GXM60 Medical Single Output 60 Watt-Range Input High Efficiency Desk-Top Supply FEATURES: • Universal Input • Small size, light weight • Medically approved to UL2601-1, CSA 22.2 No. 601.1-M90, EN60601-1, and CB certified to IEC601-1. • Ultra efficient - Up to 90% • Spill proof external enclosure • 2 year warranty • marked to LVD SPECIFICATIONS: AC INPUT 90-264 Vac, 47-63 Hz single phase. Input power is less than 75 W when output power is less than 60 W, 230 Vac in. Meets EN61000-3-2 harmonics standard. OUTPUT POWER Total continuous output power is 60 W (55 for any unit less than 8 V), 65 W peak for 60 s 10% duty cycle. Total power specified from supply output, derate for cable losses. OUTPUT REGULATION Supply regulation is +/- 1%. However, actual regulation limits are affected by cable losses on units without remote sense. Final regulation limits must include cable losses. Contact factory for assistance. INPUT CURRENT Maximum input current at minimum input voltage and output overload will be less than 2.0 A. HOLD-UP TIME Output voltage stays within regulation for 20 ms from loss of ac input at full load, from 120 Vac input. MINIMUM LOAD No minimum load required. OVERLOAD PROTECTION Fully protected against short circuit and output overload. Short circuit protection is cycling type power limit and will automatically recover after removal of fault. OUTPUT NOISE 0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode. Measured with a scope probe directly across the output connector. Load terminated with a 0.1 µF capacitor in parallel with a low Z 10 mF capacitor. TRANSIENT RESPONSE Response at supply is 1.5 ms typical for return to regulation for a 50% load step, ∆i/∆t< 0.2 A/µs. Maximum voltage deviation is 3.5%. Cable resistance will effect maximum voltage deviation. OVER VOLTAGE PROTECTION Prevents output from producing excessive voltage in the event of a single fault failure. Voltage limited typically to 130% depending upon model. OVP firing reduces output to less than 50% of nominal voltage in less than 50 ms. OVERLOAD PROTECTION Factory set to begin power limiting at approximately 75 W. EFFICIENCY 77 to 90% at full rated load depending upon model and ac line. Units are typically > 85 % @ 230 Vac. Efficiency is measured at the supply output, cabling losses are not considered and will vary with different cables. Contact factory for assistance. OVERSHOOT Less than 3% overshoot at turn-on under nominal conditions. TURN-ON TIME Less than 1.5 s at 115 Vac, 25 °C INPUT PROTECTION Both lines protected with internal ac fuses, provided on all units. Designed to blow only if a catastrophic failure occurs in the unit — Fuse does not blow on short circuit or overload. INRUSH CURRENT Inrush is limited by internal thermistor. The inrush at 240 Vac, averaged over the first ac half-cycle under cold start conditions will not exceed 60 A. EMI / EMC COMPLIANCE All models include built-in EMI filtering. Conducted Emissions EN55011, Class B; FCC Class B Static Discharge EN61000-4-2, 6 kV contact 8 kV air RF Field Susceptibility EN61000-4-3, 3V/meter Fast Transients / Bursts EN61000-4-4, 2 kV, 5 kHz Surge Susceptibility EN61000-4-5, 1 kV diff., 2 kV com. Conducted RF Susceptibility EN61000-4-6, 3V Voltage Sags & Surges EN61000-4-11 MEDICAL SAFETY AGENCY All models are approved to UL2601, CSA 22.2 No. 601.1-M90, IEC 601-1 (1988), EN 60601-1 LEAKAGE CURRENT 80 µA under normal conditions (132 Vac @ 60 Hz ). Maximum under single fault conditions (264 Vac @ 50 Hz) is 170 µA. Environmental Specification Operating Non-operating Temperature (A) 0 to 40ºC -40 to +85ºC 0 to 95% RH 0 to 95% RH Humidity (A) Shock (B) Altitude Vibration (C) 20 gpk 40 gpk -500 to 10,000 ft -500 to 40,000 ft 1.5 grms, 0.003 g2/Hz 5 grms, 0.026 g2/Hz A. Units should be allowed to warm up/operate under non-condensing conditions before application of power. B. Random vibration—10 to 2000 Hz, 6 dB/octave roll-off from 350 to 2000 Hz, 3 orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating. C. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal axes, total 6 shocks. 106 Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033 800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com GXM60 Medical Single Output Typical Standard Model Configurations: Medical Model Output (V) Current (I) Initial (1) Set Point Total Regulation Output Harness Remote Sense Ripple/ Noise GXM60-5A11 5V 11 A 2% 1% Molex Yes 1.5% GXM60-12A01 12 V 5.0 A 2% 1% 2.1 mm No 1% GXM60-15A01 15 V 4.0 A 2% 1% 2.1 mm No 1% GXM60-24A01 24 V 2.5 A 2% 1% 2.1 mm No 1% GXM60-28A01 28 V 2.2 A 2% 1% 2.1 mm No 1% GXM60-36A01 36 V 1.7 A 2% 1% 2.1 mm No 1% Notes: 1) Initial set point without load. Total regulation shown includes cable losses. When selecting cable options without remote sense, cable drop should be considered. MODEL NUMBER CONFIGURATION GXM60-XXXXX Output Cable/Plug Option Number Case Color: A = Beige B = Black Output Voltages: Units are available standard with any output voltage from 5 to 35 Vdc in 1.0 increments. Also available in 0.1 increments data custom order. Option No. Output Cord/Plug Cable/Plug Maximum Rating 01 Circular 5.6 mm O.D. , 2.1 mm I.D., 18 ga. 5 A, 80 V max. 02 Circular 5.6 mm O.D., 2.5 mm I.D., 18 ga. 5 A, 80 V max. 03 Circular 5.6 mm O.D., 2.1 mm I.D., RT angle, 18 ga. 5 A, 80 V max. 04 Circular 5.6 mm O.D., 2.5 mm I.D., RT angle, 18 ga. 5 A, 80 V max. 05 Circular 5.6 mm O.D., 2.1 mm I.D., 18 ga., shielded 5 A, 80 V max. 06 Circular 5.6 mm O.D., 2.5 mm, I.D.,18 ga., shielded 5 A, 80 V max. 07 Mini DIN 20 ga. 6 A, 36 V max. 08 Mini DIN 20 ga. shielded 6 A, 36 V max. 09 5 Pin 20 ga. DIN plug 6 A, 36 V max. 10 5 Pin 20 Ga. DIN plug, shielded 11 Molex Mini-Fit, 18 ga. 11 A, 480 V max. 12 Power DIN 11 A, 20 V max. 13 Power DIN., shielded 11 A, 20 V max. 6 A, 36 V max. MECHANICAL SPECIFICATIONS Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033 800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com 107