INFRARED GAS ANALYZER FOR STIC GAS DATA SHEET ZSK This infrared gas analyzer (ZSK) is composed of gas analyzer units, oxygen sensor and gas sampling equipment. It is capable of continuous measuring concentration of gases such as NOx, SO2, CO, CO2 and O2 contained in the flue gas generated from various boilers or garbage incinerators. FEATURES 1. Gas concentrations of 5 components measurable simultaneously and continuously The infrared ray method has integrated the measurements of NOx, SO2, CO and CO2 and the gas concentrations of 5 components can be measured simultaneously and continuously with the zirconia oxygen analyzer. 2. A high-sensitivity mass flow sensor is used in the detector unit of infrared method. Due to use of single beam system for measurement, maintenance is easy and an excellent stability is ensured for a long period of time. 3. Space-saving configuration is available with the maintenance from the front structure The main unit is downsized by half of that of the conventional product. The complete maintenance from the front makes it easy to use. 4. The analyzer has a host of capabilities Provided with a variety of functions such as O2 correction output, average value output, auto calibration, auto range selection, alarm, etc. SPECIFICATIONS Standard Specifications • Measuring principle: NOx, SO2, CO, CO2: Non-dispersive infrared absorption (NDIR) method O2; Zirconia system • Measurable component and min./max. measuring range: NOx : 0 to 200ppm/0 to 5000ppm SO2 : 0 to 200ppm/0 to 5000ppm CO : 0 to 200ppm/0 to 5000ppm CO2 : 0 to 10%/0 to 20% O2 : 0 to 10%/0 to 25% • Measuring range: Max. range ratio 1:10 (See code symbols.) • Warm-up time: 4 hours or less after power ON • Analog output signal: 4 to 20mA DC simultaneous output for each (Non-isolated or isolated: as specified by code symbols) • Instantaneous value output (NOx, SO2, CO, CO2, O2) 5 points • Instantaneous O2 correction value (NOx, SO2, CO) 3 points * Provided with O2 • O2 correction average value (NOx, SO2, CO) 3 points * Provided with O2 • Allowable load resistance: 550Ω or less (750Ω or less for insulation output) • Contact output: (1) 1a contact for each (Contact capacity: 250V AC 2A, 30V DC 3A) • Range identification for each component, analyzer unit error, calibration error, during automatic calibration, during maintenance (2) 1c contact for each (Contact capacity: 250V AC 1A, 30V DC 1A) • Instantaneous/concentration value alarm for each component (H, L, HL can be set). EDS3-136 Date Nov. 14, 2007 ZSK • Contact input: No-voltage contact (in increments of 1.5 sec or more) • Auto calibration start, average value reset • No-voltage contact (holding status) • Range selection (First range at contact closure) • Output hold, Remote pump OFF (OFF at contact closure) • Display: LCD with back light • Instantaneous value display (NOx, SO2, CO, CO2, O2) *Provided with O2 • Instantaneous O2 correction value (NOx, SO2, O2) *Provided with O2 • O2 correction average value (NOx, SO2, CO) *Provided with O2 • Each parameter setting (English) • Fluorescent lamp inside the cubicle: Provided as standard • Recorder (option): Paperless recorder (Fuji Electric’s type PHR) See EDS10-74d. • Gas extractor: Electrical heating type (filter built-in) • 40 µm-SUS316 wire gauze filter • Flange JIS 5K 65AFF • Mass: Approx. 9kg (except for gas extractor) • Power supply voltage: 100VAC 50/60Hz • Power consumption: Approx. 