NDIR TYPE INFRARED GAS ANALYZER (3-COMPONENT ANALYZER) ZKJ7 DATA SHEET This product is the replacement of our gas analyzer <model: ZRG> This gas analyzer (ZKJ7) is capable of measuring the concentrations of NO, SO2, CO2, CO, CH4 and O2 components in sample gas. NO, SO2, CO2, CO, CH4 are measured by non-dispersion infrared method (NDIR), while O2 is measured by externalmount type zirconia method sensor. A maximum of three components including O2 are simultaneously measurable. The mass flow type twin detector of high sensitivity and reliability adopted in the infrared ray method detection block makes the measurement hardly affected by interfering components. In addition, a microprocessor is built in and a large-size liquid crystal display is equipped for easier operation, higher accuracy and more functions. Optimum as an analyzer unit of gas measurement system for combustion exhaust gas from refuse incinerator and boiler, or gas from different industrial furnaces. Combination of this product and model sampling system (ZSU) is satisfied authentication test by measurement act. FEATURES 1.M e a s u r e t h r e e c o m p o n e n t s i n c l u d i n g O 2 simultaneously and continuously Simultaneously and continuously measures up to three components out of NO, SO2, CO2, CO, CH4, plus O2, or up to totally three components. 2. Hardly affected by interference by other gases The mass flow type twin detector of high sensitivity and reliability adopted makes the measurement hardly affected by interfering components of other gas, ensuring a stable operation. 3. Equipped with abundant functions O2 conversion, average value computation, automatic calibration, one touch calibration, upper/lower limit alarm, remote measurement range changeover, range identification signal output, etc. incorporated can configure applications to match particular uses. 4. Easy-to-see large LCD unit The large LCD unit adopted allows observing easily the indication of all measured components and computation values. The interactive operation facilitates setting. 5. Maximum range ratio is 1:25 Measuring ranges are changeable. 6. Drift +/-1% FS/week (more than 0 to 200ppm range) SPECIFICATIONS Standard Specifications Principle of measurement: NO, SO2, CO2, CO, CH4; Non-dispersion infrared-ray absorption method Single light source and double beams (double-beam system) O2 ; Exclusive zirconia O2 sensor (externally installed). Model: ZFK7 Measurable gas components and measuring range: NO SO2 CO2 CO CH4 O2 (External Zirconia) Minimum range 0 – 100ppm 0 – 100ppm 0 – 100ppm 0 – 100ppm 0 – 200ppm Maximam range 0 – 5000ppm 0 – 10vol% 0 – 100vol% 0 – 100vol% 0 – 100vol% 0 – 10vol% 0 – 25vol% • Max. 3 components measurement including O2. • Measuring range ratio ≤ 1:5 (O2 sensor) ≤ 1:25 (except for O2 sensor) • Measuring ranges are changeable between the specified minimum and maximum range Settable one range or two ranges *For measurable components and possible combinations of measuring ranges, refer to Tables 1-(1) to (3). EDS3-141a Date Nov. 21, 2011 ZKJ7 Measured value indication: Digital indication in 4 digits (LCD with back light) • Instantaneous value of each component • Instantaneous value after O2 conversion (only in NO, SO2, CO sensor with O2 sensor) • Average value after O2 conversion (only in NO, SO2, CO sensor with O2 sensor) Analog output signals: 4 to 20mA DC or 0 to 1V DC, non-isolated output ; 7 points max. Analog output corresponds to measured value indication in 1:1. max.load550Ω. for 4 to 20 mA DC min.load 100kΩ. for 0 to 1V DC *Refer to Table 2, for the channel No. of displayed values and analog output signals. Analog input signal: For signal input from externally installed O2 sensor. Signal requirement; (1)Signal from Fuji’s Zirconia O2 sensor (TYPE: ZFK7) (2)0 to 1V DC from an O2 sensor Input section is not isolated. (Depend on O 2 input signal, measured concentration indication and O 2 conversion.) Relay contact output: 1a contact (250V AC/2A, resistive load) Instrument error, calibration error, range identification, auto calibration status, pump ON/OFF. solenoid valve drive signal for auto calibration, auto