WBI Series - First Sensor

WBI Series
Mass flow sensors for gases
FEATURES
· Flow ranges 0...200 sccm, 0...±200 sccm,
0...1 slpm, 0...±1 slpm
· Thermal mass flow sensing
· Digital I²C bus output
· RoHS and REACH compliant
· Quality Management System according
to ISO 13485:2003 and ISO 9001:2008
MEDIA COMPATIBILITY7
To be used with dry gases only.
The WBI series is NOT designed for liquid flow and
will be damaged by liquid flow through the sensor.
SPECIFICATIONS6
ELECTRICAL CONNECTION
Maximum ratings
Supply voltage
2.7 ... 5.5 V
Temperature limits
Compensated
Operating
Storage
0 ... 50 °C
-25 ... 85 °C
-40 ... 125 °C
Humidity limits (non-condensing)
0 ... 95 %RH
Vibration1
20 g
2
Mechanical shock
30 g
Pin Connection
2
SD A
3
I / C*
4
+Vs
5
GND
6
SC L
* Internal connection.
Do not connect for any reason
E / 11354 / H
1/7
www.first-sensor.com
www.sensortechnics.com
WBI Series
Mass flow sensors for gases
FLOW SENSOR CHARACTERISTICS7
(VS = 5 ±0.01 V, TA = 20 °C, PAbs= 101.325 kPa)
Part no.
WBIM200DU...
WBIM200DB...
WBIL001DU...
WBIL001DB...
Flow range
0...200 sccm
0...±200 sccm
0...1 slpm
0...±1 slpm
Max. flow
ch an g e
Pressure drop
Max. Common
mode pressure
5.0 slpm/sec
0.1 mbar @ 200 sccm
0.5 mbar @ 1 slpm
2 5 p si
Note:
sccm denotes standard cubic centimeters per minute.
slpm denotes standard liter per minute.
PERFORMANCE CHARACTERISTICS6
(VS = 5 ±0.01 V, TA = 20 °C, PAbs= 101.325 kPa, output signal is ratiometric to VS, media = air)
Characteristics
Min.
Typ.
Accuracy3
Total accuracy (0...50 °C)4
Repeatability (incl. hysteresis)
Offset long term stability (1 year)
Noise level
Current consumption (no load)
Response time (t90)
Warm-up time8
Max.
Unit
±(2.0 % of reading +
0.25 %FSO)
±(4.0 % of reading +
0.25 %FSO)
0.25
% of reading
±0.05
10
0.1
12
5
70
%FSS
mA
ms
Digital output
Characteristics
Scale factor
Zero offset tolerance
Full scale span tolerance
Min.
WBIM200...
WBIL001...
Typ.
Max.
150
30
Unit
counts/sccm
±0.25
±2.25
% FS S
Note:
The sensor’s performance is determined by intake flow conditions which depend on mounting and environmental
effects. To ensure laminar flow through the sensor, it should be considered to insert a straight tube with a
length 10 times the inner diameter of the pneumatic connector or a laminar flow element upstream of the
sensor. Additionally, the WBI has to be mounted with both ports horizontally and pins downwards.
E / 11354 / H
2/7
www.first-sensor.com
www.sensortechnics.com
WBI Series
Mass flow sensors for gases
DIGITAL I²C BUS
The WBI complies with the following protocol (Fig. 1):
(according to the read/write bit it sent), and the slave
continues in its complementary mode (receive or
transmit, respectively).
Bus idle (A): Both the series data line and the series
clock line are HIGH.
START condition (B): When SCL is HIGH, a change of SDA
from HIGH to LOW represents a start condition that
initiates data transfer. A start condition must present
before any data transfer commands can be executed.
STOP condition (C): When SCL is HIGH, a change of SDA
from LOW to HIGH represents a stop condition that
ceases data transfer. All the data transfer commands
must be accomplished before a stop condition presents.
DATA (D): After the start condition, the series data line
must be kept steady when the series clock line is
HIGH. The series data line can change during the
period when the series clock line is LOW, and each
data bit must correspond to a clock pulse.
Slave address: The I²C-bus master-slave concept requires
a unique address for each device on the bus. The WBI
has a reserved address (00h) for broadcasting and a
second individual address preconfigured to 01h. The
sensor will listen to both slave addresses. 00h can only
be used for WRITE commands. By programming it is
possible to reset the individual adress to any number
between 1 and 127 (see Comands).
After generating a START condition the master sends
the address byte containing a 7 bit address followed
by a data direction bit (R/W). A "0" indicates a
transmission from master to slave (WRITE), a "1"
indicates a data request (READ).
DATA operation: The address and the data bytes are sent
most significant bit first.
Each data transfer will begin with a start condition and
cease after a stop condition. Every byte put on the
series data line must be 8 bits long. The number of
bytes that can be transmitted per transfer is
unrestricted. Each byte has to be followed by an
Acknowledge/not Acknowledge bit. The number of
data bytes between a start condition and a stop
condition will be decided by the bus master.
If the master wishes to write into the slave then it
repeatedly sends a byte with the slave sending an ACK
bit. (In this situation, the master is in the master transmit
mode and the slave is in the slave receive mode.)
If the master wishes to read from the slave then it
repeatedly receives a byte from the slave, the master
sending an ACK bit after every byte but the last one. (In
this situation, the master is in the master receive
mode and the slave is in the slave transmit mode.)
Acknowledge bit: The master is initially in the master
transmit mode by sending a start bit followed by the
slave address that it wishes to communicate with,
which is finally followed by a single bit representing
whether it wishes to write(0) to or read(1) from the slave.
