FUJI EDS6-132F

HYBRID ULTRASONIC FLOWMETER <Duosonics>
(Pulse Doppler method + Transit Time method)
DATA SHEET
This meter is the world’s first non-intrusive type ultrasonic
flowmeter utilizing Pulse Doppler method along with Transit
Time method. It enables measurement of velocity profile
directly resulting in high accuracy. This makes it suitable for
undeveloped flow and for short straight pipes. Pulse Doppler method requires reflectors in the liquid and is utilized on
opaque liquids while Transit Time method requires ultrasound
penetration and is ideal for clean liquids. The new hybrid
technology utilizes both methods in a complementary fashion thus enabling a wider range of applications than it is possible now. In addition, thanks to Fuji’s new state-of-the-art
algorithm, either method can be automatically switched to
accommodate for varying fluid conditions such as concentration of particles and/or air bubbles and flow velocity.
FSH, FSW, FLY
Flow transmitter
(FSH)
FEATURES
1. Automatic switchover function between Pulse Doppler
method utilizing ultrasound reflection and Transit Time
method utilizing ultrasound penetration
• Applicable to various kinds of liquids with/without air
bubbles and/or solid particles
• Applicable to liquid flow that changes in nature frequently or periodically
2. High-accuracy non-intrusive (non-contact) volumetric
flow rate measurement of liquid flow in closed pipes.
• Accuracy of 0.5% to 1% (depending on the measuring
mode and pipe size)
• Clamp-on sensor
3. Direct measurement of velocity profile in case of Pulse
Doppler method
• High accuracy of 0.5% to 1.5% (correction coefficient
unnecessary)
• Applicable to undeveloped flow (short straight pipe)
• Applicable to flow hovering in the transitional region
between laminar flow and turbulent flow
4. High speed response: 0.2sec (pulse Doppler method)/
0.5sec (transit time method)
5. Real time monitoring of velocity profile by PC in case
of Pulse Doppler method (option)
6. Dual-path option improves performance
SPECIFICATIONS
Operational specifications
System configuration:
The system is composed of one/two detectors (Model: FSW) and one Flow transmitter
(Model: FSH), realizing single-path/two-path
measurement.
Hybrid mode or transit time mode is selectable.
In case of hybrid mode, ether Pulse Doppler
method or transit time method is automatically selected depending on conditions of
measured liquid and magnitude of velocity.
Detector
(FSW)
Application: Uniform liquid in which ultrasonic waves can
propagate.
Air bubble quantity: Pulse Doppler method: 0.02 to
15% of volume at 1 m/s
Transit time method: 0 to 12%
of volume at 1 m/s
Fluid temperature: -40 to +100ºC (FSW12), -40 to
80ºC (FSW21, FSW40,FSW50)
Type of flow:
Pulse Doppler method: axisymmetric flow in a filled pipe.
Transit time method: welldeveloped turbulent or laminar
flow in a filled pipe.
Applicable flow pipe:
Material:
Plastics (PVC, FRP, etc.) or
Metals (carbon steel, SS, copper, aluminum, etc.)
Pipe size: 40 to 1000 mm
(inside
diameter)
Liner: Tar epoxy, mortar, etc.
Straight pipe length: Typically 10D for upstream and
5D for downstream. Refer to
JEMIS-032 in detail.
