SITRANS FC MC2 DN50SITRANS FC MC2 DN65SITRANS FC MC2 DN80SITRANS FC 2100 DI25SITRANS FC 2100 DI40

© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor SITRANS FCS200
■ Overview
■ Function
The flow measuring principle is based on the Coriolis effect. See
"System information SITRANS F C".
■ Integration
The complete flowmeter consists of the sensor (SITRANS
FCS200) and a transmitter SITRANS F C MASS 6000 or SIFLOW
FC070. All communication options are available for MASS 6000.
The sensor is shipped with a SENSORPROM memory unit containing all information about calibration data, device identity and
factory pre-programming of transmitter settings.
3
Installation guidelines
Siemens Flow Instruments recommends installing the sensor in
one of the following ways:
SITRANS FCS200 (DN10, DN 15 and DN 25) is a Coriolis sensor
specialized for accurate mass flow measurement of gases.
The sensor offers superior performance in terms of flow accuracy and turn down ratio. The ultra compact sensor design
makes installation, replacement and commissioning very
straight forward and easy.
■ Benefits
• High accuracy gas measurement
• Approved for use in hazardous area
• DN 10 and DN 15 is custody transfer approved, according to
OIML R 139 (Compressed gaseous fuel measuring systems
for vehicles). For custody transfer applications SIFLOW
FC070 Ex CT must be used.
• Self-draining in vertical orientation
• Pt1000 temperature measurement for optimum accuracy
• SENSORPROM enabling true "plug & play"
• Rigid enclosure design reducing influence from pipeline vibration and thermal stress
• High-pressure measurement up to 350 bar (5076 psi)
• Ultra compact sensor design with space-saving split flow
Vertical orientation with an upwards flow
■ Application
SITRANS FCS200 is designed for measurement of gases and is
suitable for use in the oil and gas industry:
• Filling of gas bottles
• CNG dispensers
• Metering of general gas applications
Horizontal installation, tubes up
■ Design
SITRANS FCS200 is available in DN 10, DN 15 and DN 25.
The sensor consists of 2 parallel measuring pipes, welded directly onto a flow splitter at each end of the sensor to eliminate a
direct coupling to the process connectors and significantly reduce effects from external vibrations. The flow-splitters are
welded directly onto a rigid sensor housing which acts as a mechanical low pass filter.
The SITRANS FCS200 DN 10 and DN 15 wetted parts material is
Hastelloy C22, and the DN 25 wetted parts material is AISI
316Ti/1.4571. The enclosure is made of stainless steel
AISI 316L/1.4404 with a grade of encapsulation of IP67.
The two black rupture discs are designed to protect the
enclosure from overpressure.
3/200
Siemens FI 01 · 2015 US Edition
Horizontal installation, tubes sideways
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor SITRANS FCS200
■ Technical specifications
Sensor size
■ Characteristic curves
DN 10
DN 15
DN 25
DN 10
Mass Flow
Accuracy [% of rate]
 0.5
Repeatability [% of rate]
 0.25
Max. zero point error
[kg/h (lb/h)]
0.25 (0.55)
Measuring range
[kg/min (lb/min)]
0 … 42
(0 … 92.6)
1.2 (2.65)
bar
2.5
3.0 (6.6)
2.0
0 … 200
0 … 500
(0 … 440.9) (0 … 1102.3)
Process temperature
-40 ... +125 °C (-40 ... +257 °F)
Ambient temperature
-40 ... +60 °C (-40 ... +140 °F)
Temperature error
Pressure [bar (psi)]
1.5
0.5 °C (0.9 °F)
350 (5076)
Enclosure grade
350 (5076)
3
1.0
0.5
214 (3104)
0
IP66/IP67 (EN 60529)
0
200
0
500 1000 1500 2000 2500 3000 3500 [kg/h]
400
600
800 1000 1200 1400 [kg/h]
Material
Measuring pipe
Hastelloy
C22/2.4602
Hastelloy
C22/2.4602
Splitter
Hastelloy
C22/2.4602
Stainless
Stainless
steel AISI
steel AISI
316L/1.4571 316L/1.4571
Enclosure and connection
(flanges)
Connection thread
Stainless
steel AISI
316L/1.4571
DN 15
bar
1.2
1.0
Stainless steel
0.8
¼" NPT
½" NPT
½" VCO
½" NPT
¾" NPT
1" NPT
¾" VCO
1" NPT
1½" NPT
1" VCO
0.6
0.4
0.2
Ex approval
• ATEX
II 1/2 G Ex ia IIC T5/T4 Ga/Gb
• IECEx
Ex ia IIC T5/T4 Ga/Gb
• FM
Class I, Div 1, Groups A, B, C and D
Weight approx.
2.8 kg
(6.2 lb)
6.0 kg
(13.2 lb)
11 kg
(24.2 lb)
Approvals Custody
transfer
DN 10/DN 15
0
DN 25
bar
3.5
PTB Germany approval nr: 5.4.11/11.22
OIML R 139 - Compressed gaseous fuel
measuring systems for vehicles
3.0
2.5
2.0
1.5
1.0
0.5
0
0
5000
10000
15000
20000
25000 [kg/h]
The pressure drop as a function of capacity for CNG with a pressure of 200 bar (2900 psi) and an ambient temperature of 20 °C
(68 °F).
Siemens FI 01 · 2015 US Edition
3/201
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor SITRANS FCS200
Operating instructions for SITRANS FCS200
Article No.
Selection and Ordering data
SITRANS F C Flow sensors
SITRANS FCS200 sensor, without heating
jacket
7 ME 4 5 0 0 7 77 7 7 - 7 7 7 7
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
3
Sensor size and material
DN 10, Hastelloy C22/2.4602
DN 15, Hastelloy C22/2.4602
DN 25, Stainless steel AISI 316Ti/1.4571
2D
2E
1F
Pressure
PN 214 (DN 25)
PN 350 (DN 10 and DN 15)
K
N
½“/VCO
¾“/VCO
1“/VCO
71
72
73
¼“/NPT pipe thread
½“/NPT pipe thread
81
82
¾“/NPT pipe thread
1“/NPT pipe thread
1½“/NPT pipe thread
83
84
85
Configuration
Standard
1
Transmitter
A
None
Cable
No cable
A
Calibration
Standard calibration
1
Order code
Selection and Ordering data
Additional information
Please add “-Z“ to Article No. and specify Order
code(s) and plain text.
Pressure testing certificate PED: 97/23/EC
C11
Material certificate EN 10204-3.1
C12
NDT-Penetrant inspection report ISO 3452
C13
Factory certificate according to EN 10204 2.2
C14
Factory certificate according to EN 10204 2.1
C15
Tag name plate, stainless steel
Y17
Accessories
Article No.
Description
3/202
5 m (16.4 ft)
FDK:083H3015
10 m (32.8 ft) FDK:083H3016
25 m (82 ft)
FDK:083H3017
50 m (164 ft) FDK:083H3018
75 m (246 ft) FDK:083H3054
150 m (492 ft) FDK:083H3055
Siemens FI 01 · 2015 US Edition
Article No.
• English
A5E02508199
• German
A5E03082574
• Spanish
A5E03082587
• French
A5E03082581
• Italian
A5E03504933
Spare parts
Process connection/flange
Cable with multiple plug
Standard blue cable between
SIFLOW FC070/MASS 6000 and
FCS200,
5 x 2 x 0.34 mm2 twisted and screened
in pairs.
Temperature range -20 °C ... +110 °C
(-4 °F ... +230 °F)
Description
Description
Article No.
Multiple plug for cable mounting
FDK:083H5056
2 kB SENSORPROM unit
(Sensor Serial No. and Article No. must
be specified by ordering)
FDK:083H4410
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor SITRANS FCS200
■ Dimensional drawings
SITRANS FCS200, DN 10 ... DN 15
3
SITRANS FCS200, DN 10 ... DN 15, dimensions in mm (inch)
Position
DN 10 with NPT DN 10 with VCO DN 15
connectors
connectors
mm (inch)
mm (inch)
mm (inch)
(1)
350 (13.78)
330 (12.99)
450 (17.72)
(2)
72 (2.84)
72 (2.84)
72 (2.84)
(3)
100 (3.94)
100 (3.94)
148 (5.83)
(4)
204 (8.03)
204 (8.03)
253 (9.96)
(5)
40 (1.57)
40 (1.57)
48 (1.89)
Siemens FI 01 · 2015 US Edition
3/203
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor SITRANS FCS200
SITRANS FCS200, DN 25
DN 25 - NPT
3
DN 25 - VCO
SITRANS FCS200, DN 25, dimensions in mm (inch)
Position
DN 25
DN 25
with NPT connection with VCO connection
mm (inch)
mm (inch)
(1)
520 (20.47)
550 (21.65)
(2)
72 (2.84)
72 (2.84)
(3)
200 (7.87)
200 (7.87)
(4)
357 (14.77)
357 (14.77)
(5)
74 (2.91)
74 (2.91)
(6)
80 (3.15)
80 (3.15)
(7)
125 (4.92)
125 (4.92)
(8)
32 (1.26)
32 (1.26)
3/204
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 1.5
■ Overview
■ Application
In many industries such as the food and beverage or pharamceutical industry, accurate recipe control means everything. The
MASS 2100 DI 1.5 has demonstrated superiour performance in
numerous applications and field trails relating to accuracy and
turn-down ratio. It is today the preferred meter for research and
development and mini-plant applications for liquid or gas measurement, where measuring small quantities is important.
The main applications for the MASS 2100 DI 1.5 sensor can be
found in:
MASS 2100 DI 1.5 is suitable for low flow measurement applications of a variety of liquids and gases.
The sensor offers superior performance in terms of flow accuracy, turn-down ratio and density accuracy. The ease of installation through a "plug & play" mechanical and electrical interface
ensures optimum performance and operation.
The sensor delivers true multi-parameter measurements i.e.:
Mass flow, volume flow, density, temperature and fraction.
■ Benefits
• High accuracy better than 0.1 % of mass flow rate
• Large dynamic turn-down ratio better than 500:1, from 30 kg/h
to a few g/h
• Densitometer performance available through a density accuracy better than 0.001 g/cm3 with a repeatability better than
0.0002 g/cm3.
• Single continuous tube design, with no internal welds, reductions or flow splitters offers optimal hygiene, safety and CIP
cleanability for food and beverage and pharmaceutical applications.
