3157-01 AC GROUNDING HiTESTER
Electronic Measuring Instrument
CE certified low-resistance measurement compliant with major safety standards
Protective ground tester indispensable for
standard certification
ISO14001
JQA-E-90091
The 3157-01 AC GROUNDING HiTESTER is designed to measure
whether the metal enclosure of an electrical equipment is connected to
the ground terminal at sufficiently low resistance levels. It also can be
used to evaluate the grounding conditions of large-scale electrical
installations. Measurement is carried out by using a high current
according to the specifications of the measurement object, and
determining the voltage drop at the measurement point. Reference
values are as set out in the various safety standards. The 3157-01 can
carry out measurements in accordance with the stipulations of multiple
standards.
http://www.hioki.co.jp/
HIOKI company overview, new products, environmental considerations and
other information are available on our website.
1
Low-resistance measurements
in accordance with all major
Main applications
safety standards
The 3157-01 passes a large AC current through the
measurement object and measures the voltage drop
according to the AC 4-terminal method, making it
possible to measure very low resistance values.
● Protective grounding checks of medical and
general electrical equipment
● Ground connection tracing of machine tools
and wiring panels
● Safeguard and equal-potential connection
checks of medical installations
● High-current behavior evaluation of connections
Major features
■ Compliant with a multitude of standards
■ “SOFT START” function
The 3157-01 allows measurement as prescribed by most
major safety standards. Using the 4-terminal method to
measure the voltage drop for a high current, the unit offers
evaluation features and a timer function to allow efficient
standard compliance testing.
The unit checks whether the probe is connected to the
measurement object, and raises the output current to the preset
value when a connection is detected. This serves to prevent
sparks caused by connecting a live probe to a measurement
object, thereby guarding against equipment damage and
ensuring operator safety.
■ Constant-current testing (max. 31.0 A) with
feedback control
■ Fluorescent tube display (VFD)
The output current is controlled by a feedback loop to achieve
stability, regardless of fluctuations in the load impedance.
The display uses an easy to read fluorescent tube. Compared
to conventional meters, the digital indication allows effortless
reading of the data.
■ Test data count function
For installations with many test points, the unit can
automatically count the number of tests, to ensure that no
points are missed.
■ Setting value store function
Up to 20 settings can be stored, allowing quick switching
between the various setups for different standards and legal
requirements.
■ Light weight and compact dimensions
Whereas conventional testing equipment required a trolley for
transport, the 3157-01 can be easily carried with one hand. Its
small dimensions, light weight, and ease of maintenance make
it ideal for use in the field.
[320 (W) × 90 (H) × 263 (D) mm
12.6" (W) × 3.56" (H) × 10.40" (D)
7 kg(247.2 oz)]
Standards supported by the 3157-01
●IEC60065
[S a f e t y r e q u i r e m e n t s f o r m a i n s o p e r a t e d
electronic and related apparatus for household
and similar general use]
●IEC60204-1
[Electrical equipment of industrial machines
-Part1,General requirements]
●IEC60335-1
[Safety of household and similar electrical
appliances - Part 1, General requirements]
●IEC60601-1
[Medical electrical equipment -Part 1, General
requirements for safety]
●IEC60950
[Safety of data processing equipment, including
office equipment]
●IEC61010-1
[Safety requirements for measurement, control,
and laboratory electrical equipment]
●UL standard
Relevant standards
(UL 1012, UL 1270, UL 1409, UL 1419, UL 1437,
UL 2601, etc.)
2
A multitude of functions in a compact body
■ Easy standard testing
Measurement current
■ Measurement range
Upper limit
Load resistance [Ω]
Measurement time
Open terminal voltage 6 V line
Preparation
1.8
Maximum output power 130 VA line
Test condition file number
1.5
Remaining time
For 6 A output, 6 V line yields
Measurement
6 (V)
6 (A)
1.0
=1.0 [ Ω ]
For 25 A output, 130 VA line yields
Result
PASS
130 (VA)
=0.208 [ Ω ]
252 (A2)
FAIL
0.5
Measurement range
Measurement value
0
Measurement value
0
* If hold is not enabled, unit reverts to READY condition after 1 second.
10
20
30
Output current [ A ]
■ Versatile functions
➀ ➁ ➂ ➃
➄ ➅ ➆ ➇
➈ ➉
➅ Test data count function (0: Disable/1: Enable)
11
Allows counting of test points for equipment with many test points.
➆ Buzzer setting
(Pressing SHIFT + STOP keys allows the user to make settings for each
function.)
➀ Output current frequency switching (0: 50 Hz / 1: 60 Hz)
1
2
3
ON
OFF
OFF
ON
Error
ON
OFF
ON
OFF
➇ Enable current control in test condition (0: No/1: Yes)
Allows changing of the output current value while a test is in
progress.
➁ PASS/FAIL hold function setting
Determines whether the condition is
held after detecting PASS or FAIL.
0
Evaluation
0
1
2
3
PASS
NO YES NO YES
FAIL
YES YES NO
➈ Momentary out
Enabling this function allows the current to be output only when
the START key is pressed.
0: Disabled (trigger operation)
1: Enabled (momentary out operation)
NO
➂ Hold function setting (0: Hold disabled / 1: Hold enabled)
Holds the condition of the unit after the preset test time has elapsed
or after the STOP key is pressed.
➉ Test mode
0: Soft start mode
1: Normal mode
2: Continuous test mode
➃ Use test lower limit setting (0: No / 1: Yes)
Disabling the setting allows only the upper limit to be set.
Enabling the setting allows also the lower limit to be set.
➄ Timer override (0: No / 1: Yes)
Determines whether a test time can be set. If test time is not set,
the test ends only when the STOP key is pressed or the result is
FAIL.
11
Print function
0: Not used
1: Automatically print PASS/FAIL result
2: Optionally print in PASS/FAIL hold condition
■ External I/O
The unit comes with I/O connectors as standard equipment.
The connectors allow external START/STOP control,
READY/TEST status checking, and PASS/FAIL result
reading. Photocouplers are used to isolate the I/O signals from
the internal circuitry.
9442 printer
■ External interface (option)
The 9518-02 GP-IB interface or 9593-02 RS-232C interface
can be installed in the unit. This allows remote control from a
computer as well as export of measurement data. The 9593-02
RS-232C interface also allows connection of the 9442 printer
for producing a hard copy of measurement data.
Printing method
Paper width
Printing speed
Power source
: Thermal serial dot printer
: 112 mm
: 52.5 cps
: 9443 AC adapter, or supplied nickel-hydride battery
(Charged through 9443; printing capability approx. 3000 lines with
full charge)
* To use the 9442 printer, an optional 9593-02 RS-232C interface, 9446 connection
cable, and AC adapter are required.
3
■ 3157-01 Specifications
● Basic specifications
: AC 4-terminal method resistance measurement
Basic functions
[ Generator section ]
Current generator
principle
Current setting
range
Accuracy
Maximum output
power
Open-terminal
voltage
Generator
frequency
SOFT START
function
: PWM constant current control
: 3.0 A - 31.0 A AC (0.1 A resolution), into 0.1W load
[ Timer section ]
: ± (1% of setting + 0.2 A) within maximum output power range
: 130 VA (at output terminals) *
Setting ON
Setting OFF
Setting range
* Subject to derating according to ambient temperature [80% at 40˚C (104˚F)]
: Max. 6 V AC
: 50 Hz or 60 Hz sine wave (selectable)
:
:
:
Setting resolution :
:
Accuracy
Counts down time after start until preset time
Shows elapsed time after start
0.5 - 999 s
0.1 s (0.5 - 99.9 s)/ 1 s (100 - 999 s)
±50 ms (0.5 - 99.9 s)/±0.5 s (100 - 999 s)
[ Other functions ]
: Apply current only after checking load connection
[ Monitor section ]
Resistance
measurement
range
Accuracy
Current monitoring
range
Accuracy
Voltage monitor : 0 - 6.00 V AC (single range 0.01 V resolution)
range
Accuracy
: ± (1% rdg.+5 dgt.)
Monitoring cycle : 0.5 s
: 0 - 1.800W (0.001W resolution)
Comparator
function
Comparator result
output
Zero-adjust
function
Zero-adjust range
Memory function
: PASS/FAIL evaluation using preset upper/lower limit
Interfaces
: 1. External I/O *
Output signals: PASS /UP FAIL /LOW FAIL /TEST
/READY .... open collector
Input signals: START /STOP /External I/O
ENABLE .... 5 - 24 V DC
2. Front EXT connector *
External START/STOP input .... contact signal
: ± (2% rdg. +4 dgt.) after zero-adjust
: 0 - 35.0 A AC (0.1 A resolution)
: Internal buzzer (PASS/FAIL, ON/OFF switchable) and
I/O output
: For measurement probe impedance cancellation
: 0 - 0.100W
: Max. 20 settings (with save/load)
: ± (1% rdg. +5 dgt.) (at 3 A or more)
● General Specifications
Display
Ambient conditions
for use
Ambient conditions
for storage
Ambient conditions
for assured
accuracy
Suitable
environments
Power supply
voltage range
: Fluorescent tube (digital display)
: 0 to +40˚C (32 to 104˚F), 30 to 90% RH (no condensation)
: -10 to +50˚C (14 to 122˚F), 10 to 95% RH (no condensation)
: 23˚C ± 5˚C (71.3˚F ± 9˚F)
30 to 90% RH (no condensation)
After 30-minute warmup period
: Indoors, altitude up to 2000 m
* When external start/stop connector is used, START key is inactive
3. RS-232C or GP-IB (option; one only)
Remote control, measurement data output
: 100 - 120 V/200 - 240 V AC (auto-switching)
(When RMT indicator is on, operation keys are locked; only LOCAL,
STOP, and external keys work)
Dimensions
Power line frequency : 50 - 60 Hz
Withstand voltage
Maximum rated
power
Fuse
Compatible
standards
: 1.35 kV AC, 20 mA, 1 min., between power supply and chassis
: 350 VA (with optional equipment)
: 250VT3. 15AL
: 1. EMC : EN55011:1991+A1:1997+A2:1996
Group1 CLASS A
EN50082-1:1992
2. Safety : EN61010-1:1993+A1:1995
Contamination 2 Overvoltage category II
(expected overvoltage category 2500 V)
Mass
Standard
accessories
: Approx. 320 (W) × 90 (H) × 263 (D)mm
Approx. 12.6" (W) × 3.56" (H) × 10.40" (D)
(Without protruding parts)
: Approx. 7 kg/247.2 oz (without options)
: Power cord, spare fuse (integrated in inlet),
shorting bar × 2 (current output - voltage sensing terminal)
3157-01 AC GROUNDING HiTESTER
*For measurement, two 9296 or one each of 9296 and 9297 are required.
■ Options
9296 CURRENT PROBE
9297 CURRENT APPLY PROBE
9296
Length: approx. 1.5 m (59.3")
9297
9518-02 GP-IB INTERFACE
9593-02 RS-232C INTERFACE
9442 PRINTER
1196 RECORDING PAPER(25m, 10 rolls)
9443-02 AC ADAPTER(for printer, EU)
9443-03 AC ADAPTER(for printer, America)
9446 CONNECTION CABLE(for printer)
Length: approx. 1.5 m (59.3")
* 9296/9297 use dual wiring up to the probe tip.