ETC PAT32/D

PORTABLE APPLIANCE TESTERS
MEGGER®
PAT32
• Wide applications,
including the testing of
Information Technology
(IT) equipment
• Easy to use – with
range sensitive
passbands displayed
• Analogue/Digital
display – accurate
digital readings and
rapid response
analogue display highlighting fluctuations
Portable Appliance Tester
DESCRIPTION
The MEGGER® PAT32 Portable
Appliance Tester will perform the
basic electrical safety tests required
for
portable
appliances
and
equipment. It can be used to test
Information Technology (IT) equipment, domestic appliances, industrial
appliances and power tools.
It performs seven tests on an
appliance:
• Circuit test
•Fuse check
• Earth continuity test
• 10 A Earth Bond
• 25 A Earth Bond
• Insulation test
• Extension leads (ELT1 required)
The PAT32 will test appliances and
tools rated at 240 V and 110 V a.c.
since it is fitted with both BS 1363
and BS 4343/CEE17 appliance test
sockets.
The tests are effected by pressing the
relevant test switch. The readings
from each test are shown on a unique
analogue/digital display. The scales
for Earth Bond and Insulation are
marked with pass bands. This
provides a rapid indication of a test
pass or failure and allows proper test
results to be kept. These records are
necessary as part of the appliance’s
safety monitoring programme.
The instrument is robust, reliable and
constructed in a strong plastic case
with a moulded carrying handle and a
detachable lid. Simple diagrammatic
instructions are attached to the inside
of the lid for quick reference.
The ELT1 accessory allows extension
leads both 110 V, 240 V and IEC mains
leads to be tested. It performs Lead
Polarity, Earth Bond and Insulation
Tests.
TEST SEQUENCE
The tests should be performed in the
correct order – starting with the
CIRCUIT test. This applies a low
voltage across the appliance and
checks the d.c. resistance, live to
neutral ensuring the relevance of
subsequent tests.
FUSE CHECK can be used to check
any suspect fuse by placing it on the
contacts.
The EARTH CONTINUITY test is a low
current test which is used where a
higher current Earth Bond test may
cause damage to the appliance.
The EARTH BOND test is used for
earthed or Class I appliances only
and is designed to test the earth lead
continuity and the earth bonding to
the metal casing of an appliance. The
crocodile clip lead supplied with the
instrument is connected between the
EARTH BOND terminal and the
appliance’s casing. A choice of two
test current levels is provided, 25 A
and 10 A. The latter is for use on
appliances with light duty cables.
The final test is an insulation test
required for both earthed and double
insulated appliances. For earthed
appliances, a test voltage of 500 V d.c.
is established between the mains
supply plug live and neutral pins
joined together and the earth pin.
No additional connection is required.
For double insulated appliances, the
test voltage is established between
the mains supply plug live and neutral
pins joined together and earth (which,
in this case, is taken to the crocodile
clip test lead attached to the EARTH
BOND terminal). The clip is attached
to the case of the appliance and the
test is then performed.
APPLICATIONS
The PAT32 will check the electrical
safety of the earthed appliances (BS
and IEC Safety Class I) and double
insulated appliances (BS and IEC
Safety Class II), symbol
.
Spheres of use for the PAT32 are:
• Periodic tests of equipment used in
factories, offices, local education
authorities, hospitals, etc.
• Routine tests before and after hiring
electrical equipment.
• Basic tests following equipment
repair.
• Tests by manufacturers and
distributors.
The instrument may be used to test
equipment originally manufactured
to a range of specifications including:
BS 3456/IEC 335-1, BS 2769, BS 4533,
BS 415 and BS 7002.
PORTABLE APPLIANCE TESTERS
SPECIFICATIONS
Tests Available
(1)Circuit test
(2)Fuse check
(3)Earth Continuity test
(4)10 A Earth Bond test
(5)25 A Earth Bond test
(6)Insulation test
(7) Extension Lead test (ELT1 required)
Pass Band Limit: 100 mΩ and 500 mΩ
Circuit Test
Meter Reading Range: 0 to 9,9 kΩ
Range:
Resolution:
(a) 0 to 1 kΩ
0,01 kΩ
(b) 1 kΩ to 2 kΩ
0,02 kΩ
(c) 2 kΩ to 5 kΩ
0,05 kΩ
(d) 5 kΩ to 9,9 kΩ
0,1 kΩ
Accuracy:
(a) 2,5% of Reading ± 0,01 kΩ
(b) 2,5% of Reading ± 0,02 kΩ
(c) 2,5% of Reading ± 0,05 kΩ
(d) 2,5% of Reading ± 0,1 kΩ
Open Circuit Voltage: 4,5 V d.c.