• Market’s biggest wall thickness, ensuring optimal life-time and
corrosion resistance and high-pressure durability
• Balanced pipe design with little mechanical energy-loss, ensures optimal performance and stability under non-ideal and
unstable process conditions (pressure, temperature, densitychanges etc.).
• 4-wire Pt1000 temperature measurement ensures optimum
accuracy on mass flow, density and fraction flow
• Multi-plug electrical connector and SENSORPROM enables
true "plug & play". Installation and commissioning in less than
10 minutes
• Intrinsically safe Ex ia design as standard
• Sensor pipe available in high-quality stainless steel AISI 316L/
1.4435 or Hastelloy C22/2.4602 offering optimum corrosion resistance
• Dual-drive pick-up and driver construction facilitate ultra lowweight pipe construction giving the markets’ smallest and
most stable zero point.
• Rugged and space-saving sensor design in stainless steel
matching all environments
• High-pressure program as standard
• The sensor calibration factor is also valid for gas measurement.
Chemical industry
Liquid and gas measurement
within Miniplant and R & D, dosing
of additives and catalysts
Cosmetic industry
Dosing of essence and fragrances
Pharmaceutical industry
High-speed dosing and coating of
pills, filling of ampuls/injectors
Food and beverage industry
Dosing of flavourings, colours and
additives, density measurement,
inline measurement of liquid or
gaseous CO2
Automotive industry
Fuel injection nozzle and pump
testing, filling of AC units, engine
consumption, paint robots, ABS
test-beds
■ Design
The MASS 2100 sensor consists of a single bent tube in a double
omega pipe configuration, welded directly to the process connectors at each end.
The sensor is available in 2 material configurations, AISI 316L/
1.4404 or Hastelloy C22/2.4602 with ¼” NPT or ¼” ISO process
connections.
The enclosure is made in stainless steel AISI 316L/1.4404 with
a grade of encapsulation of IP65/NEMA 4.
The sensor is available in either a standard version with a maximum liquid temperature of 125 °C (257 °F) or a high-temperature version, with raised electrical connector for 180 °C (356 °F).
The sensor can be installed in horizontal or vertical position. The
enclosed single quick release clamp fitting which, along with its
compact design and single multi-plug electrical connector, will
keep installation costs and time to a minimum as shown below.
Siemens FI 01 · 2015 US Edition
3/205
3
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 1.5
■ Function
■ Technical specifications
The measuring principle is based on the Coriolis effect. See
“System information SITRANS F C Coriolis mass flowmeters”.
■ Integration
The sensor can be connected to all MASS 6000 transmitters for
remote installation only.
3
All sensors are delivered with a SENSORPROM containing all information about calibration data, identity and factory pre-programming of transmitter settings
Installation guidelines MASS 2100 DI 1.5 (1/16“)
Installation of MASS 2100 sensor
• The optimal installation is horizontal.
If vertical mounting is necessary, upward flow is recommended to facilitate the removal of air bubbles. To remove the
air from the sensor the flow speed in the sensor must be at
least 1 m/s.
If there are solid particles in the liquid, especially in connection with low flow, it is recommended that the sensor be
mounted horizontally with inlet flange uppermost so that particles are more easily flushed out. To ensure that the sensor
does not become partially empty, there must be sufficient
counter-pressure on the unit min. 0.2 bar (2.9 psi).
• Mount the sensor on a vibration-free wall or steel frame.
• Locate the sensor low in the system in order to avoid an underpressure in the sensor separating air/gas in the liquid.
• Ensure that the sensor is not emptied of liquid (during normal
operation) otherwise incorrect measurement will occur.
Inside pipe diameter (sensor consists of one continuous pipe)
1.5 mm (0.06”)
Pipe wall thickness
0.25 mm (0.010”)
Mass flow measuring range
0 ... 30 kg/h (0 ... 66 lb/h)
Density
0 ... 2.9 g/cm3 (0 ... 0.10 lb/inch3)
Fraction e.g.
0 ... 100 °Brix
Temperature
Standard
-50 ... +125 °C (-58 ... +257 °F)
High-temperature version
-50 ... +180 °C (-58 ... +356 °F)
Liquid pressure measuring pipe1)
Stainless steel
230 bar (3336 psi) at 20 °C
(68 °F)
Hastelloy C22/2.4602
365 bar (5294 psi) at 20 °C
(68 °F)
Materials
Measuring pipe and connection
Stainless steel AISI 316L/1.4435
Hastelloy C22/2.4602
IP65 and stainless steel
AISI316L/1.4404
Enclosure and enclosure
material2)
Connection thread
ISO 228/1
G¼” male
ANSI/ASME B1.20.1
¼” NPT male
Cable connection
Multiple plug connection to sensor 5 x 2 x 0.35 mm2 twisted and
screened in pairs, ext. Ø 12 mm
Ex-version
II 1G Eex ia IIC T3-T6,
DEMKO 03
ATEX 135252X
c-UL-us
Ex ia IIC T3-T6
UL WYMG.E232147
Weight approx.
2.6 kg (5.73 lb)
Horizontal
1)
According to DIN 2413, DIN 17457
2)
Housing is not rated for pressure containment.
For accuracy specifications see “System information
SITRANS F C”.
Pressure drop
Viscosity [cSt]
Liquid and gas application
Δp [bar]
100
1000 500
200
100
Vertical
50
10
20
10
5
1
1.0
0.1
Liquid application (left), gas application (right)
0.01
0.1
1
10
70
Mass flow [kg/h]
MASS 2100 DI 1.5 (1/16“), pressure drop for density = 1000 kg/m3
3/206
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 1.5
Selection and Ordering data
Article No. Ord. code
Selection and Ordering data
SITRANS F C Flow sensors
7 ME 4 1 0 0 -
Additional information
MASS 2100 DI 1.5 (1/16“) sensor
7 77 7 7 - 7 7 7 7 7 7 7
Please add “-Z“ to Article No. and specify Order
code(s) and plain text.
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Order code
Pressure testing certificate PED: 97/23/EC
C11
Diameter
Material certificate EN 10204-3.1
C12
Stainless steel AISI 316L/1.4435
DI 1.5, max. 125 °C (257 °F)
DI 1.5, max. 180 °C (356 °F)
Welding certificate NDT-Penetrant: ISO 3452
C13
1A
1B
Factory certificate according to EN 10204 2.2
C14
Factory certificate according to EN 10204 2.1
C15
2A
2B
Tag name plate, stainless steel
Y17
Tag name plate, plastic
Y18
Customer-specific transmitter setup
Y20
Customer-specified, matched pair (5 x 2)
Y60
Customer-specified calibration (5 x 2)
Y61
Customer-specified, matched pair (10 x 1)
Y62
Customer-specified calibration (10 x 1)
Y63
Cleaned for oil and grease
Y80
Special version
Y99
Hastelloy C22/2.4602
DI 1.5, max. 125 °C (257 °F)
DI 1.5, max. 180 °C (356 °F)
Pressure
PN 100
PN 230 (AISI 316L/1.4404)
PN 365 (C22/2.4602)
D
L
P
Process connection/flange
Pipe thread
G ¼“ male
¼“ NPT male
10
11
Operating instructions for SITRANS F C MASS 2100 DI 1.5
Configuration
Standard
Density
Brix/Plato
Fraction (specification required)
1
2
3
9
N0Y
Transmitter compact mounted on sensor
A
MASS 6000, Ex d, stainless steel enclosure,
1 current, 1 freq./pulse and 1 relay output,
24 V AC/DC with Ex de [ia/ib] T3 -T6 Exapproval.
B
MASS 6000, IP67, Polyamide enclosure,
cable glands M20, 1 current, 1 freq./pulse
and 1 relay output, 24 V AC/DC.
C
MASS 6000, IP67, Polyamide enclosure,
cable glands M20, 1 current, 1 freq./pulse
and 1 relay output, 115/230 V AC 50/60 Hz
D
MASS 6000, IP67, Polyamide enclosure,
cable glands ½" NPT, 1 current, 1 freq./pulse
and 1 relay output, 24 V AC/DC
E
MASS 6000, IP67, Polyamide enclosure, cable
glands ½" NPT, 1 current, 1 freq./pulse and 1
relay output, 115/230 V AC 50/60 Hz, ½" NPT
F
25 m (82 ft) cable
50 m (164 ft) cable
75 m (246 ft) cable
150 m (492 ft) cable
Description
Article No.
• English
A5E03089952
This device is shipped with a Quick Start guide and a CD containing
further SITRANS F literature.
All literature is also available for free at:
No transmitter, sensor and adapter only
Cable
No cable
5 m (16.4 ft) cable
10 m (32.8 ft) cable
3
http://www.siemens.com/flowdocumentation
Accessories
Article No.
Description
Cable with multiple plug
Standard blue cable between
MASS 6000 and MASS 2100,
5 x 2 x 0.34 mm2 twisted and
screened in pairs. Temperature
range -20 °C ... +110 °C
(-4 °F ... +230 °F)
A
B
C
D
• 5 m (16.4 ft)
FDK:083H3015
• 10 m (32.8 ft)
FDK:083H3016
• 25 m (82 ft)
FDK:083H3017
• 50 m (164 ft)
FDK:083H3018
• 75 m (246 ft)
FDK:083H3054
• 150 m (492 ft)
FDK:083H3055
Spare parts
E
F
Description
Article No.
