SITRANS LVS100 Vibration Switch catalog sheet

© Siemens AG 2016
Level Measurement
Point level measurement - Vibrating switches
SITRANS LVS100
■ Overview
■ Configuration
Installation
Rigid extension with
sliding sleeve
Position tines vertically to
avoid product build-up:
use a 50 mm open-end
wrench to turn the process
connection until the tine
orientation marking faces up
or down.
L
*
Protection in case of high
material loading.
O
4
Cable gland faces
downward to avoid
water penetration.
SITRANS LVS100 is a vibrating point level switch for material
detection in bulk solids.
■ Benefits
Tine orientation marking
facing sideways.
• High resistance to mechanical forces
• Sliding sleeve options for adjustable insertion length and ease
of cleaning
• Rotatable enclosure for ease of installation and wiring
• Suitable for point level detection of materials starting at a bulk
density of 30 g/l (1.9 lb/ft3)
• Customer desired extensions up to 4 000 mm (157.48 inch)
■ Application
SITRANS LVS100 detects high, low or demand levels of dry bulk
solids in bins, silos or hoppers.
SITRANS LVS100 has a compact design and can be top, side,
or angle mounted. The vibrating fork design ensures the tines
are kept clean. The unique design of the fork and crystal assembly eliminates false high level readings even if tines become
damaged.
A signal from the electronic circuit excites a crystal in the probe
causing the fork to vibrate. If the fork is covered by material, the
change in vibration is detected by the electronic circuitry which
causes the relay to change state after a one second delay. When
the fork is free from material pressure, full vibration resumes and
the relay reverts to its normal condition.
• Key Applications: dry bulk solids in bins, silos, hoppers
4/104
Siemens FI 01 · Update 05/2016
Use angle mounting for
flowing material only. If angle
mounting is required with high
material loading, customersupplied protection from
falling material must be in
place.
*
max. deviation
from vertical O
max. length L
5°
4.0 m
45°
1.2 m
> 45°
0.6 m
SITRANS LVS100 installation, dimensions in mm (inch)
© Siemens AG 2016
Level Measurement
Point level measurement - Vibrating switches
SITRANS LVS100
■ Technical specifications
Design
Mode of Operation
Measuring principle
Vibrating point level switch
Input
Measured variable
High, low and demand
Measuring frequency
200 Hz
Material
• Enclosure
• Thread 1¼" NPT [(Taper), ANSI/
ASME B1.20.1],
R 1½" [(BSPT), EN 10226]
• Thread R 1½" [(BSPT),
EN 10226], ½" NPT [(Taper),
ANSI/ASME B1.20.1], sliding sleeve
[min. length 500 mm (19.69 inch)]
• Thread material:
stainless steel 304 (1.4301) or 316L
(1.4404) depending on configuration
Tine material
Stainless steel 316L (1.4404)
Degree of protection
IP66/Type 4/NEMA 4
Conduit entry
2 x M20 x 1.5 or 2 x ½" NPT
Weight
Standard version, no extensions:
approx. 1.7 kg (3.7 lb)
Power supply
• 19 ... 230 V AC, +10 %,
50 ... 60 Hz, 8 VA
• 19 ... 40 V DC, +10 %, 1.5 W
Certificates and approvals
•
•
•
•
•
•
Output
Relays
DPDT relay
Relay delay
From loss of vibration:
approximately 1 second
From resumption of vibration:
approximately 1 ... 2 s
Signal delay
Probe uncovered to covered:
approximately 1 s
Probe covered to uncovered:
approximately 1 ... 2 s
Relay fail-safe
High or low, switch selectable
Alarm output
Relay 8 A at 250 V AC, non-inductive
Relay 5 A at 30 V DC, non-inductive
Sensitivity
High or low, switch selectable
Rated operating conditions
Installation conditions
• Location
Indoor/outdoor
Epoxy coated aluminum
Process connection
CSA/FM General Purpose
CE
CSA/FM Dust Ignition Proof
RCM
ATEX II 1/2 D
IECex
Ambient conditions
• Ambient temperature
-40 ... +60 °C (-40 ... +140 °F)
• Installation category
III
• Pollution degree
2
Medium conditions
• Process temperature
-40 ... +150 °C (-40 ... +302 °F)
• Max. threaded bushing temperature 60 °C (140 °F)
• Max. enclosure surface temperature 90 °C (194 °F)
(Category 2D)
• Max. extension surface temperature 150 °C (302 °F)
(Category 1D)
• Pressure (vessel)
Max. 10 bar g (145 psi g)
European Pressure Directive 2014/68/
EU: Category 1
Minimum material density
Approx. 30 g/l (1.9 lb/ft3)
Siemens FI 01 · Update 05/2016
4/105
4
© Siemens AG 2016
Level Measurement
Point level measurement - Vibrating switches
SITRANS LVS100
Selection and Ordering data
Article No.
Selection and Ordering data
SITRANS LVS100, standard
Vibrating point level switch for high or low level
detection of bulk solids. Sensitivity > 30 g/l.
7ML5735-
Further Designs
Please add "-Z" to Art. No. and specify Order code(s).
77 7 77 - 07A 0
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Process temperature
Up to 150 °C (302 °F)

