2001 3159 INSULATION / WITHSTANDING HiTESTER Safety Standards Measuring Instruments Improved Efficiency for Electrical Safety Testing ISO14001 JQA-E-90091 The 3159 INSULATION / WITHSTANDING HiTESTER is a combination insulation resistance as well as voltage endurance tester. It continuously performs insulation testing and voltage endurance testing of electrical equipment and parts, doing both tests in a simpler and more efficient way. This instrument implements all the applicable safety standards, and is small, light, and inexpensive. It also comes standard with external I/O for automating production lines. http://www.hioki.co.jp/ HIOKI company overview, new products, environmental considerations and other information are available on our website. 1 One Unit Serves Two Functions -Continuous Testing of Insulation and of Voltage Endurance ● Continuous Testing of Insulation and Voltage Endurance In automatic testing mode, the 3159 continuously tests either insulation followed by voltage endurance (the ability to "withstand" voltage) (I→W), or voltage endurance followed by insulation (W→I). In manual mode it separately performs insulation testing or voltage endurance testing. ● Stores up to 10 Sets of Test Conditions ● Standards Testing Contains on-board comparator and timer functions for determining compliance, thus simplify the testing of all applicable safety standards. ● Interlock Function Based on inputs such as a starter signal, enters a state where output is blocked and testing is impossible, to guarantee safety such as during automated testing. Stores up to 10 sets of test conditions for each of voltage endurance mode and insulation mode, and can quickly switch among the test conditions. (Save/Load) Fluorescent Display Tube The display uses a bright and clear fluorescent tube. Analog Voltmeter The test voltage can be verified not only from the digital display but also in analog form. (Only for voltage endurance testing.) Danger Light The warning light flashes during testing and whenever there is a high voltage between the terminals. External Switch Start/stop may be controlled with the 9613 or the 9614. (The 9613 and 9614 are options.) 9613 Insulation Te s t i n g Two Rated Measurement Voltages: DC 500 or 1000 V 9614 Voltage Endurance Testing Transformer Capacity : 500VA Two Test Voltage Ranges : AC 0 to 2.5 or 5.0 kV ● Voltage Comparator In order to prevent test voltage errors, the timer is only activated after the voltage rises to the specified test voltage ± 5% (or ± 50V, whichever is larger). (Depending on options, it may be possible to change the voltage check just before test completion.) Measurement Range: 0.5 MΩ to 2000 MΩ ● Discharge Feature The charge on the object being tested is discharged within the tester. This discharge function avoids impacting the next test process. 2 Functions for Handling a Wide Range of Situations Automatically This tester comes standard with features for automating EXT I/O, RS-232C, and status out (relay contact output), as well as data management features. Status Out Buzzer Volume Adjustment When the output conditions set up by the dip switches are satisfied (OR condition), there is relay contact output. 1. H.V.ON Output voltage generation 2. TEST Testing in progress 3. PASS Passed 4. FAIL Failed 5. INT.LOCK Interlocked To indicate "pass" or "fail". Rear Panel Voltage Output Terminals These are normally connected to the front terminals. 6. READY Ready 7. EXT.CONT. Under external control 8. POWER ON Powers the 3159 on EXT I/O (Open Collector, Photocoupler Insulation) EXT I/O Output Signals Pin Number 1 2 3 4 5 RS-232C Allows automatic testing and reading of test results from a personal computer. EXT I/O Input Signals Pin Number Signal Name Function Signal Name READY L-FAIL U-FAIL PASS TEST 6 H.V.ON W-MODE 7 EXT-E When LO, the external I/O input signal is effect. 