DIGITAL M-ohm HiTESTER 3154

DIGITAL MΩ HiTESTER 3154
Safety Standards Measuring Instruments
For the Laboratory to the Production Line
Six Test Voltages from 25 to 1000 V
The Model 3154 DIGITAL MΩ HiTESTER provides six selectable
test voltages: 25, 50, 100, 250, 500 and 1000 V, in order to support a
wide variety of tests to evaluate electrical and electronic components
and equipment. Comparator and timer functions facilitate easy
insulation resistance testing in accordance with various safety
standards, and a complete external I/O interface is provided for
efficient automated line testing.
1
Efficient Insulation Resistance Testing
✱ Insulation testing of electrical and electronic equipment
Motors, printers, magnetic heads, etc.
✱ Insulation testing of electrical and electronic components
Transformers, coils, switches, relays, etc.
✱ Insulation resistance testing according
to various standards
Machinery, medical equipment, home appliances, office equipment, etc.
✱ High voltage resistance testing
Resistance measurements of components in actual operating circuits.
●Six Measurement Voltages Provided in a
Single Unit
Any of six measurement voltages can be selected for
use in a wide range of applications, from electrical
and electronic component testing to electrical and
electronic equipment. The comparator can be set up
for use with each measurement voltage.
●Storage for Ten Setting States
Up to ten setting states can be stored for quick
selection when testing for compliance with various
standards.
●Easy Standards Testing
Testing for compliance with various safety standards is
simplified by using the comparator and timer functions
for quick pass-fail judgment of minimum acceptable
limits.
●Automatic Electrical Discharge Function
The unit discharges any charge buildup on the measured
object as a result of electrical loading. This electrical
discharge function prevents the effects of one test from
being carried over to a subsequent procedure.
●Remote Control
Operations such as test start/stop and voltage selection
can be controlled remotely, such as for start/stop control
from a switching probe.
●Slow Sampling Selection to Minimize Instability
Selectable fast (10 sample/s) and slow (1 sample/s)
sampling rates are provided. Slow sampling minimizes
instability when measuring at highly reactive test points.
Measurement Voltages and Ranges (Auto and Manual Range Switching)
Rated Measurement Voltage
25V
Measurement Range
2.000, 20.00 or 200.0 MΩ
Accuracy
19.0 to 200.0 MΩ
No-Load Voltage
Min. Measurable Resistance
(*1)
Rated Measurement Current
(*2)
100V
250V
2.000, 20.00, 200.0 or 2000 MΩ
0 to 20.00 MΩ
±2% rdg. ±5 dgt.
±5% rdg.
50V
19.0 to 2000 MΩ
500V
1000V
2.000, 20.00, 200.0 or 4000 MΩ
0 to 100.0 MΩ
1000 to 4000 MΩ
100.1 to 2000 MΩ
1000 to 4000 MΩ
1 to 1.2 times Rated Measurement Voltage
0.025 MΩ
0.05MΩ
Short-Circuit Current
*1. Minimum resistance required to maintain rated measurement voltage.
*2. Maximum current flow through the minimum measurable resistance.
0.1MΩ
0.25MΩ
0.5MΩ
2MΩ
1 to 1.2 mA
0.5 to 0.6 mA
1.2 mA max.
0.6 mA max.
Auto Range
The range increments when the current range is exceeded by 2000
counts (except the top range).
The range decrements when the current range does not reach 190
counts (except from the 2 MΩ range).
2
Full Support for Automatic Operation
Both external I/O and RS-232C interfaces are provided
to support fully automated operation. Measurement
values can be tracked and recorded from the analog
output, and printing is supported on an external printer
using RS-232C.
●Record Variations using Analog Output
This unit provides two types of analog output. Measurements are
output linearly as voltages between 0 and +4 V DC, and insulation
resistance fluctuations can be recorded.
1. Outputs each measurement range.
2. The maximum measurement
range for each measured voltage
is output between
0 and +4 V DC.
●Analog Output Terminal
The maximum measurement range
for each measured voltage:
DC4V
0
0
200MΩ (25/50V)
2000MΩ (100/250V)
4000MΩ (500/1000V)
● Rear Panel
Measurement Terminal
(always connected to front panel
terminal)
●RS-232C Connector
Automatic calculation can be performed on a PC, and printing
can be performed using the 9442 printer.
9442 Printer
Printing method
Paper width
Printing speed
Power source
: Thermal serial dot printer
: 112 mm
: 52.5 cps
: AC adapter 9443 , or supplied nickel-hydride battery
(Charged through the 9443; printing capability approx. 3000
lines with full charge)
●External I/O Connector
Pin
I/O
Signal
Description
1•2
O
INT.GND
Connects to internal ground and power line ground lead
3
I
EXT.COM
4
O
FAIL
LOW indicates Fail judgment
Connect to ground of external device
5
O
PASS
LOW indicates Pass judgment
6
O
TEST
LOW indicates Testing state
7
I
VOLT2
8
I
VOLT1
VOLT0
Sets test voltages (including comparators)
Sets saved test conditions from No. 1 to 7
9
I
10
I
STOP
LOW = STOP key
11
I
START
LOW = START key
12
I
EXT.DCV
Power supplied from an external device
13•14
O
INT.DCV
+5V DC internal power (100 mA max.)
Note: the Cable 9444 and AC Adapter are required to connect the Printer 9442.