Multiple plug for cable
mounting
FDK:083H5056
2 kB SENSORPROM unit
FDK:083H4410
(Sensor Serial No. and Article No.
must be specified by ordering)
G
Calibration
Standard calibration 3 flow x 2 points
Standard calibration matched pair 3 flow x
2 points
Accredited calibration matched pair 5 flow x
2 points (DANAK)
1
2
Extended calibration customer-specified
select Y60, Y61, Y62 or Y63 (see additional
information)
8
Bracket
A5E02590427
3
Siemens FI 01 · 2015 US Edition
3/207
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 1.5
■ Dimensional drawings
MASS 2100 DI 1.5 (1/16“)
Ø57 (2.4)
64,5 (2.54)
4 (0.16)
120 (4.72)
NV21
G 1/4”
Ø18,8 (0.74)
Ø24 (0.94)
1/4” NPT
Ø18,8 (0.74)
Ø24 (0.94)
1/4” NPT
ANSI/ASME
B1.20.1-1983
20 (0.79)
NV21
15 (0.6)
12 (0.47)
15 (0.6)
12 (0.47)
86 (3.39)
152 (5.98)
10 (0.39)
78 (3.07)
20 (0.79)
3
Ø129 (5.08)
46 (1.81)
25 (0.98)
Ø 9 (0.35)
G 1/4”
ISO 228/1
100 (3.94)
Dimensions in mm (inch)
MASS 2100 DI 1.5 High-temperature version to 180 °C (356 °F)
Ø57 (2.4)
192,6 (7.58)
1/4” NPT
Ø18,8 (0.74)
Ø24 (0.94)
1/4” NPT
ANSI/ASME
B1.20.1-1983
4 (0.16)
120 (4.72)
NV21
G 1/4”
Ø18,8 (0.74)
Ø24 (0.94)
G 1/4”
ISO 228/1
Ø 9 (0.35)
20 (0.79)
NV21
15 (0.6)
12 (0.47)
15 (0.6)
12 (0.47)
86 (3.39)
152 (5.98)
100 (3.94)
Dimensions in mm (inch)
3/208
Siemens FI 01 · 2015 US Edition
10 (0.39)
64,5 (2.54)
78 (3.07)
20 (0.79)
Ø129 (5.08)
46 (1.81)
25 (0.98)
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor SITRANS FC300
■ Overview
• High-pressure program as standard
• The sensor calibration factor is also valid for gas measurement.
■ Application
The industry today has an increasing demand for mass flowmeters with a reduced physical size without loss of performance.
The meters must be suitable for installation in traditional process
industry environment as well as OEM equipment for instance
within automotive or appliance industry. Independent of industry
application the meter must deliver accurate and reliable measurements. The new and versatile design of the FC300 offers this
flexibility.
The main applications for the SITRANS FC300 DN 4 can be found in:
SITRANS FC300 is a compact Coriolis mass sensor suitable for
flow measurement of a variety of liquids and gases.
The sensor offers superior performance in terms of flow accuracy, turn-down ratio and density accuracy. The ease of installation through a „plug & play“ interface ensures optimum performance and operation.
A new designed encapsulation in stainless steel with a surprisingly low weight of only 3.5 kg (7.7 lb), ensures a rigid and robust
sensor performance for a wide range of applications.
■ Benefits
• High accuracy better than 0.1 % of mass flow rate
• Large dynamic turn-down ratio better than 500:1
• Densitometer performance available through a density accuracy as follows:
- For 316L/1.4404 version better than 0.0015 g/cm3
(0.000036 lb/inch3) with repeatability better than
0.0002 g/cm3 (0.0000072 lb/inch3)
- For C22/2.4602 version better than 0.0025 g/cm3
(0.000090 lb/inch3) with repeatability better than
0.001 g/cm3 (0.000036 lb/inch3)
• One tube without internal welds, reductions or flow splitters offers optimal hygiene, safety and CIP cleanability for
food and beverage and pharmaceutical applications
• Larger wall thickness, ensures optimal life-time and corrosion
resistance and high-pressure durability
• Balanced pipe design with little mechanical energy loss, ensures optimal performance and stability under non-ideal and
unstable process conditions (pressure, temperature, densitychanges etc.).
• 4-wire Pt1000 temperature measurement ensures optimum
accuracy on mass flow, density and fraction flow
• Multi-plug electrical connector and SENSORPROM enable
true „plug & play“. Installation and commissioning in less than
10 minutes.
• Intrinsically safe Ex design ia IIC as standard
• Sensor pipe available in high-quality stainless steel
AISI 316L/1.4435 or Hastelloy C22/2.4602 offering optimum
corrosion resistance.
• Rugged and space-saving sensor design in stainless steel
matching all applications.
Chemical industry
Liquid and gas measurement in
normal as well as corrosive environments
Cosmetic industry
Dosing of essence and fragrances
Pharmaceutical industry
High-speed dosing and coating
of pills, filling of ampuls/injectors
Food and beverage industry
Filling, dosing of flavorings, colors
and additives, inline density measurement
Measurement and dosing of
liquid or gaseous CO2
Automotive industry
Fuel injection nozzle and pump
testing, filling of AC units, engine
consumption, paint robots, ABS
test-beds
■ Design
The FC300 sensor consists of a single tube bent in double
omega pipe geometry, welded directly to the process connectors at each end. The sensor is available in 2 material configurations, AISI 316L/1.4404 or Hastelloy C22/2.4602 with ¼”-NPT or
G¼”-ISO process connections.
The enclosure is made of stainless steel AISI 316L/1.4409 with a
grade of encapsulation of IP67/NEMA 4. The enclosure has a
very robust design and with an overall size of
130 x 200 x 60 mm (5.12“ x 7.87“ x 2.36“) the sensor is very compact and requires only little installation space.
The sensor can be delivered in a standard version with a maximum liquid temperature of 115 °C (239 °F) or a high-temperature version, with raised electrical connector for 180 °C (356 °F).
The sensor can be installed in horizontal or vertical position. The
sensor can be mounted directly on any given plane surface or if
desired with the enclosed quick release clamp fitting which,
along with its compact design and multi-plug electrical connector, will keep installation costs and time to a minimum.
■ Function
The measuring principle is based on the Coriolis effect. See
“System information SITRANS F C Coriolis mass flowmeters”.
■ Integration
The sensor can be connected to all MASS 6000 and
SIFLOW FC070 (standard and Ex types) transmitters for remote
installation only.
All sensors are delivered with a SENSORPROM containing all information about calibration data, identity and factory pre-programming of transmitter settings
Siemens FI 01 · 2015 US Edition
3/209
3
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor SITRANS FC300
Installation guidelines for SITRANS FC300 sensor
Horizontal installation as shown in figure A is recommended with
gas or liquid applications.
This installation is also recommended when the flow velocity is
low (< 1 m/s) or the liquid contains solid particles or air bubbles.
3
■ Technical specifications
Sensor size
DN 4 (1/6“)
Mass flow
Measuring range
0 ... 350 kg/h (0 ... 772 lb/h)
Vertical installation as shown in figure B can be used for liquid or
gas applications.
Accuracy, mass flow
0.1 % of rate
Repeatability
0.05 % of rate
For liquid applications upwards flow is recommended to facilitate the removal of air bubbles and to avoid partly emptying of
the sensor.
For gas applications we recommend to place the flow inlet on the
sensor high and the outlet low to remove impurities and oil films.
• To ensure that the sensor does not become partly empty, there
must be a sufficient counter-pressure on the unit min. 0.2 bar
(2.9 psi).
• Mount the sensor on a vibration-free and plane wall or steel
frame.
• Locate the sensor low in the system in order to avoid underpressure in the sensor separating air/gas in the liquid.
• Ensure that the sensor is not emptied of liquid (during normal
operation) otherwise incorrect measurement will occur.
Max. zero point error
0.010 kg/h (0.022 lb/h)
Horizontal mounting (recommended) (fig. A)
Density
Density range
0 ... 2.9 g/cm3
(0 ... 0.105 lb/inch3)
Density error
• Stainless steel
0.007 g/cm3
(0.00025 lb/inch3)
• Hastelloy C22/2.4602
0.0025 g/cm3
(0.00009 lb/inch3)
Repeatability error
0.0002 g/cm3
(0.0000072 lb/inch3)
Temperature
Standard
-40 ... +115 °C (-40 ... +239 °F)
High-temperature version
-40 ... +180 °C (-40 ... +356 °F)
Temperature error
0.5 °C (0.9 °F)
Brix
Measuring range
0 ... 100 °Brix
Brix error
0.3 °Brix
Inside pipe diameter
Stainless steel version
3.5 mm (0.14“)
Hastelloy version
3.0 mm (0.12“)
Pipe wall thickness
Stainless steel version
0.25 mm (0.0098“)
Liquid or gas (low to high flow)
Hastelloy version
0.5 mm (0.0196“)
Vertical mounting (fig. B)
Liquid pressure measuring pipe1)
Stainless steel
130 bar (1885 psi) at 20 °C
(68 °F)
Hastelloy C22/2.4602
410 bar (5945 psi) at 20 °C
(68 °F)
Materials
Stainless steel AISI 316L/1.4435
Measuring pipe and connection
Hastelloy C22/2.4602
Enclosure2)
Material
Stainless steel AISI 316L/1.4404
Enclosure grade
IP67/NEMA4
Connection thread
ISO 228/1
G¼” male
ANSI/ASME B1.20.1
¼” NPT male
Ex approval
Ex ia IIC T3-T6
05ATEX138072X
c-UL-us Class 1 Div. 1,
Gr. A, B, C, D
Liquid or gas (medium to high flow)
3/210
Siemens FI 01 · 2015 US Edition
Weight
3.5 kg (7.7 lb)
Dimensions
135 x 205 x 58 mm
(5.31“ x 8.07“ x 2.28“)
1)
According to DIN 2413, DIN 17457
2)
Housing is not rated for pressure containment.
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor SITRANS FC300
Selection and Ordering data
Article No. Order code
Selection and Ordering data
SITRANS F C Flow sensors
7 ME 4 4 0 0 -
SITRANS FC300 DN 4 (1/6“) sensor
7 77 7 7 - 7 7 7 7 7 7 7
Additional information
Please add “-Z“ to Article No. and specify Order code(s)
and plain text.
