3154 DIGITAL MOhm HiTESTER

2000
3154 DIGITAL MΩ HiTESTER
Safety Standards Measuring Instruments
For the Laboratory to the Production Line
Six Test Voltages from 25 to 1000 V
The Model 3154 DIGITAL MΩ HiTESTER provides six selectable
test voltages: 25, 50, 100, 250, 500 and 1000 V, in order to support a
wide variety of tests to evaluate electrical and electronic components
and equipment. Comparator and timer functions facilitate easy
insulation resistance testing in accordance with various safety
standards, and a complete external I/O interface is provided for
efficient automated line testing.
ISO14001
JQA-E-90091
http://www.hioki.co.jp/
HIOKI company overview, new products, environmental considerations and
other information are available on our website.
1
Efficient Insulation Resistance Testing
✱ Insulation testing of electrical and electronic
equipment
Motors, printers, magnetic heads, etc.
✱ Insulation testing of electrical and electronic
components
Transformers, coils, switches, relays, etc.
✱ Insulation resistance testing according
to various standards
Machinery, medical equipment, home appliances, office equipment, etc.
✱ High voltage resistance testing
Resistance measurements of components in actual operating circuits.
● Six Measurement Voltages Provided in a
Single Unit
Any of six measurement voltages can be selected for
use in a wide range of applications, from electrical
and electronic component testing to electrical and
electronic equipment. The comparator can be set up
for use with each measurement voltage.
● Storage for Ten Setting States
Up to ten setting states can be stored for quick
selection when testing for compliance with various
standards.
● Easy Standards Testing
Testing for compliance with various safety standards is
simplified by using the comparator and timer functions
for quick pass-fail judgment of minimum acceptable
limits.
● Automatic Electrical Discharge Function
The unit discharges any charge buildup on the measured
object as a result of electrical loading. This electrical
discharge function prevents the effects of one test from
being carried over to a subsequent procedure.
● Remote Control
Operations such as test start/stop and voltage selection
can be controlled remotely, such as for start/stop control
from a switching probe.
● Slow Sampling Selection to Minimize Instability
Selectable fast (10 sample/s) and slow (1 sample/s)
sampling rates are provided. Slow sampling minimizes
instability when measuring at highly reactive test points.
Measurement Voltages and Ranges (Auto and Manual Range Switching)
Rated Measurement Voltage
25V
Measurement Range
2.000, 20.00 or 200.0 MΩ
±2% rdg. ±5 dgt.
Accuracy
±5% rdg.
50V
250V
500V
1000V
2.000, 20.00, 200.0 or 2000 MΩ (4000 MΩ 1000V)
0 to 20.00 MΩ
0 to 100.0 MΩ
19.0 to 200.0 MΩ
No-Load Voltage
Min. Measurable Resistance
(*1)
Rated Measurement Current
(*2)
100V
19.0 to 2000 MΩ
100.1 to 2000 MΩ
0 to 999 MΩ
1000 to 2000 MΩ
1000 to 4000 MΩ
0.5MΩ
2MΩ
1 to 1.2 times Rated Measurement Voltage
0.025 MΩ
0.05MΩ
Short-Circuit Current
*1. Minimum resistance required to maintain rated measurement voltage.
*2. Maximum current flow through the minimum measurable resistance.
0.1MΩ
0.25MΩ
1 to 1.2 mA
0.5 to 0.6 mA
1.2 mA max.
0.6 mA max.
Auto Range
The range increments when the current range is exceeded by 2000
counts (except the top range).
The range decrements when the current range does not reach 190
counts (except from the 2 MΩ range).
2
Full Support for Automatic Operation
●Record Variations using Analog Output
Both external I/O and RS-232C interfaces are provided
to support fully automated operation. Measurement
values can be tracked and recorded from the analog
output, and printing is supported on an external printer
using RS-232C.
For the maximum measurement range at 0 to each measurement
voltage, insulation resistance is output linearly as 0 to +4 V DC and
the variations can be recorded
This is not affected by the measurement
range. For a measured voltage of 100V
with a displayed resistance of 100.0 MΩ:
DC4V
4 [V ]×100.0/2000 [MΩ] = 0.2V is
output at the Analog Output terminal
●Analog Output Terminal
0
0
200MΩ (25/50V)
2000MΩ (100/250/500V)
4000MΩ (1000V)
● Rear Panel
Measurement Terminal
(always connected to front panel
terminal)
●External I/O Connector
●RS-232C Connector
Automatic calculation can be performed on a PC, and printing
can be performed using the 9442 printer.
9442 Printer
Printing method
Paper width
Printing speed
Power source
: Thermal serial dot printer
: 112 mm
: 52.5 cps
: 9443 AC adapter, or supplied nickel-hydride battery
(Charged through the 9443; printing capability approx. 3000
lines with full charge)
Pin
I/O
1•2
O
INT.GND Connects to internal ground and power line ground lead
Signal
Description
3
I
EXT.COM Connect to ground of external device
4
O
FAIL
LOW indicates Fail judgment
5
O
PASS
LOW indicates Pass judgment
6
O
TEST
LOW indicates Testing state
7
I
VOLT2
8
I
VOLT1
VOLT0
Sets the test voltage by appropriate combination
(comparators that store each voltage are set simultaneously)
9
I
10
I
STOP
LOW = STOP key
11
I
START
LOW = START key
12
I
EXT.DCV Power supplied from an external device
13•14
O
INT.DCV +5V DC internal power (100 mA max.)