100VA • Sampling tube: material and length are as specified by code symbols. SUS316 (Length: 300, 400, 600, 800, 1000mm) Titanium (Length: 600, 800, 1000mm) Sic (Length: 700, 900mm) *SUS316 is used at temperature 800°C or lower. *Titanium is used at temperature 1000ºC or lower. *Sic is used at temperature 1300ºC or lower. • Sampling gas tube: φ10/φ8mm Teflon tube or heating tube (Max. 30m) Heating tube is specified in the following cases (1) Ambient temperature is lower than -5ºC. (2) The tube length is 10m or more at SO2 measurement. (Power supply voltage: 100VAC 50/60Hz, Power consumption: 36VA/m) • Rated operation conditions: • Ambient temperature: –5 to 40ºC (as specified by code symbols) • Ambient humidity: 90% RH or less • Power supply voltage: 100,110,115,200,230V AC ±15% (±10% for fluorescent lamp) (as specified by code symbols) • Frequency: 50 or 60Hz ±0.5Hz • Power consumption: Max. 600VA (except for gas extractor and heating tube) 2 • Storage conditions: • Ambient temperature: –20 to 60ºC (Note that water in the gas conditioner is drained.) • Ambient humidity: 95% RH or less • Dry air: (Necessary when the first range of gas dryer purging for SO2 analyzer is 500ppm or more or when using oil/coal boiler.) Dew point: –20ºC DP or less Pressure: 100kPa to 400kPa Dust, mist: none • Outer dimensions (H x W x D): Indoor type: 1710 × 600 × 490mm Outdoor type: 1780 × 615 × 600mm • Mass: Approx. 200kg (except for standard gas) • Cubicle finish color: Munsell 5Y7/1 (semi-gloss) • Cubicle structure: Indoor use or outdoor use type, single swing front door Plate thickness: 2.3mm standard (both cabinet and door) • Others: Six standard gas (3.4L) cylinders can be accommodated (zero gas (10L) cylinders can be also accommodated). Note) After the warm-up time, variation up to 4 hours should be ±2% FS or less. • Serial No. for measuring method: Pending for type approval Standard Functions Functions O2 correction Contents of functions • Conversion of measured NOx,SO2,and CO gas concentrations into values at standard O2 concentration Correction formula: C = • Auto calibration • • • • • • • Auto zero calibration • • • O2 correction • average value, O2 average • value • CS (21-ON) 21-OS C : Converted concentration CS : Measured gas value of concentration OS : Measured O2 concentration ON : Standard O2 concentration (Oil based fuel 4%, gas based fuel 5%, coal based fuel 6%, refuse incinerator 12%) Setting range: 0 to 19% The result of correction can output as display and 4 to 20mA DC signal. Gas analyzer unit is auto-calibrated. Interval setting range of auto calibration: Variable within 1 to 99 hours (in increments of 1 hour) or 1 to 40 days (in increments of 1 day). Time setting range of auto calibration gas to be introduced: Variable within 60 to 599 seconds (in increments of 1sec) Contact output for auto/manual calibration error is provided when an amount of calibration exceeds 50% FS. Contact output during auto calibration and maintenance: Provided during calibration gas flow, and replacement. Also provided during maintenance. Auto calibration remote start contact input: Calibration starts at opening after short-circuit for 1.5 sec or longer. Standard gas consumption: Approx. 1 year with 3.4 L cylinder in a calibration cycle of 7 days Zero point is calibrated periodically at the predetermined cycle. This cycle is independent on “Auto calibration” cycle. Interval setting range of auto zero calibration: Variable within 1 to 99 hours (in increments of 1 hour) or 1 to 40 days (in increments of 1 day). Setting range of gas flow time: 60 to 900 sec (in increments of 1 sec) NOx, SO2 and CO values are averaged after O2 correction, and the result is indicated and output in 4 to 20 mA DC. Averaging time is settable by key operation at the front of analyzer unit. Setting range: 1 to 59 min, 1 to 4 hour (Shipping in 1H) Functions Contents of functions Remote output • The output signal values are collectivehold ly