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Pipe material and temperature
Stainless steel AISI 316L/1.4435
115 °C (239 °F)
180 °C (356 °F)
Hastelloy C22/2.4602
115 °C (239 °F)
180 °C (356 °F)
Pressure
PN 100
PN 130 (316L/C22)
PN 410 (C22)
Pressure testing certificate PED: 97/23/EC
C11
Material certificate EN 10204-3.1
C12
Welding certificate NDT-Penetrant: ISO 3452
C13
1G
1H
Factory certificate according to EN 10204 2.2
C14
Factory certificate according to EN 10204 2.1
C15
2G
2H
Tag name plate, stainless steel
Y17
Tag name plate, plastic
Y18
Customer-specific transmitter setup
Y20
Customer-specified, matched pair (5 x 2)
Y60
Customer-specified calibration (5 x 2)
Y61
Customer-specified, matched pair (10 x 1)
Y62
Customer-specified calibration (10 x 1)
Y63
Cleaned for oil and grease
Y80
Special version
Y99
D
G
Q
Process connection
Pipe thread
G ¼“ male
¼“ NPT male
Configuration
Standard
Density
Brix/Plato
Fraction (specification required)
10
11
1
2
3
9
N0Y
No transmitter, sensor and adapter only
A
MASS 6000, Ex d, stainless steel enclosure,
1 current, 1 freq./pulse and 1 relay output,
24 V AC/DC with Ex de [ia/ib] T3 -T6 Exapproval
B
MASS 6000, IP67, Polyamide enclosure, cable
glands M20, 1 current, 1 freq./pulse and 1
relay output, 24 V AC/DC
C
MASS 6000, IP67, Polyamide enclosure, cable
glands M20, 1 current, 1 freq./pulse and 1
relay output, 115/230 V AC 50/60 Hz
D
MASS 6000, IP67, Polyamide enclosure, cable
glands ½" NPT, 1 current, 1 freq./pulse and 1
relay output, 24 V AC/DC
E
MASS 6000, IP67, Polyamide enclosure, cable
glands ½" NPT, 1 current, 1 freq./pulse and 1
relay output, 115/230 V AC 50/60 Hz, ½" NPT
F
25 m (82 ft) cable
50 m (164 ft) cable
75 m (246 ft) cable
150 m (492 ft) cable
Calibration
Standard calibration 3 flow x 2 points
Standard calibration matched pair 3 flow x
2 points
Accredited calibration matched pair 5 flow x 2
points (DANAK)
Extended calibration customer-specified
select Y60, Y61, Y62 or Y63 (see additional
information)
3
Operating instructions for SITRANS F C FC300
Transmitter compact mounted on sensor
Cable
No cable
5 m (16.4 ft) cable
10 m (32.8 ft) cable
Order code
Description
Article No.
• English
A5E00698213
• German
A5E00728101
• Spanish
A5E00746629
• French
A5E00746625
This device is shipped with a Quick Start guide and a CD containing
further SITRANS F literature.
All literature is also available for free at:
http://www.siemens.com/flowdocumentation
Accessories
Article No.
Description
Cable with multiple plug
Standard blue cable between
MASS 6000 and MASS 2100,
5 x 2 x 0.34 mm2 twisted and
screened in pairs. Temperature
range -20 °C ... +110 °C
(-4 °F ... +230 °F)
A
B
C
D
E
F
• 5 m (16.4 ft)
FDK:083H3015
• 10 m (32.8 ft)
FDK:083H3016
• 25 m (82 ft)
FDK:083H3017
• 50 m (164 ft)
FDK:083H3018
• 75 m (246 ft)
FDK:083H3054
• 150 m (492 ft)
FDK:083H3055
Spare parts
G
1
2
3
Description
Article No.
Multiple plug for cable
mounting
FDK:083H5056
2 kB SENSORPROM unit
FDK:083H4410
(Sensor Serial No. and Article No.
must be specified by ordering)
8
Mounting bracket in AISI 304
A5E02590439
Demo suitcase
including MASS 6000,
FC300 (DN 4),
and HART module
A5E00789737
Siemens FI 01 · 2015 US Edition
3/211
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor SITRANS FC300
■ Characteristic curves
Pressure drop
Viscosity [cSt]
Δp [bar]
1000
10
500
200
100
50
20
10
3
5
2
1
1.0
0.1
3
10
100
500
Mass flow [kg/h]
Stainless steel 316L/1.4404
9LVFRVLW\>F6W@
P [bar]
10
1.0
0.1
3
10
Hastelloy C22/2.4602
3/212
Siemens FI 01 · 2015 US Edition
100
300 [kg/h]
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor SITRANS FC300
■ Dimensional drawings
1/4”-18 NPT
Ø18,8 (0.74)
SITRANS FC300 DN 4
15 (0.59)
12 (0.47)
3
38 (1.50)
15 (0.59)
12 (0.47)
G 1/4”
Ø18,8 (0.740)
1/4”-18 NPT (ANSI/ASME B1.20.1)
G 1/4” (ISO 228/1 )
M8
NV 5 (0.20)
190 (3.54)
14 ± 2 (0.55 ± 0.08)
121 (4.76)
77 (3.03)
19 (0.75)
205 (8.07)
20 (1.33)
NV 24 (0.95)
28,8 (1.134)
Ø27 (1.06)
18,8 (0.74)
110 (4.33)
M4
58 (2.28)
135 (5.31)
2 x M8
32 (1.26)
75 (2.95)
NV 6 (0.24)
SITRANS FC300, dimensions in mm (inch)
Siemens FI 01 · 2015 US Edition
3/213
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
■ Overview
■ Application
Coriolis mass flowmeters are suitable for measuring all liquids
and gases. The measurement is independent of changes in process conditions/parameters such as temperature, density, pressure, viscosity, conductivity and flow profile.
Due to this versatility the meter is easy to install and the Coriolis
flowmeter is recognized for its high accuracy in a wide turndown ratio which is a paramount in many applications.
The main applications of the Coriolis flowmeter can be found in all
industries, such as:
3
MASS 2100 DI 3 to DI 15 is suitable for accurate mass flow
measurement of a variety of liquids and gases.
The sensor offers superior performance in terms of flow accuracy, turn-down ratio and density accuracy. The ease of installation through a "plug & play" mechanical and electrical interface
ensures optimum performance and operation.
The sensor delivers true multi-parameter measurements i.e.:
Mass flow, volume flow, density, temperature and fraction.
Chemical and pharma
Detergents, bulk chemicals,
pharmaceuticals, acids, alkalis
Food and beverage
Dairy products, beer, wine,
soft-drinks, Brix/Plato,
fruit juices and pulps, bottling,
CO2 dosing, CIP-liquids
Automotive
Fuel injection nozzle and pump
testing, filling of AC units, engine
consumption, paint robots
Oil and gas
Filling of gas bottles, furnace control, test separators, LPG
Water and waste water
Dosing of chemicals for water
treatment
■ Benefits
• High accuracy better than 0.1 % of mass flow rate
• Large dynamic turn-down ratio better than 500:1
• Densitometer performance available through density accuracy (depending upon sensor size) ranging from 0.0005 to
0.0015 g/cm3 with a typical repeatability better than
0.0001 to 0.0002 g/cm3
• Single continuous tube design, with no internal welds, reductions or flow splitters offers optimal hygiene, safety and CIP
cleanability for food and beverage and pharmaceutical applications
• Markets’ thickest sensor walls ensure optimal life-time and
corrosion resistance and high-pressure durability
• Full bore design provides lower pressure loss due to same internal diameter throughout the entire sensor
• Balanced pipe design with little mechanical energy loss, ensures optimal performance and stability under non-ideal and
unstable process conditions (pressure, temperature, density
changes etc.)
• 4-wire Pt1000 temperature measurement ensures optimum
accuracy on mass flow, density and fraction flow
• Multi-plug electrical connector and SENSORPROM enables
true "plug & play". Installation and commissioning in less than
10 minutes
• Intrinsically safe Ex design ia IIC as standard, making service
in hazardous area possible without having to demount the
sensor if a compact Ex d transmitter needs service
• Sensor pipe available in high-quality stainless steel
AISI 316L/1.4435 or Hastelloy C22/2.4602 offering optimum
corrosion resistance
• Centre-block design decouples process noise from the environment such as vibrations, pulsations, pressure shocks etc.
making installation flexible and versatile
• Rugged and space-saving sensor design in stainless steel
matching all environments
• High-pressure program as standard
• The sensor calibration factor is also valid for gas measurement
• Uniform sensor interface matching all transmitter versions at
the same time whether it is compact IP67/NEMA 6, compact
Ex d or remote installation, one sensor fits all
3/214
Siemens FI 01 · 2015 US Edition
The wide varity of combinations and versions from the modular
system means that ideal adaptation is possible to each measuring task.
■ Design
The MASS 2100 sensor consists of a single bent tube in a double
bent pipe configuration, welded directly to the process connectors at each end.
The centre-block is brazed onto the sensor pipes from the outside acting as a mechanical low pass filter.
The sensor is available in 2 material configurations,
AISI 316L/1.4404 or Hastelloy C22/2.4602 with a wide variety of
process connections.
The enclosure is made in stainless steel AISI 316L/1.4404 with a
grade of encapsulation of IP67.
The sensor is as standard Ex ia approved, intrinsically safe.
The sensor can be installed in horizontal or vertical position. In
horizontal position the sensor is self draining.
Heating: All the sensors MASS 2100, DI 3 to DI 15, can optionally be equipped with a heating coil to avoid solidification of sensitive fluids during down-time or period between discontinuing
processes. This feature gives the user an alternative to the costly
electrical heating normally used, as it gives the freedom to
choose either hot water, superheated steam or hot oil, to maintain a constant temperature inside the sensor.
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
■ Function
The measuring principle is based on the Coriolis effect. See
“System information SITRANS F C Coriolis mass flowmeters”.
■ Integration
The sensor can be connected to all MASS 6000 transmitters for
compact and remote installation as well as SIFLOW FC070 standard and Ex type transmitters.
All sensors are delivered with a SENSORPROM containing all information about calibration data, identity and factory pre-programming of transmitter settings.
3
Installation guidelines MASS 2100 DI 3 ... DI 15 (1/8" ... ½")
Installation of sensor
In order to perform according to given specifications for flow and
density accuracy, the sensor must be installed using rigid
mounting brackets as shown in the installation examples.
If the liquid is volatile or contains solid particles, vertical mounting is not recommended.
Liquid
Gas
Horizontal
Vibration
Always locate the flowmeter as far away as possible from components that generate mechanical vibration in the piping.
Vertical
Cross talk
Cross talk between sensors mounted close to each other may
disturb the measurement. To avoid cross talk never mount more
than one meter on each frame and mount flexible hose connections between the sensors as shown.
Zero point adjustment
To facilitiate zero point adjustment a shut-off valve should always
be mounted in connection with the sensor as a proper zero point
setting is essential for a good accuarcy.
Siemens FI 01 · 2015 US Edition
3/215
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
■ Technical specifications
3
Versions (mm (inch))
DI 3 (1/8)
DI 6 (¼)
DI 15 (5/8)
DI 25 (1)
DI 40 (1½)
Inside pipe diameter (sen- mm (inch)
sor consists of one continuous pipe)
3.0 (0.12)
6.0 (0.24)
14.0 (0.55)
29.7 (1.17)
43.1 (1.70)
Pipe wall thickness
mm (inch)
0.5 (0.02)
1.0 (0.04)
1.0 (0.04)
2.0 (0.08)
2.6 (0.10)
Mass flow measuring
range
kg/h (lb/h)
0 ... 250 (0 ... 550)
0 ... 1000
(0 ... 2200)
0 ... 5600
(0 ... 12345)
0 ... 25000
(0 ... 55100)
0 ... 52000
(0 ... 114600)
Density
g/cm3
(lb/inch3)
Fraction e.g.