Note: the 9444 Cable and AC Adapter are required to connect the 9442 Printer.
Enhanced Efficiency in System Measurements
Efficient testing can be consistently performed
using a controller to conduct protective
grounding, insulation resistance and voltage
withstand testing all at once. Contact us for
system upgrades.
Controller Box
3154 DIGITAL MΩ HiTESTER
3157-01 AC GROUNDING HiTESTER
●Measurement current: Max. 31 A
●Resistance measurement range: 0 to 1.800 Ω
3158 AC WITHSTANDING VOLTAGE HiTESTER
●Voltage: Dual range: 0 to 2.5 or 5 kV
●Transformer power handling: 500 VA
Outlet Box
Operation Display Box
3
■ 3154 Specifications
● Basic Specifications
[ Test Time Timer Function] (ON/OFF selectable)
Measurement function : Insulation resistance (Applied DC voltage method)
: 10 samples/s (Fast), 1 sample/s (Slow)
Sampling rates
: Storage contents: rated measurement voltage, comparator
Memory
minimum limit, test mode, beep upon judgment, test time,
delay time, resistance range, sampling rate
Memory capacity: ten sets of settings
(supports saving and loading)
: Selectable between continuous and fail-stop
Test modes
Function
: After starting measurement, measurement is completed after
0.35s max. wait time + set time (when Delay time is OFF)
: 0.5 to 99s
Setting range
[ Delay Time Timer Function] (ON/OFF selectable)
Function
: After measurement starts, the test time timer and comparator
functions do not start operating until a wait time of max.
0.35 s plus the set delay time has elapsed.
: 0.1 to 99s
[ Comparator Functions]
Setting range
Minimum Limit
Setting
[ Response Times]
: Set by key entry; select from following table; set by
RS-232C; can be set to desired resistance
Measurement speed
Minimum settable limit (MΩ)
Rated measurement voltage
Resistance range, test voltage
25V/50V
0.1/0.2/0.5/1/2/3/4/5/10/20/30/50/100/200
100V/250V/500V
0.1/0.2/0.5/1/2/3/4/5/10/20/30/50/100/200/
300/500/1000/2000
1000V
2 or 20 MΩ
Manual ranging
200 MΩ
Manual ranging
1/2/3/4/5/10/20/30/50/100/200/300/500/1000/2000
: Pass = measured value ≥ minimum limit, Fail = measured
value < minimum limit
Judgment processing : Beep sound, PASS/FAIL indicator, output to external I/O
Beep sound selection : "Ring on Pass", "Ring on Fail", or disable beeper
: "- - - -" indication
OFF setting time
All Voltages
less than 0.5 s
25, 50 V
less than 0.6 s
100 to 1000 V less than 0.5 s
2000 or 4000 MΩ
Manual ranging
Judgments
FAST
SLOW
less than 1.5 s
100 to 1000 V less than 0.7 s
25, 50 V
Auto ranging
(including range selection)
less than 2 s
less than 5 s
100 to 1000 V less than 2.5 s
less than 6 s
General Specifications
Display type
Operating temperature
and humidity ranges
Storage temperature
and humidity ranges
Guaranteed accuracy
temp. & humidity
Suitable environment
Power supply
: LED
: 0 to 40˚C, max. 80% rh (no condensation)
: -10 to 50˚C, max. 90% rh (no condensation)
Input overload
protection
Dimensions
Mass
Conforming standards
: 23 ± 5˚C, max. 80% rh (no condensation)
: Indoors, max. 2000 m altitude
: 100 to 240 VAC
Power line frequency : 50 or 60 Hz
: 15 VA
Maximum rated
power
Withstand voltage : 1.35 kV AC for one minute between power lines and
outer case, external connectors, grounding and measurement
terminals (collectively);
1.1 kV DC for one minute between measurement terminal
and ground terminals;
500 V AC for one minute between EXT I/O terminals
and power line, ground and measurement terminals
(collectively);
Accessories
: 1100 V DC between HIGH and LOW on measurement
terminal (positive only).
: Approx. 215W × 61H × 213D mm (not including projections)
: Approx. 1.1 kg.
: EMC
: EN61326-1:1997+A1:1998
Safety
: EN61010-1:1993+A1:1995
Power
: Contamination 2
Overvoltage category II
(expected transient overvoltage 2500 V)
Measurement : Contamination 2
Overvoltage category I
(expected transient overvoltage 330 V)
: Power Cord (1)
[Not including INT DCV and INT GND terminals in the EXT I/O
terminals.]
3154 DIGITAL MΩ HiTESTER
Note: Test leads are not included: optional leads are required for use.
■ Options
9185
9294
9299
9094
9637
9638
9442
1196
TEST LEADS
TEST PROBE
SWITCHED PROBE
OUTPUT CORD (Jack to banana plug)
RS-232C CABLE (9pin-9pin/Cross/1.8 m)
RS-232C CABLE (9pin-25pin/Cross/1.8 m)
PRINTER
RECORDING PAPER (25m, 10 rolls)
9185
9294
9299
Lead length
approx. 85 cm
Lead length
approx. 1.2 m
Lead length
approx. 80 cm
9443-02 AC ADAPTER (for printer, EU)
9443-03 AC ADAPTER (for printer, America)
9444 CONNECTION CABLE (for printer)