Manual - Extech Instruments

User's Manual
Autoranging Datalogging Insulation Tester
(Megohmmeter)
with PC Interface
Model 380366
Introduction
Congratulations on your purchase of the Extech 380366 Datalogging Insulation Tester. The 380366
measures insulation resistance to 4000MΩ and includes Continuity and Voltage tests. The built-in
datalogger remotely stores approx. 16,000 readings for later transfer to a PC. The tester can also
take and record readings to a PC in real time. This professional tester, with proper care, will provide
years of safe reliable service.
Safety
1.
Circuits under test must be de-energized and isolated before connections are made (except for
voltage measurements).
2.
Circuit connections must not be touched during a test. Use extreme caution when working near
bare conductors and bus bars. Accidental contact with conductors could result in electrical
shock.
3.
Use caution when working near voltages above 60VDC or 30VACrms.
4.
After insulation tests, capacitors must be discharged.
5.
Test leads (including alligator clips) must be in good working order, clean and without broken
or cracked insulation.
6.
When servicing, use only specified replacement parts.
International Safety Symbols
Caution, refer to this manual before using this meter
Dangerous Voltages
Meter is protected throughout by double or reinforced insulation
Meter Description
1.
LCD Display
2.
Megohmmeter TEST pushbutton
3.
Test LOCK and DATA HOLD button
4.
RECORD button
5.
LCD Backlight / Wake button
6.
Function select switch
7.
Storage compartment
8.
RS-232 interface jack
9.
Test Lead input jacks (HI & LO)
10. Test Leads
2
380366-EU-EN-V1.7-4/12
Operation
Warning
Ensure that the circuit under test does not include components that can be damaged by 1000VDC; such
devices include power factor correction capacitors, low voltage mineral insulated cables, electronic light
dimmers, and ballasts/starters for fluorescent lamps.
Connecting Test Leads
For all functions (Megohmmeter, Resistance, Continuity and AC Voltage), connect the red test lead
to the HI terminal and the black lead to the LO terminal.
Test Lead Check
1.
Connect the red test lead to the HI input terminal; black lead to the LO terminal.
2.
Set the rotary switch to the resistance/ continuity position ( .
3.
Touch the test lead tips together. Resistance should be less than 1Ω and the audio tone
should sound.
Ω)
4.
With the leads not touching, the display should read “-OL-“. (over range).
5.
If any readings are displayed other than the readings just described, the test leads should be
considered faulty and must be replaced before using the meter. Failure to do so could result in
damage to equipment and electrical shock.
Insulation Resistance Measurements (Megohmmeter Tests)
Warning: Do not perform Insulation Resistance measurements if AC Voltage is present on circuit under test.
1. Connect the red test lead to the HI input terminal; black lead to the LO terminal.
2.
Set the rotary switch to the desired function.
3. Connect the tips of the test leads to the equipment under test.
4. Press and hold down the red TEST button. Release the test button to stop the test. The meter
will then hold the resistance reading. This is indicated by the H icon in the upper left corner of
the display.
5. To clear the display, press the LOCK/HOLD button twice.
Note that the charge stored on the device under test will be automatically discharged when the
testing process is finished
3-Minute Test
For hands-free operation, use the Test Lock feature.
1. Connect the red test lead to the HI input terminal; black lead to the LO terminal
2.
Set the rotary switch to the desired function.
3.
Connect the tips of the test leads to the equipment under test.
4.
Press the LOCK button. The LOCK icon will be displayed. Press the red TEST button. A 3minute continuous test will begin. The red TEST button can be pressed at any time to end the
test. If the TEST button is not pressed the meter automatically stops the test after 3 minutes.
The meter will then hold the resistance reading. This is indicated by the H icon in the upper left
corner of the display.
5.
To clear the display, press the LOCK/HOLD button twice.
