FUJI EDS3-128

CO/O2 GAS ANALYZER FOR STACK GAS
DATA SHEET
ZSW
This instrument is equipped with a non-dispersive infrared
CO sensor and a galvanic oxygen sensor and used for
measuring CO and O2 in the gas emitted from garbage incinerators.
CO instantaneous value, O 2 instantaneous value and CO
average value after O2 correction (1-hour and time-variable
averaging) can be output.
FEATURES
1. The CO sensor adopts high-performance and long-life nondispersive infrared ray (NDIR) system. The CO sensor is
not consumable, and need not be replaced in 1 or 2 years.
2. Automatic calibration function by air equipped as standard allows stable measurement. No need to worry about
the consumption of standard gas.
3. Space-saving configuration unitized into 19-inch type allows the maintenance from the front.
4. A microjet recorder (Fuji Electric’s type PHC or PHE) can
be built in for recording measured gas concentration and
temperature values.
SPECIFICATIONS
• Measuring system:
Non-dispersive infrared absorption (NDIR)
method for CO
Galvanic method for O2
• Measurable component and range:
CO ; 0~500/1000ppm
0~500/2000ppm
O2 ; 0~25vol%
• Repeatability: ± 0.5% of full scale/CO
± 1% of full scale/O2
• Zero drift:
Max. ± 2.0% of full scale/month for CO
Max. ± 2.0% of full scale/week for O2
• Span drift:
Max. ± 2.0% of full scale/week for CO
Max. ± 2.0% of full scale/month for O2
• Linearity:
Max. ± 2.0% of full scale
• Sample gas extracting rate:
Approx. 0.4R/min
• Response time: Within 150 seconds for 90% indication
(after extracting sample gas through the
inlet)
• Warm-up time: Within 4 hours after power-on
• Output signals: Each signal within range from 4 to 20 mA
DC
Non-isolated (isolated output available at
option)
CO instantaneous value
O2 instantaneous value
Moving average CO value after O 2 correction (averaging time variable)
Time setting in 1 to 59 minutes (1-minute
increment) or 1 to 4 hours (1-hour increment)
Allowable load resistance 550 Ω or less
• Pump ON/OFF contact input:
No voltage contact (Closed: Pump ON)
• Indication:
LCD with back light for indicating CO instantaneous value, O 2 instantaneous
value, CO instantaneous value after O2
correction and CO average values after
O2 correction, O2 average value (1-hour
and time-variable averaging) and parameter assignment
• Recorder (option):
100 mm-width recorder (Fuji Electric’s
type PHC or PHE) built in, max. 6 points
recordable
• Gas extractor: General type (without filter)
• Titanium probe (300, 400, 600 or 800
mm long)
• Flange JIS 5K25AFF
• Weight: Approx. 1 kg (800 mm)
Electrical heating type (filter built in)
• Filter mesh; 40µm mesh of SUS 316
stainless steel
• Probe; SUS 316 stainless steel of 300,
400, 600 or 800 mm length
• Flange; JIS 5K65AFF
• Weight; Approx. 8.5 kg
EDS3-128
Date
Feb. 20, 2004
ZSW
• Separate extraction point filter:
Glass wool filter (for general type)
Container material: PVC
• Sample inlet tube:
φ10/φ8 Teflon tube or heating tube (max.
30 m)
• Functions:
1) O2 conversion and averaging calculation
• Conversion of measured CO gas concentration into a
value at standard O2 concentration
Calculating equation: C =
C;
CS;
OS;
ON;
CS(21−ON)
21−OS
CO concentration after O2 correction
Measured CO concentration
Measured O2 concentration (%)
Standard O 2 concentration (12% for
garbage incinerator)
• Moving average calculation time
Calculation is made for 1 to 59 minutes or 1 to 4 hours.
2) Auto calibration (of CO zero and O2 span)
• Auto calibration cycle settable range:
1 to 99 hours (1-hour step) or 1 to 40 days (1-day step)
need 7 days set to keep drift spec.
3) Output hold function
Holds output signal during auto calibration.
