Download Data Sheet

DATASHEET
Rev. E
WMMPU12 SERIES
90~264VAC Input Voltage Range
Medical Approvals
12 Watt, Single Outputs
US, UK, and EU Plugs
Class II AC/DC Wall Mount Power Supplies
US Type
UK Type
EU Type
MEDICAL
FEATURES
 Class II
 Over Voltage, Over Load, and Over Temperature Protection
 RoHS Compliant
 US, UK, and EU Types Available
 Energy Star 2.0, Efficiency Level V Compliant
 US Plug Medical Approvals: UL/cUL 60601-1: 3rd Edition
 Single Outputs
 UK and EU Plug Medical Approvals: TUV/T-mark
 Wide Input Voltage Range: 90~264VAC, 47~63Hz
EN60601-1: 3rd Edition
 Optional Output Connectors
DESCRIPTION
The WMMPU12 series of Class II medical AC/DC wall mount power supplies provides up to 12 Watts of continuous output
power. This series consists of single output models from 5VDC to 27VDC with a 90~264VAC input voltage range. These
supplies also have over voltage, over load, and over temperature protection. This series has UL/cUL 60601-1: 3rd Edition
(US plug type only) and TUV/T-mark EN60601-1: 3rd Edition (UK and EU plug types only) medical approvals. All units
are RoHS and Energy Star 2.0, Efficiency Level V compliant. These units also meet new CE requirements and are well
suited for use in blood pressure measurement equipment, frequency therapy devices, and dental curing light applications.
The WMMPU12 series has three types of plugs available: US (A Type), UK (B Type), and EU (C Type). Optional output
connectors are also available for this series. Please call factory for ordering details.
11/03/2014
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 1 of 5
WMMPU12 Series
Medical Approvals
US, UK, and EU Plugs
12 Watt, Single Outputs
Class II AC/DC Wall Mount Power Supplies
Rev. E
TECHNICAL SPECIFICATIONS: WMMPU12 Series
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
INPUT SPECIFICATIONS
Operating Voltage Range
TEST CONDITIONS
Operating Input Voltage Range
Safety Approvals Input Voltage Range
Nom
90
100
47
Input Frequency
Low Line
Input Current
High Line
Low Line
Inrush Current
High Line
No Load Power Consumption
Min
Io = Full Load, Vin = 115VAC
Io = Full Load, Vin = 230VAC
Io = Full Load, 25°C, Cool Start, Vin = 115VAC
Io = Full Load, 25°C, Cool Start, Vin = 230VAC
No Load, Vin = 240VAC
0.25
0.17
14
28
Max
Unit
264
240
63
0.35
0.22
25
50
0.3
VAC
Hz
A
A
A
A
W
OUTPUT SPECIFICATIONS
Output Voltage
Load Regulation
Line Regulation
Output Power
Output Current Range
Ripple & Noise (peak to peak)
Transient Response Time
Hold-Up Time
Start-Up Time
Temperature Coefficient
PROTECTION
Over Voltage Protection
Over Current Protection
Over Temperature Protection (See Note 1)
GENERAL SPECIFICATIONS
Efficiency
Dielectric Withstanding Voltage
Safety Ground Leakage Current
Vin = 230VAC
Io = Full Load
Vin = 90 to 264VAC
0
Full Load, Vin = 90VAC
Io = Full Load to Half Load, Vin = 100VAC
Io = Full Load, Vin = 110VAC
Io = Full Load, Vin = 100VAC
Verified by design/characteristics of integrated controller
Io = Full Load, Vin = 230VAC
Primary to Secondary
Io = Full Load, Vin = 240VAC
12
0.3
-0.04
See Table
3
5
0.5
1
12
See Table
0.5
1
4
16
1
2
+0.04
112
110
-25
73.3
5656
77.7
%
%
W
%
ms
ms
s
%/°C
132
150
130
%
%
°C
85
%
VDC
mA
0.1
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
MTBF
PHYSICAL SPECIFICATIONS
Weight
Derate linearly from 100% Load at 50°C to 50% load at 70°C
Operating Temp. at 25°C, calculated per MIL-HDBK-217F
US Plug (A Type)
UK Plug (B Type)
EU Plug (C Type)
50
70
85
95
75
100,000 hours
°C
°C
%
%
4.6oz (130g)
5.3oz (150g)
4.9oz (140g)
2.52 x 2.17 x 2.10 inches
(64.0 x 55.0 x 53.3 mm)
2.52 x 2.17 x 2.37 inches
(64.0 x 55.0 x 60.2 mm)
2.52 x 2.17 x 2.70 inches
(64.0 x 55.0 x 68.6 mm)
US Plug (A Type)
Dimensions (L x W x H)
0
-40
0
0
UK Plug (B Type)
EU Plug (C Type)
SAFETY & EMI
Safety Approvals
EMI Requirements for CISPR-11
EMI Requirements for CRF 47, Part-18
EMI Requirements for EN55022
11/03/2014
US Plug (A Type)
UK & EU Plugs (B & C Types)
US Plug (A Type) Only
US Plug (A Type) Only
UK & EU Plugs (B & C Types) Only
Vin = 120VAC, 60Hz
