ASM-SENSOR 3511-50

2001
Improved with Faster Measurement !
3511-50 LCR HiTESTER
Components measuring instruments
Minimum measurement time of 5 ms, built-in comparator and ±0.08% measurement accuracy
Improved for even faster and more efficient measurements !
ISO14001
JQA-E-90091
http://www.hioki.co.jp/
HIOKI company overview, new products, environmental considerations and
other information are available on our website.
The 3511-50 LCR HiTESTER features both high
performance, highspeed measurements with a low prices.
The minimum measurement time of 5 ms and basic
accuracy of ±0.08% makes the instrument suitable for use
on production lines and laboratories. The built-in highspeed comparator significantly reduces production line tact
time and allows the construction of automatic production
lines.
The very compact body features a clearly visible LED
display that facilitates easy operation and allows settings to
be confirmed at a glance.
With its high-speed measurement, highly accurate
measurement capabilities and great cost performance, this
LCR measurement instrument is bound to satisfy the needs
of a variety of users.
1
Better Speed, Better Accuracy
■ Comparator function
Upper limit and lower limit values can
be set for both the main parameters
(any of Z or C or L or R) and subparameters (any of θ or D or Q). The
measurement results are signaled by a
buzzer and LED indication and can
also be output to an external source.
The output is separated into main- and
sub-parameter measurement results
together with AND.
■ Memory for 99 sets of
measurement conditions
Powerful Functions for Greater Line Efficiency
■ Minimum measurement time of 5 ms
Three sampling rates can be selected: FAST, NORMAL and SLOW. The minimum
measurement time of 5 ms (with 1 kHz/|Z| display) gives rapid sampling for improved
production line efficiency.
(Differs with the measurement frequency and display parameters.)
■High resolution and high measurement accuracy
The measurement resolution provides a full five digits, and the basic
measurement accuracy is ±0.08%.
■ RS-232C interface as standard feature
With the exception of turning the power on or off, all the basic functions can
be controlled from a PC. Use of a PC enables efficient data management,
processing, and setting of measurement conditions, plus a variety of other
functions. A GP-IB interface can also be installed as an option.
Up to 99 sets of measurement
conditions, including comparator
values, provide rapid response to
constantly changing components on
flexible production lines.
These conditions can also be externally
switched via the EXT.I/O.
■ Compact size
The small dimensions, 210 (W) × 100
(H) × 168 (D) mm, approximately 2.5
kg (4.00"W × 8.30"H × 6.60"D; 88 oz.
approx.), make it easy to incorporate
the instrument into production lines.
The AC power supply voltage is selectable :
100 V, 120 V, 220 V or 240 V AC.
9518-01 GP-IB interface can be fitted (optional).
■ RS-232C interface specifications
Transmission method: Start-stop synchronization. Transmission speed: 9600 bps. Data length:
8 bits. Parity: None. Stop bit: 1 bit. Delimiter: CR+LF. Handshake: Hardware. Connector shape:
D-sub 9pin (male). Connecting cable: Reverse cable
■ EXT.I/O
Trigger signals, recording ON/OFF, and loading of measurement conditions
can be externally controlled. Complete interface allows the unit to be used as
an automatic instrument where comparator results, measurement-completed
signals, etc., can be output to an external device.
EXT.I/O
Rear view
Timing chart for EXT. I/O sequencing
The following chart shows the timing sequence of the trigger
(TRIG), and end-of-measurement (EOM) signals from the
EXT. I/O connector.
EXT. I/O signals
● Outputs
• Internal DC power (+5 V output)
• Comparator result (main-/sub-
● Inputs
• External DC power supply (+5 V
Chuck ON
TRIG
Circuit response time
α* 1
500 µs
Chuck OFF
Trigger minimum interval
100 µs
Measurement time 5 ms *2
EOM
HI IN LO
to +24 V can be supplied by external device)
• External trigger signal
parameters together with AND output)
• Analog measurement completion • Memory setting selection
(including comparator conditions)
• End-of-measurement
RS-232C interface
*1 α depends on the sample and trigger delay.