(Note) JEMIS: Japan Electric
Measuring Instruments
Manufactures' Associations Standard
Velocity:
Hybrid mode: 0 to ±0.3 --- ±Maximum Velocity (depending on pipe diameter) (Note) Maximum measurement range in Hybrid mode
(see page 4)
Transit time mode: 0 to ±0.3 --- ±32 m/s
Power supply:
100 to 240 Vac+10%/-15%, 50/60Hz or 20 to
30 Vdc
EDS6-132f
Date
Apr. 1, 2011
FSH, FSW, FLY
Signal cable: Single-path system :
A pair of RF co-axial cables for ultrasound
signals and a three-core shield cable for temperature sensor,
Two-path system: Two pairs of RF co-axial
cables for ultrasound signals and a three-core
shield cable for temperature sensor,
Maximum cable length: 150m
Temperature range: 80ºC
Environment:Non-explosive environment without direct
sunlight, corrosive gas and heat radiation
Ambient temperature:
-10 to +50ºC for flow transmitter,
-20 to +80ºC for detector
Ambient humidity:
95%RH or less for flow transmitter, 100%RH
or less for detector
Grounding: Class D (less than 100 ohm)
Arrester:
Surge absorbers for outputs and power supply
incorporated as standard
Performance specifications
Accuracy :
Pulse Doppler method :
Pipe size (inside diameter)
and detector
Velocity
Accuracy
φ40mm to φ50mm or less
(Detector: FSWS12)
1.5 m/s to
Max. Velocity (Note)
±1.0% of rate
0 m/s to 1.5 m/s
±0.015m/s
1.5 m/s to
Max. Velocity (Note)
±0.5% of rate
0 m/s to 1.5 m/s
±0.0075m/s
φ50mm to φ200mm
(Detector: FSWS12)
1 m/s to
φ100mm to φ1000mm
(Detector: FSWS21,40,50) Max. Velocity (Note)
0 m/s to 1 m/s
±1.0% of rate
±0.01m/s
(Note) Maximum velocity is depend on a pipe diameter.
See Maximum measurement range in Hybrid mode (page 4).
Transit time method :
Pipe size (inside diameter)
Velocity
Accuracy
φ40mm to φ50mm or less
2 m/s to 32 m/s
±1.5% of rate
0 m/s to 2 m/s
±0.03m/s
φ50mm to φ300mm or less 2 m/s to 32 m/s
φ300mm to φ1000mm
Metric system
Velocity
m/s
ft/s
Flow rate
L/s, L/min, L/h, L/d, kL/h,
ML/d, m3/s, m3/min, m3/h,
m3/d, km3/d, Mm3/d,
BBL/s, BBL/min, BBL/h,
kBBL/d, MBBL/d
ft3/s, ft3/min, ft3/h, Mft3/d,
±1.0% of rate
English system
gal/s, gal/min, gal/h, Mgal/d,
BBL/s, BBL/min, BBL/h,
BBL/d, kBBL/d, MBBL/d
0 m/s to 2 m/s
±0.02m/s
Note: ˝gal˝ means US gal.
1 m/s to 32 m/s
±1.0% of rate
0 m/s to 1 m/s
±0.01m/s
Total display: Display of forward or reverse total
Data:up to 10 digits (decimal point to be
counted as 1 digit)
Unit: Metric/English system selectable
Response time:
Pulse Doppler method: 0.2sec
(depending on pipe diameter and
measuring condition)
Transit time method:
0.5sec
Power consumption:
20W or less
Short-term thermal stability:
140ºC, 30 min (FSWS12),
100ºC, 30 min (FSWS21, FSWS40, FSWS50)
Functional specifications
Analog output:4 to 20 mAdc (1 point)
Max. load resistance: 1k ohm
2
Digital output:+total, -total, alarm, acting range, flow switch
or total switch -- arbitrarily selectable
Mechanical relay contact:
1 point with socket (replaceable)
Normally closed/open selectable
Capacity:240 Vac/30 Vdc, 1 A
Total pulse: less than 1 p/s
(Pulse width: 50, 100 or 200 ms
selectable)
Transistor open collector: 2 points
Capacity: 30 Vdc, 0.1 A
Normally off/on selectable
Total pulse: less than 1000 p/s
(Pulse width: 0.5, 1, 2, 5, 10, 20,
50, 100 or 200 ms selectable)
Communication interface:
RS-232C equivalent / RS-485 (selectable)
Number of connectable units: one (RS-232C)/
up to 31 (RS-485)
Baud rate: 9600/19200/38400 bps selectable
Parity: none/odd/even selectable
Stop bit: 1 or 2 bits selectable
Distance: up to 15 m (RS-232C)/up to 1k m
(RS-485)
Data: velocity, flow rate, +total, -total, status
(standard), velocity profile (option)
Display device:
Graphic LCD (number of pixels: 240x64) with
back light,
Display language:
Japanese, English, French, German or Spanish selectable
Velocity/Flow rate display:
Display of velocity and/or flow rate with flow
direction
Data: up to 10 digits (decimal point to be
counted as 1 digit)
Unit: Metric/English system selectable
Metric system
English system
Total
mL, L, m3, km3, Mm3,
mBBL, BBL, kBBL
ft3, kft3, Mft3, gal, kgal,
mBBL, BBL, kBBL, ACRf
Configuration:
Fully configurable on keyboard by menudriven software
Zero adjustment:
Set zero/Clear available. (transit time method)
Damping:
0 to 100s (every 0.1s) configurable for analog
output and display
Low flow cut off:
0 to 5 m/s configurable
Hardware fault/process fault can be tied to
digital output
Burnout:
Analog output: Hold/Upper limit/Lower limit/
Zero/Not-used selectable
Total: Hold/Count selectable
Timer: 0 to 900s (every 1s) configurable Bi-directional range:
Forward and reverse ranges configurable independently
Hysteresis: 0 to 20% of working range configurable
Working range applicable to digital output
Auto-2 ranges:
Forward 2 ranges configurable independently
Hysteresis: 0 to 20% of working range configurable
Working range applicable to digital output
Flow switch: Lower and upper switching points configurable independently
Acting point applicable to digital output
Total switch: +total switching point configurable
Acting point applicable to digital output
Alarm:
Physical specifications
Enclosure protection:
Flow Transmitter: IP67,
Detector: IP67
Mounting:
Flow Transmitter: wall mount
Detector: clamped on pipe surface
Acoustic coupler:
Silicon compound (RTV)
Material:
Flow Transmitter: aluminum alloy
Detector: PBT for housing, aluminum alloy for
frame and SS for fastening belt
Sensor cable(FLY6):
RF coaxial cable (double shielded)
External sheath: Black flame-resistant vinyl
External diameter: About 7.3 mm
Terminal treatment: Water-resistant BNC connector (detector side), M3.5 amplifier
terminal (Flow Transmitter side)
Weight: About 90 g/m
Temperature sensor cable(FLY7):
3-core shield cable
External sheath: Gray flame-resistant vinyl
External diameter: About 6.9 mm
Terminal treatment: Round waterproof connector (detector side), M3.5 amplifier
terminal (Flow Transmitter side)
Weight: About 56 g/m
Dimensions: Flow Transmitter : H240 x W247 x D134 mm
Detector: H70 x W57 x L360 mm (FSWS12)
H72 x W57 x L540 mm (FSWS21)
H90 x W85 x L640 mm (FSWS40)
H82 x W71 x L258 mm (FSWS50)
Flow Transmitter: 5 kg
Detector: 1.7 kg (FSWS12), 1.9 kg (FSWS21),
5 kg (FSWS40), 1.5 kg (FSWS50)
Mass:
Conditions on straight pipe
( D : Inside diameter of pipe)
Upstream side
L
5D
10D
Detector
L
10D
or
more
90 bend
Downstream side
L
10D or more
Classification
L
10D
50D
Tee
L
30D
L
10D
L
30D
L
5D
D
Diffuser
0.5D
or
more
10D or more
1.5D
L
5D
Reducer
L
10D
Various Valve
In case that flow control valve exists on
upstream side.
Stop valve
Check valve
Pump
In case that flow control valve exists on
downstream side.
L
50D
P
(Note) The source : JEMIS-032
3
FSH, FSW, FLY
Measurement principle
Block diagram
<Pulse Doppler method>
• Ultrasound pulses are transmitted into a liquid, and
flow velocity profile is found and the flow rate is
measured by using the characteristics that Doppler
frequency of the echo from reflectors such as air
bubbles and particles in the liquid changes according
to flow velocity.
(1) Single path system (Z method)
Cable
Detector
4 to 20mA DC
Flow
Transmitter
Sensor 1
Air
bubbles
Reflection
(2) 2-path system (Z method)
Flow
velocity
Profile 1
Flow
velocity
Profile 2
Cable
Detector
Power
supply
4 to 20mA DC
Piping wall
Center
Flow
Transmitter
Piping wall
Contact
(Up to 3 points
are allowed.)
Temperature
sensor cable
<Transit Time method>
• Ultrasound pulses are propagated slanted both from the
upstream and downstream, and flow rate is measured by
detecting the time difference generated with the flow.