3
380366-EU-EN-V1.7-4/12
Note: The
symbol will appear in the upper right corner of the display and an audible tone will
sound whenever the high voltage is applied to the test leads.
Analog Bargraph Display
The bargraph can indicate up to 10GΩ. When the measured value is between 4GΩ and 10GΩ, the
LCD displays "-HI-". When the measured value is higher than 10GΩ the LCD displays "-OL-".
Low Resistance and Audible Continuity Tests
WARNING
Do not run this test unless ACV = 0. Do not use this mode to check diodes.
1. Set the Rotary switch to the audible continuity/resistance position.
2. Connect the red test lead to the HI input terminal; black lead to the LO terminal.
3. Connect the tips of the test leads to both ends of the circuit under test and read the resistance
on the LCD.
4. When the resistance of a circuit is less than approx. 40Ω an audible tone will sound.
AC Voltage Tests
1. Set the Rotary switch to the red ACV position.
2. Connect the red test lead to the HI terminal and the black test lead to the LO terminal.
3. Connect the other end of the test leads to the circuit under test.
4. Read the voltage value on the LCD.
Data Hold Function
The Data Hold function freezes the displayed reading. Data Hold works automatically on Insulation
Resistance Measurements (Megohmmeter tests) but, for the other tests, the user must press the
HOLD key to freeze the meter reading. Press HOLD again to exit the Data Hold mode.
Auto Power OFF Feature
After a 30-minute period of non-use, the meter automatically shuts down. This will conserve battery
life. To wake the meter, press the Backlight button.
Note that the meter is not completely off until the function switch is set to the OFF position. Always
turn the switch to the OFF position when not in use.
4
380366-EU-EN-V1.7-4/12
Built-in Datalogger
The meter can store approx. 16,000 readings in the built-in datalogger. Readings are grouped in
‘SETS’. Each recording session constitutes a ‘SET’. Recording a new set does not erase a
previously recorded set.
The readings remain in the datalogger memory even when power to the tester is off (as long as the
batteries are fresh). Readings can be transferred, at a later time, to a PC. Use the meter’s RECORD
button to operate the meter’s Datalogger in the following ways:
a. Clearing Datalogger memory
1. Press and hold the RECORD button while turning the meter on.
2. Release the RECORD button when &/r (clear) appears on the display. The entire datalogger
memory will now be erased.
b. Capturing one data point
Press the RECORD button for one second to capture one reading. The REC display icon will
briefly appear at the top of the LCD and the large digits will flash the data point number (the
recorded measurements are numbered). Each reading captured by this method is considered a
Set.
c. Capturing a continuous stream of data points
1. Press and hold the RECORD button for approx. 3 seconds until a long tone is heard. The
REC display icon will appear and remain on the display for as long as the recording session
lasts.
2. To stop recording, press the RECORD button again. The REC indicator will switch off. Each
group of readings captured in this manner is considered a Set.
3. Readings will be taken every 2 seconds (default sampling time). To change the sampling
time, refer to the section ‘Datalogger Software Mode’.
The following sections cover meter-to-PC data downloading & other PC interface features.
5
380366-EU-EN-V1.7-4/12
PC Interface
Minimum hardware requirements
486 PC or better with CD-ROM drive
Communications Settings
Baud Rate: 9600, Parity: None, Data bits: 8, Stop bits: 1
Connecting the meter to a PC
Connect one end of the supplied cable to the PC Com Port (DB9 connector) and the other end
(infrared connector) to the top of the meter.
Installing Software (compatible operating systems: Windows
TM
95, 98, 2000, Me, & XP)
1.
Insert the supplied CD-ROM in the PC’s CD-ROM drive.
2.
The SETUP program will run automatically. If not, press the Windows
RUN. Then type SETUP at the prompt and press the ENTER key.
3.
Follow the on-screen instructions for the setup program.
TM
START key and select
Running the Software
1.
Open the ‘Insulate’ program using the program icon.
2.
