FUJI EDS3-109E

INFRARED GAS ANALYZER FOR STACK GAS
DATA SHEET
ZSP
This instrument consists of an (or two) infrared gas analyzer(s),
a zirconia O2 sensor and a sampling device. It is used for
measuring the NOx, SO2, CO and O2 components in the flue
gas of various boilers, garbage incinerators, furnaces, etc.
through a selective combination of the components.
The instrument has passed a type approval test in compliance
with regulations on measurement and weighing (specified in
Japan) of each component. It is extensively applicable as a
gas analyzer for environmental pollution control and combustion control.
FEATURES
1. Simultaneous measurement of five components
A combination of two infrared gas analyzers plus a zirconia
0 2 sensor allows simultaneous measurement of four
components, NOx, SO 2, CO, CO2 and O2.
2. Easy operation with multiple functions
This analyzer can be equipped with a number of functions
such as auto calibration, O2 correction and self-diagnosis.
It is easier to operate and maintain than a separate type
calculator system.
3. Little interference from other gases
Measuring error due to interfering gases contained in
sample is nearly zeroed by the Permapure dryer and
interference compensating detector adopted in the analyzer section respectively.
4. Excellent measuring accuracy and stability
The optical system is so constructed as to be free from
external stress by using a single light source and double
beam paths, and therefore assures improved measuring
accuracy plus excellent stability.
5. High sensitivity and wide dynamic range
The infrared gas analyzer, which mounts a mass flow
sensor, has a high sensitivity and a broad dynamic range,
and is capable of switching the measuring range to obtain
1:20 at maximum.
6. Reliable sampling system
Simple but reliable sampling system based on Fuji's rich
experience in the field enables easy maintenance and
reduces frequency of maintenance.
SPECIFICATIONS
Type of cubicle: Indoor type or outdoor installation type
Measuring system:
Non-dispersion infrared absorption
(NDIR) method for NOx, SO 2 and C0;
zirconia method for O2
Sampling system: Dry sampling at approx. –15°C (dew
point) (Permapure dryer adopted)
Measuring range:NOx: 0 to 100…2000ppm
SO2: 0 to 100…1000ppm
CO: 0 to 200…2000ppm
CO2: 0 to 5…50%
O2: 0 to 10/25%
Dual range (except forO 2), max. range
ratio 1:20
Note: A type test-satisfied instrument having a measuring range of 50
to 2000ppm is available for each of NOx, SO 2 and CO.
± 0.5% of full scale
Zero; ± 2% of full scale/week for NOx,
SO 2 and CO
± 2% of full scale/month for O2
Span; ± 2% of full scale/week
Linearity:
± 2% of full scale
Response time: Within 3 minutes for 90% indication (after
extracting sample gas through the inlet)
Within 4 minutes for SO2
Output:
4 to 20mA DC (allowable load resistance
550Ω), non-isolated
For the number of output points, see
Table 1.
Contact output: Each NO (1a) contact (100V AC, 1 A) for
error (analyzer error, calibration error) and
auto calibration status
Indication:
Digital indication (on analyzer front panel)
Sample gas extracting rate:
Approx. 2L/min
Gas extractor: Electrical heating type (filter built in)
Filter mesh;40µm mesh of 316 stainless
steel
Probe;
316 stainless steel, length
600, 800 or l000mm
Flange;
JIS 5K65A FF
Repeatability:
Drift:
EDS3-109e
Date
Feb. 2, 2001
ZSP
Sample inlet tube:
φ 10/φ 8mm Teflon tube for CO and O 2
meter and NOx and O 2 meter
Heating tube (φ10/φ 8mm teflon tube) for
NOx, SO 2 and O2 meter and NOx, SO2,
CO and O2 meter
Standard gas (3.4L):
Standard air gas for zero calibration and
O2 span calibration
Standard analyte/remaining N2 gas for low
range span calibration and O 2 zero calibration
Ambient temperature:
–5 to +40°C
Power supply: 100 ± 10V AC, 50 or 60Hz
Power consumption:
Approx. 1.1kVA (excluding heating tube)
Mass {weight}: Approx. 300kg
Finish color:
Munsell 5Y 7/1
Standard requirements for sample gas:
Temperature; 60 to 800°C
Dust, 100 mg/ Nm3 or less
Pressure; –2.94 to +2.94kPa
Components;
SO2 0 to 1000ppm
NOx 0 to 2000ppm
CO2 several % to over 10%
CO
0 to 2000ppm
O2
0.5 to 21%
H2O
remaining percent
N2
When other component is contained in
sample gas, consult with Fuji.
Installation requirements:
Select a location free from direct sunlight
and severe vibration.