held according to external contact input. • Output is held during short-circuit. Average value • Resets output and indication of O2 reset input correction average value according to external contact input. • Output and indication are reset by a short-circuit for 1.5 sec or longer. Auto range • Automatically changed from low range selection to high range, and from high range to low range. Allows range to switch from low to high range when 90%FS or less is available in the low range. Allows range to switch from high to low range when 80%FS or less is available in the high range. Remote range • Low or high range is selectable for selection each sample component via external contact input. • Selects high range for open-circuit, and low range for short-circuit. Output for • Discrimination between low and high range identifiranges is output through a contact. cation signal • Low range at contact closure Concentration • Instantaneous value alarm is settable alarm contact for each sample component. output • High, Low, High or Low is settable by key operation at the front of analyzer unit. • Contact output hysteresis is also settable. • Contact is 1C type. Contact output • Contact output is provided when the analyzer unit is abnormal. for analyzer error Temperature • K thermocouple input x 2 (for recorder input signal available at option) Performance • Repeatability: ±0.5% FS • Zero drift: Within ±2.0% FS/week (When auto zero calibration is performed) Within ±2.0% FS/month for O2 analyzer • Span drift: ±2.0% or less/week Within ±2.0% FS/month for O2 analyzer • Linearity: Within ±1.0% FS • Response time: 90% indication after extracting sample gas through the inlet NOx : 120 sec or less SO2 : 240 sec or less CO : 120 sec or less O2, CO2 : 120 sec or less • Sample gas extracting rate: Approx. 2L/min 3 ZSK Standard Requirements for Measuring Gases • Temperature: • Dust: • Pressure: • Component: 60 to 800ºC (standard) Nonstandard: 1000ºC (Material of gas extractor: Titanium) 1300ºC (Material of gas extractor: SiC) 100mg/Nm3 or less −5k to +5kPa (selected by code symbols) SO2 : 500ppm or less NOx : 1000 ppm or less CO2 : 0 to 15% CO : 0 to 2000ppm or less O2 : 1 to 21% Hcl : 500ppm or less Remaining : N2, H2O Installation Conditions • Select a place not subjected to direct sunlight or radiation from a high-temperature object. If such a place cannot be found, a roof or cover should be prepared for protection. • Avoid a place where receives heavy vibration. • Select a place which is clean around the analyzer. SCOPE OF DELIVERY • • • • • Gas analyzer system Specified external drain separator/drain pot Specified gas extractor/probe set Specified gas inlet tube set Standard accessories OPTIONAL ITEMS 1. Standard gas, pressure controller (Type ZSY) 2. Recorder (Fuji Electric’s type PHR as required) 3. Measuring method test for each unit 4. Spare parts for one year (Type ZBN) 5. Waterproof gland at wiring port for outdoor use (A25A) Arrangement number 8641625 6. Anchor bolt 4 CODE SYMBOLS 1 2 3 4 5 6 7 8 ZSK 9 10 11 12 13 1 - Specification Measureing gas component <4th digit> NOX SO2 CO NOX,SO2 NOX,CO NOX,SO2,CO NOX,SO2,CO,CO2 P A B F H L M O2 analyzer Without With With With With 0 4 5 6 C O2 correction value <5th digit> Without 4% (Oil based fuel) 5% (Gas based fuel) 6% (Coal based fuel) 12% (Refuse incinerator) NOX measureing range <6th,7th digit>, ppm YY CE CF CG CY DE DF DG DY EF EG EH EY FG FH FY GH GY HY Without 200/500 200/1000 200/2000 200/Without 250/500 250/1000 250/2000 250/Without 500/1000 500/2000 500/5000 500/Without 1000/2000 1000/5000 1000/Without 2000/5000 2000/Without 5000/Without 1 Modification No. <8th digit> Note1) Refer to SO2 