Select PC communications port COM1 or COM2 when prompted.
3.
The Program screen will appear as a facsimile of the meter's front panel (see below).
Sampling
Time
Control
Icons
System
Time
Meter
display
Function
Switch
Com.
Status
Meter
buttons
Data Acquisition and Datalogger modes
These two modes of operation, distinctly different, are explained in detail separately.
For the Data Acquisition mode of operation, the meter must be connected to the PC. In this mode,
readings taken by the meter are displayed and stored on the PC.
In the Datalogger mode, the meter records data remotely then, at a later time, the meter is
connected to the PC and the readings are transferred. Instructions for datalogging were covered in
the previous section. Instructions for transferring data and navigating the software are given below.
6
380366-EU-EN-V1.7-4/12
Data Acquisition Software Mode
As explained above, the Data Acquisition feature allows the user to connect the meter to the PC and
take and record readings simultaneously. With the meter connected to the PC and with the software
running, follow the steps below:
System Time Set
Click the computer icon from the row of Control Icons at the top of the main software screen. This is
the SYSTEM TIME SET button. The PC system date and time will now be shared by the software
program. Note that if your PC date and time are not set correctly, do so at this time. The system
time will be shown on the main software screen.
Input ID Code
The ID Code is used as a reference number specific to a particular data acquisition session. The ID
Code does not affect the software other than allowing the user to match a set of data to a unique ID.
See the LOGGER screen at right
1. Select MEMORY from the OPTION menu on the
main software screen.
2. Highlight the number in the ID Code field and type
a new reference number. Valid numbers are 0
through 9999.
3. Click the ID Code button.
Sampling Time
The Sampling Time tells the software how often to log
a reading. The sampling time can be set from 1 to
65,400 seconds. See the LOGGER screen at right.
1.
Click on the ‘wrist-watch’ icon from the row of
Control Icons on the main software screen.
2.
Highlight the existing value in the SAMPLING
field.
3.
Type in the desired sampling time.
4.
Click the Sampling button when finished.
Creating a File to store data
1.
Click the ‘floppy disk’ icon from the row of Control Icons on the main software screen and type
a unique filename when prompted.
2.
This File can later be opened by clicking on the ‘folder’ icon (4 from left) from the row of
Control Icons on the main software screen.
3.
Saved files can be opened in spreadsheet programs as *.dat text files by opening the
spreadsheet application, selecting OPEN FILE, choosing ALL FILE TYPES, and locating the
file previously stored.
th
Start and Stop Recording
1.
nd
Click the ‘red’ icon (2 from left) from the row of Control Icons on the main software screen to
start recording. If a file has not been set up at this point, the software will prompt for a file
name.
2.
Data will now be logged every n seconds (where n is the programmed sampling time).
3.
To stop recording, click the ‘black’ icon (3 from left) in the row of Control Icons.
rd
7
380366-EU-EN-V1.7-4/12
Open an existing Data File
th
1.
Click on the ‘folder’ icon (4 from left) in the
row of Control Icons on the main software
screen.
2.
Select a file from the list of previously stored
files (if any).
3.
When a file is opened, its data list will appear
containing the date/time for each reading,
measurement, unit of measure, and low
battery indication (see diagram at right).
4.
Select PRINT from the menu heading to print
the data list.
5.
To import the data to spreadsheet or other 3
party application: Locate and open the file as
a text file (filename.dat) using the FILE OPEN
rd
commands in the 3 party application.
rd
Soft-key Remote Control
Use the four front-panel keys on the main
software screen to remotely control the meter:
TEST button
Click to start a megohmmeter test.
LOCK button
Click to lock the test for 3 minutes (read about the LOCK function earlier in this manual).
HOLD button
Click to freeze the displayed reading. The ‘H’ display icon will appear while the meter is in the Data
Hold mode. Press HOLD again to exit this mode.