}
Table 1 Number of output points and kinds of outputs
ZSP1F ZSP1P
Type
ZSP1L
Analyte
NOx, SO2 NOx,
CO, (O2) SO2,
(O2)
No. of out- 11
put points
7
ZSP1B ZSP1A ZSP1H
CO,
(O2)
SO2,
(O2)
NOx,
NOx, SO2,
CO, (O2) CO, CO2,
(O2)
3
3
3
5
NOx instantaneous value
SO 2 instantaneous value
CO instantaneous value
O2 instantaneous value
NOx instantaneous value
after O2 correction
F: SO 2 instantaneous value
after O2 correction
NOx or
CO
signal
21—On
• Cs
21—Os
Zirconia
O2 sensor
O2
converter
O2 corrected
NOx or CO
[ppm]
O2 signal
Infrared gas analyzer
Full scale equals
to instantaneous
value before
conversion.
(2) Moving average output
1 hr. moving average output (for NOx, SO2)
4 hr. moving average output (for CO)
Auto calibration
Calibration accuracy:
± 0.2% of full scale
Output hold during calibration:
Possible
Calibration gas flow mode (Example of four components):
CO zero
CO span
NOx, SO2, zero
O2 span
NOx span
SO2 span
12
A, B, C, L
(D (2points)
E, F, G, H,
J, K)
G: CO instantaneous value after
O2 correction
H: 1hr. moving average NOx value
after O2 correction
J: 1hr. moving average SO2 value
after O2 correction
K: 4hr. moving average CO value
after O2 correction
L: CO2 instantaneous value
(1) O 2 correction
(a) Calculating equation
21—On
C=
• Cs
21—Os
C : Concentration after O2 correction
Cs : NOx or CO concentration measured
Os : O2 concentration measured
On : O2 concentration as conversion basic:
4% Oil combustion boiler
2
(b) Block diagram
O2 zero
Kind of out- A, B, C, A, B,
A, (D, E) C, (D, K) B, (D) A, C, (D,
put
E, K)
(D (2points) (D, E, F,
E, F, G, H, J)
H, J, K)
A:
B:
C:
D:
E:
ZSP1M
NOx,
(O2)
5% Gas combustion boiler
6% Solid object combustion
boiler, petroleum heating
furnace
12% Garbage incinerator
Auto calibration
Calibration cycle: Once daily (variable up to 7 days)
Calibration gas flow time:
Variable from 2 to 10min on each gas
Indication:
Calibration abnormality;
Error code E — indicated when zero
or span exceeds ± 20% of full scale.
Under calibration;
CAL lamp lights up
Auto calibration contact output:
NO (1a) contact, 250V AC, 1A (resistive
load)
CODE SYMBOLS
1 2 3 4 5 6 7 8
Z S P
ORDERING INFORMATION
9 10 11 12 13
Description
4 -
Application
Combustion exhaust gas
Others
1
2
Component
NOx + (O2)
SO2+ (O2)
CO + (O2)
NOx + SO2+ (O2)
NOx +CO+ (O2)
NOx + SO2 +CO + (O2)
NOx + SO2 +CO + CO2 + (O2)
Others
P
A
B
F
H
L
M
Z
O2 meter
Without
With zirconia type (Standard)
With paramagnetic type (option)
0
1
2
1.
2.
3.
4.
5.
6.
Measuring range of analyzer
Length of sample inlet tube
Insertion length of gas extracting tube
Power supply voltage and frequency
Omission of calibration standard gases if desired
Omission of type test for each component by Japan
quality assurance organization if desired
7. Necessity of spares for 1-year measurement
8. Contents of other function if necessary
Remarks: Separate specification is required for functions such as
remote range selection, output holding via external contact,
auto calibration start via external contact, isolated output
and measured concentration alarm.
Auto calibration
With
1
1
2
Type of cubicle
Indoor installation type
Outdoor installation type
Functions of individual components
• Gas extractor (type ZBA):
With heating type stainless steel filter;
standard diameter 40µm for filter.
• Gas conditioner (type ZBH9):
For separating drain and removing sulfuric
acid mist and oil mist or fine dust particles.
Automatically extractor air to prevent drain
from being sucked when internal pressure
reaches –6.78kPa {–690mmH20}.
• Pump(typeZBG8):
A diaphragm type suction pump to
extractor sample gas (sample gas flow
rate approx. 2L/min).
• Electronic gas cooler (type ZBCA):
An electronic dehumidifier which dries
the moisture in sample gas to a dew
point of approximately 2°C.
A valve for adjusting sample gas flow rate
and to be operated while monitoring the
flowrator on the front panel.
• Converter (type ZDL):
Added to the NOx analyzer. A converter
using a special catalyst which converts
NO2 gas into NO. The catalyst must be
renewed once every year.
Temperature controller is equipped in the
unit.
• Solenoid valve (type; AB21):
Used for introducing calibration gas when
auto calibration function is provided.
• Membrane filter (type ZBBM):
Eliminates fine dust particles with a paper
filter made of glasswool or Teflon, and
permits monitoring dust adhering condition on the front panel of the gas analyzer.
• Flow rotor:
Monitors standard flow rate of sample gas.
• Standard gas (type ZBM):
Reference gas used for calibrating zero
and span of the analyzer. When using a
zirconia oxygen meter, this is used together with air for calibrating zero for NO,
SO2, and CO and for calibrating span for
O 2.
• Zirconia O 2 sensor (type; ZFK4)
Used in combination with the infrared
analyzer. Outputs about 0V at measurement of the air, and an inverse logarithm
of about 1V at measurement of 0.05% O 2.