measureing range <9th,10th digit>, ppm page 7. Without Note1) 200/500 200/1000 200/2000 200/Without 250/500 250/1000 250/2000 250/Without 500/1000 500/2000 500/5000 500/Without 1000/2000 1000/5000 1000/Without 2000/5000 2000/Without 5000/Without YY CE CF CG CY DE DF DG DY EF EG EH EY FG FH FY GH GY HY CO measureing range <11th,12th digit>, ppm YY CE CF CG CY DE DF DG DY EF EG EH EY FG FH FY GH GY HY Without 200/500 200/1000 200/2000 200/Without 250/500 250/1000 250/2000 250/Without 500/1000 500/2000 500/5000 500/Without 1000/2000 1000/5000 1000/Without 2000/5000 2000/Without 5000/Without O2 measureing range <13th digit> 0 2 1 Without 25% 10/25% 5 ZSK 1 2 3 4 5 6 7 8 ZSK 1 - 9 10 11 12 13 14 15 16 17 18 19 20 - 21 22 23 24 25 26 Specifications - CO2 measureing range <14th, 15th digit> Without 10/20% 10/Without 20/Without YY MN MY NY Y A B C D E F G H J K L M N P 0 1 2 3 4 5 6 7 0 1 2 3 4 5 6 7 Note2) O2 correction value insulating output <17th digit> (Measuring component) (Insulation output) Each component Without NOX Insulation SO2 Insulation CO Insulation NOX,SO2 Insulation NOX,CO Insulation SO2,CO Insulation NOX,SO2,CO Insulation Note2) O2 correction avarage value insulating output <18th digit> (Measuring component) (Insulation output) Each component Without NOX Insulation SO2 Insulation CO Insulation NOX,SO2 Insulation NOX,CO Insulation SO2,CO Insulation NOX,SO2,CO Insulation Note2) (Sample gas pressure) –1 to +5kPa –3 to +3kPa –5 to +1kPa –1 to +5kPa –3 to +3kPa –5 to +1kPa 1 2 3 4 5 6 Note4) Piping, wiring inlet <21th digit> (Piping) (Wiring) (External gas piping) Upper left Upper left Without Upper left Upper left Without Upper left Upper left 3 inlet on the lower left Upper left Upper left 6 inlet on the lower left Upper left Upper left 3 inlet on the lower left Upper left Upper left 6 inlet on the lower left A B C D E F Note1) (Instrumentation air inlet) Without Upper left Without Without Upper left Upper left Ambient temperature <22th digit> Standard (–5 to 40°C) Cold climate(–10 to 40°C) 2 3 Name plate Standard Standard Standard Standard A B E F Note1) to Note5) Refer to page 7. (External drain separator) <19th digit> Without Without Without With Note3) With Note3) With Note3) Locker structure <20th digit> Indoor structure Outdoor structure 1 2 6 Analog instantaneous insulation output <16th digit> (Measuring component) (Insulation output) Each component Without NOX Insulation SO2 Insulation CO Insulation NOX,SO2 Insulation NOX,CO Insulation NOX,SO2,CO Insulation NOX,SO2,CO,CO2 Insulation NOX,O2 Insulation SO2,O2 Insulation CO,O2 Insulation NOX,SO2,O2 Insulation NOX,CO,O2 Insulation NOX,SO2,CO,O2 Insulation NOX,SO2,CO,CO2,O2 Insulation 0 1 2 3 4 5 6 TAG plate Without With Without With Display language <23th digit> Japanese Japanese English English Recorder <24th digit> Without With (6-point recording: recording data 1) With (6-point recording: recording data 2) With (6-point recording: recording data 3) With (6-point recording: recording data 4) With (6-point recording: recording data 5) With (6-point recording: recording data 6) Note5) Note6) Separately purchase recorder (Type PHR) 1 2 3 4 5 6 7 8 ZSK 9 10 11 12 13 1 - 14 15 16 17 18 19 20 - 21 22 23 24 25 26 - 27 28 29 30 - 31 Contents - A B C D E F G H J K A B C D Note1) If 0-500ppm or higher is selected as SO2 first range for the 9th and 10th digits, or code D is selected for the 26th digit, select code B, E, or F with instrumentation air for the 21st digit. Note2) Only when “Without” is selected for all of the 16 to 19th digits, non-isolated output is allowed. Combination of isolated and non-isolated outputs is not allowed. Note3) Specify this code when the downward