REC button
Click to put the meter into continuous recording mode. Readings will be kept in the open file. If a
file is not open, the software will prompt for a file name.
rec button
Click to record one reading to the open file.
Backlight button
Click to activate the meter’s display backlight.
8
380366-EU-EN-V1.7-4/12
Datalogger Software Mode
Datalogging
Instructions for storing data in the meter’s built-in datalogger are provided in the ‘Built-In Datalogger’
section of this manual.
The following section details all of the datalogger functions available in the supplied software.
Datalogging Software Window (LOGGER)
rd
1. From the main software screen, click the ‘datalog’ icon (3 from right) from the row of Control
Icons. The LOGGER window will open (shown below).
2. To set the SAMPLING TIME (rate at which datalogger stores readings), type a number between
1 and 65,000 seconds in the SAMPLING field. Note that the meter must be connected to the PC
in order for the Sampling Time to be transferred to the meter. Once the sampling time is set, the
meter will datalog at the rate selected.
3. Click the TIME OF RECORDING button to download all of the data sets from the meter to the
PC. The data lists will appear as shown on the logger window diagram below.
4.
Select a set of data by clicking on the set number. The NUMBER OF SET field informs the user
as to which set is selected. The NUMBER OF RECORDS field lets the user know the number of
readings in the selected set.
5. The MEMORY and REMAINING fields inform the user as to the total memory capacity (32K)
and the remaining memory.
6. Press the SHOW DATA button to view the raw data. Once the data opens for the selected set,
the user can save or print the data by selecting SAVE or PRINT from the menu options.
Size of
overall
memory
Remaining
storage space
User ID Code
& Sample rate
System
Date &
Time
Click to
download all
stored sets
Number of data
sets in memory
List of all
loaded sets
Number of
records in
selected set
Click a
number to
select a set
Click to view
data for the
selected set
9
380366-EU-EN-V1.7-4/12
Applications
Measuring Power Tools and Small Appliances
This section applies to any device under test that
uses a line cord. For double insulated power tools,
the meter’s leads should be connected to the device’s
housing (chuck, blade, etc.) and the ground of the
power cord. Refer to the diagram at right.
Appliance
Testing AC Motors
Disconnect the motor from the line by disconnecting
the wires from the motor terminals or opening the
mains switch.
If the mains switch is opened, and the motor also has
a motor-starter, then the starter must be held in the ON position. With the mains switch opened, the
measured resistance will include the resistance of the motor wire and all other components between
the motor and the main switch. If a weakness is indicated, the motor and other components should
be checked individually. If the motor is disconnected at the motor terminals, connect one meter lead
to the grounded motor housing and the other lead to one of the motor leads. Refer to diagram at
below.
Motor
Starter
GND
Mains
Switch
LINE
Notes:
1. The STARTER must be in the ON position during
the test.
2. The motor-to-meter connection must be made to
the motor housing (ground).
3. The meter-to-main switch connection must be to the
motor side of the main switch as shown.
Testing DC Motors
1.
Disconnect the motor from the line.
2.
To test the brush rigging, field coils and armature, connect one meter lead to the grounded
motor housing and the other lead to the brush on the commutator.
3.
If the resistance measurement indicates a weakness, raise the brushes off of the commutator
and separately test the armature, field coils and brush rigging (one at a time). Leave one lead
connected to the grounded motor housing while testing the motor components. This also
applies to DC Generators.
Testing Cables
1.
Disconnect the cable under test from the line.
2.
Disconnect the opposite end of the cable to avoid errors as a result of
leakage from other equipment.
3.
Check each conductor to ground and/or lead sheath by connecting one
meter lead to ground and/or lead sheath and the other meter lead to
each of the conductors in turn.
4.
Check insulation resistance between conductors by connecting meter
leads to conductors in pairs. Refer to diagram at right. In the diagram,
note that the 3-conductor cable has two wires shorted to the ground
shield. This two-wire/shield connection is then connected to one side of the meter. The
remaining conductor is connected to the other side of the meter.