°Brix
0 ... 2.9 (0 ... 0.10)
0 ... 70 (applicable temperature range: 10 ... 99 °C (50 ... 210.2 °F))
Temperature
Standard
°C (°F)
-50 ... +180 °C (-58 ... +356 °F)
Liquid pressure measuring pipe1)
Stainless steel
bar (psi)
230 (3336)
265 (3844)
130 (1885)
110 (1595)
105 (1523)
Hastelloy C22/2.4602
bar (psi)
350 (5076)
410 (5946)
200 (2900)
185 (2683)
not available
Materials
Measuring pipe, flange and
thread connection
Stainless steel AISI 316L/1.4435
Hastelloy C22/2.4602
not available
IP67 (NEMA 4) and stainless steel AISI 316L/1.4404,
The housing is not rated for pressure containment
Enclosure and enclosure
material
Process connections2)
Flange
EN 1092-1, PN 40
DN 10
DN 15
DN 25
DN 40
ANSI B16.5, Class 150
½”
½”
1”
1½”
ANSI B16.5, Class 600
(Class 300)
½”
½”
1”
1½”
DIN 11851
DN 10
DN 15
DN 32
DN 40
ISO 2853/BS 4825 part 4
(SS3351)
25 mm
25 mm
38 mm
51 mm
25 mm
25 mm
38 mm
51 mm
Dairy screwed connection (PN 16/25/40)3)
Dairy clamp connection
(PN 16)3)
ISO 2852/BS 4825 part 3
(SMS3016)
Thread
ISO 228/1, PN 100
G¼” female
G¼” male
G½” male
G1“ male
G2“ male
ANSI/ASME B1.20.1,
PN 100
¼” NPT female
¼” NPT male
½” NPT male
1“ NPT male
2“ NPT male
Multiple plug connection to sensor 5 x 2 x 0.35 mm2 twisted and screened in pairs, ext. Ø 12 mm
Cable connection
Ex ia IIC T3-T6, DEMKO 03 ATEX 135252X
Ex-version
Weight approx.
kg (lb)
4 (8.8)
1)
Max. at 20 °C (68 °F), DIN 2413, DIN 17457
2)
Other connections to order, see “Selection and Ordering data”
3)
Material, AISI 316/1.4401 or corresponding
8 (17.6)
For accuracy specification see “System information SITRANS F C”.
3/216
Siemens FI 01 · 2015 US Edition
12 (26.5)
48 (105.8)
70 (154.5)
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
Pressure drop
D p [ bar]
Viscosity [cSt]
Δp [bar]
1000
10
500
200
100
50
20
Viscosity [cSt]
10
10
00
0
50
0
20
10
0
10
5
20
50
2
1
1
1.0
1.0
0.1
3
10
100
300
Mass flow [kg/h]
MASS 2100 DI 3 (1/8”), pressure drop for density = 1000 kg/m3
0.1
60
100
1000
MASS 2100 DI 15 (½”), pressure drop for density = 1000 kg/m3
Viscosity [cSt]
Δp [bar]
1000
10
500
6000
Mass flow [kg/h]
Viscosity [cSt]
200
100
Δp [bar]
1000
10
500
50
20
200
1
100
50
1.0
1.0
20
1
0.1
10
100
1000
Mass flow [kg/h]
MASS 2100 DI 6 (¼“), pressure drop for density = 1000 kg/m3
0.1
300
1000
10000
30000
Mass flow [kg/h]
MASS 2100 DI 25 (1”), pressure drop for density = 1000 kg/m3
Siemens FI 01 · 2015 US Edition
3/217
3
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
ANSI Flange Ratings per ASME B16.5, Group Material 3.8
Viscosity [cSt]
Δp [bar]
1000
500
200
1.0
CLASS600
100
98
90
80
70
60
50
20
CLASS300
50
49
1
40
30
20 CLASS150
0.1
500
1000
10000
50000
Mass flow [kg/h]
15
10
-20
0
30
80
130
180
Process temperature (°C)
MASS 2100 DI 40 (1½”), pressure drop for density = 1000 kg/m3
ASME flanges B16.5 Hastelloy C22/2.4602
Pressure/temperature curves
DIN 32676
ANSI Flange Ratings per ASME B16.5, Material Group 2.3
45
Process pressure (bar)
110
Process Pressure (bar)
100
90
80
CLASS600
70
60
35
30
25 PN25
25
60
50
15 PN16
16
40 CLASS300
10 PN10
10
30
5
30
-20
20
0
20
40
60
80
100 120 140 160 180
Process temperature (°C)
10 CLASS150
-20
40
20
0
30
80
130
12
180
DIN 32676 flanges stainless steel (PN 10 ... PN 25)
Process temperature (°C)
DIN 11851
ASME flanges B16.5 stainless steel
45
Process Pressure (bar)
3
100
Process Pressure (bar)
110
10
40 PN40 DN10 - DN40
40
35
30
25 PN25 DN50 - DN100
25
20
15 PN16 DN125 - DN150
16
10
5
-20
0
20
40
60
80
100 120 140 160 180
Process temperature (°C)
DIN 11581 flanges stainless steel (PN 25 ... PN 40)
3/218
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
ISO 228, Pipe thread
NPT - ANSI/ASME B1.20.1, Pipe thread
Metric flange ratings, EN 1092-1, Material group 13E0
360
100 PN100
90
80
82
70
60
PN63
Process Pressure (bar)
Process pressure (bar)
110
310
260
PN265
3
PN230
50
52
210
40 PN40
188
33
30
160
164
20
PN130
10
-20
0
30
80
130
110
180
Process temperature (°C)
92
PN100
71
60
EN 1092 flanges stainless steel (PN 40 ... PN 100)
-20
0
30
80
130
180
Process temperature (°C)
Flange EN 1092-1, UNS NO6022
ISO 228 and NPT pipe thread stainless steel (PN 100 ... PN 265)
100 PN100
ISO 228, Pipe thread
NPT - ANSI/ASME B1.20.1, Pipe thread UNS NO6022
90
420
80
79
70
60
PN63
50
50
40 PN40
30
Process Pressure (bar)
Process Pressure (bar)
110
PN410
370 PN365
PN350
320
322
270
287
275
31
20
10
-20
0
30
80
130
180
220
Process temperature (°C)
EN 1092 flanges Hastelloy C22/2.4602 (PN 40 ... PN 100)
PN200
PN185
170
157
145
120
-20
0
30
80
130
180
Process temperature (°C)
ISO 218 and NPT pipe thread stainless steel (PN 185 ... PN 410)
For further information on the PED standard and requirements,
see page 9/6.
Siemens FI 01 · 2015 US Edition
3/219
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
Selection and Ordering data
Article No. Ord. code
SITRANS F C sensors
Article No. Ord. code
SITRANS F C sensors
MASS 2100 without heating jacket
7 ME 4 1 0 0 -
MASS 2100 without heating jacket
7 ME 4 1 0 0 -
MASS 2100 heated, DN 15 connection
7 ME 4 2 0 0 -
MASS 2100 heated, DN 15 connection
7 ME 4 2 0 0 -
MASS 2100 heated, ½ inch, ANSI B16.5
connection
7 ME 4 2 1 0 -
MASS 2100 heated, ½ inch, ANSI B16.5
connection
7 ME 4 2 1 0 -
7 77 7 7 - 7 7 7 7 7 7 7
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
3
Selection and Ordering data
7777 7 - 7 7 7 7 7 7 7
Dairy screwed connection DIN 11851
Diameter
Stainless steel AISI 316L/1.4435
DI 3 (PN 100/PN 230)
1C
DI 6
DI 15
1D
1E
Hastelloy C22/2.4602
DI 3 (PN 100/PN 350)
DI 6
2C
2D
DN 10 (PN 40)
DN 15 (PN 40)
DN 25 (PN 40)
40
41
42
DN 32 (PN 40)
DN 40 (PN 25)
DN 50 (PN 25)
43
44
45
DN 65 (PN 25)
46
Dairy clamp connection ISO 2852 (DIN 32676)
Cone down the sensor in order to obtain selfdrainage with connectors ISO 2852
Pressure
PN 16 (DI 6, DI 15)
PN 25 (DI 6, DI 15)
PN 40 (DI 6, DI 15)
A
B
C
PN 100 (DI 3, DI 6, DI 15)
PN 130 (DI 15, ½“, AISI 316L/1.4404)
D
G
PN 200 (DI 15, ½“, Hastelloy C22/2.4602)
PN 230 (DI 3, ¼“, AISI 316L/1.4404)
PN 265 (DI 6, ¼“, AISI 316L/1.4404)
K
L
M
PN 350 (DI 3, ¼“, Hastelloy C22/2.4602)
PN 410 (DI 6, ¼“, Hastelloy C22/2.4602)
N
Q
Class 150 (DI 6, DI 15)
Class 600 (DI 6, DI 15)
R
S
25 mm (PN 16)
38 mm (PN 16)
51 mm (PN 16)
50
51
52
Dairy screwed connection ISO 2853
25 mm (PN 16)
38 mm (PN 16)
51 mm (PN 16)
60
61
62
Configuration/calibration type
Standard
Density
Brix/Plato
1
2
3
Fraction (specification required)
9
N0Y
Transmitter compact mounted on sensor
Process connection/flange
No transmitter, sensor and adapter only
A
G ¼“
¼“ NPT
G ½“
10
11
12
MASS 6000, Ex d, stainless steel enclosure,
1 current, 1 freq./pulse and 1 relay output,
24 V AC/DC with Ex de [ia/ib] T3 -T6 Exapproval
B
½“ NPT
G1
1“ NPT
13
14
15
MASS 6000, IP67, Polyamide enclosure,
cable glands M20, 1 current, 1 freq./pulse
and 1 relay output, 24 V AC/DC
C
G 2“
2“ NPT
16
17
MASS 6000, IP67, Polyamide enclosure,
cable glands M20, 1 current, 1 freq./pulse
and 1 relay output, 115/230 V AC 50/60 Hz
D
MASS 6000, IP67, Polyamide enclosure,
cable glands ½" NPT, 1 current, 1 freq./pulse
and 1 relay output, 24 V AC/DC
E
MASS 6000, IP67, Polyamide enclosure, cable
glands ½" NPT, 1 current, 1 freq./pulse and 1
relay output, 115/230 V AC 50/60 Hz
F
Pipe thread
Flange EN1092-1 Form B
DN 10 (PN 40/PN 100)
DN 15 (PN 40/PN 100)
DN 25 (PN 40/PN 100)
20
21
22
DN 40 (PN 40/PN 100)
DN 50 (PN 40/PN 100)
23
24
Flange ASME/ANSI B 16.5
½“ (class 150/class 600)
¾“ (class 150/class 600)
1“ (class 150/class 600)
30
31
32
1 ½“ (class 150/class 600)
2“ (class 150/class 600)
33
34
Cable
No cable
5 m (16.4 ft) cable
10 m (32.8 ft) cable
A
B
C
25 m (82 ft) cable
50 m (164 ft) cable
75 m (246 ft) cable
D
E
F
150 m (492 ft) cable
G
Calibration/verification
3/220
Siemens FI 01 · 2015 US Edition
Standard calibration 3 flow x 2 points
Stand. calibration matched pair 3 flow x 2 points
Accredited calibration matched pair 5 flow x
2 points (DANAK to ISO 17025)
1
2
3
Extended calibration customer-specified
select Y60, Y61, Y62 or Y63 (see additional
information)
8
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
■ Selection and Ordering data
Dairy MLFB example
7 ME 4 1 0 0 - 77777 7777
MASS 2100
1E
Sensor size DI 15,
AISI 316L/1.4435
Accessories
Description
Dimension
Article No.