4) Other functions
• Temperature input signal: K thermocouple input x 2 (input for recorder)
• Standard requirements for sample gas:
• Temperature; 60 to 800˚C
• Dust;
0 to 300mg/Nm3 or less
• 0 to 50mg/Nm3:
Use general type extractor.
• 0 to 150mg/Nm3:
Use general type extractor with separate extraction point filter.
• 0 to 300mg/Nm3:
Use heating extractor.
• Pressure;
–2 to +2kPa
• Components; SO2 500 ppm or less
NOx 1000 ppm or less
CO2 0 to 15%
CO
0 to 2000 ppm
O2
0 to 21%
HCR 1000 ppm or less
N2, H2O Remaining percent
• Rated operating conditions:
• Ambient temperature;
0 to +40°C or -5 to 40°C,
-10 to 40°C
• Ambient humidity; 90% RH or less
• Power supply voltage;
100 V AC ± 10 V, 50 or
60 Hz, 115 V AC ± 10%
60 Hz, 230 V AC ± 10%
50 Hz
• Power supply frequency;
Rating ± 0.5 Hz
• Power consumption; Max. 400 VA
(without gas extractor)
2
• Installation requirements:
1) Selection of a place which does not
receive direct sunlight or radiation
from hot substances
2) Avoidance of a place under heavy vibration
3) Clean atmospheric air
• External dimensions (H x W x D):
Indoor type
1550 x 600 x 650 mm
Outdoor type
1640 x 615 x 765 mm
• Weight:
Approx. 180 kg (excluding standard gas)
• Cubicle finish color:
Munsell 5Y7/1 semi-gloss
• Other:
One standard gas (3.4R) cylinders
accommodable
(2 cylinders accommodable)
SCOPE OF DELIVERY
•
•
•
•
•
•
•
Gas analyzer system
Specified external drain separator
Specified gas extractor/probe set
Specified gas inlet tube set
Specified standard gas (with pressure adjustor) set
Specified recorder to be accommodated in cubicle
Standard accessory set
ORDERING INFORMATION
1. Code symbols
2. Necessity of spares for 1-year measurement
3. Type of recorder (option)
Type
• For 6-point continuous recording
Type: PHC62043-NA0YY-B
Scale specifications: 0 to 500ppmCO, TK7K0750C2
• For 2-point continuous recording
Type: PHE2BB12-660NY-B
Scale specifications: 0 to 500ppmCO, TK7K0750C4
CODE SYMBOLS
1 2 3 4 5 6 7 8
Z SW
9 10 11 12 13
1 -
14 15 16 17
Description
-
Indication (4th code)
Japanese
English
1
3
Measurement range (5th code)
0~500 / 1000 ppmCO
0~25% O2
0~500 / 2000 ppmCO
0~25% O2
C
D
Power supply (6th code)
AC 100V 50Hz
AC 100V 60Hz
AC 115V 60Hz
AC 230V 50Hz
3
4
6
7
Cubicle structure (7th code)
Indoor type
Temperature of installation site : 0°C or higher
Outdoor type
Indoor type
Temperature of installation site : – 5°C or higher
Outdoor type
Indoor type
Temperature of installation site : – 10°C or higher
Outdoor type
1
2
A
B
C
D
Gas pressure at extraction point (9th code)
– 2~+2kPa
4
External drain separator (10th code)
Not provided
Provided (Note 1)
Y
A
Recorder (11th code)
Not provided
With 6-point recorder PHC
With 2-point recorder PHE
Y
C
E
(Note 2) (Recorder hardware needs to be ordered
simultaneously.)
