MEANWELL RPT-75B

R P T- 7 5
75W Triple Output Medical Type
series
Features :
Universal AC input / Full range
Low leakage current<200uA
Protections: Short circuit / Overload / Over voltage
Free air convection for rated power and 23.5CFM forced air
convection for peak load
UL60601-1/IEC60601-1/EN60601-1 medical safety approved
Fixed switching frequency at 65KHz
3 years warranty
EN60601-1
IEC60601-1
SPECIFICATION
OUTPUT
INPUT
CH1
CH2
CH3
CH1
CH2
CH3
CH1
CH2
CH3
DC VOLTAGE
5V
12V
-5V
5V
12V
-12V
5V
15V
-15V
RATED CURRENT
6A
3A
0.5A
6A
3A
0.5A
6A
2.3A
0.5A
CURRENT RANGE
0.6 ~ 8A
0.2 ~ 4A
0.1 ~ 1A
0.6 ~ 8A
0.2 ~ 4A
0.1 ~ 1A
0.6 ~ 8A
0.1 ~ 3A
0.1 ~ 1A
RATED POWER
68.5W
PEAK LOAD (23.5CFM)
93W
150mVp-p
150mVp-p
120mVp-p
80mVp-p
100W
120mVp-p
120mVp-p
80mVp-p
5.0%
2.0%
6.0%
5.0%
2.0%
8.0%
5.0%
LINE REGULATION
0.5%
1.0%
1.0%
0.5%
1.0%
1.0%
0.5%
1.0%
1.0%
LOAD REGULATION
1.5%
3.0%
1.0%
1.5%
3.0%
1.0%
1.5%
3.0%
1.0%
SETUP, RISE TIME
500ms, 30ms/230VAC
HOLD UP TIME (Typ.)
80ms/230VAC
VOLTAGE RANGE
90 ~ 264VAC
FREQUENCY RANGE
EFFICIENCY(Typ.)
47 ~ 63Hz
76%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.5A/115VAC
1A/230VAC
COLD START 25A/115VAC
50A/230VAC
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
WITHSTAND VOLTAGE
500ms, 30ms/115VAC at full load
20ms/115VAC at full load
127 ~ 370VDC
77%
77%
Earth leakage current <200uA / 264VAC, Patient leakage current <100uA/264VAC
140 ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
CH1: 5.75 ~ 6.75V
Protection type : Shut down o/p voltage, re-power on to recover
-20 ~ +70
(Refer to output load derating curve)
20 ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH
0.03%/ (0 ~ 45 )
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
UL60601-1, TUV EN60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms/500VDC
EMI CONDUCTION & RADIATION Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B
Compliance to EN61000-3-2,-3
HARMONIC CURRENT
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61000-6-2, EN61204-3, heavy industry level, EN61204-3 medical level, criteria A
EMS IMMUNITY
ISOLATION RESISTANCE
MTBF
NOTE
100W
120mVp-p
CH1:4.75 ~ 5.5V
VOLTAGE TOLERANCE Note.3 2.0%
6.0%
OVER VOLTAGE
OTHERS
72W
VOLTAGE ADJ. RANGE
OVERLOAD
(Note 4)
72W
RIPPLE & NOISE (max.) Note.2 80mVp-p
PROTECTION
EMC
RPT-75C
OUTPUT NUMBER
LEAKAGE CURRENT
SAFETY &
RPT-75B
RPT-75A
MODEL
DIMENSION
PACKING
521.2K hrs min. MIL-HDBK-217F (25
127*76.2*31mm (L*W*H)
0.26Kg; 63pcs/17.4Kg/1.35CUFT
)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
5. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
File Name:RPT-75-SPEC 2008-03-25
R P T- 7 5
75W Triple Output Medical Type
series
Features :
Universal AC input / Full range
Low leakage current<200uA
Protections: Short circuit / Overload / Over voltage
Free air convection for rated power and 23.5CFM forced air
convection for peak load
UL60601-1/IEC60601-1/EN60601-1 medical safety approved
Fixed switching frequency at 65KHz
3 years warranty
EN60601-1
IEC60601-1
SPECIFICATION
RPT-75D
MODEL
OUTPUT
INPUT
CH1
CH2
CH3
CH1
CH2
CH3
DC VOLTAGE
5V
24V
12V
3.3V
5V
12V
RATED CURRENT
5A
1.5A
1A
6A
6A
1A
CURRENT RANGE
0.6 ~ 7A
0.1 ~ 2A
0.1 ~ 1A
0.7 ~ 7A
0 ~ 8A
0 ~ 1.5A
RATED POWER
73W
PEAK LOAD (23.5CFM)
95W
120mVp-p
120mVp-p
CH1:4.75 ~ 5.5V
VOLTAGE TOLERANCE Note.3 2.0%
8.0%
8.0%
6.0%
+10,-6%
0.5%
1.0%
1.0%
1.0%
1.0%
1.5%
LOAD REGULATION
1.5%
3.0%
3.0%
+3,-4%
+5,-4%
6.0%
SETUP, RISE TIME
500ms, 30ms/230VAC
HOLD UP TIME (Typ.)
80ms/230VAC
VOLTAGE RANGE
90 ~ 264VAC
FREQUENCY RANGE
EFFICIENCY(Typ.)
47 ~ 63Hz
79%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.5A/115VAC
1A/230VAC
COLD START 25A/115VAC
50A/230VAC
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
WITHSTAND VOLTAGE
500ms, 30ms/115VAC at full load
20ms/115VAC at full load
127 ~ 370VDC
74%
Earth leakage current <200uA / 264VAC, Patient leakage current <100uA/264VAC
140 ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
CH1: 5.75 ~ 6.75V
CH1: 3.8 ~ 4.45V
Protection type : Shut down o/p voltage, re-power on to recover
-20 ~ +70
(Refer to output load derating curve)
20 ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH
0.03%/ (0 ~ 45 )
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
UL60601-1, TUV EN60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms/500VDC
EMI CONDUCTION & RADIATION Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B
Compliance to EN61000-3-2,-3
HARMONIC CURRENT
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61000-6-2, EN61204-3 heavy industry level, EN61204-3 medical level, criteria A
EMS IMMUNITY
ISOLATION RESISTANCE
MTBF
NOTE
80mVp-p
LINE REGULATION
OVER VOLTAGE
OTHERS
120mVp-p
-----4.0%
OVERLOAD
(Note 4)
81W
200mVp-p
VOLTAGE ADJ. RANGE
PROTECTION
EMC
61.8W
RIPPLE & NOISE (max.) Note.2 80mVp-p
LEAKAGE CURRENT
SAFETY &
RPT-7503
OUTPUT NUMBER
DIMENSION
PACKING
521.2K hrs min. MIL-HDBK-217F (25
127*76.2*31mm (L*W*H)
0.26Kg; 63pcs/17.4Kg/1.35CUFT
)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
5. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
File Name:RPT-75-SPEC 2008-03-25
R P T- 7 5
75W Triple Output Medical Type
Mechanical Specification
series
Unit:mm
FAN
5.7
15cm
23.5CFM min.
CN2
1
2
3
4
5
6
7
8
3
2
1
64.75
76.2
CN1
4115.8
5.6
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No.
1
2
3
Assignment
AC/N
No Pin
AC/L
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
DC Output Connector (CN2) : JST B8P-VH or equivalent
3.8
Pin No.
1,2
3,4,5
6,7
8
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
3max.
31
127
Assignment
V1
COM
V2
V3
Block Diagram
RECTIFIERS
&
FILTER
EMI
FILTER
I/P
POWER
SWITCHING
V3
RECTIFIERS
&
FILTER
+V2
RECTIFIERS
&
FILTER
+V1
COM
DETECTION
CIRCUIT
FG
CONTROL
O.L.P.
fosc : 65KHz
RECTIFIERS
&
FILTER
O.V.P.
Derating Curve
Output Derating VS Input Voltage
Ta=25
100
133
23.5CFM Forced air Cooling
100
90
80
Convection Cooling
70
LOAD (%)
LOAD (%)
80
50
20
-20
0
10
20
30
40
50
AMBIENT TEMPERATURE ( )
60
70
(HORIZONTAL)
60
50
40
90
95
100
115
120
140
160
180
200
220
240
264
INPUT VOLTAGE (VAC) 60Hz
File Name:RPT-75-SPEC 2008-03-25