If the slave exists on the bus then it will respond with
acknowledge (ACK) bit (active low for acknowledged)
for that address. The master must provide an extra
SCL pulse for each ACK bit. The master then
continues in either transmit or receive mode
(A)
(B)
(D)
The master then ends transmission with a stop bit, or it
may send another start bit if it wishes to retain control
of the bus for another transfer (a "combined message").
Note:
The WBI sensor can hold SCL LOW after each data
byte before ACK. The transaction cannot continue until
SCL is HIGH again and therefore the master has to wait.
(D)
(D)
(C)
(A)
SCL
SDA
START
condition
Fig. 1:
Data
valid
Data
allowed
to change
STOP
condition
I²C bus protocol
E / 11354 / H
3/7
www.first-sensor.com
www.sensortechnics.com
WBI Series
Mass flow sensors for gases
DIGITAL I²C BUS (cont.)
Commands: The WBI series flow sensors use a
communication mode based on the command
interpretation mechanism. The data accesses are
accomplished through various commands (Fig. 2):
Read instant flow index
0
1
S slave address R/W A 83h A S slave address R/W A data byte 1 A data byte 2 A data byte 3 A data byte 4 A P
This command is used for enquiring the current instant flow index. The index consists of four 8-bit values,
which are combined to give a 32-bit value as follows:
Flow index = (data byte 1× 16777216 ) + (data byte 2 × 65536 ) + (data byte 3 × 256 ) + (data byte 4 )
The actual flow value can be calculated as the following:
Actual flow value =
flow index
scale factor × 1000
Negative numbers are represented by the two's complement.
Example for WBIM200DBH5:
Sensor output = FF FE 1D C0
Flow index = -123456
Actual flow value =
−123456
= − 0.823 sccm
150 × 1000
Read I²C address
0
1
S slave address R/W A 85h A S slave address R/W A slave address 0 A P
Write I²C address
0
S slave address R/W A 05h A new slave address 0 A P
Auto zero
0
S slave address R/W A 1Ch A 00h A P
generated by master
S = START condition
generated by slave
A = Acknowledge
P = STOP condition
Fig. 2:
WBI commands
E / 11354 / H
4/7
www.first-sensor.com
www.sensortechnics.com
WBI Series
Mass flow sensors for gases
DIGITAL I²C BUS (cont.)
I²C Interface Parameters
Parameter
Symbol
Input high level
Input low level
Output low level
Pull-up resistor
Load capacitance @ SDA
Input capacitance @ SDA/SCL
SCL clock frequency
Bus free time between STOP and START condition
Hold time (repeated) START condition, to first clock pulse
LOW period of SCL
HIGH period of SCL
Setup time repeated START condition
Data hold time
Data setup time
Rise time of both SDA and SCL
Fall time of both SDA and SCL
Setup time for STOP condition
tSU;DAT
tHD;DAT
Min.
Typ.
90
0
Max.
Unit
100
10
10
% of V s
10
C SD A
CI2C_IN
FSC L
tBUF
tHD.STA
tLOW
tHIGH
tSU.STA
tHD.DAT
tSU.DAT
tR
tF
tSU.STO
kΩ
400
10
100
4.7
4.7
4.7
4.0
4.7
0
0.25
pF
kHz
µs
1
0.3
4
tSU;STA tHD;STA
tR
tSU;STO tBUF
tHD;STA
tF
SCL
SDA
tHIGH
tLOW
SCL
SDA
Fig. 3:
Timing characteristics
E / 11354 / H
5/7
www.first-sensor.com
www.sensortechnics.com
WBI Series
Mass flow sensors for gases
OUTLINE DRAWING
Note:
Positive flow direction is
defined as proceeding
from P1 to P2 and results
in positive output.
third angle projection
dimensions in mm
E / 11354 / H
6/7
www.first-sensor.com
www.sensortechnics.com
WBI Series
Mass flow sensors for gases
GAS CORRECTION FACTORS9
Gas type
Ai r
Oxygen (O2)
Nitrogen (N2)
Argon (Ar)
Hydrogen (H2)
Carbon dioxide (CO2)
Gas correction factor
1.0
1.0
1.0
1.18
*
0.67
* For Hydrogen applications, the actual H2 calibration is performed whenever possible.
Specification notes:
1.
2.
3.
4.
5.
6.
7.
8.
9.
Sweep 20 to 2000 Hz, 8 min, 4 cycles per axis, MIL-STD-883, Method 2007.
5 shocks, 3 axes, MIL-STD-883E, Method 2002.4.
Accuracy is the combined error from offset and span calibration, linearity, hysteresis and repeatability.
Total accuracy is the combined error from offset and span calibration, linearity, hysteresis, repeatability and temperature effects.
Full Scale Span (FSS) is the algebraic difference between the output signal for the highest and lowest specified flow.
Specification is preliminary. Data sheet is based on Pre-Series sample verification.
A 5 µm filter is recommended to protect the sensing element from dust particles which may be present in some applications.
Warm-up time is the time from power on to the first stable reading.
To obtain the real flow rates in a specific gas, multiply the readings from the sensor by the gas correction factor in the table. The
factors are approximate and should be used as guidelines only. Sensor performance strongly depends on gas dynamics and has
to be evaluated in the respective application.
ORDERING INFORMATION
Series
Options
WBI
Flow range
Gas
Flow direction
M200 200 sccm D* Dry air
B Bidirectional
L 001
U Unidirectional
1 slpm
Grade
H High
Calibration
5 5V
(VS=2.7...5.5 V)
* other calibration
gases on request
Example:
WBI
M200
D
U
H
5
First Sensor reserves the right to make changes to any products herein.
First Sensor does not assume any liability arising out of the application
or use of any product or circuit described herein, neither does it convey
any license under its patent rights nor the rights of others.
E / 11354 / H
7/7
www.first-sensor.com
www.sensortechnics.com