Datasheet

LAMP
LONG LIFE XENON LAMP SERIES
75 W: L10725/-01/-02, L10726/-01/-02
150 W: L11033, L11034
OVERVIEW
Our long life xenon lamp series now have a drastically extended service life1
compared to our previous lamps, thanks to the use of an innovative new electrode. This significant increase in service life helps reduce time-consuming
maintenance tasks such as lamp replacement and lamp position alignment.
Other benefits from using the long life xenon lamp include saving natural
resources and a smaller burden on the environment.
1Guaranteed service life (75 W: twice that of our conventional models, 150 W: about 1.7 times
that of our conventional models)
75 W Xenon Lamp
APPLICATIONS
FEATURES
●Semiconductor Inspection Equipment Light Source
●Scanner Light Sources
●Microscope Light Sources
●Spectrophotometer Light Sources
●High Performance Liquid Chromatograph (HPLC)
Light Sources
●Long Life 75 W Xenon Lamp Guaranteed Life: 2000 h
Average Life : 3000 h
150 w Xenon Lamp Guaranteed Life: 3000 h
Average Life : 4000 h
●High Stability Fluctuation (p-p) : 0.2 % Typ. (1.0 % Max.)
Drift
: ±0.5 %/h Typ.
●Point Light Source
LIGHT OUTPUT INTENSITY AND OPERATING TIME
110
DETECTOR: S1226
TLSXB0180EB
RELATIVE LIGHT OUTPUT INTENSITY (%)
RELATIVE LIGHT OUTPUT INTENSITY (%)
●75 W Xenon Lamp
100
90
80
70
60
50
40
30
LONG LIFE LAMP
CONVENTIONAL TYPE
20
10
0
0
500
1000
1500
2000
2500
3000
110
WAVELENGTH: 365 nm
TLSXB0179EB
100
90
80
70
60
50
40
30
LONG LIFE LAMP
CONVENTIONAL TYPE
20
10
0
0
500
1000
110
DETECTOR: S1226
TLS B0004EA
100
90
80
70
60
50
40
30
LONG LIFE LAMP
CONVENTIONAL TYPE
20
10
0
0
500
1000
1500
2000
2500
OPERATING TIME (h)
3000
3500
4000
RELATIVE LIGHT OUTPUT INTENSITY (%)
RELATIVE LIGHT OUTPUT INTENSITY (%)
●150 W Xenon Lamp
1500
2000
2500
3000
OPERATING TIME (h)
OPERATING TIME (h)
110
WAVELENGTH: 365 nm
TLS B0003EA
100
90
80
70
60
50
40
30
LONG LIFE LAMP
CONVENTIONAL TYPE
20
10
0
0
500
1000
1500
2000
2500
OPERATING TIME (h)
3000
3500
4000
SPECIFICATIONS
Parameter
Window Material
(Transmission wavelength)
Lamp Rating
Arc Length
Lamp Current
Lamp Voltage
Light Output Fluctuation (p-p) (Max.)
Stability
Drift (Typ.)
Operating Guaranteed Life 2
Average Life
Orientation
Cooling Method
Weight
Long Life Lamp 75 W
Long Life Lamp 150 W
L10725/-01/-02
L10726/-01/-02
L11033
L11034
Fused Silica
Ozone-free Silica
Fused Silica
Ozone-free Silica
(185 nm to 2000nm)
(240 nm to 2000nm)
(185 nm to 2000nm)
(240 nm to 2000nm)
Approx. 75 W
Approx. 150 W
1.0 mm ± 0.1 mm
2.0 mm ± 0.1 mm
5.7 A ± 0.3 A
8.5 A ± 0.5 A
Approx. 13.5 V
Approx. 17 V
1.0 %
±0.5 %/h
2000 h
3000 h
3000 h
4000 h
Vertical ±15 degree or Horizontal ±15 degree
Convection Cooling
Approx. 15 g
Approx. 45 g
2The life end is defined as the time when the light output intensity falls to 50 % of its initial value or when the output fluctuation (p-p) exceeds 1.0 %
RADIANT INTENSITY (µW·cm-2 ·nm-1) at 50 cm
SPECTRAL DISTRIBUTION (TYPICAL DATA: 75 W Xenon Lamp)
102
TLSXB0181EA
101
100
10-1
10-2
FUSED SILICA BULB
OZONE-FREE SILICA BULB
10-3
200
400
600
800
1000
1200
1400
1600
1800
L11033
L11034
2000
WAVELENGTH (nm)
15
DIMENSIONAL OUTLINES (Unit: mm)
M4
UP
L10725
L10726
L10725-01
L10726-01
2
L10725-02
L10726-02
12
M2.3
2
METAL BASE
12
58.0 ± 0.50
95 MAX.
9
13
11
7.5
36.85 ± 0.50
90 MAX.
82
13
2
36.85 ± 0.50
82
9
125
14
14
82
14
90 MAX.
9
9
15
9
150 MAX.
UP
UP
36.85 ± 0.50
13
13
13
20
UP
METAL BASE
2
TLSXA0122EA
METAL BASE
METAL BASE
TLSXA0123EA
M2.3
TLSXA0007EE
M4
TLS A0004EA
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are
subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2011 Hamamatsu Photonics K.K.
HAMAMATSU PHOTONICS K.K.
www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Electron Tube Division
314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: [email protected]
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: [email protected]
France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: [email protected]
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: [email protected]
North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: [email protected]
TLS 1008E01
Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: [email protected]
JUL. 2011 IP