Sensor 1
Flow
Sensor 2
Transmission
Sensor 2
1
Sensor 1
2
Time difference
Receiving
∆T
Maximum measurement range in hybrid mode
When stainless steel is selected as pipe material, nominal wall thickness is Sch20s, and the fluid is water
<Maximum measurable flow velocity>
Caliber
40A
50A
65A
80A
90A
100A
125A
150A
200A
250A
300A
350A
400A
450A
500A
550A
600A
650A
700A
750A
800A
850A
900A
1000A
4
Contact
(Up to 3 points
are allowed.)
Temperature
sensor cable
Flow
Sensor 2
Flow velocity
Power
supply
FSWS12
6.56
6.52
5.31
4.65
4.12
3.69
3.08
2.63
2.04
FSWS21
7.25
6.08
5.20
4.05
3.30
2.78
2.51
2.20
FSWS40
7.77
6.38
5.41
4.90
4.31
3.80
3.48
Unit: m/s
FSWS50
3.48
3.17
2.91
2.71
2.52
2.35
2.21
2.08
1.97
1.77
<Maximum measurable flow rate>
Unit: m3/h
FSWS12
FSWS21
FSWS40
FSWS50
33.6
52.7
72.1
86.5
102
118
231
147
289
179
354
239
474
908
604
1168
735
1428
820
1598
951
1858
2118
2358
2358
2618
2879
3096
3357
3618
3879
4140
4400
4902
CODE SYMBOLS
<Flow transmitter>
1 2 3 4 5 6 7 8
F SH
1
Y
<Signal cable>
9 10 11 12
1 2 3 4 5 6 7 8
SYY
Description
Type (4th digit)
Standard
E
Velocity profile output (5th digit)
None
Available
Y
A
1
F L Y
6
7
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
1
1
1
1
1
1
Use (6th digit)
Single path or Changeover two-path (Note)
Note: 2 sets of detectors and coaxial cables
(FLY6) needed for two-path system.
Y
Power supply (7th digit)
100 to 240 Vac, 50/60 Hz
20 to 30 Vdc
1
4
Modification No. (8th digit)
Mark 1
1
Case structure (9th digit)
IP67
S
Conduit connection (10th digit)
G1/2 and G3/8 (female screw) with
water-proof connection
Y
For use with explosion-proof detector
(11th digit)
None
Y
Y
A
B
C
Parameter setting, Tag Plate (12th digit)
None
With setting
With setting and Tag Plate
With Tag Plate
0
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
9
0
1
2
3
4
5
Cable length (5th to 7th digit)
5m
10m
15m
20m
25m
30m
35m
40m
45m
50m
55m
60m
65m
70m
75m
80m
85m
90m
95m
100m
110m
120m
130m
140m
150m
Length in m to be designated with 3 digits
5
0
5
0
5
0
5
0
5
0
5
0
5
0
5
0
5
0
5
0
0
0
0
0
0
1
<Detector>
1 2 3 4 5 6 7 8
0 1
F SW S
9 10 11
Description
Y
Type (4th digit)
Standard
S
1
2
4
5
Kind of detector (5th and 6th digit)
Small diameter detector (φ 40 to 200 mm)
Small size detector (φ 100 to 400 mm)
Middle size detector (φ 200 to 500 mm)
Large size detector (φ 500 to 1000mm)
2
1
0
0
Use (7th digit)
Standard (IP67)
0
Modification No. (8th digit)
Mark 1
1
Signal cable (9th digit)
None
Y
Acoustic coupler (10th digit) (Note)
None
Silicone compound
Y
A
Y
A
Option (11th digit)
None
Tag Plate
(Note) Select silicone compound (A) for acoustic
coupler in ordinary cases. Silicon compound is
supplied in a tube (150g). If one or more
detectors one ordered, silicon compound may
be selected onece every 5 orders or so.
Description
Kind of cable (4th digit)
Coaxial cable (for ultrasonic sensors)
Three-core cable (for temperature sensor)
Modification No. (8th digit)
Mark 1
Loader software for PCs
Equipped as standard
• Works on PC/AT compatible machines.
• Operation on PC98-series machines (NEC) cannot be
guaranteed.
• Operation on self-made PCs or shop-brand PCs cannot
be guaranteed.