Vin = 120VAC, 60Hz
Vin = 230VAC, 50Hz
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
UL/cUL 60601-1:3rd Edition
TUV/T-mark EN60601-1:3rd Edition
B
Class
B
Class
B
Class
Page 2 of 5
WMMPU12 Series
Medical Approvals
US, UK, and EU Plugs
12 Watt, Single Outputs
Class II AC/DC Wall Mount Power Supplies
Rev. E
MODEL SELECTION TABLE
Input Voltage Range
Output Voltage
Output Current
Total Regulation (2)
Output Power
WMMPU12A-102
90 ~ 264 VAC
5 ~ 6 VDC
2.00 ~ 1.66 A
5%
10W
WMMPU12A-103
90 ~ 264 VAC
6 ~ 8 VDC
2.00 ~ 1.50 A
5%
12W
WMMPU12A-104
90 ~ 264 VAC
8 ~ 11 VDC
1.50 ~ 1.09 A
5%
12W
WMMPU12A-105
90 ~ 264 VAC
11 ~ 13 VDC
1.09 ~ 0.92 A
5%
12W
WMMPU12A-106
90 ~ 264 VAC
13 ~ 16 VDC
0.92 ~ 0.75 A
5%
12W
WMMPU12A-107
90 ~ 264 VAC
16 ~ 21 VDC
0.75 ~ 0.57 A
3%
12W
WMMPU12A-108
90 ~ 264 VAC
21 ~ 27 VDC
0.57 ~ 0.45 A
3%
12W
WMMPU12B-102
90 ~ 264 VAC
5 ~ 6 VDC
2.00 ~ 1.66 A
5%
10W
WMMPU12B-103
90 ~ 264 VAC
6 ~ 8 VDC
2.00 ~ 1.50 A
5%
12W
WMMPU12B-104
90 ~ 264 VAC
8 ~ 11 VDC
1.50 ~ 1.09 A
5%
12W
WMMPU12B-105
90 ~ 264 VAC
11 ~ 13 VDC
1.09 ~ 0.92 A
5%
12W
WMMPU12B-106
90 ~ 264 VAC
13 ~ 16 VDC
0.92 ~ 0.75 A
5%
12W
WMMPU12B-107
90 ~ 264 VAC
16 ~ 21 VDC
0.75 ~ 0.57 A
3%
12W
WMMPU12B-108
90 ~ 264 VAC
21 ~ 27 VDC
0.57 ~ 0.45 A
3%
12W
WMMPU12C-102
90 ~ 264 VAC
5 ~ 6 VDC
2.00 ~ 1.66 A
5%
10W
WMMPU12C-103
90 ~ 264 VAC
6 ~ 8 VDC
2.00 ~ 1.50 A
5%
12W
WMMPU12C-104
90 ~ 264 VAC
8 ~ 11 VDC
1.50 ~ 1.09 A
5%
12W
WMMPU12C-105
90 ~ 264 VAC
11 ~ 13 VDC
1.09 ~ 0.92 A
5%
12W
WMMPU12C-106
90 ~ 264 VAC
13 ~ 16 VDC
0.92 ~ 0.75 A
5%
12W
WMMPU12C-107
90 ~ 264 VAC
16 ~ 21 VDC
0.75 ~ 0.57 A
3%
12W
WMMPU12C-108
90 ~ 264 VAC
21 ~ 27 VDC
0.57 ~ 0.45 A
3%
12W
Model Number
Plug Type
US
UK
EU
NOTES
1. Over Temperature Protection: Thermal shutdown by junction temperature controller; the adapter will automatically recover when the
fault condition is removed.
2. Models WMMPU12-102~103 needs to use AWG#18/6FT output cable in order to meet the total regulation specified.
Models WMMPU12-104~108 needs to use AWG#20/6FT output cable in order to meet the total regulation specified.
The regulation and efficiency will change if a different output cable is used.
3. Optional output connectors are available. Please call factory for ordering details.
Output Power (%)
DERATING CURVE
Ambient Temperature (°C)
11/03/2014
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 3 of 5
WMMPU12 Series
Medical Approvals
US, UK, and EU Plugs
12 Watt, Single Outputs
Class II AC/DC Wall Mount Power Supplies
Rev. E
MECHANICAL DRAWING
US TYPE (SUFFIX “A”)
Unit: inches (mm)
2.10±0.08 (53.3±2.0)
1.77±0.04 (45.0±1.0)
0.25 (6.3)
0.31 (8.0)
1.77±0.04 (45.0±1.0)
2.52±0.04 (64.0±1.0)
2.17±0.04 (55.0±1.0)
POWER
1.22±0.04
(31.1±1.0)
LED (OPTIONAL)
0.19±0.04
(4.7±1.0)
1.81±0.04 (46.0±1.0)
UK TYPE (SUFFIX “B”)
Unit: inches (mm)
2.37±0.08 (60.2±2.0)
1.89±0.04 (48.0±1.0)
2.52±0.04 (64.0±1.0)
1.85±0.04 (47.0±1.0)
2.17±0.04 (55.0±1.0)
POWER
1.22±0.04
(31.1±1.0)
LED (OPTIONAL)
0.25±0.04
(6.4±1.0)
1.81±0.04 (46.0±1.0)
EU TYPE (SUFFIX “C”)
Unit: inches (mm)
2.70±0.08 (68.6±2.0)
2.52±0.04 (64.0±1.0)
2.17±0.04 (55.0±1.0)
POWER
LED (OPTIONAL)
11/03/2014
1.22±0.04
(31.1±1.0)
0.73±0.04
(18.5±1.0)
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
1.81±0.04 (46.0±1.0)
Page 4 of 5
Rev. E
WMMPU12 Series
Medical Approvals
US, UK, and EU Plugs
12 Watt, Single Outputs
Class II AC/DC Wall Mount Power Supplies
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development
engineers will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008
certification is just one example of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working
closely with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to
provide ongoing support for your project above and beyond the design and production phases. Give us a call today to discuss
your future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
11/03/2014
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
37 Industrial Drive
Exeter, NH 03833
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 5 of 5