25 A Earth Bond Test
Meter Reading Range: 0 to 1990 mΩ
Range:
Resolution:
(a) 0 to 999 mΩ
1 mΩ
(b) 900 mΩ to 1990 mΩ 10 mΩ
Short Circuit Current: 1 mA d.c.
Fuse Check
5 V d.c. 0,5 mA Typical
Earth Continuity Test
Range:
Resolution:
(a) 0 to 999 mΩ
1 mΩ
(b) 900 mΩ to 9990 mΩ 10 mΩ
Accuracy
(a) 2,5% of Reading ± 5 mΩ
(b) 2,5% of Reading ± 10 mΩ
Open Circuit Voltage:
Typical 100 mV d.c.
Constant Current (≤ 1 Ω):
Typical 100 mA d.c.
10 A Earth Bond Test
Meter Reading Range: 0 to 1990 mΩ
Range:
Resolution:
(a) 0 to 600 mΩ
1 mΩ
(b) 600 mΩ to 999 mΩ 1 mΩ
(c) 900 mΩ to 1990 mΩ 10 mΩ
Accuracy:
2,5% of Reading ± 5 mΩ
(0 to 600 mΩ range only)
Open Circuit Voltage:
Typically 3 V rms a.c. 50 Hz
Spec. Point Current:
10 A into 100 mΩ at 240 V
8,7 A into 100 mΩ at 230 V
Short Circuit Current:
Typically 14 A
Accuracy:
(a) 2,5% of Reading ± 5 mΩ
(b) 2,5% of Reading ± 10 mΩ
Short Circuit Current:
Typically 2 mA
Extension Lead Test
Polarity Test
Display will indicate:
“YES” if polarity Correct
“POL” if connections crossed
“S.C.” if short-circuit L–N
“O.C.” if open-circuit L or N
“BAD” if leakage L–N
Earth Bond Test (10 A and 25 A)
Insulation Test
Ranges as the PAT32
Pass Band Limit: 100 mΩ and 500 mΩ
Temperature Range
Operation: 5°C to +40°C
Storage: -25°C to + 65°C
Open Circuit Voltage:
Typically 6 V rms a.c. 50 Hz
Humidity Range
Operation: ≤ 90% RH at 25°C
Spec. Point Current:
Typically 25 A into 100 mΩ (10 A into
500 mΩ)
Short Circuit Current:
Typically 38 A
Insulation Test
Meter Reading Range: 0 to 99 MΩ
Range:
Resolution:
(a) 0 to 10 MΩ
0,1 MΩ
(b) 10 MΩ to 20 MΩ
0,2 MΩ
(c) 20 MΩ to 50 MΩ
0,5 MΩ
(d) 50 MΩ to 99 MΩ
1 MΩ
Accuracy:
(a) 2,5% of Reading ± 0,1 MΩ
(b) 5% of Reading ± 0,2 MΩ
(c) 5% of Reading ± 0,5 MΩ
(d) Indication Only
Safety
The PAT32 meets the requirements
of IEC 1010-1(1990) Safety Class I,
Installation Category II, Pollution
Degree 2, Altitude up to 2000 m.
Supply Voltage
230 V 50 Hz 300 VA (supply voltage
not to exceed +10%/-6% of the
nominal voltage)
Dimensions
PAT32: 314L x 152H x 206W mm
(12,5L x 6H x 8W in approx.)
ELT1: 123L x 58H x 64W mm
(4,75L x 2,25H x 2,5W in approx.)
Weight
PAT32: 3,75 kg (8,25 lb approx.)
ELT1: 350g (0,75 lb approx.)
Pass Band Limit:
2 MΩ (Class I) 7 MΩ (Class II)
Open Circuit Voltage: ≤ 680 V d.c.
Spec. Point Current:
500 V into 2 MΩ
ORDERING INFORMATION
Item (Qty)
Order Code
Portable Appliance Tester ........................................PAT32
Included Accessories
Earth Bond test lead with crocodile clips.......... 6231-043
Accessory Pouch, PVC ..........................................6420-089
Operating instruction book ..................................6171-728
Optional Accessories
Extension lead tester, ELT 1 .................................6111-130
Earth Bond probe, EP 1 .........................................6320-225
Safebloc adaptor lead ............................................6331-230
Appliance safety log book, complete...................6131-813
Test sheets for log book (pk of 30) ......................6171-417
Test stickers for log book (pk of 72)....................6171-418