*2 Reference value for 1 kHz measurement frequency,
FAST mode, |Z| measurement.
Measurement time differs with measurement conditions.
Decision result
2
... and Better Size !
Basic Performance
■ Seven parameters measured
■ Measurement signals
The seven parameters |Z|, R, θ, C, L, D, and Q can be
measured. The main- and sub-displays can be combined
in five ways: |Z|-θ, C-D, L-D, L-Q, R.
Measurement frequency: 120 Hz/1 kHz. Signal level:
50 mV, 500 mV, 1 Vrms settable.
■ Printer output
■ Easy operation by simple selections
and LED display
Measurement values and comparator results can be
printed out on the optional 9442 Printer by connecting
this via the standard RS-232C interface. This is
convenient for attaching data to inspection reports, etc.
To operate, simply select from the items displayed on
the panel. Selected measurement conditions are
indicated by illuminated LEDs allowing settings to be
checked at a glance. Measurement results are also
displayed by LED indication that makes it easy to check
the values even in dark locations.
(The optional 9444 Connection Cable and AC adapter are necessary for connecting the printer.)
Printout example
■ DC bias measurement
Using the optional 9268/9269 DC BIAS UNIT, voltage
and current bias measurements are simple.
The 9268 can be used for
voltages up to a maximum
of DC±40 V. The 9269 can
be used for currents up to a
maximum of DC±2 A.
■ 9442 PRINTER specifications
●Printing method : Thermal serial dot printer●Recording width : 112 mm (4.41”)●Printing
speed : 52.5 cps●Power supply : 9443 AC ADAPTER or supplied Ni-MH battery pack
(prints 3000 lines on full charge from 9443 AC ADAPTER)●Dimensions and mass: 160W
× 66.5H × 170D mm; 580 g approx. (6.30"W × 2.62"H × 6.70"D; 20.46 oz. approx. )
Resulting measurement data can be output not only to a printer, but also
other media such as a PC or sequencer. Using the RS-232C interface
makes transferring the inspection data simple and convenient.
Example of connecting the 9262 and 9268 / 9269
■ Specifications
Measurement parameters
|Z|, C, L, R, θ, D, Q
* Five possible display combinations: |Z|-θ, C-D, L-D, L-Q, R.
Measurement frequency (±0.01%)
120 Hz
1 kHz
Measurement time (typical values
for displaying |Z|)
Excluding time for open/short circuit
compensation, evaluation.
FAST : 13 ms, NORMAL : 80 ms,
SLOW : 400 ms
FAST : 5 ms, NORMAL : 60 ms,
SLOW : 300 ms
Measurement
ranges
|Z|, R
10 mΩ to 200.00 MΩ
C
9.40 pF to 999.99 mF
0.940 pF to 99.999 mF
L
14.00 µH∼200.00 kH
1.600 µH∼20.000 kH
θ
-90.00˚ to +90.00˚
D
0.0001 to 1.9900
Q
0.85 to 999.99
Basic accuracy
Measurement signal levels
Equivalent circuit mode
Output impedance
Z : ±0.08% rdg.
θ ± 0.05˚
50 mV/500 mV/1 V rms (±10% ±5 mV)
Serial- and parallel equivalent circuit mode, automatic/manual
50 Ω
Display method/Max. count
LED (5-digit display, full-scale count depends on range)
No. of measurement
condition memory retention
Max. 99 (including comparator conditions)
Comparator comparison
method
Any of the main parameters (any of |Z| or C or L or R) and
sub-parameters (any of θ or D or Q) can be set to upper limit and lower
limit value settings. The measurement results are signaled by LED
indication and a buzzer and EXT.I/O output (main- and sub-parameter
evaluation results, AND output).
DC bias
Possible when the optional 9268 (±40 V max.) or 9269 (±2 A max.) is used.
External printer
External interfaces
9442 PRINTER (option)
RS-232C, (GP-IB is option), EXT.I/O for sequence use.
