RPS-160 160W Single Output Medical Type series Authorized Distributor: Total Power International, Inc. Toll Free:877-646-0900(US) www.total-power.com Features : Universal AC input / Full range Built-in active PFC function Protections: Short circuit / Overload / Over voltage / Over temperature 5"x3" compact size Free air convection for 110W and 160W with 20.5 CFM forced air With power good and fail signal output Built-in remote sense function for 5~15V No load power consumption under 0.75W by PS-ON control (Optional) Standby 5V@1A with Fan/0.6A without Fan (Optional) 3 years warranty RPS G 160 12 G: With 5Vsb Blank: Without 5Vsb EN60601-1 IEC60601-1 SPECIFICATION RPS MODEL OUTPUT RPS -160-12 RPS -160-15 RPS -160-24 RPS 12V 15V 24V 48V -160-48 RATED CURRENT (20.5CFM) 30A 12.9A 10.3A 6.5A 3.25A CURRENT RANGE (convection) 0 ~ 20A CURRENT RANGE (20.5CFM) 0 ~ 30A 0 ~ 9.1A 0 ~ 7.3A 0 ~ 4.6A 0 ~ 2.3A 0 ~ 12.9A 0 ~ 10.3A 0 ~ 6.5A 0 ~ 3.25A RATED POWER (convection) Note.7 103W 112.2W 112.5W 113.4W 113.4W RATED POWER (20.5CFM) Note.8 155W RIPPLE & NOISE (max.) Note.2 100mVp-p 159.8W 159.5W 161W 161W 100mVp-p 120mVp-p 150mVp-p 250mVp-p CH1:10.8 ~ 13.2V 3.0% CH1:13.5 ~ 16.5V 3.0% CH1:22 ~ 27V 2.0% CH1:43.2 ~ 52.8V 2.0% VOLTAGE ADJ. RANGE VOLTAGE TOLERANCE Note.3 LINE REGULATION LOAD REGULATION CH1:4.5 ~ 5.5V 4.0% 0.5% 0.5% 0.5% 0.5% 0.5% 1.0% 1.0% 1.0% 1.0% 1.0% SETUP, RISE TIME 1200ms, 30ms/230VAC HOLD UP TIME (Typ.) 16ms/230VAC/115VAC at full load VOLTAGE RANGE INPUT -160-5 5V DC VOLTAGE Note.6 90 ~ 264VAC 2500ms, 30ms/115VAC at full load 127 ~ 370VDC FREQUENCY RANGE 47 ~ 63Hz POWER FACTOR (Typ.) PF>0.93/230VAC EFFICIENCY (Typ.) 85% AC CURRENT (Typ.) INRUSH CURRENT (Typ.) 2A/115VAC 1.1A/230VAC COLD START 35A/115VAC 70A/230VAC PF>0.98/115VAC at full load 87% 87% 87% LEAKAGE CURRENT Earth leakage current <300uA / 264VAC, Patient leakage current <100uA/264VAC OVERLOAD 105 ~ 135% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed OVER VOLTAGE PROTECTION OVER TEMPERATURE 5.75 ~ 6.75V 13.8 ~ 16.2V 17.25 ~ 20.25V 27.6 ~ 32.4V 88% 55.2 ~ 64.8V Protection type : Shut down o/p voltage, re-power on to recover 110 (5V),105 (12V,15V,24V,48V) (TSW1 : detect on heatsink of power transistor) 95 (5V),90 (12V,15V,24V,48V) (TSW2 : detect on heatsink of power transistor) Protection type : (TSW1)Shut down o/p voltage, recovers automatically after temperature goes down Protection type : (TSW2)Shut down o/p voltage, re-power on to recover FUNCTION PS-ON INPUT SIGNAL (OPTIONAL) Power on: PS-ON = "Hi" or ">2V" ; Power off: PS-ON = "Low" or "<0.5V" PF>1ms POWER GOOD / POWER FAIL 500ms>PG>10ms 5 ~ 15V REMOTE SENSE WORKING TEMP. WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY TEMP. COEFFICIENT VIBRATION SAFETY STANDARDS WITHSTAND VOLTAGE SAFETY & EMC (Note 4) NOTE (Refer to output load derating curve) -40 ~ +85 , 10 ~ 95% RH 0.03%/ (0 ~ 50 ) 10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes UL60601-1, TUV EN60601-1 approved I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC ISOLATION RESISTANCE I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH EMI CONDUCTION & RADIATION Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B HARMONIC CURRENT Compliance to EN61000-3-2,-3 EMS IMMUNITY MTBF OTHERS -20 ~ +70 20 ~ 90% RH non-condensing DIMENSION PACKING Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A 230.5Khrs min. MIL-HDBK-217F (25 ) 127*76.2*34.6mm (L*W*H) 0.32Kg; 36pcs/12.5Kg/0.79CUFT 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. 5. HS1,HS2 & HS3 can not be shorted. 6. Derating may be needed under low input voltages. Please check the derating curve for more details. 