11 8 START When LO, it functions as a "Start" key. 12 I-MODE 9 STOP When LO, it functions as a "Stop" key. 13 W-FAIL 10 INT.LOCK Interlocked when open. 15 to 18 ISO.COM Ground inputs for external devices. 14 33 to 36 I-FAIL ISO.DCV Function LO when in "ready state" LO when in "fail state" for the lower bound LO when in "fail state" for the upper bound LO when in "pass state" LO when in "test state" LO when a voltage is being generated to the output terminals LO during voltage endurance ("withstanding") testing LO during insulation testing LO when in "fail state" for voltage endurance testing LO when in "fail state" for insulation testing 15 V (0.1 A) outputs Wide Range of Functions 1. Pass Hold Function (0: No Hold, 1: Hold) The pass state is held when it occurs. This is convenient for verifying the decision value. ➀ ➁ ➂ ➃ 2. Fail Hold Function (0: No Hold, 1: Hold) The fail state is held when it occurs. This is convenient for temporarily stopping the test process. Various functions can be specified with the SHIFT + STOP keys. 3. Hold State (0: No Hold, 1: Hold) This saves the state when the stop key is pressed during a test in order to unconditionally end the test. 4. Momentary Out (0: Not Specified, 1: Specified) This function outputs a voltage only when the start key is being pressed. The start key is effective both as EXT SW and external I/O. ➄ ➅ ➆ ➇ ➈ ➉ 8. Interlock Function (0: Not Specified, 1: Specified) This specifies whether the interlock terminal for external I/O should be effective. 9. Voltage Comparison Time (0: Start of Test, 1: End of Test) When the voltage comparator is on during voltage endurance testing, this specifies whether the comparison should be done at the start or the end of the test. 10. Insulation Resistance Measurement Range 5. Double Action (0: Not Specified, 1: Specified) This function starts the test if the start key is pressed less than about 0.5 seconds after the stop key. 6. Fail Mode (0: Not Specified, 1: Specified) This function means that hold state can be released only by the stop key on the main body. 7. RS Command "START" (0: Not Specified, 1: Specified) This specifies whether the RS command "START" should be effective. (0: Fixed Range, 1: Automatic Range) This specifies whether the measurement range for the insulation resistance test should be a fixed range or an automatic range. 11. Insulation Resistance Test End Mode 0 : Test for the specified time 1 : Stop when "pass" is determined 2 : Stop when "fail" is determined This specifies the method of ending the insulation resistance test. 3 3159 Specifications ■ Voltage Endurance Testing ■ Decision Function [ Test Voltage ] Decision method Decision results Output voltage Voltage output method Transformer capacity Voltage adjustment method Voltmeter Waveform Frequency : : : : Two ranges: AC 0 to 2.5 or 5.0 kV Zero input switch 500 VA (rating: 30 minutes) Manual adjustment : Average value rectified root-mean-square display Digital: AC 0.00 kV to 5.00 kV (full scale) Accuracy: ± 1.5% f.s. Analog: AC 0 to 5 kV (full scale) Accuracy: ± 5% f.s : Same as the power supply waveform : Same as the power supply frequency Decision processing Specification ranges : Window comparison method (digital specification). : UPPER-FAIL: The measured current (insulation resistance value) exceeded the specified upper bound. PASS: The measured current (insulation resistance value) was between the specified upper and lower bounds and the specified time elapsed LOWER-FAIL: The measured current (insulation resistance value) was less than the specified lower bound. : For each decision result, output the display portion, the buzzer sound, and EXT I/O signal : Voltage endurance testing : 0.1 mA to 120 mA (upper bound) / 0.1 mA to 119 mA (lower bound) [ Current Detection ] Insulation testing : Current measurement : 0.01 mA to 120 mA range Indicated valueange : Average value rectified root-mean-square display (digital) Measurement : 0.01 mA (2 mA or 8 mA range) resolution 0.1 mA (32 mA range) 1 mA (120 mA range) Measurement : ( ± 3% f.s. + 20 µA) over the entire range accuracy (Assumes a power supply waveform distortion ratio of 5% maximum.) 