Features
• 2-channel display box for strain rings DSRC
• Measurement value of each sensor,
Average and peak value of both sensors
• Bending measurement by individual S/G display
• A/C or Battery operation
• Display in µe, kN, t
• 2 analog outputs
• Reset with keypad or InspectMaster
Order Code
Electrical Data
Connection
2 channels for
2 x 1/4 S/G bridge (350 Ω)
Display
Sensor A or B
peak value, average value
Delivery Contents
Measuring range ±1000 µε (calibrated)
Resolution
1 µε
• Display box
Characteristic
curve deviation
< 0,25% FS
Reset/operate offset
< ±0,1% FS
Bridge completion
resistors Rc
350 Ω
Reset
Zeroing is performed by pressing the reset button or by the software InspectMaster
Measuring rate
InspectMaster/
analog output 250/sec
Analog output
±1 V calibrated at ±1000 µε
Display refresh rate
2/sec
Battery
Maintenance-free Li-Ion battery
USB connection
USB 2.0, type B
DDBF 2-SC
• Power adapter (100 - 240 VAC)
• Software InspectMaster
• USB connecting cable
• Ground connection cable with clip
Mechanical Data
Sensor connection
4 pin Binder (series 712)
Enclosure
Aluminum, lacquered
Environmental Conditions
Operating temp. range +5...+50 °C
Storage temperature
-20...+60 °C
Protection class IP 40
7.4
www.baumer.com
3
DDBF 2-SC
110
35
Analog
OUT1
Analog
OUT2
CH A
CH D
Error
DDBF
RESET disable
Analog OUT2
F1
Analog OUT1
CH C
F3
Esc
ON/OFF
CH D
Enter
Prog.
35
CH B
F2
CH A CH B
100
CH A
212
View Connector Side
110
Analog
OUT2
CH C
Displa ybox
Dimensions (mm)
Analog
OUT1
CH B
RESET
110
View Bottom Side, under the top cover
Displa ybox
Error
85
RESET disable
212
DDBF
Analog
OUT1
Analog
OUT2
CH A
USB-Connection
F2
CH C
CH D
F3
Esc
Enter
Prog.
Display batteryError
chargeDDBF
status RESET
RESET
Displa ybox
disable
212
ON/OFF
100
F1
CH B
Connection for power adapter, 15-33 VDC
85
F1
Electrical Connections
3
1
4
F3
Esc
S/G Bridge
Pin
Signal
1
S/G 1
EXC,+
2
S/G 1
SIG.-
3
S/G 2
EXC.+
4
S/G 2
SIG.-
DMS 1 / S/G 1
100
ON/OFF
Sensor connection
2
F2
Enter
Prog.
DDBF
RESET
EXC.+
1
R
2
7
85
R
DMS 2 / S/G 2
R
EXC.+
3
R
4
R
R
Pro Kanal / Per channel
Analysis Software InspectMaster
Functions:
-Display in µe, N, kN, t
-Display of deviation of tie bar load distribution in %
-Cycle function with auto reset mode
-Display of graphs
-Saving of measured data
-Export function (Text file may be exported and
processed in Excel)
Requirements:
-PC with Windows, 2000, NT, XP
minimum 500 MHz and USB interface required
www.baumer.com
7.5