3
±5%
±5%
±5%
±5%
±5%
+12 V
-12 V
-15 V
+12 V
Output #3
Min.
Max.
current current
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
0.1 A
0.7 A
0.1 A
0.7 A
0.1 A
0.7 A
0.1 A
0.7 A
Tol.
±10%
±10%
±10%
±10%
Max. Output
Power
60 W
60 W
60 W
60 W
60 W
60 W
60 W
40 W
30 W
40 W
40 W
40 W
NOTES:
1.
The output voltages of a multiple output model may go outside of the stated tolerance when an output load current is out of
stated limits. All models may be operated at no-load without damage.
2.
Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage
and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the output.
MECHANICAL SPECIFICATIONS
OUTPUT POWER DERATING CURVE
NOTES:
1.
Dimensions shown in inches [mm]
2.
Tolerance 0.02 [0.5] maximum
3.
Weight: 600 grams (1.33 lbs.) approx.
4.
Output connector is 5 pin DIN plug, mating with Switchcraft P/N 57GB5F receptacle or equivalent.
5.
Refer to Section titled “OPTIONAL OUPUT CONNECTORS“ for optional output connectors. Add the suffix assigned for a selected
connector to a wanted model number, e.g. PMP60-12-B2, for ordering.
6.
To order a model with on / off switch, add suffix “ S “ to the model number, e.g. PMP60-12-B2-S
PIN CHART
PIN
MODEL
PMP60-12 PMP60-14
PMP60-13 PMP60-16
PMP60-13-1
PMP60-17
PMP60-18
PMP60-23
PMP60-30
PMP60-31
PMP60-32
1
2
3
4
5
V1 Return
V1 Return
+V1
V1 Return
+V1
Common Return
Common Return
V1
N.C.
V2
Common Return
Common Return
V1
V3
V2
PMP60-36
48