POWERBOX PMP55-32

MEDICAL & MILITARY
PMP55 Series
4 0 - 5 5 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Features
• Input surge current protection.
• Overvoltage protection.
• Overcurrent protection.
• Short-circuit protection with auto recovery.
• 100% burn-in at full rated load.
Specifications
OPERATING
INPUT
Input Voltage
85-264Vac
Frequency
47Hz–67Hz
Input Current
1.50A (rms) for 115Vac.
0.90A (rms) for 230Vac.
MTBF
450,000 hrs @ 25°C
Efficiency
65-70% minimum.
ENVIRONMENTAL
Operating Temperature
0°C to 70°C
Leakage current
90µA max. @ 115Vac. 60Hz.
150µA max. @ 230Vac. 50Hz.
Derating
Derate to 50% at 70°C
Isolation
Input – Output: 4000Vac
Input – Ground: 1500Vac
Output – Ground: 500Vac
Storage Temperature
-40°C to 85°C
Relative Humidity
5%–95% non-condensing
Inrush Current
15A @ 115V, 30A @ 230V
OUTPUT
STANDARDS AND APPROVALS
Safety standards
UL2601-1, CSA C22.2, TUV EN60601-1
C-Tick
AS/NZS CISPR11 Group 1, Class A
EMC standards
EN60601-1-2: 2001,EN61000-3-2, -3,
EN61000-4-2, -3, -4, -5, -6, -8, -11
Output Voltage
See table
Output Current
See table
Ripple & Noise
1% peak to peak
EMI standards
EN55011 Class B, FCC Class B, VCCI Class B
Line Regulation
±0.5% maximum at full load.
MEPS compliance
TBC
Load Regulation
See table
MECHANICAL
Transient Response
Max. deviation 4% recovering to 1% within 500µs
after 25% load charge
Dimensions
178x95x60mm
Weight
0.7kgs approx.
Overvoltage Protection
Provided on V1 only. Set at 110%–132% of
nominal.
Overcurrent Protection
All outputs protected for short circuit conditions
Temperature Coefficient
All outputs ±0.04% per °C.
Holdup Time
Typically 10ms at 110VAC.
Your dependable power partner – www.powerbox.com.au
361
MEDICAL & MILITARY
PMP55 Series
4 0 - 5 5 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Selection Table
MAX.
MODEL
OUTPUT #1
OUTPUT #2
VNOM
IMIN
IMAX
OUTPUT #3
TOL
VNOM
IMIN
IMAX
OUTPUT
NUMBER
VNOM
IMIN
IMAX
TOL
TOL
PMP55-10
5–6V
0A
8.0A
5%
40W
PMP55-10-1
6–8V
0A
7.5A
5%
45W
PMP55-11
9–11V
0A
5.0A
5%
50W
PMP55-12
11–13V
0A
4.6A
5%
55W
PMP55-13
13–17V
0A
3.7A
3%
55W
PMP55-13-1
17–21V
0A
3.1A
3%
55W
PMP55-14
21–27V
0A
2.3A
3%
55W
PMP55-16
27–33V
0A
1.9A
3%
55W
PMP55-17
33–39V
0A
1.6A
3%
55W
PMP55-18
48V
0A
1.2A
3%
55W
PMP55-23
+5V
1A
5.0A
5%
+12V
0.5A
3.0A
5%
45W
PMP55-24
+5V
1A
5.0A
5%
+15V
0.4A
2.5A
5%
45W
PMP55-25
+5V
1A
5.0A
5%
+24V
0.3A
2.0A
5%
PMP55-31
+5V
1A
5.0A
5%
+12V
0.5A
3.0A
5%
-12V
0.1A
0.5A
10%
45W
PMP55-32
+5V
1A
5.0A
5%
+15V
0.4A
2.5A
5%
-15V
0.1A
0.5A
10%
45W
45W
NOTES: The output voltage of a multiple output model may go outside of the stated tolerance when an output load current is of out of stated limits.
All models may be operated at no load without damage
Pin Chart
MODEL/PIN
PMP55-10
PMP55-10-1
PMP55-11
PMP55-12
PMP55-13
PMP55-23
PMP55-24
PMP55-13-1
PMP55-14
PMP55-16
PMP55-17
PMP55-18
PMP55-25
PMP55-.31
PMP55-13
Technical Illustration
362
POWER
1
2
3
4
5
Return
Return
Output
Return
Output
Common
Return
Common
Return
Output
1
Common
Return
Output
2
Common
Return
Common
Return
Output
1
Output
3
Output
2