3
ZSP
Sampling system block diagram
O2
Gas extractor
Zirconia O2 sensor
Flow
mater
Membrane filter
NO2/NO
converter
Membrane
filter
Gas inlet tube
Gas conditioner
Sample gas inlet
Pump
CO,
CO2
NOx,SO2
Infrared gas
analyzer
Infrared gas
analyzer
Electronic
gas cooler
Solenoid
valve
Dain, gas
outlet
Pressure
regulator
Air
CO
/N2
CO2
/N2
NO
/N2
Standard gas
4
SO2
/N2
O2
/N2
Standard accessories
Quantity
No. Name
Type/part No.
NOx, SO 2 NOx
SO 2
Others
Remarks
1
Filter paper for membrane filter
TK701837C6
—
1 pack
—
1 pack
1 pack contains 25 sheets. Except for
SO2 meter
2
Fluoropore filter for membrane
filter
TK741833P3
2
—
2
—
For meter including SO2
3
Fuse (for instrument SW)
Various types
2
2
2
2
4
Tool for electric cooler
TK725417P7
1
1
1
1
5
Joint
TK7F7627P1
1 set
1 set
1 set
1 set
PT1/4 φ6mm, for pressure regulator
6
Hose band
TK539474C410
1 set
1 set
1 set
1 set
For φ6mm tube and manual calibration pipe
7
Toaron tube 0.3m
TK727528P1
1
1
1
1
8
Polyethylene tube 5m
415966P4
1
1
1
1
9
Anchor bolt
M12 x 160 x 50
4
4
4
4
Option
10
Instruction manual
INZ-TN4ZSP
INZ-TN2ZRG
1 set
1 set
1 set
1 set
1 copy each
11
Analyzer accessory
3 x fuse 2A
1 set
1 set
1 set
1 set
12
Water bottle
TK7H3493P1
1
1
1
1
Spare parts for 1-year measurement (option)
Quantity
No. Name
NOx, SO2 NOx
SO2
Other
Remarks
1
Spare parts for converter
1
1
—
—
ZBNK0001
2
Filter element for gas conditioner
2
2
2
2
TK7H8043P1
3
O-ring for gas conditioner
2
2
2
2
4
Filter paper for membrane filter
—
1 pack
—
1 pack
G65 chloroprene
25 sheet pack, ZBNC60
Except for SO2 meter
5
Fluoropore filter for membrane filter 6
—
6
—
TK741833P3, for meters including SO2
6
O-ring for membrane filter
O-ring for membrane filter
1
1
1
1
1
1
P49 chloroprene
7
1
1
8
Fuse
6
6
6
6
2A x 4, 3.2A x 2
9
Fixed restrictor
1
1
1
1
TK729264C6 (ø1 mm)
10 Fixed restrictor
1
1
1
1
TK729264C7 (ø1.4 mm)
11 Diaphragm membrane for pump
1
1
1
1
TK725417P5
12 Diaphragm valve for pump
1
1
1
1
TK725417P6
13 Spare parts for gas extractor
1
1
1
1
TK7J6183C1
G65 chloroprene
For instrument equipped with gas
extractor
5
ZSP
OUTLINE DIAGRAM
(Unit:mm)
For outdoor installation type
Anchor plan and door
opening diagram
1000
815
Sample
gas inlet
Rc3/8
Cable inlet (60x120 rectangular hole)
with plastic resin-cover
Rear
Front
70
760
70
30°
120
620
1400
1800
A–140–2–1
(Key No.200)
Anchor bolt
32
35
100
100
50
Drain and
gas outlet
2–Rc1/2
770
(4–M12x160x50)
760
620
800
800
4–ø15 hole
For indoor installation type
Anchor plan and door
opening diagram
800
620
Sample
gas inlet
Rc3/8
Rear
Front
Cable inlet (60x120 rectangular hole)
with plastic resin-cover
70
760
70
30°
800
1740
1400
1610
A–140–2–1
(Key No.200)
850
Anchor bolt
770
50
100
100
200
6
760
620
800
800
4–ø15 hole
35
(4–M12x160x50)
32
Drain and
gas outlet
2–Rc1/2
H2
NOx instantaneous value
A1
NOx average
value
A3
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Heating tube
SO2 instantaneous value
SO2 average
value
O2 instantaneous value
O2 corrected
NOx
O2 corrected
SO2
O2 instantaneous value
A2
A4
B1
B2
B3
B4
C1
C2
C3
C4
C5
C6
C1A
C2A
CONNECTION DIAGRAM
J2
H1
8
D22
J1
7
Auto calibration
contact output
D21
D12
Extractor
6
D11
G
5
Failure contact
output
C8
L2
4
C7
L1
3
O2 corrected
CO
E4
Power supply
100 AC
2
CO average
value
E3
E2
1
E1
25 26 27 28 29 30 31 32 33 34 35 36
CO instantaneous value
7
ZSP
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan
http://www.fesys.co.jp/eng
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http: //www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
Printed in Japan