144L...-PCB Series Signal conditioned precision pressure transducers FEATURES · 0...5 to 0...50 mbar, 0...1 to 0...20 "H2O gage or differential pressure (custom calibrations available) · 0...5 V output · Single power supply · Internal supply regulation · Precision temperature compensated and calibrated SERVICE Non-corrosive, non-ionic working fluids, such as dry air and dry gases SPECIFICATIONS ELECTRICAL CONNECTION Maximum ratings Supply voltage Maximum load current source sink Temperature limits Storage Operating Compensated Lead temperature (10 sec soldering) Humidity limits pressure inlets only 7.5...24 V 20 mA 10 mA -25 to 85°C -10 to 70°C 0 to 50°C 300°C 144L...-PCB 7.5 V to 24 V Vout V 0 - 95 %RH Proof pressure1 144LP...-PCB 144LU...-PCB 300 mbar 4 psi Common mode pressure 144LP...-PCB 144LU...-PCB 600 mbar 8 psi Januar 2005 / 004 +Vs -Vs 1/4 www.first-sensor.com www.sensortechnics.com 144L...-PCB Series Signal conditioned precision pressure transducers PERFORMANCE CHARACTERISTICS Ω, tamb = 25°C) (unless otherwise noted VS = 8 V, RL > 100 kΩ Characteristics Operating pressure Min. 144LP05D-PCB 144LP10D-PCB 144LP20D-PCB 144LP50D-PCB 144LU01D-PCB 144LU02D-PCB 144LU05D-PCB 144LU10D-PCB 144LU20D-PCB 2 Zero pressure offset Full scale output Typ. 0 0 0 0 0 0 0 0 0 Max. 5 10 20 50 1 2 5 10 20 Unit mbar "H2O -0.05 0 0.05 4.9 5.0 5.1 V 5.0 Full scale span 3 Thermal effects (0 to 50°C) 4 Offset Span 144LU01D-PCB 144LU02D-PCB 144LP05D-PCB all other devices 144LU01D-PCB all other devices Non-linearity and hysteresis (BSL)5 ±0.05 ±0.02 ±0.02 ±0.01 ±0.05 ±0.02 ±0.15 ±0.07 ±0.07 ±0.03 ±0.10 ±0.04 0.1 0.5 %FSO/°C %FSO Long term stability6 ±0.2 Response time (10 to 90%) 200 µs 0.5 0.1 %FSO/g 4.2 mA 0.05 0.03 %FSO/V Position sensitivity 144LU01D-PCB all other devices Current consumption Power supply rejection Offset Span Specification notes: 1. 2. 3. 4. 5. 6. 2/4 Proof pressure is the maximum pressure which may be applied without causing damage to the sensing element. The output signal is proportional to the pressure applied to port B, relative to port A, e.g. the output signal increases when vacuum is applied to port A relative to port B. Full scale span is the algebraic difference between the positive full scale output and the zero pressure offset. Non-linearity refers to the Best Straight Line fit measured for offset pressure, full scale pressure and 1/2 full scale pressure. Thermal effects tested and guaranteed from 0 to 50°C relative to 25°C. All specifications shown are relative to 25°C. Change in output after one year or 1 million pressure cycles. Januar 2005 / 004 www.first-sensor.com www.sensortechnics.com 144L...-PCB Series Signal conditioned precision pressure transducers OUTLINE DRAWING 32.0 14.0 9.1 3 terminals 0.56 diam. 3.5 diam. 2 x mounting holes mass: 20 g Januar 2005 / 004 22.7 4.8 diam. 52.4 60.0 10.7 2 x 5.08 typ. 18.3 30.0 port B2 (high pressure board) NOM 6.0 +Vs -Vs Vout 12.6 dimensions mm 3/4 www.first-sensor.com www.sensortechnics.com 144L...-PCB Series Signal conditioned precision pressure transducers ORDERING INFORMATION Operating pressure Order number 0...5 mbar 144LP05D-PCB 0...10 mbar 144LP10D-PCB 0...20 mbar 144LP20D-PCB 0...50 mbar 144LP50D-PCB 0...1 "H2O 144LU01D-PCB 0...2 "H2O 144LU02D-PCB 0...5 "H2O 144LU05D-PCB 0...10 "H2O 144LU10D-PCB 0...20 "H2O 144LU20D-PCB Custom calibrations available First Sensor reserves the right to make changes to any products herein. First Sensor does not assume any liability arising out of the application or use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others. 4/4 Januar 2005 / 004 www.first-sensor.com www.sensortechnics.com