Process connection
Threaded
R 1½" [(BSPT), EN 10226]

1¼" NPT [(Taper), ANSI/ASME B1.20.1]

R 1½" [(BSPT), EN 10226] DIN 2999 thread,

sliding sleeve - min. length 500 mm (19.69 inch)2)
1½" NPT [(Taper), ANSI/ASME B1.20.1] ,

sliding sleeve [min. length 500 mm (19.69 inch)]2)
A
Signal bulb inserted in M20 cable gland1)
A20
Operating Instructions
Multi-language
All literature is available to download for free, in a
range of languages, at http://www.siemens.com/
processinstrumentation/documentation
This device is shipped with the Siemens Level and
Weighing manual DVD containing the ATEX Quick
Start and Operating Instructions library.
Article No.
7ML1998-5FT63
Spare Parts
Replacement Electronics Module LVS100 DPDT
Relay (19 ... 253 V AC, 19 ... 55 V DC)
R 1½" [(BSPT), EN 10226] DIN 2999 thread,
sliding sleeve
1½" NPT [(Taper), ANSI/ASME B1.20.1],
sliding sleeve [min. length 500 mm (19.69 inch)]
A
B
C
D
Extension length
Stainless steel 316L (1.4404)
Standard length, 170 mm (6.69 inch)

11
Add Order code Y01 and plain text:
"Insertion length ... mm"
Stainless steel 304 (1.4301)
• 300 ... 500 mm (11.81 ... 19.69 inch)
• 501 ... 1 000 mm (19.72 ... 39.37 inch)
• 1 001 ... 1 500 mm (39.41 ... 59.06 inch)
• 1 501 ... 2 000 mm (59.09 ... 78.74 inch)




12
13
14
15
1)


7ML1830-1NT
7ML1830-1NU
■ Dimensional drawings
60
(2.36)
A
B
C
D
120 (4.72)
Tine orientation
marking
Only available with the following configurations 7ML5735-2AA11-0AA0 or
7ML5735-2AB11-0AA0
Not available with extension length options 11, 12
 We can offer shorter delivery times for configurations designated with the
Quick Ship Symbol . For details see page 9/5 in the appendix.
Threaded bushing
[max. temp.
80 °C (176 °F)]
150 (5.91)
1)
2)
7ML1830-1NS
Available only with approval CE
16
17
18
20
• 2 001 ... 2 500 mm (78.78 ... 98.43 inch)
• 2 501 ... 3 000 mm (98.46 ... 118.11 inch)
• 3 001 ... 3 500 mm (118.15 ... 137.80 inch)
• 3 501 ... 4 000 mm (137.83 ... 157.48 inch)
Approvals
CSA/FM General Purpose, CE, RCM
CSA/FM Class II, Div. 1, Group E, F, G, Class III,
ATEX II 1/2 D, RCM
IEC-Ex Ex t IIIC T-- Da/Db IP6X
EAC Ex ta/tb IIIC Da/Db
Y01
Zone 21
(cat. 2)
Customer supplied
process flange
Tine
SITRANS LVS100, dimensions in mm (inch)
4/106
Siemens FI 01 · Update 05/2016
170 ... 4 000 (6.69 ... 157.48):
Customer specified rigid
Zone 20
(cat. 1)
125 (4.92):
Standard fork
4
 1
 2
Total insertion length: Enter the total insertion length
in plain text description, max. (50 mm increments)
60
(2.36)
75 (2.95)
Input Voltage
DPDT Relay: 19 ... 230 V AC, 19 ... 40 V DC
DPDT Relay: 19 ... 230 V AC, 19 ... 40 V DC
(stocked version)1)
Order code
© Siemens AG 2016
Level Measurement
Point level measurement - Vibrating switches
SITRANS LVS100
■ Schematics
Universal voltage (DPDT relay)
Alarm
output
relay
FSH/FSL
adjustment
Sensitivity adjustment:
B = factory setting,
A = decreased
876 543
LED
21
4
AC: Terminal 1: L
Terminal 2: N
19 ... 230 V AC, +10 %, 50 ... 60 Hz, 8 VA
DC: Terminal 1: +
Terminal 2: 19 ... 50 V DC, +10 %, 2 W
SITRANS LVS100 connections
Siemens FI 01 · Update 05/2016
4/107