calibration end. 1c contact (250V AC/2A, resistive load selectable 6 outputs) High/Low limit alarm contact output. * All relay contacts are isolated mutually and from the internal circuit. Contact input: No-voltage contact (ON/0V, OFF/5V DC, 5mA flowing at ON) * For ZRG (ON/5V, OFF/0V) Remote range switch, auto calibration remote start, remote holding, average value reset. Isolated from the internal circuit with photocoupler. Contact inputs are not isolated from one another. Power supply: Voltage rating ; 100V to 240V AC Allowable range; 85V to 264V AC Frequency ; 50Hz/60Hz Power consumption; 250VA max. Operating conditions: Ambient temperature;-5°C to 45°C Ambient humidity ;90% RH max., non-condensing Storage conditions: Ambient temperature; -20°C to 60°C Ambient humidity ;95% RH max., non-condensing Dimensions (H x W x D): Analyzer main unit; 835 x 218 x 202mm Mass: Approx. 16 kg Finish color: Front panel; Off-white (Munsell 10Y7.5/0.5 or equivalent) 2 Enclosure: Steel casing, for indoor use Material of gas-contacting parts: Gas inlet/outlet; Teflon Sample cell; SUS304,chloroprene rubber Infrared-ray transmitting window; CaF 2 O2 sensor sample cell : SUS316 Internal piping; Toaron, Teflon Gas inlet/outlet: Rc1/4 or ø6 hose end Purge gas flow rate:1L/min ( when required) Standard Functions Output signal holding: Output signals are held during manual and auto calibrations by activation of holding (turning “ON” its setting). The values to be held are the ones just before start calibration mode or setting value. It is selectable. Indication of instantaneous values will not be held. Remote output holding: Output signal is held at the latest value or setting value by short-circuiting the remote output holding input terminals. Holding is maintained while the terminals are short-circuited. Indication of instantaneous values will not be held. Switch ranges : The switch ranges is available in manual, auto, and remote modes. Only preset switch method is effective. Manual: Allows range to switch by key operation. Auto: Allows range to switch from low to high range when 90%FS or more is available in the low range. Allows range to switch from high to low range when 80%FS or less is available in the low range. Remote: No-voltage contact input (for measurable components) Allows range to switch via an external signal when remote range switch input is received. When the contact input terminals for each component are short-circuited, the first range is selected, and it is switched to the second range when the terminals are open. Range identification signal: The present measuring range is identified by a contact signal. The contact output terminals for each component are short-circuited when the first range is selected, and when the second range is selected, the terminals are open. Auto calibration: Auto calibration is carried out periodically at the preset cycle. W h e n a s t a n d a r d g a s c y l i n d e r fo r calibration and a solenoid valve for opening/closing the gas flow line are prepared externally by the customer, calibration will be carried out with the solenoid valve drive contacts for zero calibration and each span calibration turned on/off sequentially at the set auto calibration timing. Auto calibration cycle setting: Auto calibration cycle is set. Setting is variable within 1 to 99 hours (in increments of 1 hour) or 1 to 40 days (in increments of 1 day). Gas flow time setting: The time for flowing each calibration gas in auto calibration is set. Settable within 60 to 900 seconds (in increments of 1 second) Auto calibration remote start: Auto calibration is carried out only once according to an external input signal. Calibration sequence is settable in the same way as the general auto calibration. Auto calibration is started by opening the auto calibration remote start input terminals after short-circuiting for 1.5 seconds or longer. Auto zero calibration: Auto zero calibration is carried out periodically at the preset cycle. This cycle is independent on “Auto calibradion” cycle. When zero calibration gas and solenoid valve for opening/closing the calibration gas flow line