Mating parts for hygienic fittings
DIN 11851
Includes:
• 2 unions
• 2 mating parts (for welding in)
• 2 EPDM gaskets
DN 10
FDK:085U1016
DN 15
FDK:085U1017
DN 25
FDK:085U1019
DN 32
FDK:085U1020
DN 40
FDK:085U1021
DN 50
FDK:085U1022
DN 65
FDK:085U1023
Mating parts for hygienic clamp
ISO 2852
Includes:
• 2 clamps
• 2 mating parts
• 2 EPDM gaskets
25 mm
FDK:085U1029
40 mm
FDK:085U1031
50 mm
FDK:085U1032
2 EPDM gaskets with collar for
mounting set DIN 11851
DN 10
FDK:085U1006
DN 15
FDK:085U1007
DN 25
FDK:085U1009
DN 32
FDK:085U1010
Addtional information
DN 40
FDK:085U1011
Please add “-Z“ to Article No. and specify Order
code(s) and plain text.
DN 50
FDK:085U1012
DN 65
FDK:085U1013
C
PN 40
41
DN 15 connector
1
Standard configuration/calibration
D
MASS 6000 IP67 compact mounted
A
No cable
1
Standard calibration, 3 flow x 2 points
Order code
Selection and Ordering data
Pressure testing certificate PED: 97/23/EC
C11
Material certificate EN 10204-3.1
C12
NDT- X-ray inspection report: EN 1435
DI3 sensor only: NDT-Penetrant inspection report
ISO 3452.
C13
Factory certificate according to EN 10204 2.2
C14
Factory certificate according to EN 10204 2.1
C15
Tag name plate, stainless steel
Y17
Tag name plate, plastic
Y18
Customer-specific transmitter setup
Y20
Customer-specified, matched pair (5 x 2)
Y60
Customer-specified calibration (5 x 2)
Y61
Customer-specified, matched pair (10 x 1)
Y62
Customer-specified calibration (10 x 1)
Y63
Cleaned for oil and grease
Y80
Special version
Y99
Operating instructions for
SITRANS F C MASS 2100 DI 3 to DI 40
Description
Article No.
5 m (16.4 ft)
FDK:083H3015
10 m (32.8 ft) FDK:083H3016
25 m (82 ft)
FDK:083H3017
50 m (164 ft)
FDK:083H3018
75 m (246 ft)
FDK:083H3054
150 m (492 ft) FDK:083H3055
Spare parts
Description
Article No.
Adapter for MASS 2100
FDK:083L8889
FDK:083H5056
Description
Article No.
Multiple plug for cable mounting
• English
A5E02896535
• German
A5E03073519
• Spanish
A5E03073549
2 kB SENSORPROM unit, including
programming
(Sensor Serial No. and Article No. must
be specified by ordering)
• French
A5E03073539
This device is shipped with a Quick Start guide and a CD containing
further SITRANS F literature.
Length
Cable with multiple plug
Standard blue cable between
MASS 6000 and MASS 2100,
5 x 2 x 0.34 mm2 twisted and
screened in pairs. Temperature range
-20 °C ... +110 °C (-4 °F ... +230 °F)
FDK:083H4410
All literature is also available for free at:
http://www.siemens.com/flowdocumentation
Siemens FI 01 · 2015 US Edition
3/221
3
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
■ Dimensional drawings
MASS 2100 sensor
155 (6.10)
130 (5.12)
L2
225 (8.86)
L2
88 (3.5)
D1
D4
H1
D2
77 (3.03)
B1
D3
284 (11.2)
3
L1
D5
L1
Flange EN 1092-1
ANSI B16.5
L1
Clamp
ISO 2852
L1
Dairy screwed connection DIN 11851
(Connection)
L1
Screwed connector for Di 3 ISO 228/1-G 1/4
and ANSI/ASME B1.20.1-1983 1/4 NPT
17.5 (0.69)
Dimension in mm (inch)
For not listed variants please contact product support
Sensor
size
Connections
DI (inch)
Type
Pressure rating Size
DI 3
(1/8)
Pipe thread ISO 228/1 - G¼
Pipe thread
ANSI/ASME B 1.20.1 - ¼” NPT
PN 100
PN 100
DI 6
(¼)
Flange EN 1092-1
Flange EN 1092-1
Flange ANSI B16.5
Flange ANSI B16.5
Screwed connection DIN 11851
Clamp ISO 2852
Flange EN 1092-1
Flange EN 1092-1
Flange ANSI B16.5
Flange ANSI B16.5
Screwed connection DIN 11851
Clamp ISO 2852
Flange EN 1092-1
Flange EN 1092-1
Flange ANSI B16.5
Flange ANSI B16.5
Screwed connection DIN 11851
Clamp ISO 2852
Flange EN 1092-1
Flange EN 1092-1
Flange ANSI B16.5
Flange ANSI B16.5
Screwed connection DIN 11851
Clamp ISO 2852
PN 100
PN 40
Class 150
Class 600
PN 40
PN 16
PN 100
PN 40
Class 150
Class 600
PN 40
PN 16
PN 100
PN 40
Class 150
Class 600
PN 40
PN 16
PN 100
PN 40
Class 150
Class 600
PN 25
PN 25
DI 15
(½)
DI 25
(1)
DI 40
(1½)
3/222
Siemens FI 01 · 2015 US Edition
L1
mm
L2
mm
L3
mm
H1
mm
B1
mm
D1
mm
D2
mm
D3
mm
D4
mm
D5
mm
¼”
¼”
400
400
280
280
75.5
75.5
60
60
0
0
21.3
21.3
104
104
-
-
-
DN 10
DN 10
½“
½“
DN 10
25 mm
DN 15
DN 15
½“
½“
DN 15
25 mm
DN 25
DN 25
1“
1“
DN 32
38 mm
DN 40
DN 40
1½“
1½“
DN 50
51 mm
580
560
624
608
532
570
634
620
639
660
586
624
970
934
967
992
922
940
1100
1063
1100
1128
1090
1062
390
390
390
390
390
390
444
444
444
444
444
444
700
700
700
700
700
700
850
850
850
850
850
850
62.0
62.0
62.0
62.0
62.0
62.0
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
75.5
40
40
40
40
40
40
44
44
44
44
44
44
126
126
126
126
126
126
180
180
180
180
180
180
12
12
12
12
12
12
20
20
20
20
20
20
25
25
25
25
25
25
0
0
0
0
0
0
17.0
17.0
17.0
17.0
17.0
17.0
21.3
21.3
21.3
21.3
21.3
21.3
33.7
33.7
33.7
33.7
33.7
33.7
48.3
48.3
48.3
48.3
48.3
48.3
104
104
104
104
104
104
129
129
129
129
129
129
219
219
219
219
219
219
273
273
273
273
273
273
100
90.0
88.9
95.3
105
95.0
88.9
95.3
140.0
115.0
108.0
124.0
170.0
150.0
127.0
155.4
-
70.0
60.0
60.5
66.5
75.0
65.0
60.5
66.5
100.0
85.0
79.2
88.9
125.0
110.0
98.6
114.3
-
14.0
14.0
15.7
15.7
14.0
14.0
15.7
15.7
18.0
14.0
15.7
19.1
22.0
18.0
15.7
22.4
-
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
For not listed variants please contact product support.
