Model 36M

Moore 36M Multi-Point Recorder
FEATURES & BENEFITS
High resolution, four-color vacuum fluorescent display
(VFD) provides information with exceptional color and
clarity
Six-color fiber tip dotting system delivers clear tracing
Two, three, or six universal inputs, up to 6 contact event
inputs, up to 12 relay outputs, and up to 4 analog outputs provide real flexibility
Control Solutions - Recorders
Easy front panel configuration simplifies start-up
50
PCMCIA memory card drive affords convenient data
storage
Die-cast door and steel case ensure operation in harsh
industrial environments
Requires only 9.29" (236 mm) depth behind the panel
DESCRIPTION
The Moore 36M Strip Chart Recorder displays information
on a high resolution, four-color VFD, and delivers clear traces
with a six-color fiber tip dotting system. Front panel configuration allows easy set-up, and a variety of input, output,
and performance options provide extensive system flexibility. Moreover, its rugged steel case and die cast door are designed to meet the rigors of industrial environments.
The four-color VFD creates exceptional color, vividly displaying five 12 mm characters for process value, twenty 4
mm characters for text and three 1 mm bargraphs. The
display also shows the measured value of each channel
with its associated descriptor or scale, as well as bargraph
indication of the three channels’ values. Information is recorded in distinctive colors on the chart using six different
fiber tip pens.
Inputs are fully universal with V, mV, TC, RTD and contact inputs. Recorder inputs can also be used as contact
inputs to trigger internal recorder actions. A dedicated six
contact event input card and up to 12 relay outputs are
also available. Up to four channels can be retransmitted as
linearized currents or voltage signals to other instruments.
The 36M is fully configurable from the front panel using local push buttons and text prompts. This allows direct
access to operator changes, and access to password protected
configurations.
Data stored on a Type I PCMCIA data card may be formulated for analysis, presentation, storage, and archiving using standard spreadsheet packages. Stored data can include
the recorder’s configuration that can be saved to the data
card for archiving or to transfer it to another recorder.
A variety of math, timer, totalizer, and counter functions offer full integrating and counting capabilities. These options
make it possible to carry out even the most complex application-specific function, such as environmental parameter
monitoring or gas flow compensation.
The Moore 36M Multi-Point Recorder supports MODBUS® communications to ensure communication with
other devices.
Moore Process Automation Solutions
www.moore-solutions.com
SPECIFICATIONS
Memory Configuration
EEPROM
ELECTRICAL
Custom Linearization
1 table of 32 point pairs
Power Requirements
Line Power: 45 to 65 Hz; 90 to 264 Vac
DC Power: 24/48 Vdc
Consumption (max): 100 VA, 100W
Fuse Type: None
Interrupt Protection: 40 msec @ 75% instrument
load
ENVIRONMENTAL
Temperature Limits
Operating: 0 to 50°C
Storage: -20 to +70°C
Protection
Door and Bezel: IP54 (similar to NEMA 3)
Case Sleeve: IP31 (similar to NEMA 1)
Shock/Vibration: IEC348, IEC873 recovers from
2G peak @ 10Hz to 150Hz
EMC
Electromagnetic Emissions: EN50081-2
Safety: To EN61010;Installation II; Pollution
Category 2
Static: IEC801-2 level 4 to 15KV(air discharge,
panel mounted, open door)
Fast Transients: IEC801-3 (instrument within
specification)
IEC801-4 (automatic recovery from any error/
malfunctions)
R.F. Immunity: IEC801-3 to 10V/m (level 3)
+ 1% on 0 to 100°C
PROPERTIES
Printing System
Pen Type: 6 nib cartridge
Pen Resolution: 0.2 mm
Pen Colors: Ch. 1: violet; Ch. 2: red; Ch. 3: black;
Ch. 4: green; Ch. 5: blue; Ch. 6: brown
Pen Life: 1.5 x 106 dots per color
Update Rate: 2 Hz
Response Time: 1 pass every 1.5 seconds
Paper Transport
Type: Stepper motor driving sprocket tube
Chart Speeds: 0 to 1,200 mm/hr (user selectable)
Chart Type: 100 mm calibrated width, 16 meter
Z-fold
Input to Paper Accuracy: 0.5% of span
Inputs
Maximum Number: 2, 3, or 6 inputs
Types
Thermocouple: B, C, D, E, G2, J, K, L, N, R, S, T,
U, Ni/NiMo, Platinel II
RTD: 2/3-wire resistance temperature detector
