INSULATION/VOLTAGE WITHSTAND TEST INSTRUMENT series

INSULATION / VOLTAGE WITHSTAND TEST
INSTRUMENT SERIES
Safety Standards
Measuring Instruments
3159
3153
3158
3154
This electrical safety test instrument series is designed
for insulation resistance and voltage withstand testing of
electrical devices and components according to various
safety standards. A multitude of automation and laborsaving features are provided to ensure effective testing
for a wide variety of requirements and test conditions.
Select the most appropriate model for your applications.
3153 AUTOMATIC INSULATION /
WITHSTANDING HiTESTER
3159 INSULATION / WITHSTANDING HiTESTER
3158 AC WITHSTANDING VOLTAGE HiTESTER
3154 DIGITAL MΩ HiTESTER
1
Model 3153 Automatic Insulation Voltage Withstand Testing
Standards-Based Testing
Voltage Control from a PC
Full Remote Control
Comparator/Timer
All test parameters can be controlled by RS-232C or GP-IB,
including test voltage, cut-off current, resistance threshold
and timer durations. Start-stop control can be provided
with the 9613 single hand remote control or 9614 twohand remote control.
Includes built-in pass-fail comparator and timer functions
for easy compliance testing to various safety standards
such as those for Electrical Appliance Safety Regulations.
Settable Ramp Up/Down Test Voltage Timing
Test Status Read/Write
Program Function
Raising and lowering of
test voltage can be set for
voltage withstand testing.
Test Voltage
Ramp Timer Functions
Rise
Test
Fall
Test conditions can be stored and recalled to support
complete automation by sequencer. Up to 32 files can be
stored with up to 50 steps per file.
Up to 32 files
Time
Fluorescent Display Tube
The display uses a bright, easy-toread fluorescent tube.
Analog Voltmeter
The test voltage can be verified not
only on the digital display but also on
the analog meter.
Danger Lamp
The warning light flashes
during testing and
whenever high voltage is
present at the terminals.
External Switch
Start/stop may be controlled with the
9613 or the 9614 .
(The 9613 and 9614 are options.)
Test Mode Selection
Three test modes are selectable:
1. Manual test modes: ACW, DCW, I
2. Automatic test modes: W→ I, I → W
3. Programmed test modes: testing by preprogrammed functions.
● Automation Support
● Save 10 Sets of Test Conditions
● Interlock Protect Function
● Auto Discharge Feature
● Zero-V Switching
● PWM Switching Technique
Automation features include programmable
testing modes, EXT I/O, RS-232C, GP-IB,
connection scanning and various data
management functions.
Any charge on the object under test is
discharged by the test instrument, so there is
no residual charge after testing. (DC voltage
Store up to 10 sets of test conditions for
voltage withstand and insulation resistance
modes, so you can quickly switch among the
test conditions. (Save/Load)
Test voltage on/off switching can be forced to
occur only at sine wave zero-crossings.
(AC voltage withstand testing)
To ensure safety during automatic testing, test
output can be interrupted and testing inhibited
by input signals from automatic sensing
devices.
Enhanced accuracy is obtained by preventing
variations in supply voltage from affecting test
voltage.
2
Wide Range of Functions for Various Conditions
1. Pass Hold Function (0: No Hold, 1: Hold)
The pass state is held when it is activated.This is
convenient for verifying the decision value.
2. Fail Hold Function (0: No Hold, 1: Hold)
The fail state is held when it is activated.This is
convenient for temporarily stopping the test process.
3. Hold State (0: No Hold, 1: Hold)
This saves the state when the Stop key is pressed during
a test to unconditionally end the test.
4. Momentary Out (0: Disabled, 1: Enabled)
This function outputs a voltage only when the Start key
is being pressed. The Start key is effective both for EXT
SW and external I/O.
5. Double Action (0: Disabled, 1: Enabled)
This function allows testing to start only if the Start key
is pressed within a half second after the Stop key.
6. Fail Mode (0: Disabled, 1: Enabled)
This function allows the Hold state to be released only
by the Stop key on the instrument panel.