7. The rated power includes 5Vsb @ 0.6A. 8. The rated power includes 5Vsb @ 1A. Authorized Distributor: Total Power International, Inc. Toll Free:877-646-0900(US) www.total-power.com 160W Medical Type with PFC Function Universal AC input / Full range Built-in active PFC function Protections: Short circuit / Overload / Over voltage / Over temperature Free air convection for 105W and 160W with 20.5 CFM forced air (RPS-160); Free air convection for 100W and 150W with 20.5 CFM forced air (RPD / T-160) 5"x3" compact size With power good and fail signal output Built-in remote sense function for 5~15V (RPS-160) Standby 5V@1A (Optional) No load power consumption under 0.75W by PS-ON control (Optional) 3 years warranty EN 60601-1 IEC 60601-1 NE W 127x 76.2x 33mm AC input voltage range ....... 90~264VAC; 127~370VDC AC inrush current ............... Cold start, 35A at 115VAC, 70A at 230VAC DC adjustment range .......... RPS-160: ±10% rated output voltage RPD/T-160: -5~+10% for CH1 Overload protection ............ 105~135% hiccup mode, auto-recovery Over voltage protection .......CH1: 115%~135% shut off, re-power on to recover Leakage current ................. Less than 300uA at 240VAC Setup, rise, hold up time ..... 1200ms, 30ms, 16ms at full load and 230VAC Withstand voltage ...............I/P - O/P: 4KVAC, I/P - FG: 1.5KVAC, 1 minute Working temperature .......... -20ºC~70ºC (refer to output derating curve) Safety standards ............... UL60601-1, TUV EN60601-1 approved EMC standards .................. EN55011/EN55022 class B, EN61000-4-2,3,4,5,6,8,11, EN61000-3-2,3, EN60601-1-2, ENV50204 Connection ........................ 3+8+2(4)P/3.96mm pitch JST B3P / B8P-VH, B2B (B4B)-XH Sing le O u tpu t Stock No. 12841 Model No. RPS-160-5 Output 5V, 0~30.0A Tol. ±2% R&N 100mV Effi. 85% 12842 RPS-160-12 12V, 0~12.9A ±1% 100mV 87% 12843 RPS-160-15 15V, 0~10.3A ±1% 100mV 87% 12844 RPS-160-24 24V, 0~6.5A ±1% 150mV 87% 12845 RPS-160-48 48V, 0~3.25A ±1% 250mV 88% D u al O ut pu t Stock No. 12851 Model No. RPD-160B Output 5V, 1.0~12A Tol. ±2% R&N 80mV 24V, 0.2~3.6A ±6% 200mV Effi. Max. 85% 151W Tr i ple O ut put Stock No. Model No. 12861 RPT-160A 12862 RPT-160B 12863 RPT-160C 12864 RPT-160D Output Tol. R&N 5V, 0.6~14A ±2% 80mV 12V, 0.2~5.5A ±5% 120mV -5V, 0.1~1.0A -5%,+7% 120mV 5V, 0.6~14A ±2% 80mV 12V, 0.2~5.0A ±5% 180mV -12V, 0.1~1.0A -4%,+7% 120mV 5V, 0.6~14A ±2% 100mV 15V, 0.1~3.6A ±4% 150mV -15V, 0.1~1.0A -5%,+7% 150mV 5V, 0.3~11A ±2% 100mV 12V, 0.2~5.0A ±7% 120mV 24V, 0.15~1.2A ±7% 200mV Effi. 84% Max. 146W 84% 147W 83% 144W 84% 148W Energy Saving We care about energy saving. This logo represents that this model has "low no load power consumption"! RPS-160 160W Single Output Medical Type series Authorized Distributor: Total Power International, Inc. Toll Free:877-646-0900(US) www.total-power.com Mechanical Specification Unit:mm FAN AC Input Connector (CN1) : JST B3P-VH or equivalent 5cm Pin No. 1 2 3 5 20.5CFM min. SVR1 CN3 66.2 76.2 HS1 1 2 3 4 1 2 3 CN1 Pin No. 1,2,3,4 5,6,7,8 1 2 3 4 5 6 7 8 HS3 FS2 CN2 JST VHR or equivalent JST SVH-21T-P1.1 or equivalent 117 Assignment +V -V Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent Pin No. Status Mating Housing Terminal 1 2 3 4 PG GND -S +S JST XHP or equivalent JST SXH-001T-P0.6 or equivalent CN901 4- 5 Terminal Power Good Connector(CN3):JST B4B-XH or equivalent LED CN901 Mating Housing DC Output Connector (CN2) : JST B8P-VH or equivalent HS2 FS1 Assignment AC/N No Pin AC/L 1234 3.5 5VSB Connector(CN901) : JST B-XH or equivalent 127 Pin No. 1 2,4 3 PS/ON 34.6 5mA(max.) GND Assignment Mating Housing Terminal JST XHP or equivalent JST SXH-001T or equivalent PS/ON GND 5VSB 3 max. HS1,HS2,HS3 can not be shorted Block Diagram fosc :100KHz(5V) 70KHz(12~48V) +5VSB DETERCTION CIRCUIT RECTIFIERS & FILTER POWER SWITCHING PWM CONTROL RECTIFIERS & PFC EMI FILTER I/P +5VSB +S +V -V -S PG RECTIFIERS & FILTER POWER SWITCHING PG DETECTION CIRCUIT FG CONTROL O.L.P. O.V.P. PS/ON CONTROL PS/ON Output Derating VS Input Voltage Derating Curve Ta=25 100 100 With FAN 90 80 80 70 Without FAN LOAD (%) LOAD (%) 60 40 20 -20 0 10 20 30 40 50 AMBIENT TEMPERATURE ( ) 60 70 (HORIZONTAL) 60 50 40 90 95 100 115 120 140 160 180 200 220 INPUT VOLTAGE (VAC) 60Hz 240 264