are prepared externally by the customer, zero calibration will be carried out with the solenoid valve drive contact for zero calibration turned on/off at the set auto zero calibration timing. Auto zero calibration cycle setting: Auto zero calibration cycle is set. Setting is variable within 1 to 99 hours (in increments of 1 hour) or Setting is variable within 1 to 40 days (in increments of 1 day) Gas flow time setting: The timing for flowing zero gas in auto zero calibration is set. Settable 60 to 900 seconds (in increments of 1 second) High/Low limit alarm: Alarm contact output turns on when measurement value reach to the preset high or low limit alarm value. Contacts close when the channel value of each channel becomes larger than the high alarm limit value or smaller than the low alarm limit value. Instrument error contact output: Contacts close at occurrence of analyzer error No. 1, 3 or 10. Calibration error contact output: Contacts close at occurrence of manual or auto calibration error (any of errors No. 4 to 9). Auto calibration status contact outputs: Contacts close during auto calibration. Pump ON/OFF contact output: During measurement, this contact close. While calibration gas is flowing, this contact open. This contact is connected in power supply of pump, and stop the sample gas while calibration gas flowing. Average value reset: Ave r a g e va l u e a ft e r O 2 c o nve r s i o n is started under preset condition by opening the average value reset input terminals after short-circuiting for 1.5 seconds or longer. Reset is carried out by short-circuiting. Restart is carried out by opening. 3 ZKJ7 Auto calibration interlocking function: When these two products are lined up and installed, output the auto calibration synchronized signal to second product. Contact output during auto calibration: While auto calibration is carried out, this contact is closed. Auto calibration end contact output: Cont act is closed for 1.5 seconds after finishing to flow the gas of auto calibration. Optional Functions O2 conversion: Conversion of measured NO, SO 2 and CO gas concentrations into values at standard O2 concentration Conversion formula: C = 21–On × Cs 21–Os C : Sample gas concentration after O 2 conversion Cs: Measured concentration of sample gas Os: Measured O2 concentration (Limit settable, 1 to 20%O 2) On: Standard O 2 concentration (value changeable by setting; 0 to 19%O2) Average value after O2 conversion : The result of O 2 conversion or instantaneous O 2 value can be outputted as an average value in the preset period of time. Used for averaging is the moving average method in which sampling is carried out at intervals of 30 seconds. (Output is updated every 30 seconds. It is the average value in the determined period of time just before the latest updating.) Averaging time is settable within 1 to 59 minutes (in increments of 1 minute) or 1 to 4 hours (in increments of 1 hour). Communication function: RS-232C (9pins D-sub) Half-duplex bit serial Start-stop synchronization ModbusTM protcol Contents: Read/Wright parameters Read measurement concentration and instrument status Remark: When connecting via RS-485 interface, a RS-232C RS485 converter should be used. 4 Performance : ±0.5% of full scale : ±1% of full scale : ±1% of full scale/week (±2% of full scale/week; range be tween 0 to 100ppm and 0 to 200ppm) Span drift : ±2% of full scale/week Response time : (for 90% FS response) 15 sec electrical response Within 60 seconds including replacement time of sampling gas (when gas flow rate is 0.5L/min) Gas replacement time depends on the number of measuring components and measuring range Repeatability Linearity Zero drift Standard Requirements for Sample Gas : 0.5L / min ±0.2L / min : 0 to 50°C : 10 kPa or less (Gas outlet side should be open to the atmospheric air.) Dust : 100µg/Nm 3 or less in particle size of 1µm or less Mist : Unallowable Moisture : Below a level where saturation occurs at 2°C (condensation unallowable). Corrosive component: 1 ppm or less Standard gas for calibration: Zero gas ;Dry N2 Span gas ;E a ch s a m p l e g a s h av i n g concentration 90 to 