Sensor
size
Connections
DI (inch)
Type
Pressure rating Size
DI 3
(1/8)
Pipe thread ISO 228/1 - G¼
PN 100
Pipe thread
ANSI/ASME B 1.20.1 - ¼” NPT
PN 100
DI 6
(¼)
Flange EN 1092-1
Flange EN 1092-1
DI 15
(½)
DI 25
(1)
DI 40
(1½)
L1
inch
L3
inch
H1
inch
B1
inch
D1
inch
D2
inch
D3
inch
D4
inch
D5
inch
¼”
15.75 11.02 2.97
2.36
0
0.84
4.09
-
-
-
¼”
15.75 11.02 2.97
2.36
0
0.84
4.09
-
-
-
PN 100
DN 10
22.83 15.35 2.44
1.57
0.47
0.67
4.09
3.94
2.76
0.55
PN 40
DN 10
22.05 15.35 2.44
1.57
0.47
0.67
4.09
3.54
2.36
0.55
Flange ANSI B16.5
Class 150
½“
24.57 15.35 2.44
1.57
0.47
0.67
4.09
3.5
2.38
0.62
Flange ANSI B16.5
Class 600
½“
23.94 15.35 2.44
1.57
0.47
0.67
4.09
3.75
2.62
0.62
Screwed connection DIN 11851 PN 40
DN 10
20.94 15.35 2.44
1.57
0.47
0.67
4.09
-
-
-
Clamp ISO 2852
PN 16
25 mm
22.44 15.35 2.44
1.57
0.47
0.67
4.09
-
-
-
Flange EN 1092-1
PN 100
DN 15
24.96 17.48 2.97
1.73
0.79
0.84
5.08
2.95
4.13
0.55
Flange EN 1092-1
PN 40
DN 15
24.41 17.48 2.97
1.73
0.79
0.84
5.08
3.74
2.56
0.55
Flange ANSI B16.5
Class 150
½“
25.16 17.48 2.97
1.73
0.79
0.84
5.08
3.5
2.38
0.62
Flange ANSI B16.5
Class 600
½“
25.98 17.48 2.97
1.73
0.79
0.84
5.08
3.75
2.62
0.62
Screwed connection DIN 11851 PN 40
DN 15
23.07 17.48 2.97
1.73
0.79
0.84
5.08
-
-
-
Clamp ISO 2852
PN 16
25 mm
24.57 17.48 2.97
1.73
0.79
0.84
5.08
-
-
-
Flange EN 1092-1
PN 100
DN 25
38.19 27.56 2.97
4.96
0.98
1.33
8.62
3.94
5.51
0.71
Flange EN 1092-1
PN 40
DN 25
36.77 27.56 2.97
4.96
0.98
1.33
8.62
4.53
3.35
0.55
Flange ANSI B16.5
Class 150
1“
38.07 27.56 2.97
4.96
0.98
1.33
8.62
4.25
3.12
0.62
Flange ANSI B16.5
Class 600
1“
39.06 27.56 2.97
4.96
0.98
1.33
8.62
4.88
3.50
0.75
Screwed connection DIN 11851 PN 40
DN 32
36.30 27.56 2.97
4.96
0.98
1.33
8.62
-
-
-
Clamp ISO 2852
PN 16
38 mm
37.01 27.56 2.97
4.96
0.98
1.33
8.62
-
-
-
Flange EN 1092-1
PN 100
DN 40
43.31 33.46 2.97
7.09
0
1.9
10.75 4.92
6.69
0.87
Flange EN 1092-1
PN 40
DN 40
41.85 33.46 2.97
7.09
0
1.9
10.75 5.91
4.33
0.71
Flange ANSI B16.5
Class 150
1½“
43.31 33.46 2.97
7.09
0
1.9
10.75 5
3.88
0.62
Flange ANSI B16.5
Class 600
1½“
44.41 33.46 2.97
7.09
0
1.9
10.75 6.12
4.50
0.88
Screwed connection DIN 11851 PN 25
DN 50
42.91 33.46 2.97
7.09
0
1.9
10.75 -
-
-
Clamp ISO 2852
51 mm
41.81 33.46 2.97
7.09
0
1.9
10.75 -
-
-
PN 25
L2
inch
Siemens FI 01 · 2015 US Edition
3/223
3
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MASS 2100 DI 3 to DI 15
MASS 2100 sensor with “heating jacket”
H3
L5
D6
D7
B2
H3
3
D8
Dimensions in mm (inch)
Sensor size Connections heated
L5
H3
B2
D6
D7
D8
DI (inch)
Type
Pressure rating Size
mm (inch)
mm (inch)
mm (inch)
mm (inch)
mm (inch)
mm (inch)
DI 3 (1/8)
EN 1092-1
PN 40
DN 15
234 (9.21)
122 (4.8)
22 (0.87)
95 (3.74)
65.0 (2.56)
14.0 (0.55)
ANSI B16.5
Class 150
½"
234 (9.21)
131.6 (5.18)
22 (0.87)
88.9 (3.5)
60.5 (2.38)
15.7 (0.62)
EN 1092-1
PN 40
DN 15
234 (9.21)
112 (4.41)
22.7 (0.89)
95 (3.74)
65.0 (2.56)
14.0 (0.55)
ANSI B16.5
Class 150
½"
234 (9.21)
121.6 (4.79)
22.7 (0.89)
88.9 (3.5)
60.5 (2.38)
15.7 (0.62)
EN 1092-1
PN 40
DN 15
234 (9.21)
126.5 (4.98)
31.5 (1.24)
95 (3.74)
65.0 (2.56)
14.0 (0.55)
ANSI B16.5
Class150
½"
234 (9.21)
136.1 (5.36)
31.5 (1.24)
88.9 (3.5)
60.5 (2.38)
15.7 (0.62)
EN 1092-1
PN 40
DN 15
420 (16.54)
213.6 (8.41)
60 (2.36)
95 (3.74)
65.0 (2.56)
14.0 (0.55)
ANSI B16.5
Class 150
½"
420 (16.54)
223.2 (8.79)
60 (2.36)
88.9 (3.5)
60.5 (2.38)
15.7 (0.62)
EN 1092-1
PN 40
DN 15
500 (19.68)
267.5 (10.53) 43 (1.69)
95 (3.74)
65.0 (2.56)
14.0 (0.55)
ANSI B16.5
Class 150
½"
500 (19.68)
277.1 (10.91) 43 (1.69)
88.9 (3.5)
60.5 (2.38)
15.7 (0.62)
DI 6 (¼)
DI 15 (½)
DI 25 (1)
DI 40 (1½)
MASS 2100 and MASS 6000 Ex d compact version
MASS 2100 and MASS 6000 IP67 compact version
Ø130 (5.12)
+
H5
H6
+
255.1 (10.04)
L3
/
Dimensions in mm (inch)
Dimensions in mm (inch)
Sensor size L3
[mm (inch)]
[DI (inch)]
H5
H6
H5 + H6
[mm (inch)] [mm (inch)] [mm (inch)]
Sensor size L3
H5
H6
[DI (inch)]
[mm (inch)] [mm (inch)] [mm (inch)]
H5 + H6
[mm (inch)]
3 (1/8)
75 (2.95)
82 (3.23)
247 (9.72)
329 (12.95)
3 (1/8)
75 (2.95)
82 (3.23)
306 (12.04)
388 (15.28)
6 (¼)
62 (2.44)
72 (2.83)
257 (10.12)
329 (12.95)
6 (¼)
62 (2.44)
72 (2.83)
316 (12.44)
388 (15.28)
15 (½)
75 (2.95)
87 (3.43)
267 (10.51)
354 (13.94)
15 (½)
75 (2.95)
87 (3.43)
326 (12.83)
413 (16.26)
25 (1)
75 (2.95)
173 (6.81)
271 (10.67)
444 (17.48)
25 (1)
75 (2.95)
173 (6.81)
330 (13.00)
503 (19.80)
40 (1½)
75 (2.95)
227 (8.94)
271 (10.67)
498 (19.61)
40 (1½)
75 (2.95)
227 (8.94)
330 (13.00)
557 (21.93)
3/224
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MC2
■ Overview
SITRANS F C MC2 is available in sizes DN 100 and DN 150 (4"
and 6").
The MC2 sensor is suitable for accurate mass flow measurement
of a variety of liquids.
The sensor offers superior performance in terms of flow accuracy, turn-down ratio and density accuracy and delivers true
multi-parameter measurements i.e.: mass flow, volume flow,
density, temperature and fraction flow.
The very compact sensor construction makes installation and
commissioning of even the largest sizes very straight forward
and easy.
■ Benefits
• High accuracy better than 0.15 % of mass flow rate
• Large dynamic turn-down ratio
• Densitometer performance available through density accuracy better than 0.001 g/cm3
• Space-saving split-flow sensor design facilitating low pressure loss
• Parallel S-tube design and optimal oriented inductive sensors
enhances accuracy and turn-down ratio.
• Self-draining in both horizontal and vertical position
• Rigid enclosure design reduces the influence from pipeline vibration and thermal stress
• 4-wire Pt100 temperature measurement ensures optimum accuracy on mass flow, density and fraction flow
• SENSORPROM enables true "plug & play" - installed and commissioned in less than 10 minutes.
• Safe Ex design Ex em [ib] IIC
• Sensor pipe available in high-quality stainless steel
AISI 316Ti/1.4571 or Hastelloy C4/2.4610 offering optimum
corrosion resistance.
• CIP cleanability for food and beverage and pharmaceutical
applications
The main applications of the Coriolis flowmeter can be found in all
industries, such as:
Chemical and pharma
Detergents, bulk chemicals, pharmaceuticals, acids, alkalis
Food and beverage
Dairy products, beer, wine, softdrinks, Plato/Brix, fruit juices and
pulps, bottling, CO2 dosing, CIPliquids
Oil and gas
Liquid measurement, furnace control, test separators, LPG, oil bunkering
Water and waste water
Dosing of chemicals for water
treatment
The wide variety of combinations and versions from the modular
system means that ideal adaptation is possible to each measuring task. MC2 ist not recommended for gas applications.
■ Design
The MC2 sensor consists of 2 parallel measuring pipes, welded
directly onto a flow-splitter at each end to eliminate a direct coupling to the process connectors and significantly reduce effects
from external vibrations.
The flow-splitters are welded onto a rigid sensor housing which
acts as a mechanical low-pass filter.
The sensor is available in 2 material configurations, AISI 316L/
1.4436 or Hastelloy C4/2.4610 with a wide variety of process
connections.
The enclosure is made of stainless steel AISI 304/1.4301 with an
encapsulation grade of IP67/NEMA 4.
The sensor is Ex-approved Ex em [ib] IIC.
It can be installed in horizontal or vertical position, and is selfdraining in both positions.
The MC2 Ex version sensor is based on a different Ex concept
than MASS 6000. Therefore the MC2 Ex version sensor can only
be connected to MASS 6000 IP67, MASS 6000 19" or SIFLOW
FC070 standard versions, which have to be remote mounted in
the safe area. MASS 6000 Ex d, MASS 6000 19" Ex and SIFLOW
FC070 Ex can not be used with MC2 Ex sensors.
■ Application
Coriolis mass flowmeters are suitable for measuring all liquids.
The measurement is independent of changes in process conditions/parameters such as temperature, density, pressure, viscosity, conductivity, and flow profile.
Due to this versatility the meter is easy to install and the Coriolis
flowmeter is recognized for its high accuracy in a wide turndown
ratio which is paramount in many applications.
Hazardous area
Zone 1 + 2
Safe area
MC2 sensors are not designed or approved for flow measurement of gaseous process media.
The product is manufactured by ABB Automation Products
GmbH and distributed by Siemens.
Siemens FI 01 · 2015 US Edition
3/225
3
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MC2
Avoid cross talk
■ Function
The measuring principle is based on the Coriolis effect. See
“System information Coriolis mass flowmeters”.