inclination of the sample inlet tube from the gas extraction point to the analyzer gas inlet is less than 15°C or when moisture content of the sample gas is higher than 30%. Note4) Special specification for cubicle structure Specified as “Z” at the 20th digit only for cubicle materials, plate thickness, door with window, finished color, special coating film. Note5) The contents to be recorded with a 6-point recorder are assigned and connected as specified in the above table for its delivery. Recorder type: PHR *Specify separately for other than those in the following table. Note6) When 1 to 6 is set for the 24th digit, C to K cannot be selected for the 25th digit. Note7) When the measurement method test or the CO2 analyzer is provided, code 3 is specified. Code Recording data NOx instantaneous value average value O2 correction value SO2 instantaneous value average value O2 correction value CO instantaneous value average value O2 correction value O2 instantaneous value Combustion temperature Dust precipitator temperature 1 2 { { { { { { { { { { { { 3 1 { { { { { { 2 3 { { { { { { { { { { { { { { <Specify when placing an order> (1) Code symbols (2) Written contents when TAG plate is provided Power supply <25th digit> 100V AC 50Hz 100V AC 60Hz 110V AC 50Hz 110V AC 60Hz 115V AC 50Hz 115V AC 60Hz 200V AC 50Hz 200V AC 60Hz 230V AC 50Hz 230V AC 60Hz Note6) Application <26th digit> Refuse, industrial waste incineration, etc. Gas boiler Sludge incineration Coal-oil boiler Note1) Measuring method test <27th digit> Without NOX SO2 CO NOX,SO2 NOx,CO NOX,SO2,CO NOX,O2 SO2,O2 CO,O2 NOX,SO2,O2 NOx,CO,O2 NOX,SO2,CO,O2 Y A B C D E F G H J K L M Zero gas <28th digit> Note7) Instrumentation air Atmospheric air Standard gas ... Separately purchase standard gas (Type ZSY). 1 2 3 Gas extractor <29th digit> Extractor Sampling tube Insertion material length Without Without Without With Without Without With SUS316 300mm With SUS316 400mm With SUS316 600mm With SUS316 800mm With SUS316 1000mm With SUS316 1200mm With SUS316 1500mm With SUS316 2000mm With Titanium 600mm With Titanium 800mm With Titanium 1000mm With SiC 700mm With SiC 900mm Y 1 A B C E G H J K P Q R D F Sample inlet tube type Without φ10/φ8mm Teflon tube φ10/φ8mm Teflon tube φ10/φ8mm Teflon tube φ10/φ8mm Teflon tube φ10/φ8mm Teflon tube φ10/φ8mm Teflon tube φ10/φ8mm Teflon tube Heating tube Heating tube Heating tube Heating tube Heating tube Y A B C D E F G H J K L M Z Sampling point temperature — — 800°C or lower 800°C or lower 800°C or lower 800°C or lower 800°C or lower 800°C or lower 800°C or lower 800°C or lower 1000°C or lower 1000°C or lower 1000°C or lower 1300°C or lower 1300°C or lower Length <30th digit> Without 5m 10m 15m 20m 25m 30m 50m 10m 15m 20m 25m 30m Non standard specification <31th digit> Other specification 7 ZSK GAS SAMPLING SYSTEM DIAGRAM 1 (Refuse/industrial waste disposable incineration, SO2 analyzer is not included) (1) Gas extractor e or ° 15 Air inlet *1 (7) Flow checker rm o (10) 3-way solenoid valve 2-way solenoid valve (3) Gas aspirator Sample gas inlet Atmospheric air (2) Gas conditioner *1 Needle valve (6) Membrane filter Infrared gas analyzer CO (12) NO2/NO converter 30kPa 0.6L/min (4) Electronic gas cooler (5) Solenoid valve Infrared gas analyzer NO ZERO NO CO O2 Pressure adjuster (9) Zirconia type O2 analyzer Air Drain NO /N2 CO /N2 O2 /N2 (Air) (8) Calibration gas Filter regulator 30kPa Instrumentation air inlet Exhaust Functions of Individual Components (1) Gas extractor: Gas extraction, with heating type stainless steel filter having a standard diameter of 40µm (2) Gas conditioner: Removes drain, mist and dust, and monitors the gas pressure. (3) Gas aspirator: Aspirates sample gas (Flow rate of sample gas: Approx. 