100% of its measuring component range (recommended). Gas beyond concentration 100%FS is unusable. In case a zirconia O 2 analyzer is installed externally and calibration is carried out on the same calibration gas line: Zero gas ;Dry air or atmospheric air (provided without CO 2 sensor) Span gas ; Except O 2 measurement, e a ch s a m p l e g a s h av i n g concentration 90 to 100% of its measuring range. For O 2 sensor, O 2 gas of 1 to 2vol%. Flow rate Temperature Pressure Installation Requirements • Indoor use. (Select a place where the equipment does not receive direct sunshine, draft/rain or radiation from hot substances. If such a place cannot be found, a roof or cover should be prepared for protection.) • Avoide a place where receives heavy vibration • Select a place where atmospheric air is clean EC Directive Compliance The product conforms to requirement of the Low Voltage Directive and EMC directive. It conforms to following standards for product safety and electromagnetic compatibility ; EN61010-1 : 2001, EN62311 : 2008 Safet y requirements for electrical equipment for measurement, control and laboratory use. EN61326-1~2006 EN61326-2-3 : 2006, EN61000-3-2 : 2006, A1: 2009, A2: 2009, EN61000-3-3 : 2008 Electrical equipment for measurement, control and laboratory use – EMC requirements. *The product mounted in a steel enclosure conforms to the requirements of EMC directive. ZRG ZKJ7 differences ZRG Contact input DC5V ZKJ7 No-voltage contact Zirconia O2 analyzer ZFK3, 4 ZFK7 Average value Calculation is always carried out even during holding. Calculation is suspended during holding Calibration error contact Auto calibration status error Calibration status error (Auto/manual) 5 ZKJ7 Principle diagram of NDIR type measurement (For NO, SO2, CO2, CO) Motor Distribution cell (interference filter) Trimmer Interference compensating detector Reference cell Rotary sector Sample cell Sample gas inlet Sample gas outlet Gain changeover Infrared (single) light source AC amplifier Sector sync. signal O2 sensor (option) A/D CPU Output 0 to 1V or 4 to 20mA DC 6 Indicator Indication D/A Operating keys External input / Contact output RAM ROM Example configuration of gas sampling system The following illustrates a typical system configuration for five component gas measurement for monitoring combustion exhaust gas from boiler, refuse incinerator, etc. Contact FUJI ELECTRIC for system configuration matching the particular use or further information. In the case infrared gas analyzer (Model: ZRG) is replaced, Ziruconia O 2 sensor should also be replaced. (1) Gas extractor ZKJ7 Unit Gas inlet tube (2) Mist filter (7) Membrane filter (8) Flowmeter (4) Gas aspirator 15° or larger (11) NO2/NO converter for NOx measurement (5) Electronic gas cooler (7) Membrane filter Infrared gas analyzer main unit NO, SO2 (3) Drain trap (6) Solenoid valve ZERO NO SO2 O2 Pressure reducing valve Gas outlet N2 or Air NO /N2 O2 /N2 (10) Zirconia O2 sensor (ZFK7) (9) Standard gas Drain Functions of Individual Components (1)Gas extractor:Gas extractor with a heating type stainless steel filter of standard mesh 40µm (2)Mist filter: For separation of drain and removal of dust and mist (3)Safety drain trap: Prevention of drain from being sucked and composite operation of constantpressure bubbler (4)Gas aspirator:For aspiration of sample gas (sample gas flow rate approx. 2L/min) (5)Electronic gas cooler: Dries the moisture in sample gas to a dew point of approx. 3°C. (6)Solenoid valve:Used for introducing calibration gas. (7)Membrane filter: PTFE filter used to eliminate fine dust particles and permit monitoring of dust adhering condition on the front panel of the gas analyzer. SO2 /N2 O2 (8)Flowmeter: Adjusts and monitors the flow rate of sample gas. (9)Standard gas: Reference gas used for calibrating zero and span of the analyzer. (10)Zirconia O2 sensor: External zirconia oxygen sensor used for measuring the oxygen concentration (0 to 25%) in sample gas. In the case ZFK3-4 is used, ZFK7 should also be replaced. (11)Converter: Added to NOx analyzer. A special catalyst material for efficient conversion of NO2 gas to NO is used. *(Note) For each gas sampling device, refer to the separate Data Sheet for each gas sampling device. 