ISL 0501

ISL 05x Inductive Conductivity Transmitter
Wetted parts in acid-proof, stainless steel and PEEK
Compact, food compatible, hygienic design
3A approved / FDA and EHEDG compliant
Process temperature -20...130°C
4 configurable measuring ranges
Adjustable, active temperature compensation
ISL052
Insensitive to polarisation, adhesion and solids
LCD display for conductivity and temperature
4...20 mA output for conductivity and temperature
ISL051
Description
The ISL05x is a sensor for inductive measurement of conductivity. The
compact design in all stainless steel enables installation in pipes from
DN40 and upwards.
Precise, configurable temperature compensation and remote setting
of the four pre-configured measuring ranges make the ISL ideal for a
wide range of conductivity measurements.
The integrated display for mS/cm and °C offers the user instant local
supervision, which is an advantage e.g. in manually operated cleaning
systems.
www.baumerprocess.com
Design and specifications subject to change without notice
A high operating temperature limit is an advantage in SIP systems.
The optimised flow geometry and the fast response time makes the
ISL05x particularly suitable in applications for separation of medias
and measurements of cleaning agents in CIP equipment.
The accuracy is excellent even at very low conductivity and flow rates.
The comprehensive range of accessories makes hygienic installation
easy (Refer to separate data sheet).
Data Sheet ISL-1
Page 1
Technical Data
Sensor
Electrical data
Inductive
2 toroidal-core transformers
Power supply
18...36 Vdc; max. 180 mA
Process connection
G1 hygienic
Response time
T90 < 3 sec.
Insulating material
PEEK
LCDisplay
All output data also valid for the display
Electrical connection
Conductivity output
Cable gland M16
Plast
Measuring ranges
4 ranges, remote selectable
Plug M12
Nickel-plated brass
Ranges
0.5...999 mS/cm, configurable
Mechanical data
Temp. compensation
0...5%/K, configurable per range
Housing
Stainless Steel, W1.4301/AISI 304
Temperature drift
< 110 ppm/°C
Process connection
Stainless Steel, W1.4404/AISI 316 L
Accuracy
±1% of the selected range
Protection class
IP67
Repeatability
0.2 % f.s.
Media pressure
Max. 10 bar
Output
Vibrations
IEC 68-2-6, GL test2
4...20 mA; max. 500 Ohm
galv. isolated from power supply
Approval
3A
Overrange
21.6mA
Adapters
Refer to “Accessories” data sheet
Temperature output
EMC data
Accuracy
< ±0.2°C (20...50°C)
< ±1.5°C (-20...140°C)
0.1°C
Immunity
EN 61326
Resolution
Emission
EN 61326
Repeatability
0.2 % f.s.
Temperature drift
< 150 ppm/°C
Output
4...20 mA; max. 500 Ohm
galv. isolated from power supply
Under/overrange
2.4/21.6mA
Operational conditions
Process temperature
-20...130°C (140°C <1 hour)
Amb. temperature
-20...60°C
Relative humidity
<100%, condensing
Disposal of product and packing
According to national laws or by returning to Baumer
Functional Principle
A
AC
Proc.
D
D
1
2
1
2
A
3
4...20 mA
outputs
Inductive conductivity measurement is based on the principle of a
transformer. The primary side of the transformer is controlled by an AC
voltage generator.
The liquid flowing through the channel bore (3) in the measuring head
forms a conductor loop, which links between the primary side of the
transformer (1) and the secondary side of the transformer (2).
The output current is proportional with the conductivity of the media.
Signal conditioning, amplification and conversion provides a 4...20 mA
signal output from the galvanically isolated D/A converter.
The fast-response temperature sensor in the tip (4) compensates
for the temperature in the liquid resulting in maximum accuracy and
reliability.
4
Ordering Details - ISL 05x
ISL 05x x
Type (Excl. welding part or adapter)
Standard sensor shaft (40 mm)
Extended sensor shaft (87 mm)
Standard sensor shaft (40 mm)
Approval
6´ digit
0
1
2
3A
Gland
Cable gland, M16
Plug, M12
7´ digit
1
2
3.1.b material certificate, type number 5509-227
www.baumerprocess.com
Design and specifications subject to change without notice
Data Sheet ISL-1
Page 2
Electrical Installation
1
+
2
-
3
+
4
-
5
+
6
7
-
Range
8
R2
R1
Remote range control
See table
R1
R2
Volt
Volt
1
0/open
0/open
2
24
0
3
0
24
4
24
24
18...36 Vdc
4...20 mA
4...20 mA
18...36Vdc (+)
mS/cm (-)
R1
°C (-)
4
1
2
3
1
4
2
3
°C
mS/cm
mS/cm (+)
18...36Vdc (-)
1: Brown
2: White
3: Blue
4: Black
Jog
shuttle
range
Resolution
for display
mS/cm
mS/cm
0...0.5
0.001
0...1
0.001
0...2
0.01
0...3
0.01
0...5
0.01
0...10
0.1
0...20
0.1
0...30
0.1
0...50
0.1
0....100
1.0
0...200
1.0
0...300
1.0
0...500
1.0
0...999
1.0
°C (+)
R2
Dimensional Drawings
LCDisplay
[mm]
ø89
999.9 mS/cm
139.9 °C
Max. indication
98
Min. indication
AF36
Jog shuttle
PUSH: SW version
TURN: Select menu
PUSH: Config menu
TURN: Select option
Left: Decrease
Right: Increase
G1
Refer to installation manual
M16
35
0.001 mS/cm
000.1 °C
M12plug
ø6 Channel boring
L
26
Temperature sensor tip
ø25.5
Insert length L = 40: Standard sensor shaft
Insert length L = 87: Extended sensor shaft
www.baumerprocess.com
Design and specifications subject to change without notice
Data Sheet ISL-1
Page 3
3A Approval
The ISL 052x is approved by 3A providing it is mounted in a 3A
approved counter part and installed according to the guidelines given
in the installation manual.
The 3A approved products fulfill the FDA demands and follow the
EHEDG guidelines regarding design, materials and finishing.
Refer to the 3A marked counter parts in the data sheet “Accessories”.
Angle > 135 deg.
PEEK natura
UK/2007-11-26 This data sheet may only be reproduced in full.
R > 6.35 mm
www.baumerprocess.com
Design and specifications subject to change without notice
Data Sheet ISL-1
Page 4