2L/min) (4) Electronic gas cooler: Dries the moisture in the sample gas. (5) Solenoid valve: Used for introducing calibration gas. (6) Membrane filter: PTFE filter used to eliminate fine dust particles and permit monitoring of dust adhering condition on the gas analyzer. (7) Flow checker: Controls and monitors the sample gas flow rate (it can be controlled by the separate needle valve.) (8) Standard gas: Reference gas used for calibrating zero and span of the analyzer. Up to 6 gases (Zero gas air, span gas NOx, SO2, CO, CO2 and O2) can be used. Used for measuring the oxygen con(9) O2 sensor: centration (0 to 25%) in sample gas. (10) Atmospheric air solenoid valve: Can be built in for using the atmospheric air instead of standard air. 8 (11) Switching box: 6 power ON-OFF switches of the following equipments are built in. • Gas aspirator • Ventilator • Fluorescent lamp and service outlet (Max. 2A) • Electronic gas cooler, built-in recorder, and isolation signal converter • Converter (at NOx measurement) • Heater for the gas conditioner (12) Converter: Added to NOx analyzer. A special catalyst material for efficient conversion of NO2 gas to NO is used. (13) Pressure control valve: Pressure adjustor to keep the sample gas pressure at a fixed level. (14) Gas dryer: Semi-permeable membrane type dehumidifier to dry the moisture in the sample gas to dew point -15ºC or less. (15) Mist catcher: Removes sulfate mist in the sample gas. Replaced every 4 months when SO3 concentration value is 30ppm. Added when SO2 value is more than 0 to 500ppm or when oil/coal boiler is used. GAS SAMPLING SYSTEM DIAGRAM 2 (SO2 measurement (first range 500 ppm or less), gas boiler or sludge incineration) (1) Gas extractor re Air inlet 2-way solenoid valve *1 (7) Flow checker Infrared gas analyzer SO2,CO2 o ° 15 rm o (10) 3-way solenoid valve *3 (3) Gas aspirator Sample gas inlet 50kPa 0.6L/min (6) Membrane filter (12) NO2/NO converter (14) Gas dryer 30kPa 0.6L/min Atmospheric air (2) Gas conditioner *1 (4) Electronic gas cooler Infrared gas analyzer NO,CO (5) Solenoid valve (13) Pressure control valve ZERO NO CO SO2 CO2 O2 Pressure adjuster (9) Zirconia type O2 analyzer *3 Air NO /N2 SO2 /N2 CO /N2 CO2 /N2 O2 /N2 (Air) (8) Calibration gas Drain Filter regulator 30kPa Exhaust Instrumentation air inlet GAS SAMPLING SYSTEM DIAGRAM 3 (SO2 measurement (first range 500 ppm or higher), oil/coal boiler) (1) Gas extractor e Air inlet or ° 15 or *1 2-way solenoid valve *3 m (10) 3-way solenoid valve (7) Flow checker RE2 Flow meter 50kPa (1L/min) 0.6L/min (6) Membrane filter (12) NO2/NO converter (14) Gas dryer Sample gas inlet (3) Gas aspirator Atmospheric air (2) Gas conditioner *1 30kPa 0.6L/min (15) Mist catcher (4) Electronic gas cooler Infrared gas analyzer NO,CO (5) Solenoid valve (13) Pressure control valve ZERO NO SO2 CO CO2 O2 Pressure adjuster Air Drain Infrared gas analyzer SO2,CO2 *3 NO /N2 SO2 /N2 CO /N2 CO2 /N2 O2 /N2 (Air) (9) Zirconia type O2 analyzer (8) Calibration gas Filter regulator 30kPa Instrumentation air inlet Exhaust 9 ZSK EXTERNAL TERMINAL DIAGRAM Terminal block TB1 Grounding terminal Main power supply (AC power) 1 Grounding for internal devices (do not wire.) 3 Power supply for gas extractor (M6 screw) Power supply for heating tube 2 4 5 6 7 8 9 Temperature input 1 (Option) 10 Temperature input 2 (Option) 12 11 (M4 screw) Terminal block 1 <TN1> Ch1 Output Ch2 Output Ch3 Output Ch4 Output Ch5 Output Ch6 Output + 1 13 – + 2 14 3 15 – + 4 16 5 17 – + 6 18 7 19 – + 8 20 9 21 – + 10 22 11 23 – 12 24 + – + – + – + – + Terminal block 2 <TN2> Ch7 Output Remote hold input Ch8 Output Average