• Major functions: Setting/changing of various parameters
for the main unit
If no flow velocity profile output is selected,
the following functions are not available.
”Detailed setting” and ”flow velocity profile
display” in pulse Doppler measurement
”Detailed setting” and ”receved signal display” in Transit time measurement
• O/S: Windows2000/XP
• Memory requirement: 128MB or more
• Disk unit: Windows2000/XP-compatible CD-ROM drive
• Hard disk drive capacity: Free space of 52MB or more
N o t e : P C l o a d e r c o m m u n i c a t i o n c a b l e ( t y p e
ZZP*TK4H6253, Specifications: D-sub 9 pin receptacle,
cable length 3m) is separately required.
Detector frame installation fixture
Installation fixture is provided to facilitate the positioning of
the frame to the piping.
Select a desired type from the following according to the
detector to be used.
Type
Applicable detector
ZZP TK7M7071C1
FSWS12
ZZP TK7M7071C2
FSWS21
ZZP TK7M7071C3
FSWS40
Note: The installation fixture cannot be used for detector type
FSWS50, which is not provided with a frame.
5
FSH, FSW, FLY
OUTLINE DIAGRAM (Unit:mm)
<Flow transmitter (type: FSH)>
Mtg. plate
9
(273)
240
332
352
30
4-φ
Duosonics
3
134
72
233
33 33
14
30 30 30
48
48
24
48
CONNECTION DIAGRAM
Power supply terminals (AC)
1
2
3
4
L
N
1
B
2
B
3
A
4
100 to 240V AC
Cable
(Black)
–
Cable
(Black)
Cable
(Yellow/
Green)
To temp sensor
20 to 30V DC
2
3
4
5
6
7
8
1
HF GND HF GND HF GND HF GND
UP STR DOWN STR UP STR DOWN STR
LINE1
6
7
LINE2
1
2
+
–
DO1
–
9
Iout
SERIAL
No.
RS485
RS232C
4
5
6
TRX2 (–)
TRX1 (+)
SHILD
TXD
RXD
COM
3
+
4
5
–
+
DO2
Open collector
6
–
DO3
Relay contact
* Use LINE1 terminals in case of single measuring path.
6
8
+
SERIAL
Power supply terminals (DC)
1
2
3
4
+
5
7
Communication cable
(option)
Cap (red)
Cap (yellow)
Cap (black)
OUTLINE DIAGRAM (Unit:mm)
<Detector (type: FSWS12, 21)>
Connector
W
Temp. sensor
Absorber unit
Sensor unit
(with Temp. sensor)
Frame
Cursor
L
(H)
Sensor unit
(without Temp. sensor)
TYPE
PIPE SIZE
L
H
W
MASS APPROX
FSWS12
ø40∼ø200
360
70
57
1.7
FSWS21
ø100∼ø400
540
72
57
1.9
<Detector (type: FSWS40)>
Temp. sensor
Absorber unit
Sensor unit
Cursor
Frame
(with Temp. sensor)
85
Connector
640
(90)
Sensor unit
(without Temp. sensor)
7
FSH, FSW, FLY
OUTLINE DIAGRAM (Unit:mm)
<Detector (type: FSWS50)>
BNC Connector
Sensor unit
Absorber unit
73
Temp sensor
258
8
82
<Signal cable (type: FLY6)>
<Signal cable (type: FLY7)>
φ15
φ21.5
BNC Connector
Connector
φ11
φ7.3
L ± 1 (m)
L ± 1 (m)
φ6.9
3-φ6
.6
3-φ6
.6
L:According to the designation of the 5th, 6th, and the 7th
digits of the Code Symbols.
<Loader cable: ZZP*TK4H6253>
34
Connector
3.2 ± 0.2 (m)
67
46
To personal computer
SCOPE OF DELIVERY
•Flow transmitter (Type: FSH):
Flow transmitter
CD-ROM (Instruction manual,
Loader software)
•Detector (Type: FSW):
φ6.9
Sensor unit
Mounting belt
Silicon compound (option)
•Signal cable (Type: FLY6): Cable (2 wires)
To Flow transmitter
Terminal cap
(Red)
3φ6
.6
Terminal cap (Yellow)
Terminal cap (Black)
Red
TXD
Yellow Black
RXD
RS-232C
COM
•Signal cable (Type: FLY7): Cable for temperature sensor (1)
ITEMS DESIGNATED ORDERING
•Flow transmitter code symbols
•Detector code symbols
•Signal cable code symbols
9
FSH, FSW, FLY
<Parameter specification table>
No.