■ Integration
Installation guidelines MC2 DN 100 and DN 150
Installation of sensor
The optimal installation orientation is a vertical installation with an
upward flow as shown in the following figure. This has the advantage that any solids contained in the fluid will settle downward
and gas bubbles will move upward out of the meter tube when
the flow rate is zero. Additionally, it is easy to drain the meter
tube. Deposits can thereby be avoided.
Installation in a drop line
Mount with reduction (A) or orifice (B) to prevent partially draining (min. back pressure: 0.2 bar).
Vertical orientation:
A
B
Vertical installation self-draining (upward flow)
Horizontal orientation, self-draining
α
3
Rigid mounting brackets must be used when installing the sensor. The brackets must be installed as close to the sensor as
possible, attached to the piping outside the process connections.
Installation in a drop line
Avoid vibrations
3/226
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MC2
■ Technical specifications
150 (6)
Inside pipe diameter
mm (inch) 43.1 (1.69)
76.1 (2.99)
Pipe wall thickness
mm (inch) 2.6 (0.10)
3.2 (0.13)
Fraction e.g. Brix
°Brix
10 0
0.5 ... 3.5 (0.18 ... 0.126)
DN
g/cm3
(lb/inch3)
602 000
(1 327 181)
100
3
(6 ˝
)
Density
203 500
(448 640)
p [mbar]
1000
0 ... 100
(on request)
Not possible
DN
Mass flow measurkg/h
ing range at pressure (lb/h)
drop of 2 bar (29 psi)
at 1 g/cm3
(0.036 lb/inch3)
Pressure drop
(4˝ )
100 (4)
150
Versions (mm (inch))
Temperature
Standard-version
-50 ... +200 °C (-58 ... +392 °F)
Ex-version
-50 ... +200 °C (-58 ... +392 °F)
10
1
10
100
1000
10000
0DVVIORZ>NJPLQ@
Liquid pressure
measuring pipe
Stainless steel
(DIN 2413, 20 °C
(68 °F))
bar (psi)
40 (580)
40 (580)
Materials
Measuring pipe
Stainless steel AISI 316Ti/1.4571
or Hastelloy C4/2.4610
IP67
Enclosure
Enclosure material/
connection box
AISI 304 (1.4301)/aluminum,
max. pressure 40 bar (580 psi)
Process connections
See dimensional drawings
Electrical connections
Screw terminals, M 20
5 x 2 x 0.35 mm2 twisted and
screened in pairs, ext. Ø 12 mm
Cable
10, 25, 75 or 150 m
(32.8, 82, 246 or 492 ft.)
Cable length
Ex-version
ATEX 1443X
Weight approx.
II 2G Ex em [ib] IIC T2-T6
kg (lb)
91 (201)
261 (573)
For accuracy specifications see „System information Coriolis
mass flowmeters“.
Siemens FI 01 · 2015 US Edition
3/227
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MC2
Selection and Ordering data
Article No. Ord. code
SITRANS F C flow sensors MC2
7 ME 4 3 0 0 7 77 7 7 - 777 7 77 7
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Dairy MLFB example
MC2 sensor
Article No.
7 ME 4 3 0 0 - 77777 - 7777
Sensor size DN 100. AISI 316Ti/1.4571
Nominal pressure: Clamps
DIN 11851, DN 100, PN 25
1D
F
43
Nominal diameter
Stainless steel AISI 316Ti/1.4571
1D
1E
DN 100
DN 150
3
2D
2E
DN 100
DN 150
Nominal pressure
PN 40
Class 150
Class 300
Clamps/screwed-connections
1
Configuration/calibration type: flow
and density (5 kg/m3 [0.31 lb/ft3])
Without Ex appproval
No cable
Standard calibration
Hastelloy C4/2.4610
A
C
D
F
Selection and Ordering data
A
A
1
Order code
Additional information
Please add “-Z“ to Article No. and specify Order
code(s) and plain text.
Process connections
Pressure testing certificate PED: 97/23/EC
C11
Flange EN 1092-1
DN 80 (PN 40, PN 100)
DN 100 (PN 40)
DN 150 (PN 40)
Material certificate EN 10204-3.1
C12
Material certificate according to NACE
C16
Tag name plate, stainless steel
Y17
Tag name plate, plastic self-adhesive
Y18
Customer-specified, matched pair (5 x 2)
On request
Customer-specified calibration (5 x 2)
On request
Customer-specified, matched pair (10 x 1)
On request
Customer-specified calibration (10 x 1)
On request
Flange ASME/ANSI B16.5
3“ (class 150/300/600)
4“ (class 150/300)
6“ (class 150/300)
Dairy screwed connection to DIN 11851
DN 80 (PN 25)
DN 100 (PN 25)
22
23
24
32
33
34
42
43
Operating instructions for SITRANS F C MC2
Dairy clamp connection DIN 32676
(ISO 2852) Tri-clamp
81 mm clamp (PN 10)
100 mm clamp (PN 10)
52
53
Aseptic nut flange DIN 11864-2 form A for
pipes dimensioned by DIN 11866
DN 80 (3“)
DN 100 (4“)
63
64
Configuration
Flow (0.15% of rate) and density
(5 kg/m3 [0.31 lb/ft3])
Flow (0.15% of rate) and density
(1 kg/m3 [0.06 lb/ft3])
Ex-approval and cable gland
Non-Ex, M20 x 1.5
ATEX, M20 x 1.5
Cable
No cable
All literature is also available for free at:
http://www.siemens.com/flowdocumentation
1
5
A
B
A
Calibration
Standard
3/228
This device is shipped with ABB documentation and an installation/connection instruction in four languages (Article No. A5E34730442).
1
Siemens FI 01 · 2015 US Edition
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MC2
■ Dimensional drawings
Remote design, flanged construction, DIN EN/ANSI
$
3
)
*
)ODQJHV
(1
$16,%
,62
)ORZGLUHFWLRQ
/
%
Process L
connec- [mm (inch)]
tion size
inch DN inch DN DIN 11864-2
form A
4
100 3
80 1618 (63.70)
4
100 1463 (57.60)
6
150 N/A
6
150 6
150 N/A
G1)
F
B
A
Weight
[mm
[mm
[mm
[kg (lb)]
[mm
(inch)] (inch)] (inch)] (inch)]
Meter
size
1)
EN 1092-1
PN 40
1640 (64.57)
1480 (58.27)
1778 (69.92)
2040 (80.31)
EN 1092-1
PN 100
1680 (66.14)
1530 (60.24)
N/A
N/A
ANSI B16.5
CL 150
1660 (65.35)
1500 (59.06)
1806 (71.10)
2070 (81.50)
ANSI B16.5
CL 300
1680 (66.14)
1520 (59.84)
1826 (71.89)
2090 (82.28)
ANSI B16.5
CL 600
1702 (67.01) 500
1568 (61.73) (19.69)
N/A
N/A
613
(24.13)
215
(8.46)
131
(5.16)
170
(6.69)
285
190
(11.22) (7.84)
260
(9.84)
84 (185)
91 (201)
120 (265)
260 (573)
For Ex add 54 mm
Remote design, food industry fittings, DIN 11851
Food industry fittings DIN 11851
Ø80
Ø80
Threaded
stubs
Threaded stubs
F
A
G
R
Flow direction
L
B
Meter size
Process connection size
inch
4
inch
3
4
1)
DN
100
g
DN
80
100
Rd 110 x 1/6
Rd 130 x ¼
L [mm
(inch)]
g [mm
(inch)]
1618 (63.70) 8 (0.31)
1463 (57.60) 10 (0.39)
L
G1) [mm
(inch)]
F [mm
(inch)]
500 (19.69)
215 (8.46) 131 (5.16) 170 (6.69) 401 (15.79)
314 (12.36)
B [mm
(inch)]
A [mm
(inch)]
R [mm
(inch)]
Weight
[kg (lb)]
82 (180)
86 (190)
For Ex add 54 mm
Siemens FI 01 · 2015 US Edition
3/229
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MC2
Remote design, Tri-clamp DIN 32676 (ISO 2852)
80
80
G
R
F
A
3
Flow direction
L
B
Dimensions in mm (inch)
L
G1)
F
B
A
R
[mm (inch)] 3 [mm (inch)] [mm (inch)] [mm (inch)] [mm (inch)] [mm (inch)]
Weight
[kg (lb)]
80
1598 (62.91)
440 (17.32)
71 (157)
100
1448 (57.01)
365 (14.37)
69 (152)
Meter size
Process connection
size
inch
DN
inch
DN
4
100
3
4
1)
For Ex add 54 mm
3/230
Siemens FI 01 · 2015 US Edition
500 (19.69) 215 (8.46)
131 (5.16)
170 (6.69)
© Siemens AG 2015
Flow Measurement
SITRANS F C
Flow sensor MC2
The max. allowable operating pressure is a function of the process connection type, the fluid temperature, the bolts and the
gaskets.
3UHVVXUH>EDU@
Pressure/temperature curves
Process Connections
• Flanges EN 1092-1/ANSI B16.5
• Tri-Clamp DIN 32676 (ISO 2852)
- DN 100: Series 1
• Food Industry fittings DIN 11851
80
40
PN 40
3
20
PN 16
0
-50
Housing as secondary containment
• Max. 40 bar
Pressure Equipment Directive 97/23/EG
• Conformity evaluation category III, fluid group 1
PSmax.
at 20 °C
(68 °F)
TSmax.
TSmin.
3UHVVXUH>EDU@
Material strength for process connections
Size
0
50
100
150
7HPSHUDWXUH>r&@
DIN-Flanges stainless steel AISI 316Ti/1.4571 to DN 100 (4”)
Corrosion resistance of measuring pipe material to measuring
medium has to be considered.
120
100
80
60
DN
inch
bar (psi g) °C (°F)
100
4
25 (363)
140 (284) -40 (-40)
Tri-Clamp acc. 100
DIN 32676
4
10 (145)
120 (248) -40 (-40)
Thread acc.
DIN 11851
100
60 PN 63
Pressure Rating
• PN 16, PN 40
Class 150, Class 300
Process
connection
120
°C (°F)
40
20
300 lb
150 lb
0
-50
0
50
100
150
7HPSHUDWXUH>r&@
ASME-Flanges stainless steel AISI 326Ti/1.4571 to DN 100 (4”)
For further information on the PED standard and requirements,
see page 9/6.
Siemens FI 01 · 2015 US Edition
3/231