10
380366-EU-EN-V1.7-4/12
Application Note for Large Installations
In large wiring installations where the insulation of outlets is being tested, more than one insulation resistance
measurement may have to be made to take into account parallel resistances. Divide large systems into subgroups
and test subgroups individually. Also, in large installations, the capacitance of the insulation will be high, thereby
taking longer to charge when being tested. Care must be taken not to finish a measurement until the insulation
capacitance is fully charged (a steady, stable reading is an indication that this is the case). The charge stored in
the insulation after a test will be discharged automatically when the test button is released. Be careful not to turn
the range switch while the test button is pressed.
Maintenance
Battery Replacement
When the low battery symbol ( - + ) appears on the lower left side of the LCD the eight 1.5V AA batteries must be
replaced.
1.
Ensure that the meter is powered down and that the test leads are not connected
2.
Remove the two large Phillips head screws on the rear of the meter
3.
Remove the battery compartment cover
4.
Replace the eight AA batteries ensuring proper polarity
5.
Affix the rear cover and secure the rear screws
You, as the end user, are legally bound (Battery ordinance) to return all used batteries and
accumulators; disposal in the household garbage is prohibited!
You can hand over your used batteries / accumulators at collection points in your community or
wherever batteries / accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the end of its
lifecycle
Cleaning
Periodically wipe the case with a dry cloth. Do not use solvents or abrasives to clean this instrument.
Specifications
General specifications
Display
Backlit 3-3/4 digit (4000 count) LCD with analog indicator
Sampling rate
2.5 samples/sec. (digital); 10 samples/sec. (analog)
Audible Continuity
Low resistance range with audible continuity < 40Ω
Over range indicator
"OL” displayed when measurement exceeds range limits
Low battery indicator
" - +" symbol displayed when battery voltage is low
Datalogger memory
16,000 readings
Datalogger Sample Rate
Programmable: 1 to 65,400 sec/sample or single manual record
Power source
Eight 1.5V AA cells
Power consumption
20 to 90mA depending upon function (40 hours typical)
Environmental
CAT III; Altitude: 2000 meters; for indoor use only
Operating conditions
0 to 40oC (32 to 104oF); < 80% RH
Storage conditions
-10 to 60oC (14 to 140oF); < 80% RH
Dimensions/Weight
190 x 140 x 77mm (7.5 x 5.5 x 3”)/ 900g (2 lbs)
Environmental specifications
•
•
•
•
Installation Category III 600V
Pollution Degree 2
Altitude up to 2000 meters
Intended for indoor use only
11
380366-EU-EN-V1.7-4/12
Range Specifications
Resistance / Audible Continuity
Range
400 Ω
Resolution
Accuracy
0.1 Ω
±1%rdg + 5d
Max. open
circuit voltage
Protection
Overload
12.8V
220Vrms
Audible tone when Resistance <40Ω;
Short circuit current: 280mA (min. 200mA)
AC Voltage (40 to 500Hz)
Range
Resolution
Accuracy
Input
impedance
Overload
Protect.
600VAC
1V
±1.5%rdg + 3d
4.5MΩ
750Vrms
Megohm (Auto Range)
Voltage
250V
500V
Ranges
4, 40, 400,
4000MΩ
Maximum
Resistance
Voltage
Resolution
Accuracy
Accuracy
1kΩ
1000V
Voltage
Range
±3%rdg + 5d (<2000MΩ)
±5%rdg + 5d (>2000MΩ)
1mA Test Current
250V
500V
+20% ~ -0%
+20%~ -0%
+10%~ -0%
Short Circuit
Current
@250KΩ
4000MΩ
@500KΩ
< 1.5mA
1000V
@1MΩ
o
o
Note: Accuracies are specified as % reading + digits at (73°F ± 9°F) 23 C ±5 C < 80% RH