XFPM Data sheet Pre-amplified/5V Excitation/Gauge ■Features On-chip amplification and temperature compensations Pre-calibration of offset voltage and span Dual-in-line package (DIP) ■Ordering Information XFPM-025KP G R Terminal leadsdirection Model No Mark: Rated Pressure KP:kPa MP:MPa Pressure type R : G:Gauge(Positive pressure) GV:Gauge Pressure(Negative pressure) GW:Gauge Pressure(Positive/Negative pressure) Weight (grams) RoHS compliance 1.2 Measurable pressure range(kPa) -100 to 100 0 to -100 0 to 25 0 to 50 0 to 100 0 to 200 0 to 1000 Part number XFPM-100KPGW XFPM-100KPGV XFPM-025KPG XFPM-050KPG XFPM-100KPG XFPM-200KPG XFPM-001MPG XFPM-100KPGWR XFPM-100KPGVR XFPM-025KPGR XFPM-050KPGR XFPM-100KPGR XFPM-200KPGR XFPM-001MPGR ■Specifications 100KPGW Model Recommended operating conditions Pressure type Rated pressure Measurable pressure range Temperature range Pressure media Supply voltage(constant) +/-100 -100 to100 100KPGV -100 0 to -100 025KPG 050KPG 100KPG 200KPG Gauge pressure 50 100 200 0 to 50 0 to 100 0 to 200 0 to 85 Non-corrosive gases only (No liquid) 5+/-0.25 25 0 to 25 Absolute maximum rating Maximum load pressure Twice of rated pressure 8 -40 to 125 -40 to 125 30 to 80 (Non dew condition) Electrical characteristics (Excitation voltage Vcc=5.0V constant ,ambient temperature Ta=25deg.C) Power consumption 10mA max. Output impedance 10Ω max. Source current 0.2mA max. Sink current 2mA max. Response time 2 (for the reference) Output span voltage 4.5 Offset voltage *1 *2 *3 0.2+/-0.1125 (at 0 kPa ) Output voltage at full scale *1 *4 *5 4.7+/-0.1125 (at rated pressure) Accuracy *1 +/-2.5 Maximum excitation voltage Operating temperature Storage temperature Operating humidity *1 Excluding input voltage error. *2 0.2+/-0.1125V(at -100kPa ) *4 4.7+/-0.1125V(at +100kPa) 0-85deg.C *3 0.2+/-0.1125V(at 0kPa) *5 4.7+/-0.1125V(at -100kPa) 1 001MPG 1000 0 to 1000 1.5 times of rating pressure Unit kPa kPa deg.C VDC VDC deg.C deg.C %RH mA Ω mA mA msec. V V V %FS/0-85deg.C XFPM Data sheet Pre-amplified/5V Excitation/Gauge ■Outline dimensions■ 2 XFPM Data sheet Pre-amplified/5V Excitation/Gauge ■Connection diagram■ View from Lavel Marking ■Transfer Function■ Vout = Vs×( P×α+β) ± (Pressure Error×Temperature Error Multiplier×α×Vs) Vs=Vcc=5.0V P=Input pressure(kPa) Model XFPM-100KPGW XFPM-100KPGV XFPM-025KPG XFPM-050KPG XFPM-100KPG XFPM-200KPG XFPM-001MPG XFPM-100KPGWR XFPM-100KPGVR XFPM-025KPGR XFPM-050KPGR XFPM-100KPGR XFPM-200KPGR XFPM-001MPGR Measurable pressure range(kPa) -100 to 100 0 to -100 0 to 25 0 to 50 0 to 100 0 to 200 0 to 1000 Temperature Error Multiplier 3 α β Pressure Error (kPa) 0.0045 -0.009 0.036 0.018 0.009 0.0045 0.0009 0.49 0.04 0.04 0.04 0.04 0.04 0.04 5 2.5 0.625 1.25 2.5 5 25 XFPM Data sheet Pre-amplified/5V Excitation/Gauge ■Output characteristics■ ↑Output voltage(mV) XFPM-100KPGV XFPM-025KPG XFPM-050KPG XFPM-100KPG XFPM-200KPG XFPM-001MPG XFPM-100KPGVR XFPM-025KPGR XFPM-050KPGR XFPM-100KPGR XFPM-200KPGR XFPM-001MPGR Output voltage at full scale 4.7V Output span voltage Offset voltage 0.2V Rated pressure 0kPa →Pressure(kPa) Atomospheric pressure XFPM-100KPGW ↑Output voltage(mV) XFPM-100KPGWR Output voltage at full scale 4.7V Output span voltage Offset voltage 0.2V -100kPa 0kPa Atomospheric pressure Note ; +100kPa →Pressure(kPa) Please read instruction “Notes” before using the sensor. Fujikura reserves the right to change specifications without notice. Please keep the sensors sealed using static shielding bags on storage.The pins of the sensor are plated by Ag. If the sensors expose to an atmosphere, the pins will be black by sulfuration. Please set Zero-calibration function up your products. The offset voltage may be shifted some mechanical stress such as mounting, installation and etc. over longtime using. If you have any questions regarding technical issues or specifications, please contact us. Fujikura Ltd. Sensor Department 5-1 Kiba 1-chome, Koto-ku, Tokyo 135-8512, Japan Phone +81-(0)3-5606-1072 E-mail : [email protected] 4