(not channel 1 if any other channel is a thermocouple input); Pt100, Pt1000, Ni100, Ni1000
Vdc: 38mV; 150mV; 1V; 10V dc milliamps
(mA): External shunt required
Ohms (Ω): 0 to 150Ω, 0 to 600Ω, 0 to 6KΩ
Contact Closure: Not channel 1
Minimum: 250 msec
Type Mix: Freely configurable
Noise Rejection
Common mode: >140dB (48 to 62 Hz ch to ch and
ch to gr)
Series mode: >60dB
Maximum Comm. Mode
250V continuous
Maximum Series Mode
45mV @ lowest range; 12V peak @ highest range
Isolation
300V RMS (dc to 65Hz; EN61010)
(or dc ch to ch and ch to gr)
Dielectric Strength
1350 Vac for 1 min (ch to gr)
2300 Vac for 1 min (ch to ch)
Insulation Resistance
>10MΩ @ 500Vdc
Control Solutions - Recorders
Humidity Limits
Operation: 5% to 80% RH (non-condensing)
Storage: 5% to 90% RH (non-condensing)
GENERAL PARAMETERS
Input Impedance
38mV, 150mV, 1V range: >10MΩ
10V range: 68.8KΩ
Overload Protection
50V peak
Open Circuit Detection
+ 57 nA maximum
Recognition Time: 500 msec
Minimum Break Resistance: 10MΩ
Data Storage Format
PCMCIA data card (Type I)
Americas +1 215 646 7400 ext. 6613
Asia Pacific +65 299 6051
Europe +44 1935 706262
51
Physical
Panel (Bezel) Size: 5.7" x 5.7" (144 mm x 144
mm)
Panel Cutout: 5.4" x 5.4" (138 mm x 138 mm)
Depth Behind Panel: 9.29"/236 mm with terminal
cover
Panel Mounting: DIN43700; Vertical + 30°
Weight: 7.75 lbs. (3.5 kg)
OPTIONS
Control Solutions - Recorders
Communications
RS422/RS485 Gould Modicon MODBUS® RTU
52
Math Pack
Number of Derived Channels: 16 configurable
Basic Functions: Off, const., add, subtract,
multiply, divide, modulus
Advanced Functions: (Includes Basic)
Sq rt, ch avg, DV group avg, rolling avg, ex, 10x,
in, log10, rate-of-change, sample and hold,
ch min, DV group latch min, DV group cont min,
ch max, DV group latch max, DV group cont max,
3rd order polynominal, F°, relative humidity,
mass flow: linear; mass flow: sq rt; ZrO2 probe,
high select, low select, switch, stopwatch,
timestamp, O2 correction, percentile
Event Input Board
Maximum Number Inputs: 6 discrete contact
closures or 16 coded (using binary codes on
inputs 1-4, strobed @ input 5)
Outputs
Relay Outputs: Up to 12
Ratings for Resistive Loads
(derate for inductive loads)
Switching Voltage: 250Vac, 30Vdc
Switching Power: 500VA, 60Ω
Break Current: 2A within referenced voltage &
power limits
Retransmission Outputs
Configurable Output Ranges: 0-20mA;
4-20mA; 1-5 V; 0-10V
Transmitter Power Supply
Outputs: 6 isolated 24Vdc
Current: 20mA each output
Permissible Load: 600Ω maximum
ACCESSORIES
Part number
4LA249556
GD201125
GD204080U050
IF247152U102
IF247152U104
IF247152U106
IF247152U107
LA207028U004
LA207028U006
Description
36M Multipoint Printhead
36P,M Chart, 50 line, Z-fold
36P,M Chart, 50 line, Roll
36P,M—128K SRAM Card
36P,M—512K SRAM Card
36P,M—2 Meg SRAM Card
36P,M—4 Meg SRAM Card
(4)-250 ohm shunts for 36V/P/M
(6)-250 ohm shunts for 36V/P/M
INSTALLATION DRAWINGS
Moore Process Automation Solutions
www.moore-solutions.com
MODEL NUMBER
Chart Paper Recorder with Chart Annotation
Power
1
90-264Vac (Set for 120 Vac)
2
24/28 Vdc
3
90-264 Vac (Set for 220/240 Vac)
Pens
1
2 Isolated Universal Input Channels
2
3 Isolated Universal Input Channels
3
6 Isolated Universal Input Channels
Chart Type and PCMCIA Drive Type
1
Z-fold and No Drive
2
Z-fold and Packed Data Storage Drive
3
Roll and No Drive
4
Roll and Packed Data Storage Drive
Option Slot #1
N
None
1
Relays-4NC
2
Relays-4NO
3
Relays-3 Form C
4
Retransmission-2 Outputs
5
Event Contacts-6 Inputs
Option Slot #2
N
None
1
Relays-4NC
2
Relays-4NO
3
Relays-3 Form C
4
Retransmission-2 Outputs
Option Slot #3
N
None
1
Relays-4NC
2
Relays-4NO
3
Relays-3 Form C
Transmitter Supply (not available with 24 Vdc Operating Voltage Option)
N
No Supply
1
3 Channel Supply
2
6 Channel Supply
Math
1
Basic Math
2
Advanced
3
Advanced + 32pt Characterizer
4
Advanced + 20 Messages
5
Advanced + 32pt Characterizer + 20 Messages
Advanced Functions
N
None
1
2 Totalizers
2
4 Totalizers
3
6 Totalizers
4
6 each Timers, Counters
5
6 each Timers, Counters, Totalizers
Communications
N
None
A
RS 485 Modbus
Reserved for Factory Use
N
Certifications
1
CE
2
CE and CSA
36M
1
2
1
N
N
N
N
1
N
Americas +1 215 646 7400 ext. 6613
N
N
1
Control Solutions - Recorders
36M
Sample Model Number
Asia Pacific +65 299 6051
Europe +44 1935 706262
53