7 ZKJ7 CODE SYMBOLS Description Digit 4 <Custom specifications> Replecement of ZRG type 5 <Measurable component (NO, SO2, CO2, CO, CH4)> 1st component 2nd component NO SO2 CO2 CO CH4 NO SO2 CO2 CO Others 6 <Measurable component (O2)> None External zirconia type O2 sensor (ZFK7 type) External O2 analyzer without external indication (input the signal for O2 conversion externally) 7 <Gas inlet/outlet> Rc1/4 (with purging inlet) Teflon φ6(none purging inlet) 8 <Revision code> 9 10 <Indication> In Japanese In English 11 <Measuring range> 1st component.1st range 0-100ppm 0-200ppm 0-250ppm 0-500ppm 0-1000ppm 0-2000ppm 0-5000ppm 0-1% 0-2% 0-5% 0-10% 0-20% 0-50% 0-100% Others 12 <Measuring range> 1st component.2nd range None 0-200ppm 0-250ppm 0-500ppm 0-1000ppm 0-2000ppm 0-5000ppm 0-1% 0-2% 0-5% 0-10% 0-20% 0-50% 0-100% Others 8 note 1 2 3 4 5 6 7 8 9 10 11 12 13 ZKJ 7 Y 4 YYYY 7 note 1 note 2, 9 P A D B E F G Z Y A B D 1 4 4 note 3 note 3 14 15 16 17 18 19 20 Y J E B C D E F G H J K L M N P R Z Y C D E F G H J K L M N P R Z 21 22 23 Digit No. of code Digit Description 13 <Measuring range> 2nd component.1st range None 0-100ppm 0-200ppm 0-250ppm 0-500ppm 0-1000ppm 0-2000ppm 0-5000ppm 0-1% 0-2% 0-5% 0-10% 0-20% 0-50% 0-100% Others 14 <Measuring range> 2nd component.2nd range None 0-200ppm 0-250ppm 0-500ppm 0-1000ppm 0-2000ppm 0-5000ppm 0-1% 0-2% 0-5% 0-10% 0-20% 0-50% 0-100% Others 15 16 17 18 19 <O2 sensor range> 20 Minimun range Maximum range None None 0-10% 0-25% 0-25% None Others 21 <Output> 4 to 20mA DC 0 to 1V DC 4 to 20mA DC + communication function 0 to 1V DC + communication function 22 <O2 conversion> None With O2 conversion output 23 <Ajustment> For combustion exhaust gas Others note note 3 note 3 1 2 3 4 5 6 7 8 9 10 11 12 13 ZKJ 7 YJ 4 14 15 16 17 18 19 20 21 22 23 YYYY Digit No. of code Y B C D E F G H J K L M N P R Z Y C D E F G H J K L M N P R Z YY YY note 4 note 5, 9 note 6 note 7 note 8 YY MV VY ZZ A B C D Y A B Z Note 1 a) when “B” is specified at the 6th digit, O2 sensor signal should be set as 0-1VDC linear corresponding to full scale. b) External zirconia O2 sensor and external O2 analyzer are not included in the scope of supply. Note 2 When two products are lined up and installed, please refer to the corresponding table for measured value to specify the digit for second product. (Please also refer to note 9) Note 3 Please refer to the appendix, for possible combination of measuring components and range in the data sheet. Note 4 When "Y", "D" is specified at the 6th digit, Only "YY" should be selected. Note 5 When two products are lined up and installed, Only "VY" should be selected for both products. (Please also refer to note 9) Note 6 Only measuring value of NO, SO2, CO are calculated as O2 calculation, O2 converted average value are outputted at the same time. Note 7 When "Y" is specified at the 6th digit, Only "Y" should be selected. Note 8 When "Z" is specified at the 23rd digit, gas composition table of actual measured gas has to be sent to Fuji with your purchase order. 9 ZKJ7 Note 9 Precaution to observe when performing installation of two analyzers with external O2 analyzer • When two ZKJ7 are lined up side by side and installed with external O2 analyzer, Be sure to observe connection of external O2 analyzer shown following diagram on the right side. (with converted value/converted average value) In this case O2 indication can not be conducted with second ZKJ7 (due to limitation of measurement ) Please refer to "Connecting method/analog output component" for connection to the terminal. [Composition for ZRG] [Composition for ZKJ7] ZAG ZKG1, 2 ZKG1, 2 ZFK7 ZFK3, 4 ZRG (First) ZRG (Second) Replace ZKJ7 (First) ZKJ7 (Second) *First analyzer: This analyzer is connected to O2 signal directly and indicate O2 indication. Second analyzer: This analyzer is connected to O2 instantaneous value from first analyzer and could not indicate O2 indication. • O2 range is fixed 0-25%. • With these connection component for second analyzer should be NO sensor, SO2 sensor