Isolated output (12th code)
Non-isolated output
Isolated output (3 outputs)
0
3
Fluorescent lamp (13th code)
Not provided
Provided
Y
1
Standard gas (14th code)
Not provided
CO/N2 (3.4L, cylinders 1)
1
4
Gas extractor, Flange dimension (15th code)
Without gas extractor
General type, Flange JIS 5K 25A FF
General type (With separate extraction point filter), Flange JIS 5K 25A FF
Heating extractor, Flange JIS 5K 65A FF
Y
1
2
3
Gas extractor insertion length (16th code)
Without gas extractor
300mm
400mm
600mm
800mm
Y
A
B
C
D
Y
A
B
C
D
E
F
G
H
J
K
L
Sample inlet tube type
Length (17th code)
Not provided
Not provided
φ10/ φ8 Teflon tube
5m
φ10/ φ8 Teflon tube
10m
φ10/ φ8 Teflon tube
15m
φ10/ φ8 Teflon tube
20m
φ10/ φ8 Teflon tube
25m
φ10/ φ8 Teflon tube
30m
Heating tube
10m
Heating tube
15m
Specify when the ambient temperature
Heating tube (Note)
20m
may fall to –5˚C or lower
Heating tube
25m
Heating tube
30m
(Note) Specify heating extractor.
(Note 1) Specify when the inclination of the gas inlet tube from the gas extraction point to the gas inlet of the main unit is
less than 15° downward. If there is a possibility of freezing, separately take measures against freezing.
(Note 2) Recorder assignment is as follows.
In 2-point recording: CH1; Average CO value after O2
In 6-point recording: CH1; CO instantaneous value
correction
CH2; O2 instantaneous value
CH2; O2 instantaneous value
CH3; vacant
CH4; 1-hour average CO value after O2 correction
PHE of 2-point continuous recording type only can be
CH5; K thermocouple 1
specified.
CH6; K thermocouple 2
3
ZSW
Extraction point filter
(option)
0 to 90°
Sampling module
Gas extractor
Flow checker
(0.4L/min)
Gas
aspirator
Sample gas inlet
15° or more
Mist filter
Air inlet
Rc1/4
Needle
valve
Membrane
filter
2-direction
Electronic solenoid
gas cooler valve
2-direction
solenoid valve
Solenoid
pump
Infrared
CO sensor
Galvanic
O2 sensor
Gas
outlet
Rc1/4
CO SPAN
O2 ZERO
(With filter)
CO ZERO
O2 SPAN
Safety drain trap
Pressure
adjuster
Drain
CO/ 3.4L
N2 Standard gas
Rc1/4
Functions of Individual Components
• Gas extractor: Extracts sample gas.
General type (without filter), Used when
dust concentration is 50mg/Nm3 or lower.
General type (with separate extraction
point filter): Used when dust concentration is 150mg/Nm 3 or lower.
Heating type (with filter, Filter diameter:
40µ): Used when dust concentration is
300mg/Nm3 or lower.
• Extraction point filter:
Removes dust. Used when dust concentration is in the range from 50 to 150mg/
Nm3
• Mist filter:
Removes dust and mist.
• Safety drain trap:
Functions to prevent suction of drain pot
and drain.
(Suction prevention operating pressure:
6kPa, Water seal pressure: Approx. 2kPa)
4
• Solenoid pump:
Operated at the time of calibration (atmospheric aspiration)
• Gas aspirator: Aspirates sample gas (Approx. 0.4L/min.)
• Needle valve: Adjusts sample gas flow rate.
• Electronic gas cooler:
Dehumidifies sample gas.
• Membrane filter:
Removes minute dust.
• Flow checker: Monitors flow rate.
• Standard gas (CO/CO2):
For CO span and O2 zero calibration
• Pressure adjuster:
Depressurizes standard gas.
Standard Accessories
Name
Type of extractor General extractor General extractor
with extraction
Heating extractor
or without
point filter
extractor
Filter paper for membrane filter (Spare)
2 sheets
2 sheets
2 sheets
Fuse with nail (3A)
2
2
2
Standard gas joint
1
1
1
Hose band for fixing standard gas cylinder
6
6
6
Toalon tube for standard gas connection, 0.3m and φ9/φ6mm
1
1
1
Polyethylene tube for standard gas connection, 1m and φ6/φ4mm 1
1
1
Fixed restrictor for standard gas connection
1
1
1
Water bottle for injection
1
1
1
Gas extractor flange packing
1 pc. or none
1 pc.
1 pc.
