SHEATH TYPE PLATINUM RESISTANCE THERMOMETER BULB DATA SHEET FTH This resistance thermometer bulb is firmly sealed with highpurity oxidized magnesium (insulating material) in a fine metallic protective tube. Unlike ordinary resistance thermometer bulbs, it provides excellent sensitivity and vibration resistance, and is designed to permit easy bending work. FEATURES SPECIFICATIONS • Applied standard: JIS C 1604 -1997 • Resistance element: Nominal resistance; Pt100 Class; A, B Rated current; 2mA Allowable error due to temperature; Class Allowable error A ±(0.15+0.002t)˚C B ±(0.3+0.005t)˚C Note:1) "t" is measured temperature expressed in (˚C) independent of +/– symbols. 2) Allowable error is the maximum limit of resistance value of resistance thermometer bulb obtained by subtracting measured temperature from the temperature calculated according to the reference resistance table. • Protective tube material: SUS316 • Sheath outside diameter: φ4.8mm or φ6.4mm • Mounting method: Insertion type or screw-in type (with mounting fitting) Mounting fitting screw size; R1/8(PT1/8) or R1/4(PT1/4) • Insertion length: 100 to 9000mm(specified in the unit of 10mm) • Terminal structure: Extension cable type, connector type, exposed terminal type, extension cable (with connector) type • Operating temperature range: –200 to +500˚C • Element structure: Single-core, 3-wire system • Minimum bending radius: 5 times the sheath outside diameter, not bent up to 100mm from tip • Response characteristic: See next page. Code symbols F TH W Description 5 - Element Single-core, 3-wire system (JPt100. class B) Single-core, 3-wire system (JPt100, class A) 1 7 Sheath outside diameter φ4.8mm φ6.4mm 3 4 Sheath material SUS316 W Mounting method Insertion type With mounting fitting R1/8(PT1/8) With mounting fitting R1/4(PT1/4) 0 1 2 Insertion length Enter insertion length in cm unit (minimum length; 10cm, maximum length; 900cm). Example: 055 / 55cm 0 1 0 ∼ 1. Resistance thermometer bulb is very small in size, providing excellent sensitivity. 2. Adoption of integrated structure has improved the vibration resisting characteristic. 3. Easy bending work allows fabricating a long-sized resistance thermometer bulb for easy measurement even in a place where ordinary type resistance thermometer bulb cannot be used. 4. Various types of terminals such as extension cable type, connector type, exposed type, terminal box type, etc., are also available for selection according to applications. 5. Measurement range covers –200 to +500˚C. 9 0 0 Terminal structure Extension cable type Exposed terminal type Terminal box type Connector type Extension cable(with connector)type L D G P Q 0 1 3 Cable length None (12th code D, G, P) With 1m cable With 3m cable (12th code L, Q) EDS2-24a Date Jun. 14, 2002 FTH < Resistance bulb time-lag > 55 L (as specified) 90 Mounting fitting shown at lower left is supplied as specified. φ4.8 sheath type 80 φ6.4 sheath type 0 FTH 1 3 W 1 5 7 4 2 70 D0 Ordinary type, φ12 SUS304 protective tube 50 74 L (as specified) Mounting fitting shown at lower left is supplied as specified. G3/8 (PF3/8) 30 20 ø7.5 0 FTH 1 3 W 1 5 7 4 2 10 0 1 2 T=15sec G0 3 T=1min10sec 70 Time lapse (min) L (as specified) ø8 0 ø4.8 or ø6.4 (as specified) 40 (82) Temperature (˚C) 63.2 60 ø4.8 or ø6.4 (as specified) 50 100 Resistance thermometer bulb time-lag largely varies with measurement conditions. The above shows an example of measurement in 100°C warm water with sheath type resistance thermometer bulb and ordinary type resistance thermometer bulb. ø4.8 or ø6.4 (as specified) Note: "T" :Time constant ø17 T=8sec Connector 0 FTH 1 3 W 1 5 7 4 2 P0 Outline diagram (Unit: mm): L (as specified) 40 Cabtyre cable PVC cabtyre cable Mounting fitting (supplied as specified) R1/8 (PT1/8) or R1/4 (PT1/4) 0 FTH 1 3 W 1 5 7 4 2 ø10 ø8 L (as specified) H (as specified) 70 ø6.5 ø17 40 ø4.8 or ø6.4 (as specified) 1000 ø10 ø6.5 ø4.8 or ø6.4 (as specified) ø4.3 hole Connector 0 FTH 1 3 W 1 5 7 4 2 Q 1 3 L1 3 Caution on Safety *Before using this product, be sure to read its instruction manual in advance. Head Office Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fesys.co.jp/eng Instrumentation Div. International Sales Dept. No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187 http://www.fic-net.jp/eng Information in this catalog is subject to change without notice. Printed in Japan