0.2 MΩ to 2000 MΩ (same for the upper and lower bounds). Specification resolution : Voltage endurance testing : 0.1 mA (0.1 mA to 9.9 mA) / 1 mA (10 to 120 mA) Insulation testing : 0.01 MΩ (0.2 MΩ to 2 MΩ), 0.1 MΩ (2.1 MΩ to 20 MΩ) 1 MΩ (21 MΩ to 200 MΩ), 10 MΩ (210 MΩ to 2000 MΩ) ■ Timer Specification range Action After starting, a countdown from the specified time is displayed ■ Insulation Resistance Testing When OFF is specified: Displays the elapsed time since the start. [ Test Voltage and Measurement Range ] Rated voltage Unloaded voltage Rated measured current Short circuit current Measurement range and accuracy Measured resistance range Specification resolution and accuracy Nondeterministic interval : DC 500 V or 1000 V : From 1 to 1.2 times the rated voltage : 1 mA to 1.2 mA : 4 mA to 5 mA (500 V) / 2 mA to 3 mA (1000 V) : 0.5 MΩ to 999 MΩ (500 V), 1 MΩ to 999 MΩ (1000 V) / ± 4% rdg 1000 MΩ to 2000 MΩ / ± 8% rdg. : 2 MΩ, 20 MΩ, 200 MΩ, 2000 MΩ (500 V) 4 MΩ, 40 MΩ, 400 MΩ, 2000 MΩ (1000 V) EXT I/O RS-232C ■ General Specifications Power supply frequency : : : : Fluorescent display tube (digital display) Output voltage, detected current, measured resistance 2 times/sec. minimum 0 ˚C to 40 ˚C, 80% rh maximum (no condensation) Voltage endurance Maximum rated power Dimensions : -10 ˚C to 50 ˚C, 90% rh maximum (no condensation) Mass : 23 ˚C ± 5 ˚C, 80% rh maximum (no condensation), after warming up for at least 5 minutes : Indoor, altitude 2000 m maximum : AC 100 V (3159), AC 120 V (3159-01), AC 220 V (3159-02), AC 230 V (3159-03), AC 240 V (3159-04), : 50 Hz to 60 Hz 3159 INSULATION / WITHSTANDING HiTESTER 3159-01 INSULATION / WITHSTANDING HiTESTER 3159-02 INSULATION / WITHSTANDING HiTESTER 3159-03 INSULATION / WITHSTANDING HiTESTER 3159-04 INSULATION / WITHSTANDING HiTESTER (100V) (120V) (220V) (230V) (240V) : 0.1 sec. (0.5 to 99.9 sec.) ± 50 msec 1 sec. (100 to 999 sec.) ± 0.5 sec : 0.5 sec. (Mask time until the determination begins during insulation resistance testing.) ■ Interfaces EXT SW Display Monitor functions Monitor period Operating temperature range Storage temperature range Temperature and humidity range for assured accuracy Operating locations Power supply voltage : 0.5 to 999 sec : When ON is specified: Applicable standards Standard accessories Options : Output signal: Open collector output Maximum loaded voltage DC 30 V Maximum output current DC 100 mA / 1 signal Input signal: Active low input / maximum applied voltage DC 30 V : Input signal (contact point input) START, STOP, SW.EN (front terminal switch is effective) : Start-stop synchronization, full duplex, 9600 bps : AC 1.35 kV, 10 mA, for one minute, between power supply and chassis : 800 VA : Approx. 320 (W) × 155 (H) × 330 (D) mm (not including protrusions.) : Approx. 18 kg (3159), Approx. 20.5 kg (3159-01), Approx. 21.5 kg (3159-02 to 3159-04) : EMC: EN 61326-1:1997 + A1:1998 class A Safety: EN 61010-1:1993 + A1:1995 Contamination level 2, overvoltage category (expected overvoltage category 2500V) : 9615 H.V. test lead (high voltage side and return, one each), power cord, extra fuse : 9613 REMOTE CONTROL BOX (SINGLE), 9614 REMOTE CONTROL BOX (DUAL), 9616 WARNING LAMP ■ Options 9613 REMOTE CONTROL BOX (SINGLE) 9614 REMOTE CONTROL BOX (DUAL) 9616 WARNING LAMP 9637 RS-232C CABLE (1.8 m) (9pin-9pin/Cross) 9638 RS-232C CABLE (1.8 m) (9pin-25pin/Cross) DISTRIBUTED BY HEAD OFFICE : 81 Koizumi, Ueda, Nagano, 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail: os-com@ hioki.co.jp HIOKI USA CORPORATION : 6 Corporate Drive, Cranbury, NJ 08512 USA TEL +1-609-409-9109 / FAX +1-609-409-9108 E-mail: hioki@ hiokiusa.com All information correct as of Mar. 9, 2001. All specifications are subject to change without notice. ■ Inter net HIOKI website http://www.hioki.co.jp/ F3159E1-13B-05P Printed in Japan