PMK150-D series

PMK150-D
1 5 0 Wa t t E n c l o s ed with DIN Rail mount and PFC
series
150 Watts Output Power
Single and Dual Outputs
Universal 90-264VAC Input
5VDC to 48VDC Outputs
4242VDC Input to Output Isolation
Active Power Factor Correction
3.2”W x 5.5”L x 1.6”H
Model Number
Output Voltage
Output Amps (max) Line Regulation (max) Ripple & Noise
SINGLE OUTPUT
PMK150S-5D
PMK150S-12D
PMK150S-15D
PMK150S-24D
PMK150S-48D
5 VDC
12 VDC
15 VDC
24 VDC
48 VDC
30
12.5
10
6.3
3.2
±1%
±1%
±1%
±1%
±1%
150mV pk-pk
150mV pk-pk
150mV pk-pk
150mV pk-pk
250mV pk-pk
DUAL OUTPUT
PMK150D-AD
PMK150D-BD
PMK150D-CD
PMK150D-DD
5/12 VDC
5/24 VDC
12/24 VDC
12/48 VDC
15/7
15/3.5
7/3.5
7/1.5
±0.5/1%
±0.5/2%
±1/2%
±1/4%
50/100mV pk-pk
50/200mV pk-pk
100/200mV pk-pk
100/400mV pk-pk
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
PMK150-D
1 5 0 Wa t t E n c l o s ed with DIN Rail mount and PFC
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
Power Factor Correction
Inrush Current, typ:
Leakage Current
ENVIRONMENTAL SPECIFICATIONS
90-264 VAC
47-63 Hz
0.96 min.
30A @ 115VAC
60A @ 230VAC *
<300UA @ 264VAC, 50Hz
OUTPUT SPECIFICATIONS
Voltage and Current (Note 6)
See Selection Chart
Line Regulation (Note 8)
See Selection Chart
Load Regulation (10%-FL, Note 7)
Singles:
± 1%, max.
Duals:
±3/5%
Preset Accuracy (Note 9)
±1%, typ
Ripple/Noise (Notes 1, 3, 10)
See Selection Chart
Over Voltage Protection
Auto Recover after fault
condition is removed *
Short Circuit Protection
Auto Recover after fault
condition is removed *
Hold Up Time
20mS, typ (Nom I/P, FL)
GENERAL SPECIFICATIONS
Isolation (Note 4)
Efficiency
Safety UL/cUL: UL-EU:
CE: CB: series
I/P-O/P: 4242VDC
I/P-Ground: 2828VDC
O/P-Ground: 707VDC
75%, min.
UL60950-1 2nd ed./
C22.2 60950-1 2nd ed.
EN60950-1 2nd ed.
EN60601-1-2
IEC60950-1 2nd ed. All specifications are typical at nominal input, full load, and 25°C unless otherwise noted
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
WWW.ASTRODYNE.COM
Oper. Temperature
Storage Temperature
Relative Humidity
EMC
MTBF
-10 to +70°C (See Derate Curve)
-25 to +85°C *
0% to +95%, non-cond *
EN55011 Class B
270,000 Hrs
Mil Std 217, 25°C
PHYSICAL SPECIFICATIONS
Size
Construction
Weight
3.2” x 5.5” x 1.6”
Enclosed DIN Rail
1.4 lb, (635g)
NOTES
1. All measurements should be made directly at the terminals of the power supply
2. All specifications typical @ 25°C, unless otherwise noted, at nominal line and load.
3. Ripple and noise dependent upon output voltage as specified per particular model.
4. Isolation for up to 1 minute duration.
5. Specified for free air convection cooling.
6. Minimum load is not required for proper operation.
7. Load Regulation is measured by change ±40% of measured output load from 60% full load, with the other output set to 60% full load.
8. Line Regulation measured from 90-264VAC. 100VAC minimum required for full load start.
9. Preset Accuracy measured at nominal load, 120VAC input.
10. O/P Noise measured directly at the pins/terminals at nominal load, 0.1uF bypass and 47uF electrolytic, pk-pk @ 20MHz bandwidth.
Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
PMK150-D
1 5 0 Wa t t E n c l o s ed with DIN Rail mount and PFC
series
MECHANICAL DIMENSIONS
65 .70 m m [2.58 7 in]
134 .00 m m [5.276 in]
13 9.70 mm [5.500 in]
28.20 mm [1.110 in]
5.70 mm [0.224 i n]
15.10 mm [0.594 in]
12.00 mm [0.472 in]
24.00 mm [0.945 in]
M3 hole size
80.40 mm [3.165 in]
65.30 mm [2.571 in]
7_ M3x0.5 - 6H
( The L ength : 3 mm )
40.60 mm [1.598 in]
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
PMK150-D
1 5 0 Wa t t E n c l o s ed with DIN Rail mount and PFC
series
98.90 m m [3.894 in]
WWW.ASTRODYNE.COM
6.50 mm [0.256 in]
1.50 m m [0.059 in]
6.50 mm [0.256 in]
8.80 mm [0.346 in]
139.70 mm [5.500 i n]
80.40 mm [3.165 in]
40.60 m m [1.598 in]
50.90 mm [2.004 in]
20.20 m m [0.795 in]
60.20 mm [2.370 in]
M3 hole size
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 5 0 Wa t t E n c l o s ed with DIN Rail mount and PFC
PMK150-D
series
OUTPUT DERATING CURVE - SINGLE OUTPUT
OUTPUT DERATING CURVE - DUAL OUTPUT
1. Forced air colling 10 CFM
2. Free air convection
- Open Frame, U-Channel
3. Free air convection
- Enclosed
WWW.ASTRODYNE.COM
4112
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458