Enhanced Efficiency in System Measurements
Efficient testing can be consistently performed
using a controller to conduct protective grounding,
insulation resistance and voltage withstand testing
all at once. Contact us for system upgrades.
Controller Box
DIGITAL MΩ HiTESTER 3154
AC GROUNDING HiTESTER 3157-01
●Measurement current: Max. 31 A
●Resistance measurement range: 0 to 1.800 Ω
AC WITHSTANDING VOLTAGE HiTESTER 3158
●Voltage: Dual range: 0 to 2.5 or 5 kV
●Transformer power handling: 500 VA
Outlet Box
Operation Display Box
3
■ 3154 Specifications
● Basic Specifications
Measurement function : Insulation resistance (Applied DC voltage method)
: 10 samples/s (Fast), 1 sample/s (Slow)
Sampling rates
: Storage contents: rated measurement voltage, comparator
Memory
minimum limit, test mode, beep upon judgment, test time,
delay time, resistance range, sampling rate
Memory capacity: ten sets of settings
(supports saving and loading)
: Selectable between continuous and fail-stop
Test modes
[ Comparator Functions] Minimum Limit
Setting
Resistance range
2 MΩ
20 MΩ
200 MΩ
2000 MΩ
Function
Setting range
Minimum settable limit (MΩ)
0.1/0.2/0.3/0.4/0.5/0.6/0.7/0.8/0.9/1/2
3/4/5/6/7/8/9/10/20
30/40/50/60/70/80/90/100/200 300/400/500/600/700/800/900/1000/2000/3000/4000
Limited to test voltage 500V,1000V for 3000/4000
: Pass = measured value ≥ minimum limit, Fail = measured
value < minimum limit
Judgment processing : Beep sound, PASS/FAIL indicator, output to external I/O
Beep sound selection : "Ring on Pass", "Ring on Fail", or disable beeper
: "- - - -" indication
OFF setting time
Judgments
: After starting measurement, measurement is completed after
0.35s max. wait time + set time (when Delay time is OFF)
: 0.5 to 99s
[ Delay Time Timer Function] (ON/OFF selectable)
Function
Setting range
: Set by key entry; select from following table; set by
RS-232C; can be set to desired resistance
(4000 MΩ range for 500V,1000V)
[ Test Time Timer Function] (ON/OFF selectable)
: After measurement starts, the test time timer and comparator
functions do not start operating until a wait time of max.
0.35 s plus the set delay time has elapsed.
: 0.1 to 99s
[ Response Times] Measurement speed
FAST
Resistance range, test voltage
2 or 20 MΩ
Manual ranging
200 MΩ
Manual ranging
All Voltages
less than 0.5 s
25, 50 V
less than 0.6 s
100 to 1000 V less than 0.5 s
2000 or 4000 MΩ
Manual ranging
SLOW
less than 1.5 s
100 to 1000 V less than 0.7 s
Auto ranging
(including range selection)
25, 50 V
less than 2 s
100 to 1000 V less than 2.5 s
less than 5 s
less than 6 s
● General Specifications
Display type
Operating temperature
and humidity ranges
Storage temperature
and humidity ranges
Guaranteed accuracy
temp. & humidity
Suitable environment
Power supply
: LED
: 0 to 40˚C, max. 80% rh (no condensation)
: -10 to 50˚C, max. 90% rh (no condensation)
: 23 ± 5˚C, max. 80% rh (no condensation)
: Indoors, max. 2000 m altitude
: 100 to 240 VAC
Power line frequency : 50 or 60 Hz
: 15 VA
Maximum rated
power
Withstand voltage : 1.35 kV AC for one minute between power lines and
outer case, external connectors, grounding and measurement
terminals (collectively);
1.1 kV DC for one minute between measurement terminal
and ground terminals;
500 V AC for one minute between EXT I/O terminals
and power line, ground and measurement terminals
(collectively);
Input overload
protection
Dimensions
Mass
Conforming standards
Accessories
: 1100 V DC between HIGH and LOW on measurement
terminal (positive only).
: Approx. 215W × 61H × 213D mm (not including projections)
: Approx. 1.1 kg.
: EMC : EN61326
Safety
: EN61010
: Power Cord (1)
[Not including INT DCV and INT GND terminals in the EXT I/O
terminals.]
3154 DIGITAL MΩ HiTESTER Note: Test leads are not included: optional leads are required for use.
■ Options
TEST LEADS 9185 TEST PROBE 9294 SWITCHED PROBE 9299 OUTPUT CORD (Jack to banana plug) 9094 RS-232C CABLE (9pin-9pin/Cross/1.8 m) 9637 RS-232C CABLE (9pin-25pin/Cross/1.8 m) 9638 PRINTER 9442 RECORDING PAPER (25m, 10 rolls) 1196 CONNECTION CORD(1.3 m) 9257
9185
Lead length
approx. 85 cm
9294
Lead length
approx. 1.2 m
9299
Lead length
approx. 80 cm
AC ADAPTER (for printer, EU) 9443-02 AC ADAPTER (for printer, America) 9443-03 CONNECTION CABLE (for printer) 9444 DISTRIBUTED BY
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@ hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@ hiokiusa.com
All information correct as of Jan. 28, 2010. All specifications are subject to change without notice.
■ Internet HIOKI website http://www.hioki.co.jp/
3154E4-01E