or NO/SO2 sensor. Please refer to the "correspondence table for measured value" "Code symbols" for details. • When ZRG is replaced, two analyzers should be replaced at the same time. 10 Table 1. Measurable component and range – availability check table – (*) Range code shows settable combination of the maximum range rate. (1) Single component analyzer (NO, SO2, CO2, CO, CH4) ☆ : NO Measuring range ○ : CO Measuring range 2st range 1st range B 0 ~ 100ppm C 0 ~ 200ppm D 0 ~ 250ppm E 0 ~ 500ppm F 0 ~ 1000ppm G 0 ~ 2000ppm H 0 ~ 5000ppm J 0 ~ 1% K 0 ~ 2% L 0 ~ 5% M 0 ~ 10% N 0 ~ 20% P 0 ~ 50% R 0 ~ 100% C 0~ 200ppm ☆□◎○ □ : SO2 Measuring range △ : CH4 Measuring range D 0~ 250ppm ☆□◎○ ☆□◎○△ E 0~ 500ppm ☆□◎○ ☆□◎○△ ☆□◎○△ F 0~ 1000ppm ☆□◎○ ☆□◎○△ ☆□◎○△ ☆□◎○△ G 0~ 2000ppm ☆□◎○ ☆□◎○△ ☆□◎○△ ☆□◎○△ ☆□◎○△ ◎ : CO2 Measuring range H 0~ 5000ppm J K L M N P R 0 ~ 1% 0 ~ 2% 0 ~ 5% 0 ~ 10% 0 ~ 20% 0 ~ 50% 0 ~ 100% ☆□◎○△ ☆□◎○△ ☆□◎○△ □◎○△ ☆□◎○△ □◎○△ □◎○△ ☆□◎○△ □◎○△ □◎○△ □◎○△ □◎○△ □◎○△ □◎○△ □◎○△ □◎○△ □◎○△ □◎○△ ◎○△ □◎○△ □◎○△ ◎○△ ◎○△ □◎○△ ◎○△ ◎○△ ◎○△ ◎○△ ◎○△ ◎○△ ◎○△ ◎○△ ◎○△ (2) Double-component analyzer (NO/SO2) ○ : Double-component analyzer Measuring range (1st range) SO2 NO B C D E F G 0 ~ 100ppm 0 ~ 200ppm 0 ~ 250ppm 0 ~ 500ppm 0 ~ 1000ppm 0 ~ 2000ppm B 0~ 100ppm ○ ○ ○ ○ ○ ○ C 0~ 200ppm ○ ○ ○ ○ ○ ○ D 0~ 250ppm ○ ○ ○ ○ ○ ○ E 0~ 500ppm ○ ○ ○ ○ ○ ○ F 0~ 1000ppm ○ ○ ○ ○ ○ ○ G 0~ 2000ppm ○ ○ ○ ○ ○ ○ H 0~ 5000ppm ○ ○ ○ ○ ○ ○ * 2nd range: Max. NO (0-200ppm), SO2 (0-5000ppm), Selectable range up to 25 times of 1st. range (3) Double-component analyzer (CO2/CO) ①~⑤ : Double-component analyzer Measuring range (1st range) CO CO2 B C D E F G H J K L M N P R B 0~ 100ppm 0 ~ 100ppm ① 0 ~ 200ppm ① 0 ~ 250ppm ① 0 ~ 500ppm ① 0 ~ 1000ppm ① 0 ~ 2000ppm ① 0 ~ 5000ppm ① 0 ~ 1% 0 ~ 2% 0 ~ 5% 0 ~ 10% ② 0 ~ 20% ② 0 ~ 50% 0 ~ 100% C 0~ 200ppm ① ① ① ① ① ① ① D 0~ 250ppm ① ① ① ① ① ① ① ② ② ② ② E 0~ 500ppm ① ① ① ① ① ① ① ② ② F G H 0~ 0~ 0~ 1000ppm 2000ppm 5000ppm ① ① ① ① ① ① ① ① ① ① ① ① ① ① ① ① ① ① ① ① ① ② ② ② ② J K L M N 0 ~ 1% 0 ~ 2% 0 ~ 5% 0 ~ 10% 0 ~ 20% ③ ③ ③ ③ ③ ④ ④ ④ ③ ④ ④ ④ ⑤ ⑤ ⑤ ⑤ ⑤ ⑤ ⑤ ⑤ ② ② P R 0 ~ 50% 0 ~ 100% ⑤ ⑤ ⑤ ⑤ ⑤ ⑤ ⑤ ⑤ * Max. measuring range as 2nd range is following. Selectable range up to 25times of 1st range. ① : CO (0-5000ppm), CO2 (0-5000ppm) ② : CO (0-5000ppm), CO2 (0-20%) ③ : CO (0-50%), CO2 (0-20%) ④ : Selectable range up to 25 times. ⑤ : CO (0-100%), CO2 (0-100%) 11 ZKJ7 OUTLINE DIAGRAM (Unit: mm) <Analyzer main unit> <Upper> <Mounting size> Sample gas inlet Rc1/4 or φ6 PURGE Sample gas outlet Rc1/4 or φ6 INLET1 (It is the same as the ZRG type) OUTLET1 813 Purge gas inlet Rc1/4 or φ6 Communication connector AUTO CAL OPTION2 OPTION1 A/O 140 A/O Terminal OPTION2 Terminal AUTO CAL terminal 3 - φ12 M10 screw is needed formounting to main unit OPTION1 Terminal <Side> 11.5 φ12 (12) <Front> ZKJ GAS ANALYZER Operation panel POWER MODE ZERO SPAN ENTER (597.1) ESC 835 813 (787) Power supply (M4 screw) E Power supply AC100V to AC240V 50/60Hz Power Terminal ON F Power switch USE PE FUSE 250V T 3.15A N L AC V POWER SOURCE Fuse 140 218 12 3.2 12 43.2 202.2 EXTERNAL CONNECTION DIAGRAM A/O 1st component 2nd component 0 to 1V or 4 to 20mA DC (instantaneous value) 0 to 1V or 4 to 20mA DC (O2 converted average value) + 1 11 + - 2 12 - + 3 13 + - 4 14 - 0 to 1V or 4 to 20mA DC (instantaneous value) + 5 15 + - 6 16 - 0 to 1V or 4 to 20mA DC (O2 converted average value) + 7 17 + - 8 18 - 9 O2 sensor input O2 Output 0 to 1V or 4 to 20mA DC (instantaneous value) 1st component O2 converted value output ***) 0 to 1V DC or 4 to 20mA DC 2nd component O2 converted value output **) 0 to 1V DC or 4 to 20mA DC 19 FAULT 10 1st component Upper limit alarm contact output (option) 2nd component Upper limit alarm contact output (option) O2 Upper limit alarm contact output (option) Unassigned (*) 20 OPTION 1 Remote output holding input ①-④ + + 1st component Remote range input 1 11 2 12 3 13 + + R1 Auto calibration end contact output ⑤-⑧ 1st component Range identification contact output + O2 Remote range input R1 4 14 5 15 6 16 7 17 8 18 9 19 10 20 2nd component Range identification contact output (option) O2 Range identification contact