value reset input Ch9 Output Auto calibration remote start input Ch10 Output Auto zero calibration start input Ch11 Output Pump ON/OFF input – 1 11 2 12 3 13 4 14 5 15 6 16 7 17 8 18 9 19 10 20 Remote range changeover input (Ch1) Remote range changeover input (Ch2) Remote range changeover input (Ch3) Remote range changeover input (Ch4) Remote range changeover input (Ch5) (M3.5 screw) For internal connection (M3.5 screw) Terminal block 3 <TN3> For internal connection 1 11 2 12 Instrument error output 3 13 4 14 Calibration error output 5 15 6 16 7 17 8 18 9 19 10 20 Output during auto calibration Output during maintenance (M3.5 screw) 10 Range identification output (Ch1) Range identification output (Ch2) Range identification output (Ch3) For internal connection Alarm output (Ch1) Alarm output (Ch2) Range identification output (Ch4) Range identification output (Ch5) Alarm output (Ch3) Terminal block 4 <TN4> 1 11 2 12 3 13 4 14 5 15 6 16 7 17 8 18 9 19 10 20 (M3.5 screw) For internal connection Alarm output (Ch4) Alarm output (Ch5) Alarm output (Ch6) CORRESPONDENCE BETWEEN MEASUREMENT CHANNEL AND MEASURED VALUE The contents of each output signal which corresponds to the code symbols are listed below. Code symbols Contents 4th digit 5th digit P 0 Ch1: NOX A 0 Ch1: SO2 B 0 Ch1: CO F 0 Ch1: NOX, Ch2: SO2 H 0 Ch1: NOX, Ch2: CO L 0 Ch1: NOX, Ch2: SO2, Ch3: CO M 0 Ch1: NOX, Ch2: SO2, Ch3: CO2, Ch4: CO P 4 to C Ch1: NOX, Ch2: O2, Ch3: Correction NOX, Ch4: Correction NOX Average A 4 to C Ch1: SO2, Ch2: O2, Ch3: Correction SO2, Ch4: Correction SO2 Average B 4 to C Ch1: CO, Ch2: O2, Ch3: Correction CO, Ch4: Correction CO Average F 4 to C Ch1: NOX, Ch2: SO2, Ch3: O2, Ch4: Correction NOX, Ch5: Correction SO2, Ch6: Correction NOX Average, Ch7: Correction SO2 Average H 4 to C Ch1: NOX, Ch2: CO, Ch3: O2, Ch4: Correction NOX, Ch5: Correction CO, Ch6: Correction NOX Average, Ch7: Correction CO Average L 4 to C Ch1: NOX, Ch2: SO2, Ch3: CO, Ch4: O2, Ch5: Correction NOX, Ch6: Correction SO2, Ch7: Correction CO, Ch8: Correction NOX Average, Ch9: Correction SO2 Average, Ch10: Correction CO Average M 4 to C Ch1: NOX, Ch2: SO2, Ch3: CO2, Ch4: CO, Ch5: O2, Ch6: Correction NOX, Ch7: Correction SO2, Ch8: Correction CO, Ch9: Correction NOX Average, Ch10: Correction SO2 Average, Ch11: Correction CO Average STANDARD ACCESSORIES No 1 Name Q’ty Remarks Teflon filter for membrane filter/spare (Teflon) 4 Provided with SO2 analyzer Filter paper for membrane filter (25 sheets) /spare (Glass fiber) 1 Not provided with SO2 analyzer 2 Joint for standard gas Rc1/4−φ6mm 1 3 Hose band for fixing standard gas cylinder 1 4 Toalon tube for standard gas connection 1m φ9/φ5mm 1 5 Polyethylene tube for standard gas connection 6m φ6/φ4mm 1 6 Anchor bolt for installing locker (Option) M12×160×50 4 7 Water bottle for injection 1 8 Flange packing for gas extractor 1 9 Mounting bolt and nut for gas extractor (M12×60mm) 1 10 Support fixture for heating tube 1 11 Instruction manual 1 Provided with gas extractor Provided with heating tube 11 ZSK SPARE PARTS FOR ONE YEAR • • • • • • • • • • • • • • • • • • Teflon filter for membrane filter (4 sheets) × 1 (Note 1) O-ring for membrane filter (P49) × 2 O-ring for membrane filter (P3) × 2 Filter element for mist filter × 2 O-ring for mist filter (G65) × 2 Diaphragm for gas aspirator × 1 Gas aspirator valve × 1 Fuse (2A) × 2 Fuse (3.2A) × 4 Capillary for 50kPa/0.6L ×1 ...............Added in cases of SO2 measurement (first range 0 – 550ppm or higher), gas boiler, sludge incineration, oil/coil boiler O-ring for gas extractor (G45) × 1 Added when O-ring for gas extractor (G50) × 1 gas extractor is Wire gauze filter for gas extractor × 1 provided Wire gauze filter packing for gas extractor × 1 NO2/NO converter catalyst × 1 Added when NOx Glass wool for above analyzer is provided Joint for above × 2 Mist catcher × 3 ........ Added in cases of SO2 measurement (first range 500ppm or higher) or oil/coal boiler Note1) Provided with filter paper for membrane filter (25 sheets) x 1 for other than SO2 analyzer. CODE SYMBOLS FOR SPARE PARTS FOR ONE YEAR 1 2 3 4 5 6 7 8 Z BN1 S 2 K 1 0 1 2 3 A B C D E F G H 12 Contents (Applications) Refuse incinerator, gas boiler Oil/coal boiler (Gas extractor) (NOX analyzer) (SO2 analyzer) Without Without Without With Without Without Without With Without With With Without Without Without With (500ppm or less) With Without With (500ppm or less) Without With With (500ppm or less) With With With (500ppm or less) Without Without With (500ppm or more) With Without With (500ppm or more) Without With With (500ppm or more) With With With (500ppm or more) CODE SYMBOLS FOR STANDARD GAS/PRESSURE ADJUSTER 1 2 3 4 5 6 7 8 ZSY 9 10 11 2 - Contents NOX measureing 1st range (4th digit) Without 200ppm 250ppm 500ppm 1000ppm 2000ppm 5000ppm 0 2 3 4 5 6 7 SO2 measureing 1st range (5th digit) Without 200ppm 250ppm 500ppm 1000ppm 2000ppm 5000ppm 0 2 3 4 5 6 7 CO measureing 1st range (6th digit) Without 200ppm 250ppm 500ppm 1000ppm 2000ppm 5000ppm 0 2 3 4 5 6 7 CO2 measureing 1st range (7th digit) Without 10% 20% O2 span gas (9th digit) Without 1.8 to 2%O2/N2 Y B C 0 1 Zero gas (10th digit) Without Air cylinder 3.4L (Not Tested) Air cylinder 3.4L (Tested) Air cylinder 10L (Not tested)* Air cylinder 10L (Tested)* Y A B E F Y A B C D E F G H J K L M Tested (11th digit) Without NOX analyzer SO2 analyzer CO analyzer NOX,SO2 analyzer NOX,CO analyzer NOX,SO2,CO analyzer NOX,O2 analyzer SO2,O2 analyzer CO,O2 analyzer NOX,SO2,O2 analyzer NOX,CO,O2 analyzer NOX,SO2,CO,O2 analyzer *Note) Range 0 to 500pm is included in zero gas, and 10L cylinder of code E, F is recommended to use when you perform auto zero calibration. 13 ZSK OUTLINE DIAGRAM (Unit : mm) <Gas extractor> φ155 Rc1/2 gas outlet UP Filter replacement handle Cover (750) 357 Heating pipe support (Mounting for heating tube specifications) 212 293 Terminal box Wiring port G1/2 (76) Rc 3/4 Main nameplate Packing Flange JIS 5K 65AFF SUS316 sampling tube 20A 4-φ15 φ130 φ155 7th digit code Y: None A (300) G (1000) B (400) H (1200) C (600) J (1500) E (800) K (2000) Z: as specified Packing SiC sampling tube 7th digit code Y (not-supplied) D (700) F (900) φ30 Gas inlet Gas inlet ø27 Electric heating type (applicable to high-temperature) <Drain separator> <Drain pot> φ10/φ8 teflon tube Drain exhaust port Joint (teflon) Gas inlet 10 60 30 142 Approx. 96 70 φ67 90 Drain 14 Rc1/2 Approx. 34 555 Approx. 1550 2-φ9 hole φ60 φ39 hole Gas outlet 5 Rc1/4 40 (35) 2-Rc1/2 φ22 Mass: Approx. 0.8kg Mass: Approx. 0.8kg (Unit : mm) Sample gas inlet Rc 3/8 Instrumentation air inlet Rc 1/4 615 600 30° 70 90 Air sunction port Rc 1/2 70 OUTLINE DIAGRAM 4-φ34 grommet 1380 1710 1780 (for outdoor use) (For outdoor version, must be protected from water by wire brushing, etc.) 25 865 490 100 120 Drain Rc 1/2 Exhaust Rc 1/4 450 480 550 600 Left side view Front view Anchor plan/door switching 4-φ15 hole Anchor bolt (option) (4-M12×160×50) 550 35 450 32 Rear Front 15 ZSK Caution on Safety *Before using this product, be sure to read its instruction manual in advance. Head Office Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fesys.co.jp/eng Instrumentation Div. International Sales Dept. No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187 http://www.fic-net.jp/eng Information in this catalog is subject to change without notice. Printed in Japan