Gas extractor fastening bolt and nut
1 set or none
1 set
1 set
Extraction point filter fastening bolt and nut
–
1 set
–
O-ring (G50) for gas extractor (spare)
–
–
1
Heating tube support (only when heating tube is provided)
–
–
1 set
Instruction manual
1 copy
1 copy
1 copy
Spare Parts for 1-Year Measurement
Name
Type of extractor General extractor General extractor
or without
with extraction
Heating extractor
extractor
point filter
Type
ZBN1SW12
ZBN1SW22
ZBN1SW32
Filter paper for membrane filter (25 sheets)
1 bag
1 bag
1 bag
Rubber ring for membrane filter
2
2
2
O-ring (G65) for membrane filter
2
2
2
Fuse (3A)
4
4
4
Element for mist filter
2
2
2
O-ring (G65) for mist filter
2
2
2
Diaphragm for gas aspirator
1
1
1
Gas aspirator valve
1
1
1
Filter for extraction point filter
–
12
–
O-ring (G65) for extraction point filter
–
2
–
Mesh filter for heating extractor
–
–
1
Packing for heating extractor
–
–
1 pc.
Seal metal packing for heating extractor
–
–
1 pc.
O-ring (G50) for heating extractor
–
–
1
Recorder (Option)
• Standard accessory:
Ink cartridge × 1
Recording paper × 1 roll
• Spares for 1 year:
PHE type recorder
Ink cartridge (Type: PHZH2002) × 2
Recording paper (Type: PEX00DL1-5000B) × 12 rolls
PHC type recorder
Ink cartridge (Type: PHZH1002) × 2
Recording paper (Type: PEX00DL1-5000B) × 12 rolls
5
ZSW
CONNECTION DIAGRAM
Terminal block TB1
TB2
1
1
Main power
100V AC supply
2
2
3
Ground
3
4
Option
(Heating extractor)
5
Option
(Heating tube)
7
8
6
+
Analog output CO
instantaneous value –
9
7
4
Internal
connection
6
5
10
+
Analog output O2
instantaneous value –
11
Analog output (CO average –
value after correction)
+
13
Pump ON/OFF contact input (DRY) –
(Closed: Pump ON)
+
15
8
12
9
Temperature +
input 1
–
10
+
11
–
12
14
Temperature
input 2
16
(M4 screw)
OUTLINE DIAGRAM (Unit: mm)
615
765 (For outdoor use)
Sample gas inlet
(with through
connector for
φ10 Teflon tube)
100
300
Anchor plan and door
opening diagram
420
100
Air inlet
Rc1/4
610
Rear side
Exhaust port
Rc1/4
1550
1295
6
Door
Anchor bolt
(4-M12 × 160 × 50)
100
610
650
420
(675)
600
Left side view
Front view
4-φ15
35
Drain
Rc1/4
150
32
70 70 70
170
3-φ34
(With grommet)
(Waterproofing
measures with
through wire
metal, etc.
required for
outdoor
specifications.)
1640 (For outdoor use)
Front side
OUTLINE DIAGRAM (Unit: mm)
<Heating gas extractor>
φ155
UP
Gas outlet Rc1/2
Handle exchange filter
Cover
297
357
(750)
Heating tube support
(Mounted for heating
tube specifications.)
(309.5)
300, 400, 600, 800
(Insertion length depends on code symbol.)
(76)
210
Cable hole G1/2
100V AC, 100W
Terminal box
Rc3/4
Flange
JIS 5K 65A FF
Sampling pipe 20A
4-f15
φ130
φ155
φ27
Gas inlet
7
ZSW
<General type gas extractor>
Insertion length depends on
code symbols.
4-φ12 (Even)
φ75
Approx.67
Gas inlet
Sampling pipe 15A
10
Flange
JIS5K 25A FF
95
33
Gas outlet
(with joint for Rc1/2)
(φ10/φ8 Teflon tube)
<Extraction point filter>
200
Maintenance space
10
INLET
Gas inlet
30
55
18
16
φ100
25°
150
80
55
2-MTG. Holes 7 DIA.
Clear
Joint for φ10/φ8 Teflon tube
2-Rc1/4
φ60
Gas outlet
100
105
OUTLET
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan
http://www.fesys.co.jp/eng
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http: //www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
Printed in Japan