output (M3.5 screw) OPTION 2 AUTO CAL 11 2 12 3 13 4 14 5 15 6 16 7 17 8 18 9 19 10 20 1st component Lower limit alarm contact output (option) Auto calibration contact output Zero gas contact output 2nd component Lower limit alarm contact output (option) O2 Lower limit alarm contact output (option) Unassigned (*) Ground Power supply Span gas-1 contact output Span gas-2 contact output 2 12 3 13 4 14 5 15 6 16 7 17 8 18 9 19 10 20 Calibration error contact output + Auto calibration remote start input - Unassigned (*) Span gas-3 contact output Unassigned (*) (M3.5 screw) Connector <CN2> For serial communicetion 1 2 3 4 *) Do not use the terminal for relay. **) When two analyzers are lined up and installed with O2 converted value and converted average value, by First analyzer O2 instantaneous value (0-1V DC: 0-25% range equivalent) is outputted. ***) When two analyzers are lined up and installed and first analyzer is used as CO2/CO sensor, CO converted value is outputted to 1st component O2 converted value output. 11 Pump ON/OFF contact (M3.5 screw) Power supply (M4 screw) 1 2nd component Remote range input Contact output during auto calibration ⑮-⑱ (M3.5 screw) 1 Average value reset input ⑪-⑭ 5 6 7 8 9 13 ZKJ7 Connecting method/analog output component Measurment of NO/SO2/CO/O2 sensor [Example connection] (with converted value and converted average value) ZKG1,2 ZFK7 ZKJ7 (First) A/O terminal CO Instantaneous value CO Converted average value unassigned unassigned 14 ZKJ7 (Second) A/O terminal + ① ⑪ + – ② ⑫ – + ③ ⑬ + – ④ ⑭ – + ⑤ ⑮ + – ⑥ ⑯ – + ⑦ ⑰ + ⑧ ⑱ – ⑨ ⑩ – NO Instantaneous value O2 Instantaneous value NO Converted average value CO converted value SO2 Instantaneous value + ① ⑪ + – ② ⑫ – + ③ ⑬ + – ④ ⑭ – + ⑤ ⑮ + – ⑥ ⑯ – + ⑦ ⑰ + ⑧ ⑱ – ⑲ ⑨ ⑲ ⑳ ⑩ ⑳ O2Instantaneous SO2 value Converted (0-1V DC only) average value – unassigned NO converted value SO2 converted value Measurment of NO/SO2/CO2CO/O2 sensor [Example connection] (with converted value and converted average value) ZKG1,2 ZFK7 ZKJ7 (First) A/O terminal CO2 Instantaneous value unassigned CO Instantaneous value CO Converted average value ZKJ7 (Second) A/O terminal + ① ⑪ + – ② ⑫ – + ③ ⑬ + – ④ ⑭ – + ⑤ ⑮ + – ⑥ ⑯ – + ⑦ ⑰ + ⑧ ⑱ – ⑨ ⑩ – NO Instantaneous value O2 Instantaneous value NO Converted average value CO converted value SO2 Instantaneous value + ① ⑪ + – ② ⑫ – + ③ ⑬ + – ④ ⑭ – + ⑤ ⑮ + – ⑥ ⑯ – + ⑦ ⑰ + ⑧ ⑱ – ⑲ ⑨ ⑲ ⑳ ⑩ ⑳ O2Instantaneous SO2 value Converted (0-1V DC only) average value – unassigned NO converted value SO2 converted value 15 ZKJ7 Table 2. Correspondence between measurement channels and measured value The following table gives measurement channels and their contents according to the code symbols. 1. In case of using only one analyzer. Code symbol 5th digit 6th digit 22nd digit Y Y P Y Y A Y Y D Y Y B Y Y E Y Y F Y Y G A, B Y P A, B Y A A, B Y D A, B Y B A, B Y E A, B Y F A, B Y G A, B A P A, B A A A, B A B A, B A F G 16 A, B A Contents Ch1: NO Ch1: SO2 Ch1: CO2 Ch1: CO Ch1: CH4 Ch1: NO, Ch2: SO2 Ch1: CO2, Ch2: CO Ch1: NO, Ch2: O2 Ch1: SO2, Ch2: O2 Ch1: CO2, Ch2: O2 Ch1: CO, Ch2: O2 Ch1: CH4, Ch2: O2 Ch1: NO, Ch2: SO2, Ch3: O2 Ch1: CO2, Ch2: CO, Ch3: O2 Ch1: NO, Ch2: O2, Ch3: Converted NO, Ch4: Converted NO average Ch1: SO2, Ch2: O2, Ch3: Converted SO2, Ch4: Converted SO2 average Ch1: CO, Ch2: O2, Ch3: Converted CO, Ch4: Converted CO average Ch1: NO, Ch2: SO2, Ch3: O2, Ch4: Converted NO, Ch5: Converted SO2, Ch6: Converted NO average, Ch7: Converted SO2 average Ch1: CO2, Ch2: CO, Ch3: O2, Ch4: Converted CO, Ch5: Converted CO average 2. In case of using two analyzers installed. 