Setting item
Settable range
Initial value
1
Outer diameter
10.00 to 6200.00mm
(0.393 to 244.100 inch)
60.00mm
(2.362 inch)
2
Pipe material
12 menus
PVC
3
4
5
6
PiPing specification
Pipe S.V. : 1000 to 3700m/s
(3280 to 12140 ft/s)
Wall thickness
0.10 to 100.00mm
(0.003 to 3.940 inch)
4.00mm
(0.157 inch)
Lining material
8 menus
No lining
Lining S.V. : 1000 to 3700m/s
(3280 to 12140 ft/s)
Lining thickness
0.01 to 100.00mm
(0.000 to 3.940 inch)
Kind of Fluid
17 menus
[mm, inch]
Carbon steel, Stainless steel, PVC, Copper,
Castiron, Aluminum, FRP, Ductileiron, PEEK,
PVDF, Acrylic
Others (Sound velocity :
[m/s, ft/s])
[mm, inch]
No lining, Tar epoxy, Mortar, Rubber,
Teflon, Pyrex glass, PVC,
Others (Sound velocity :
[m/s, ft/s])
[mm, inch]
Water
Water, Seawater, DIST. water, Ammonia,
Alcohol, Benzene, Bromide, Ethanol, Glycol,
Kerosene, Milk, Methanol, Toluol,
Lube oil,
Fuel oil, Petrol,
Others (Sound velocity :
[m/s, ft/s])
(Kinematic viscosity
[x10-6m2/s, ft2/s])
Fluid S.V. : 500 to 2500m/s
(1641 to 8203 ft/s)
Kinematic viscosity :
0.001 to 999.9999 x 10-6m2/s
(0.0107 to 10763.9088 x 10-6ft2/s)
Range unit
19 menus
m/s
(ft/s)
m/s, L/s, L/min, L/h, L/d, kL/d, ML/d, m3/s,
m3/min, m3/h, m3/d, km3/d, Mm3/d, BBL/s,
BBL/min, BBL/h, BBL/d, KBBL/d, MBBL/d,
(ft/s, ft3/s, ft3/min, ft3/h, ft3/d, kft3/d, Mft3/d,
gal/s, gal/min, gal/h, gal/d, kgal/d, Mgal/d,
BBL/s, BBL/min, BBL/h, BBL/d, kBBL/d,
MBBL/d)
8
Range type
4 menus
Single
Single, Auto 2, Bi-dir, Bi-dir Auto 2
9
Full scale or
Full scale 1
In terms of flow velocity 0.00 ···
0.30 to 32.00m/s
(0.98 to 104.98 ft/s)
2.00m/s
(6.56 ft/s)
[
]
Full scale 2
In terms of flow velocity 0.00 ···
0.30 to 32.00m/s
(0.98 to 104.98 ft/s)
4.00m/s
(13.12 ft/s)
[
]
Range HYS.
0.00 to 20.0%
10.00%
%
Output limit LO.
-20 to 0%
-20%
%
13
Output limit HI.
100 to 120%
120%
%
14
Output burnout
5 menus
Hold
15
Burnout timer
0 to 900sec
10sec
16
Rate limit
0.00 to 5.00m/s
(0.00 to 16.40 ft/s)
in terms of flow velocity
0.00m/s
(0.00 ft/s)
17
Rate limit timer
0 to 900sec
0sec
sec
0.0 to 100.0sec
5.0sec
sec
1 : Display kind
7 menus
Flowrate (m3/s)
2 : Display kind
7 menus
Flow velocity (m/s) Flow velocity, Flowrate, Total forward,
Total reverse,
F : Total pulse, R : Total pulse, Flow rate (%)
0.00 to 5.00m/s
(0.00 to 16.40 ft/s)
in terms of flow velocity
0.01m/s
(0.03 ft/s)
12
18
19
20
21
Damping
Display setting
11
Output setting
7
10
10
Settable value
Low flow cut
Not use, Hold, Upper, Lower, Zero
sec
[
]
Flow velocity, Flowrate, Total forward,
Total reverse,
F : Total pulse, R : Total pulse, Flow rate (%)
[
]
No.