7. "START" Interface Command (0: Disabled, 1: Enabled)
This specifies whether the "START" command is
enabled.
8. Interlock Function (0: Disabled, 1: Enabled)
This specifies whether the interlock terminal for external
I/O is enabled.
9. Maximum Output Voltage
Sets the upper limit of the test voltage.
10. Insulation Resistance Measurement Range
(0: Fixed Range, 1: Automatic Range)
This specifies whether the measurement range for
insulation resistance testing should be fixed or
automatically determined.
1 2 3 4
5 6 7 8
10111213 1415
Address specified for GP-IB control
9
(Various functions can be specified with the SHIFT + STOP keys)
11. Insulation Resistance Test End Mode
0: Test for the specified time
1: Stop when "pass" is detected
2: Stop when "fail" is detected
This specifies the method of ending insulation
resistance tests.
12. Ramp Time Setting
0: No judgment during ramp-up
1: Judgment during ramp-up
This specifies whether the judgment is enabled during
ramp-up. Valid only during voltage withstand testing.
13. PC Interface
0: RS-232C (PC, 9600 bps)
1: RS-232C (PC, 19200 bps)
2: GP-IB
This specifies the type of PC interface to use.
14. Electrical Discharge Function
(0: Disabled, 1: Enabled)
This specifies whether the electrical discharge function
is enabled at the end of testing.
15. Test Signal Output
0: ON also when TEST indicator is flashing
1: OFF when TEST indicator is flashing
2: ON only when TEST indicator is flashing
(excluding ramp down time)
This specifies whether the TEST signal of the external I/
O should be outputwhen the TEST indicator is flashing.
Enhanced System Measurements
Maximum 32-Channel Multi-Point Testing
Simultaneous Protective Ground Continuity Testing
Model 3930 HIGH VOLTAGE SCANNER
Safety Inspection System
Combine Model 3153 with the 3930 HIGH VOLTAGE
SCANNER to perform insulation withstand testing
easily. Single-end inputs test up to 8 points (between
any 4 points) per instrument, and can connect up to 4
instruments together.
Model 3930 Specifications
No. of Channels
Operating Voltage
Multi-Mode: 4 High-Low channels
Single-End Mode: 8 High-Common channels
5 kV AC, 7 kV DC
Action/Recovery Time 6 ms or less
Supply Voltage
24 V DC ±5% (at control signal input connector)
Size & Mass
Approx. 320(W) × 90(H) × 250(D) mm, 3 kg.
Combine Model 3153 with the 3157-01 AC GROUNDING
HiTESTER and a general-purpose sequencer for a simple
safety test inspection system that includes protective
ground continuity and insulation withstand testing.
Model 3157-01
AC GROUNDING HiTESTER
Settable current
ranges:
3.0 to 31 A AC
Max. output power:
130VA
Resistance
measurement range:
0 to 1.800 Ω
3
A Full Line-up of Models to Suit Various Needs
Model 3153
AUTOMATIC INSULATION /
WITHSTANDING HiTESTER
■ Insulation and AC/DC continuous voltage withstand testing
■ Safety standards compliance testing
■ Stores up to 10 sets of test conditions
(10 each for insulation and voltage withstand)
■ Zero-V Switching (voltage withstand testing)
■ Electrical Discharge function
(DC voltage withstand, insulation testing)
■ Automated testing support
■ Interlock Protect function
■ Remote controllable
EXT I/O
GP-IB
RS-232C
Scanner Connector
Voltage Withstand Testing
Insulation Testing
AC: 0.20 to 5.00 kV, 500 VA
DC: 0.20 to 5.00 kV, 50 VA
0.01 to 100.0 mA
DC: 50 to 1200 V
0.10 to 9999 MΩ
Model 3159
INSULATION / WITHSTANDING
HiTESTER
Output Terminal
■ Insulation and voltage withstand testing
■ Safety standards compliance testing
■ Stores up to 10 sets of test conditions
(10 each for insulation and voltage withstand)
■ Voltage comparator functions (voltage withstand testing)
■ Zero-V Switching (voltage withstand testing)
■ Electrical Discharge function (insulation testing)
■ Automated testing support
■ Interlock Protect function
Status Out
■ Remote controllable
Beeper Volume
Adjust
Output Terminal
Voltage Withstand Testing
Insulation Testing
AC: 0 to 2.5 kV/5.0 kV, 500 VA
DC: 500/1000 V
0.01 to 120.0 mA
0.5 to 2000 MΩ
RS-232C
Model 3158
AC WITHSTANDING VOLTAGE HiTESTER
EXT I/O
■ Safety standards compliance testing
■ Stores up to 20 sets of test conditions
■ Voltage comparator functions
■ Zero-V Switching
■ Electrical Discharge function
■ Automated testing support
■ Interlock Protect function
■ Selectable remote controls
Beeper Volume Adjust
Voltage Withstand Testing
AC: 0 to 2.5 kV/5.0 kV, 500 VA
0.01 to 120.0 mA
RS-232C
EXT I/O
4
EXT I/O Output Signals
External control can be provided by various signals (signal lines have photocoupler isolation)
Pin
I/O
Signal
Function
1
OUT READY LO when in "ready state"
3
OUT U-FAIL LO when in "fail state" for the upper bound
2
4
5
6
7
OUT L-FAIL LO when in "fail state" for the lower bound
OUT
TEST
LO when in "pass state"
LO when in "test state"
OUT H.V.ON LO when voltage is present at the output terminals
IN
8
EXT-E When LO, external I/O input signals are enabled
IN
9
10
PASS
OUT
START When LO, it functions as a "Start" key
IN
STOP When LO, it functions as a "Stop" key
IN INT.LOCK Interlock engaged when open
11
OUT W-MODE LO during voltage withstand testing
13
OUT W-FAIL LO when in "fail state" for voltage withstand testing
12
14
15-16
OUT I-MODE LO during insulation testing
OUT
Pin
I/O
Signal
Function
17-18 IN EXT.COM Common terminals for external devices
19
20
21
22
23
24
25
26
OUT STEP-END LO when at the end of a step
OUT FILE-END LO when at the end of a file
IN
FILE-E LO when FILE 0 to 4 is in use
IN
FILE-1
IN
IN
IN
IN
FILE-0
File selection
FILE-2
File selection
FILE-3
FILE-4
File selection
File selection
File selection
33-34 OUT ISO.DCV Internal power 5V DC (60 mA)
35-36 IN EXT.DCV External power supply (5 to 30V DC)
I-FAIL LO when in "fail state" for insulation testing
IN ISO.GND Ground inputs for external devices
Pin
Status Out
When the output conditions set by
the DIP switches are satisfied (OR
condition), output is provided at
relay contacts.
1. H.V.ON
Output voltage generation
3. PASS
Passed
2. TEST
4. FAIL
5. INT.LOCK
6. READY
7. EXT.CONT.
8. POWER.ON
3
OUT U-FAIL LO when in "fail state" for the upper bound
2
4
5
6
7
8
9
Failed
12
Under external control
Powers the 3159 on
OUT L-FAIL LO when in "fail state" for the lower bound
OUT
OUT
PASS
TEST
LO when in "pass state"
LO when in "test state"
OUT H.V.ON LO when voltage is present at the output terminals
IN
IN
IN
EXT-E When LO, external I/O input signals are enabled
START When LO, it functions as a "Start" key
STOP
When LO, it functions as a "Stop" key
IN INT.LOCK Interlock engaged when open
11
OUT W-MODE LO during voltage withstand testing
13
OUT W-FAIL LO when in "fail state" for voltage withstand testing
14
15-18
OUT I-MODE LO during insulation testing
OUT
I-FAIL LO when in "fail state" for insulation testing
IN ISO.COM Ground inputs for external devices
33-36 OUT ISO.DCV Internal power 15V DC (100 mA)
Pin
I/O
Signal
OUT READY LO when in "ready state"
3
OUT U-FAIL LO when in "fail state" for the upper bound
4
5
6
7
8
9
10
15-18
OUT L-FAIL LO when in "fail state" for the lower bound
OUT
OUT
PASS
TEST
LO when in "pass state"
LO when in "test state"
OUT H.V.ON LO when voltage is present at the output terminals
IN
IN
IN
EXT-E When LO, external I/O input signals are enabled
START When LO, it functions as a "Start" key
STOP
1. PASS Hold function
2. FAIL Hold function
3. Hold function
4. Momentary out
5. Double actions
6. FAIL mode
7. "START" RS command
8. Interlock function
9. Voltage Comparator position
10.Insulation Resistance
measurement range
11.Insulation Resistance Test
End mode
Function
1
2
1. PASS Hold function
2. FAIL Hold function
3. Hold function
4. Momentary out
5. Double actions
6. FAIL mode
7. "START" interface command
8. Interlock function
9. Maximum Output Voltage
10.Insulation Resistance
measurement range
11.Insulation Resistance Test
End mode
12.Ramp Timer setting
13.PC Interface
14.Electrical Discharge function
15.TEST signal output
Function
OUT READY LO when in "ready state"
10
Ready
Signal
1
Testing in progress
Interlocked
I/O
Various Function Settings
When LO, it functions as a "Stop" key
IN INT.LOCK Interlock engaged when open
IN ISO.COM Ground inputs for external devices
33-36 OUT ISO.DCV Internal power 15V DC (100 mA)
1. PASS Hold function
2. FAIL Hold function
3. Hold function
4. Momentary out
5. Double actions
6. FAIL mode
7. "START" RS command
8. Interlock function
9. Voltage Comparator position
5
Specifications
■ Voltage Withstand Testing
Model 3153
AUTOMATIC INSULATION / WITHSTANDING HiTESTER
Output voltage
Voltage output
method
Transformer capacity
Model 3159
Model 3158
INSULATION / WITHSTANDING HiTESTER
AC WITHSTANDING VOLTAGE HiTESTER
AC
DC
AC
0.20 to 5.00 kV
PWM switching method
(zero-switching)
0.20 to 5.00 kV
Two ranges: AC 0 to 2.5 or 5.0 kV
PWM switching method
Zero-switching
500 VA (rated 30 minutes)
...
500 VA (rated 30 minutes)
...
50 VA (continuous)
Output capacity
Voltage adjustment
method
Output voltage
accuracy
...
Digital setting (0.01 kV setting resolution)
Manual adjustment
±1.5% of setting voltage ±2 dgt.
...
Voltage change rate
±7% or less
±16% or less
(max. 5 kV at 100 mA →
(max. 5 kV at10 mA → unloaded:
unloaded: with resistive load)*2
with resistive load)*2
...
Voltage waveform
Sine wave
(5% or less distortion, unloaded)
...
Power waveform
Voltage frequency
50 or 60 Hz ±0.2%
...
Power synchronization
Output ripple voltage
6% of output voltage or less
...
Output current
100 mA *1
10 mA (continuous)
Average rectified effective value display
Average display
Voltmeter
Digital: 0.00 to 5.00 kV (full scale)
Accuracy: ±1.5% f.s.
Analog: 0 to 5 kV (full scale)
Accuracy: ±5% f.s.
...
(at 5 kV DC, 10 mA, resistive load)
...
Average rectified effective value display
Digital: 0.00 to 5.00 kV (full scale)
Accuracy: ±1.5% f.s.
Analog: 0 to 5 kV (full scale)
Accuracy: ±5% f.s.a
Current measurement
range
0.01 to 100.0 mA
0.01 to 10.0 mA
0.01 to 120.0 mA
Indicated value range
10 or 100 mA
10 mA
2, 8, 32 or 120 mA
Measurement
resolution
Current measurement
accuracy
0.00 to 10.00 or 0.01 mA (10-mA range)
10.1 to 100.0 or 0.1 mA (100-mA range, AC only)
± (2% rdg. + 5 dgt.) common to each range *3
*1. Time vs. Output Voltage (at 23˚C ambient)
Current Measurement Range
Max. Test Time
Standby Time
60 mA < 1 < 100 mA
15 minutes
15 minutes
1 < 60 mA
0.01 mA (2- or 8-mA range), 0.1 mA (32-mA range),
1 mA (120-mA range)
continuous
± (3% f.s. + 20 µA) for all ranges (at 5% power distortion or less)
*2. Unloaded = 40 MΩ load (instrument input impedance)
*3. Plus scanner accuracy, when used.
none
■ Insulation Resistance Testing
Model 3153
Model 3159
Model 3158
Test voltage
Output voltage: Positive polarity 50 to 1200V DC
Voltage adjustment method: Digital setting (1V resolution)
Output voltage accuracy: ±1.5% ±2 dgt. of setting level
Rated voltage: 500 or 1000V DC
Unloaded voltage: 1 to 1.2 times rated
voltage
...
Rated measurement
current
1 mA
1 to 1.2 mA
...
Short-circuit current
200 mA or less
4 to 5 mA (500V)
2 to 3 mA (1000V)
...
Voltmeter
Average display
Digital: 0 to 1200V DC (full scale)
Accuracy: ±1.5% rdg. ±2 dgt.
Analog: 0 to 1200V DC
Accuracy: ±5% f.s. (5 kV full scale)
Measurement range/
accuracy
0.100 to 1.049 MΩ
1.05 to 10.49 MΩ*1
10.5 to 104.9 MΩ*1
105 to 9999 MΩ*1
Fundamental accuracy: ±4% rdg.*2
*1. Measurement range changes according to test voltage.
*2. Plus scanner accuracy, when used.
Average display
Digital: 0 to 1200V DC (f.s.)
...
Analog: not applicable
0.5 to 999 MΩ (500V)/±4% rdg.
1 to 999 MΩ (1000V)/±4% rdg.
1000 to 2000 MΩ /±8% rdg.
...
6
■ Decision Function
Decision method
Decision results
Decision processing
Model 3153
Model 3159
Window comparison method (digital specification)
Model 3158
UPPER-FAIL: Measured current (insulation resistance value) exceeded the specified upper bound.
PASS: Measured current (insulation resistance value) was between the specified upper and lower bounds during the specified time elapsed
LOWER-FAIL: Measured current (insulation resistance value) was less than the specified lower bound
(Note: Model 3158 has no insulation resistance function)
For each decision result, output the display portion, the beeper sound, and EXT I/O signal
Specification ranges
Voltage withstand testing:
ACV: 0.1 to 100 mA (upper bound) / 0.1 to 99 mA (lower bound)
DCV: 0.1 to 10 mA (upper bound) / 0.1 to 9.9 mA (lower bound)
Insulation testing: 0.10 to 9999 MΩ (same upper/lower bounds)
Voltage withstand testing:
0.1 to 120 mA (upper bound) / 0.1 to 119 mA (lower bound)
Insulation testing (Model 3159 only):
0.2 to 2000 MΩ (same upper/lower bounds)
Specification
resolution
Voltage withstand testing:
0.1 mA (0.1 to 9.9 mA), 1 mA (10 to 100 mA)
Insulation testing:
0.01 MΩ (0.10 to 9.99 MΩ), 0.1 MΩ (10.0 to 99.9 MΩ),
1 MΩ (100 to 9999 MΩ)
Voltage withstand testing:
0.1 mA (0.1 to 9.9 mA), 1 mA (10 to 120 mA)
Insulation testing (Model 3159 only):
0.01 MΩ (0.2 to 2 MΩ), 0.1 MΩ (2.1 to 20 MΩ),
1 MΩ (21 to 200 MΩ), 10 MΩ (210 to 2000 MΩ)
■ Timers
Model 3153
Testing timer
Specification range: 0.3 to 999 s
Specification resolution: 0.1 s (0.3 to 99.9 s), 1 s (100 to 999 s)
Accuracy: ±0.5% of specified value
Model 3159
Model 3158
Specification range: 0.5 to 999 s
Specification resolution/accuracy:
0.1 s (0.5 to 99.9 s), ±50 ms; 1 s (100 to 999 s) ±0.5 s
Action: (when ON is specificed) after starting, the countdown from the specified time is displayed.
(when OFF is specified) displays the elapsed time from starting
Ramp timer
(withstand test time)
Specification range: 0.1 to 99.9 s ramp-up and -down specified separately
Specification resolution/accuracy: 0.1 s, ±0.5% of specified value
Delay timer
(insulation resistance
test time)
Specification range: 0.1 to 99.9 s
Specification resolution/accuracy: 0.1 s, ±0.5% of specified value
Action: specify a delay time after testing is set to begin to inhibit
decisions during that time
...
Non-deterministic interval: 0.5 s
...
(Mask time until determination begins
during insulation resistance testing)
■ Interfaces
Model 3153
Model 3159
EXT I/O
Open-collector outputs, active low, max. 30V DC loaded voltage, all signal lines photocoupler-isolated
EXT SW
START, STOP, SW.EN (panel terminal switch enabled), connection point inputs
RS-232C
Start-stop synchronization, full duplex, 9600 or 19200 bps
GP-IB
IEEE 488.2 (1987) compliant
Model 3158
Start-stop synchronization, full duplex, 9600 bps
...
■ General Specifications
Model 3153
Display
Fluorescent display tube (digital display), analog meter
Monitor functions
Output voltage, detected current, measured resistance
Monitor period
2 times per second minimum
Model 3159
Model 3158
Output voltage, detected current
Operating temperature range 0 to 40 ˚C, 80% RH maximum (non-condensating)
Storage temperature range -10 to 50 ˚C, 90% RH maximum (non-condensating)
Temperature and humidity
range for guaranteed accuracy 23 ± 5 ˚C, 80% RH maximum (non-condensating) (after 10-min. warm-up for 3153, or 5-min. warm-up for 3158 and 3159)
Operating environment
Indoors, up to 2000m ASL
Power supply voltage
100 to 240V AC (installed fuse depends on actual voltage, so specify
100V AC (3159), 120V AC (3159-01),
supply voltage when ordering)
220V AC (3159-02), 230V AC (3159-03),
100 to 120V AC: installed fuse 250V T10AL
240V AC (3159-04)
200 to 240V AC: installed fuse 250V T5AL
120V AC (3158-01), 220V AC (3158-03),
230V AC (3158-04), 240V AC (3158-05)
Power supply frequency 50 or 60 Hz
Max. power consumption 1000 VA
800 VA
Dimensions
Approx. 320 (W) × 155 (H) × 480 (D) mm
Approx. 320 (W) × 155 (H) × 330 (D) mm
Approx. 320 (W) × 155 (H) × 263 (D) mm
Mass
Approx. 18 kg
Approx. 18 kg (3159),
20.5 kg (3159-01),
21.5 kg (3159-02 to 3159-04)
Approx. 16 kg (3158-01),
18 kg (3158-03 to 3159-05)
Supplied accessories
9615 H.V. TEST LEADS (high voltage side and return, one each), POWER CORD, EXTRA FUSE
Options
9613 REMOTE CONTROL BOX (SINGLE), 9614 REMOTE CONTROL BOX (DUAL), 9637 RS-232C CABLE (9-pin Dsub to 9-pin Dsub),
9638 RS-232C CABLE (9-pin Dsub to 25-pin Dsub), 9267 SAFETY TEST DATA MANAGEMENT SOFTWARE
3930 HIGH VOLTAGE SCANNER
9151-02 GP-IB CONNECTOR CABLE (2m)
9151-04 GP-IB CONNECTOR CABLE (4m)
9616 WARNING LAMP
7
■ Six test voltages from 25 to 1000V
■ Stores 10 setting states
■ Easy standards testing with comparator timer function
■ Automatic electrical discharge feature
■ Minimize instability with slow sampling
■ Record changes with analog output
Model 3154
DIGITAL MΩ HiTESTER
[Measurement Voltage/Range (Auto/Manual range switching)]
Measurement voltage
Measurement range
25V
50V
±2% rdg. ±5 dgt.
±5% rdg.
19.0 to 200.0 MΩ
Unloaded voltage
1000V
0 to 100.0 MΩ
19.0 to 2000 MΩ 100.1 to 2000 MΩ
0 to 999 MΩ
1000 to 4000 MΩ
0.025 MΩ
0.05 MΩ
0.1 MΩ
0.25MΩ
0.5MΩ
2MΩ
Rated meas. current
1 to 1.2 mA
0.5 to 0.6 mA
Short-circuit current
1.2 mA or less
0.6 mA or less
■ Options
3930 HIGH VOLTAGE SCANNER
9613 REMOTE CONTROL BOX (SINGLE)
9614 REMOTE CONTROL BOX (DUAL)
9151-02 GP-IB CONNECTOR CABLE (2m)
9151-04 GP-IB CONNECTOR CABLE (4m)
9637 RS-232C CABLE (1.8 m) (9pin-9pin/Cross)
9638 RS-232C CABLE (1.8 m) (9pin-25pin/Cross)
9267 SAFETY TEST DATA MANAGEMENT SOFTWARE
WITHSTANDING
WITHSTANDING
WITHSTANDING
WITHSTANDING
500V
2.000, 20.00, 200.0 and 4000 MΩ
1 to 1.2 times the measurement voltage
3153 AUTOMATIC INSULATION / WITHSTANDING HiTESTER 3158-01 AC
3158-03 AC
3158-04 AC
3158-05 AC
250V
2.000, 20.00 and 2000 MΩ
0 to 20.00 MΩ
Accuracy
Min. meas. resistance
100V
2.000, 20.00 and 200.0 MΩ
3159 INSULATION / WITHSTANDING HiTESTER (100V AC) 3159-01 INSULATION / WITHSTANDING HiTESTER (120V AC) 3159-02 INSULATION / WITHSTANDING HiTESTER (220V AC) 3159-03 INSULATION / WITHSTANDING HiTESTER (230V AC) 3159-04 INSULATION / WITHSTANDING HiTESTER (240V AC) ■ Options
9613 REMOTE CONTROL BOX (SINGLE)
9614 REMOTE CONTROL BOX (DUAL)
9616 WARNING LAMP
9637 RS-232C CABLE (1.8 m) (9pin-9pin/Cross)
9638 RS-232C CABLE (1.8 m) (9pin-25pin/Cross)
9267 SAFETY TEST DATA MANAGEMENT SOFTWARE
VOLTAGE HiTESTER (120V AC)
VOLTAGE HiTESTER (220V AC)
VOLTAGE HiTESTER (230V AC)
VOLTAGE HiTESTER (240V AC)
9613 ■ Options
REMOTE CONTROL BOX (SINGLE)
9613 REMOTE CONTROL BOX (SINGLE)
9614 REMOTE CONTROL BOX (DUAL)
9637 RS-232C CABLE (1.8 m) (9pin-9pin/Cross)
9638 RS-232C CABLE (1.8 m) (9pin-25pin/Cross)
9267 SAFETY TEST DATA MANAGEMENT SOFTWARE
9616
WARNING LAMP
(100V AC, 0.1A)
9615
H.V. TEST LEADS
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
1904 Shanghai Times Square Office, 93 Huai Hai Zhong Road
Shanghai, P.R.China POSTCODE: 200021
TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360
E-mail: [email protected]
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: [email protected]
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: [email protected]
9614 REMOTE CONTROL BOX (DUAL)
When using Model 9616 with Models
3153 or 3158, please contact HIOKI for
specific details.
DISTRIBUTED BY
Beijing Office :
A-2602 Freetown, 58 Dong San Huan Nan Road
Beijing, P.R.China POSTCODE: 100022
TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090
E-mail: [email protected]
Guangzhou Office :
Room 303, Profit Plaza, No.76, West Huangpu Road
Guangzhou, P.R.China POSTCODE: 510623
TEL +86-20-38392673/2676 FAX +86-20-38392679
E-mail: [email protected]
All information correct as of Jun. 29, 2007. All specifications are subject to change without notice.
3153E4-76E
Printed in Japan