1st analyzer Code symbol 5th digit 6th digit B Y D Y E Y G Y B Y D Y E Y G Y B Y D Y E Y G Y B Y D Y E Y G Y B Y D Y E Y G Y B Y D Y E Y G Y D A, B 22nd digit Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Ch1: CO Ch1: CO2 Ch1: CH4 Ch1: CO2, Ch2: CO Ch1: CO Ch1: CO2 Ch1: CH4 Ch1: CO2, Ch2: CO Ch1: CO Ch1: CO2 Ch1: CH4 Ch1: CO2, Ch2: CO Ch1: CO Ch1: CO2 Ch1: CH4 Ch1: CO2, Ch2: CO Ch1: CO Ch1: CO2 Ch1: CH4 Ch1: CO2, Ch2: CO Ch1: CO Ch1: CO2 Ch1: CH4 Ch1: CO2, Ch2: CO Ch1: CO2 E A, B Y Ch1: CH4 D A, B Y Ch1: CO2 E A, B Y Ch1: CH4 D A, B Y Ch1: CO2 E A, B Y Ch1: CH4 B A, B A G A, B A B A, B A G A, B A B A, B A G A, B A Ch1: CO, Ch2: O2 Ch3: Converted CO Ch4: Converted CO average Ch1: CO2, Ch2: CO, Ch3: O2 Ch4: Converted CO Ch5: Converted CO average Ch1: CO, Ch2: O2 Ch3: Converted CO Ch4: Converted CO average Ch1: CO2, Ch2: CO, Ch3: O2 Ch4: Converted CO Ch5: Converted CO average Ch1: CO, Ch2: O2 Ch3: Converted CO Ch4: Converted CO average Ch1: CO2, Ch2: CO, Ch3: O2 Ch4: Converted CO Ch5: Converted CO average Contents Second analyzer Code symbol Contents 5th digit 6th digit 22nd digit Y Y P Ch1: NO A Y Y Ch1: SO2 F Y Y Ch1: NO, Ch2: SO2 P A, B Y Ch1: NO, Ch2: O2 A A, B Y Ch1: SO2, Ch2: O2 F A, B Y Ch1: NO, Ch2: SO2 Ch3: O2 P A, B A A A, B A F A, B A P D A Ch1: NO Ch2: O2 Ch3: Converted NO Ch4: Converted NO average Ch1: SO2 Ch2: O2 Ch3: Converted SO2 Ch4: Converted SO2 average Ch1: NO Ch2: SO2 Ch3: O2 Ch4: Converted NO Ch5: Converted SO2 Ch6: Converted NO average Ch7: Converted SO2 average Ch1: NO Ch2: Converted NO Ch3: Converted NO average A D A Ch1: SO2 Ch2: Converted SO2 Ch3: Converted SO2 average F D A Ch1: NO Ch2: SO2 Ch3: Converted NO Ch4: Converted SO2 Ch5: Converted NO average Ch6: Converted SO2 average 17 ZKJ7 Example of Code symbol for replacement [ZRG] Component CO, CO2, O2 NO, SO2, O2 Example of code symbol ZRG6GBB2-0B0ND-FF1F5FY ZRG6FBB2-0B0ND-FF1F5FY 1st analyzer Component CO, CO2, O2 Example of code symbol ZKJ7G A 14-YJBFB-FYYYY VY -CAB O2 range 0-25% fixed External zirconia O2 sensor 2nd analyzer NO, SO2 ZKJ7F D 14-YJBFB-FYYYYYY-CAB without external O2 indication 1st analyzer 2nd analyzer [ZKJ7] SCOPE OF DELIVERY ORDERING INFORMATION •Gas analyzer ... 1 unit •Spare fuses (250V, 3.15A AC, delay type) ... 2 pcs •Instruction manual ... 1 copy 1. Code symbols 2. Application and composition of sample gas Items to be prepared separately •Various sampling devices (refer to Data Sheets for the sampling devices) •Dedicated zirconia O2 sensor (see Page 16) 18 Exclusive Zirconia O2 Sensor (to be purchased separately) This sensor should be used with ZKJ7. Measuring method: Zirconia system Measurable component and measuring range: Measurable component Range Oxygen O2 0 to 25vol% Repeatability: Within ± 0.5% of full scale Linearity: Within ± 1% of full scale Zero drift: Within ± 1% of full scale/week Span drift: Within ± 2% of full scale/week Response time: Approx. 20 seconds (for 90% response) Measured gas flow rate: 0.5 ± 0.25L / min Remark: T h e Z i r c o n i a s y s t e m , d u e t o i t s principle, may produce a measuring error due to relative concentration versus the com-bustible O2 gas concentration. Also, a corrosive gas (SO 2 of 250 ppm or more, etc.) may affect the life of the sensor. Enclosure: Steel casing, for indoor application Indication: Temperature indication (LED) Temperature alarm output: Contact output 1a contact, Contact capacity 220V, 1A AC (resistive load) Outer dimensions (H x W x D): 141 x 170 x 190mm Mass {weight}: Approx. 3kg Finish color: Munsell 5Y 7/1 CODE SYMBOLS 1 2 3 4 5 6 7 8 Z F K Description Y 0 Y Y Measuring method Zirconia method 7YY Power supply 100 to 115V AC 50/60Hz(Standard) 200 to 240V AC 50/60Hz(Standard) 200 to 240V AC 50/60Hz(CE mark) 9 B C Gas inlet/outlet size: Rc1/4 or NPT1/4 Power supply: Rated voltage ;100 to 115V AC or 200 to 240V AC Rated frequency ;50Hz/60Hz Max. rated power;215VA (during power ON) 65VA (during steadystate operation) OUTLINE DIAGRAM 9 10 11 12 13 4 - Gas inlet/outlet size Rc 1/4 NPT 1/4 1 8 (Unit:mm) 17 170 172.5 4 152 OXYGEN ANALYZER US 2 1 AC250V T3.15A 3 4 5 6 7 E F 110 131 141 TEMP. L AC INPUT N OUT ALM 9TH DIGIT 1:RC1/4 8:NPT1/4 48 37 23 ø5 EXTERNAL CONNECTION DIAGRAM 1 2 3 4 5 + AC power supply E 6 INLET OUTLET 7 Output to ZKJ Temperature alarm output 19 ZKJ7 Caution on Safety *Before using this product, be sure to read its instruction manual in advance. International Sales Div Sales Group Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fujielectric.com Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425 http://www.fujielectric.com/products/instruments/ Information in this catalog is subject to change without notice. Printed in Japan