Measurement range (Auto/Hold range, 5-digit display)
|Z|, R :
100 m/1/10/100/1 k/10 k/100 k/1 M/
10 M/200 MΩ
C (120 Hz) :
145 p/1.45 n/14.5 n/145 n/1.45 µ/14.5 µ/
145 µ/1.45 m/14.5 m/1 F
C (1 kHz) :
17 p/170 p/1.7 n/17 n/170 n/1.7 µ/17 µ/
170 µ/1.7 m/100 mF
L (120 Hz) :
130 µ/1.3 m/13 m/130 m/1.3/13/130/
1.3 k/13 k/200 kH
L (1 kHz) :
15.5 µ/155 µ/1.55 m/15.5 m/155 m/1.55/
15.5/155/1.55 k/20 kH
Dimensions, mass : 210H × 100W × 168D mm, 2.5 kg approx.
(8.30"H × 4.00"W × 6.60"D ; 88 oz. approx.)
Power supply :
100 V/120 V/220 V/ 240 V AC ± 10%
(selectable), 50/60 Hz
Max. rated power : 20 VA max.
Supplied accessories :
Power cord, spare fuse for power supply
(in accordance with the ordered power
specifications, either 100/120 VAC 1 A,
220/240 VAC 0.5 A)
Conformity : EMC EN61326-1:1997+A1:1998
EN61000-3-2:1995+A1:1998+A2:1998
EN61000-3-3:1995
Safety EN61010-1:1993+A2:1995
Power supply; Pollution degree 2 Overvoltage Category II
(anticipated transient overvoltage 2500 V)
Test terminals; Pollution degree 2 Overvoltage Category I
(anticipated transient overvoltage 330 V)
3
Measurement accuracy and range
Conditions of guaranteed accuracy :
Temperature and humidity 23˚C±5˚C (73˚F±9˚F), less than 80% RH (no condensation), following 60 min. warm-up after power is turned ON,
after open/shut calibration, use of 9261 Test Fixture, measurement signal level 1 Vrms, measurement speed set to SLOW.
The various accuracy specifications presume that θ< ±6˚C for R, D≤0.1 for C-D, D≤0.1 for L-D, Q≥10 for L-Q.
Q accuracy is defined by the calculation of 1/D.
Measurement range and accuracy differ with the used Test Fixture, measurement signal level and measurement speed.
Frequency
Range
(|Z|-θ and
R have
common
frequency)
IZI-θ
R
100 mΩ
1Ω
10 Ω
100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
200 MΩ
IZI
± (1.00+0.15/ZL)%
±1.80%
±0.35%
±0.08%
±0.08%
±0.11%
±0.14%
±0.30%
±(0.15+0.16✕ZH)%
±(2.00+0.11✕ZH)%
θ
±(0.10+0.09/ZL)˚
±1.00˚
±0.18˚
±0.08˚
±0.05˚
±0.08˚
±0.10˚
±0.19˚
±(0.10+0.09✕ZH)˚
±(0.70+0.08✕ZH)˚
-
±(1.00+0.21/RL)%
±2.10%
±0.39%
±0.10%
±0.09%
±0.13%
±0.16%
±0.34%
±(0.15+0.20✕RH)%
±(2.00+0.16✕RH)%
120 Hz
1F
14.5 mF 1.45 mF
145 µF
14.5 µF
1.45 µF
145 nF
14.5 nF
1.45 nF
145 pF
1 kHz
100 mF
1.7 mF
170 µF
17 µF
1.7 µF
170 nF
17 nF
1.7 nF
170 pF
20 pF
C
±(0.60+1.50✕f✕CH)%
±2.10%
±0.39%
±0.10%
±0.09%
±0.13%
±0.16%
±0.34%
D
±(0.0015+0.0108✕f✕CH) ±0.0179
±0.0034
±0.0016
±0.0011
±0.0016
±0.0020
±0.0036 ±{0.0020+0.264/(f✕CL)} ±{0.0120+0.25/(f✕CL)}
13 mH
130 mH
1.3 H
13 H
130 H
1.3 kH
13 kH
200 kH
1.55 kH
20 kH
±{0.17+30/(f✕CL)}% ±{1.70+30/(f✕CL)}%
C-D
120 Hz
130 µH
1.3 mH
1 kHz
15.5 µH
155 µH 1.55 mH 15.5 mH 155 mH
1.55 H
15.5 H
155 H
L
±{0.90+30/(f✕LL)}%
±2.10%
±0.39%
±0.10%
±0.09%
±0.13%
±0.16%
±0.34%
±{0.0021+0.264/(f✕LL)} ±0.0179
±0.0034
±0.0016
±0.0011
±0.0016
±0.0020
±0.0036 ±(0.0020+0.0110✕f✕LH) ±(0.0120+0.0100✕f✕LH)
±(0.17+1.17✕f✕LH)% ±(2.00+1.00✕f✕LH)%
L-D
D
* ZL is the sample impedance [Ω], ZH is the sample impedance [MΩ], RL is the sample resistance [Ω], RH is the sample resistance [MΩ], CH is the sample capacitance [mF],
CL is the sample capacitance [pF], LL is the sample inductance [µH], LH is the sample inductance [kH], and f is the measurement frequency [kHz]. ( |Z|, R, C, L : ± %rdg.)
■ Options for a wide range of applications
9140 FOUR-TERMINAL
PROBE
DC to 100 kHz
9143 PINCHER PROBE
DC to 5 MHz
9261 TEST FIXTURE
DC to 5 MHz
9262 TEST FIXTURE
DC to 5 MHz
9263 SMD TEST
FIXTURE
DC to 5 MHz
* All cable lengths are 1 m (39.37”).
3511-50 LCR HiTESTER
[Standard accessories: power cord, spare power fuse (1 A for 100/120 V
AC rating, 0.5 A for 220/240 VAC rating) ]
Test fixtures are not supplied with the unit.
Select an optional test fixture when ordering.
■ Options
9140 FOUR-TERMINAL PROBE
9143 PINCHER PROBE
9261 TEST FIXTURE
9262 TEST FIXTURE (direct connection type)
9263 SMD TEST FIXTURE (direct connection type)
9268 DC BIAS VOLTAGE UNIT
9269 DC BIAS CURRENT UNIT
9268 DC BIAS VOLTAGE UNIT
Maximum applied voltage: ± 40 V DC
9269 DC BIAS CURRENT UNIT
Maximum applied current: ± 2 A DC
9165 CONNECTION CORD (for 9268/9269; BNC to BNC; 1.5 m/59.05”)
9166 CONNECTION CORD (for 9268/9269; BNC to clips; 1.5 m/59.05”)
9518-01 GP-IB INTERFACE
9151-02 GP-IB CONNECTION CABLE (2 m/78.74”)
9151-04 GP-IB CONNECTION CABLE (4 m/157.48”)
9442 PRINTER
9443-01 AC ADAPTER (for 9442, Japan)
9443-02 AC ADAPTER (for 9442, EU)
9443-03 AC ADAPTER (for 9442, USA)
9443-02
9443-01
9444 CONNECTION CABLE (for 9442 / 1.5 m/59.05”)
1196 RECORDING PAPER (for 9442 / 25 m/984.25”, 10 rolls)
DISTRIBUTED BY
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@ hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@ hiokiusa.com
Europe Representative Office :
Meineckestrasse 48, 40474 Dusseldorf, Germany
TEL / FAX +49-221-4544153
E-mail: hioki-eu@ doitsu.de
Shanghai Representative Office :
1108 Union Building, 100 Yan An Road (East),
Shanghai, 200002, P.R.China
TEL +86-21-6328-9947/4938
FAX +86-21-6328-2064
E-mail: [email protected]
All information correct as of Jul. 19, 2001. All specifications are subject to change without notice.
3511-50E1-17M-05P
Printed in Japan