Setting item
Settable range
Initial value
Settable value
Total mode
3 menus
Total stop
Total stop, Total run, Total reset
23
Total unit
8 menus
mL
mL, L, m3, km3, Mm3, mBBL, BBL, kBBL,
ft3, kft3, Mft3, kgal, gal, mBBL, BBL, kBBL,
ACRF
24
Total rate
0.000 to 999999.999
0.000
[
]
25
F : Total preset
0.000 to 9999999999.999
0.000
[
]
26
F : Total SW
0.000 to 9999999999.999
0.000
[
]
R : Total preset
0.000 to 9999999999.999
0.000
[
]
28
R : Total SW
0.000 to 9999999999.999
0.000
[
]
29
Output burnout
2 menus
Hold
30
Burnout timer
0 to 900sec
10sec
31
Pulse width 1
3 menus
50ms
50, 100, 200
32
Pulse width 2
9 menus
50ms
0.5, 1.0, 2.0, 5.0, 10.0, 20.0, 50.0, 100.0,
200.0
33
Flow sw high
In terms of flow velocity
0.00 to 32.00m/s
(0.00 to 104.98 ft/s)
0.00m/s
[
]
Flow sw low
In terms of flow velocity
0.00 to 32.00m/s
(0.00 to 104.98 ft/s)
4.00m/s
[
]
35
Flow sw HYS.
0 to 20%
10%
36
Output DO1
15 menus
Not use
Not use, Signal error, F : Total pulse,
R : Total pulse, F : Total alarm,
R : Total alarm, F : Total overflow,
R : Total overflow, Flow SW high,
Flow SW Low, Full scale2, AO range over,
Pulse range over, R : Flow direction,
Device error
37
Mode DO1
2 menus
Normal
Normal, Reverse
38
Output DO2
15 menus
Not use
Not use, Signal error, F : Total pulse,
R : Total pulse, F : Total alarm,
R : Total alarm, F : Total overflow,
R : Total overflow, Flow SW high,
Flow SW Low, Full scale2, AO range over,
Pulse range over, R : Flow direction,
Device error
39
Mode DO2
2 menus
Normal
Normal, Reverse
40
Output DO3
15 menus
Not use
Not use, Signal error, F : Total pulse,
R : Total pulse, F : Total alarm,
R : Total alarm, F : Total overflow,
R : Total overflow, Flow SW high,
Flow SW Low, Full scale2, AO range over,
Pulse range over, R : Flow direction,
Device error
41
Mode DO3
2 menus
Normal
Normal, Reverse
Status output
34
Flow switch
27
Total
22
Not use, Hold
sec
%
11
FSH, FSW, FLY
No.
Setting item
Settable range
Initial value
Settable value
42
System unit
2 menus
Metric
Metric, English
43
Language
5 menus
English
Japanese, English, German, French,
spanish
COM. speed
3 menus
38400BPS
9600BPS, 19200BPS, 38400BPS
COM. parity
3 menus
None
None, Even, Odd
COM. stop bit
2 menus
1 bit
1 bit, 2bits
Serial method
2 menus
RS232C
RS232C, RS485
Station No.
31 menus
1
1 to 31
Measurement
mode
2 menus
1 Path
1 Path, 2 Path
AO Definition
3 menus
Line 1
Average, Line 1, Line 2
4 menus
FSW12
FSW12, FSW21, FSW40, FSW50
46
47
System
45
Serial com.
44
49
50
51
Measuremant
mode
48
Sensor Type
Note1: When total pulse output has been selected for DO1, DO2 or DO3 specify total pulse value and total pulse width so that conditions 1
and 2 shown below are satisfies.
Condition 1 :
Condition 2 :
Flow span-1*[m3/s]
total pulse value*[m3]
1000 [In the case of DO1 and DO2]
1 [In the case of DO3]
Flow span-1*[m3/s]
1000
total pulse value*[m3]
2 × total pulse width [ms]
* In the case of 2 ranges, perform calculations using either flow span-1 or flow span-2, whichever is greater.
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan