MEANWELL RPS-75-15

RPS-75
75W Single Output Medical Type
series
Features :
Universal AC input/Full range
Low leakage current<200uA
Protections: Short circuit / Overload / Over voltage
Free air convection for rated power and 23.5CFM forced air
convection for peak load
UL60601-1/IEC60601-1/EN60601-1 medical safety approved
No load power consumption<0.75W
Fixed switching frequency at 65KHz
3 years warranty
EN60601-1
IEC60601-1
SPECIFICATION
MODEL
OUTPUT
RPS-75-3.3
RPS-75-5
RPS-75-12
RPS-75-15
RPS-75-24
RPS-75-36
DC VOLTAGE
3.3V
5V
12V
15V
24V
36V
48V
RATED CURRENT
15A
14A
6.3A
5A
3.2A
2.1A
1.6A
CURRENT RANGE
0 ~ 20A
0 ~ 18.7A
0 ~ 8.3A
0 ~ 6.7A
0 ~ 4.2A
0 ~ 2.8A
0 ~ 2.1A
RATED POWER
49.5W
70W
75.6W
75W
76.8W
75.6W
76.8W
PEAK LOAD (23.5CFM)
66W
94W
99.6W
100.5W
100.8W
100.8W
100.8W
80mVp-p
120mVp-p
150mVp-p
240mVp-p
300mVp-p
300mVp-p
4.75 ~ 5.5V
11.4 ~ 13.2V
13.5 ~ 16.5V
22.8 ~ 27.6V
34.2 ~ 39.6V
45.6 ~ 52.8V
RIPPLE & NOISE (max.) Note.2 80mVp-p
VOLTAGE ADJ. RANGE
VOLTAGE TOLERANCE Note.3
2.0%
2.0%
1.0%
1.0%
1.0%
1.0%
1.0%
LINE REGULATION
0.5%
0.5%
0.5%
0.5%
0.5%
0.5%
0.5%
1.5%
1.5%
1.0%
1.0%
1.0%
1.0%
1.0%
86%
86%
41.4 ~ 48.6V
55.2 ~ 64.8V
LOAD REGULATION
INPUT
SETUP, RISE TIME
500ms, 30ms/230VAC
HOLD UP TIME (Typ.)
80ms/230VAC
VOLTAGE RANGE
90 ~ 264VAC
FREQUENCY RANGE
47 ~ 63Hz
73%
78%
1.5A/115VAC
1A/230VAC
LEAKAGE CURRENT
Earth leakage current <200uA / 264VAC, Patient leakage current <100uA/264VAC
OVERLOAD
140 ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed.
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
(Note 4)
OTHERS
SAFETY STANDARDS
WITHSTAND VOLTAGE
COLD START 25A/115VAC
3.8 ~ 4.46V
5.75 ~ 6.75V
83%
82%
85%
50A/230VAC
13.8 ~ 16.2V
17.25 ~ 20.25V
27.6 ~ 32.4V
Protection type : Shut down o/p voltage, re-power to recover
-20 ~ +70 (Refer to output load derating curve)
20 ~ 90% RH non-condensing
-40 ~ +85
, 10 ~ 95% RH
0.03%/ (0 ~ 45
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
UL60601-1, TUV EN60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH
ISOLATION RESISTANCE
EMI CONDUCTION & RADIATION Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B
HARMONIC CURRENT
Compliance to EN61000-3-2,-3
EMS IMMUNITY
MTBF
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61000-6-2, EN61204-3, heavy industry level, EN61204-3 medical level, criteria A
DIMENSION
PACKING
NOTE
20ms/115VAC at full load
127 ~370VDC
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
OVER VOLTAGE
EMC
500ms, 30ms/115VAC at full load
EFFICIENCY(Typ.)
PROTECTION
SAFETY &
2.9 ~ 3.6V
RPS-75-48
446.8K hrs min. MIL-HDBK-217F (25
127*76.2*31mm (L*W*H)
0.26Kg; 63pcs/16.3Kg/1.35CUFT
)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
6. Heat Sink HS1,HS2,HS3 can not be shorted.
File Name:RPS-75-SPEC 2011-03-31
RPS-75
75W Single Output Medical Type
Mechanical Specification
series
Unit:mm
FAN
5.7
15cm
23.5CFM min.
HS3
CN2
3
2
1
64.75
76.2
1
2
3
4
5
6
7
8
1 CN3
2
HS1
CN1
HS2
AC FUSE
T2.5/250V
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No.
1
2
3
Assignment
AC/N
No Pin
AC/L
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
DC Output Connector (CN2) : JST B8P-VH or equivalent
FS2 FS1
4115.8
5.6
Pin No.
1,2,3,4
5,6,7,8
3.8
127
Assignment
+V
-V
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Remote Sense(CN3) : JST B2B-XH or equivalent
31
Pin No.
1
2
Assignment
RS+
RS-
Mating Housing
Terminal
JST XHP
or equivalent
JST SXH-001T-P0.6
or equivalent
3max.
HS1,HS2,HS3 can not be shorted
Block Diagram
fosc : 65KHz
+S
RECTIFIERS
&
FILTER
EMI
FILTER
I/P
RECTIFIERS
&
FILTER
POWER
SWITCHING
+V
-V
-S
DETECTION
CIRCUIT
FG
O.L.P.
CONTROL
O.V.P.
Derating Curve
Output Derating VS Input Voltage
Ta=25
100
133
23.5CFM Forced air Cooling
90
Others
100
80
Convection Cooling
70
3.3V,5V
LOAD (%)
LOAD (%)
80
50
20
-20
0
10
20
30
40
50
AMBIENT TEMPERATURE ( )
60
70
(HORIZONTAL)
60
50
40
90
95
100
115
120
140
160
180
200
220
240
264
INPUT VOLTAGE (VAC) 60Hz
File Name:RPS-75-SPEC 2011-03-31