TRUMPower 130 WATT MEDICAL POWER ADAPTERS TMP130 SERIES D E SCR IPTION (MDD+EMC) The TMP130 series of AC/DC switching power adapters are for 100 to 130 watts of continuous output power. They are designed for medical applications that require external power. The units are enclosed in a 94V-1 rated polyphenylene-oxide case with an IEC 320 inlet to mate with interchangeable cord for world-wide use. An on/off switch can be added at request. All models meet CISPR 11 and FCC class B emission limits, and comply with UL, CSA, IEC and CE requirements. F EATU R E S ‧ In compliance with EN61000-3-2, EN61000-3-3 and EN61000-1-1-2 ‧ Optional output connectors ‧ Optional on/off sw itch ‧ Optional Power Fail Detect (PFD) signal ‧ 100% burn-in at full load ‧ Auto-ranging input ‧ Input surge current protection ‧ O v e r v o l ta g e p r o t e c t i o n ‧ Thermal protection S af e t y Sta nd ar d A pp ro va l s: UL 2601-1 CSA C22.2 No. 601.1 INPUT SPECIFICATIONS I n p u t v o l ta g e : 9 0 t o 1 3 2 VA C o r 1 8 0 t o 2 6 4 VA C (auto-ranging) Input frequency: 47 to 63 Hz Input current: 3 . 0 A ( r m s ) f o r 11 5 VA C 1 . 7 A ( r m s ) f o r 2 3 0 VA C L e a k a g e c u r r e n t : 9 0 µ A m a x . @ 11 5 VA C , 6 0 H z 1 5 0 µ A m a x . @ 2 3 0 VA C , 5 0 H z GENERAL SPECIFICATIONS OUTPUT SPECIFICATIONS Output voltage/current: Total output power: Ripple and noise: Overvoltage protection: See rating chart 130 watts maximum 1% peak to peak maximum Provided on output: set at 112 – 132% of its nominal output voltage Overcurrent protection: All outputs protected to short circuit conditions Temperature coefficient: All outputs ±0.04%/ºC maximum Transient response: Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500µs after a 25% step load change E N V IR O N M ENTA L SP E C IF ICATI O N S Operating temperature: Storage temperature: Relative humidity: Derating: TÜV EN60601-1 0ºC to +70ºC -40ºC to +85ºC 5% to 95% non-condensing Derate from 100% at +50ºC linearly to 50% at +70ºC Switching frequency: 68 Khz ±8Khz 70% minimum on Vo ≥12V Efficiency: 65% minimum on others 10 msec minimum at 115VAC Hold-up time: ±0.5% maximum at full load Line regulation: 15amps @115VAC, or 30amps Inrush current: @230VAC, at 25ºC and cold start 4000VAC from input to output Withstand voltage: 1500VAC from input to ground 5000VAC from output to ground 400,000 hours minimum at full load MTBF: at 25ºC ambient, calculated per MIL-HDBK-217F, excluding DC fan. EMC Performance (EN60601-1-2: 2001) EN55011: Class B conducted, class B radiated FCC: Class B conducted, class B radiated VCCI: Class B conducted, class B radiated EN61000-3-2: Harmonic distortion, class A EN61000-3-3: Line flicker EN61000-4-2: ESD, ±8KV air and ±6KV contact EN61000-4-3: Radiated immunity, 3V/m for 80~2500MHz EN61000-4-4: Fast transient/burst, ±2KV EN61000-4-5: Surge, ±1KV diff, ±2KV com. EN61000-4-6: Conducted immunity, 3Vrms EN61000-4-8: Magnetic field immunity, 3A/m EN61000-4-11: Voltage dips, 30% reduction for 500ms, 60% reduction for 100ms and >95% reduction for 10ms Tumbler Technologies + TRUMPower 3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com MEDICAL GRADE TMP130 SERIES OU TPUT VOLTAGE/CURR ENT RATING CHAR T Imin. Imax. Tol. Maximum Output Power Output Product No. Vnom. TMP130-05 5V 0A 20A 5% 100W TMP130-07 7V 0A 14A 5% 100W TMP130-10 10V 0A 10A 5% 100W TMP130-12 12V 0A 10.8A 5% 130W TMP130-15 15V 0A 8.6A 4% 130W TMP130-18 18V 0A 7.2A 4% 130W TMP130-24 24V 0A 5.4A 3% 130W TMP130-30 30V 0A 4.3A 3% 130W 130W 130W TMP130-36 36V 0A 3.6A 2% TMP130-48 48V 0A 2.7A 2% Notes: (1) All models are with built-in fan. (2) Models except TMP130-05, TMP130-07 and TMP130-10 are without remote sense. Consult factory for applications requiring remote sense. M E C H A N ICA L S P E C I F IC AT IO N S NOTES: 1. Dimensions shown in inch [mm] 2. Tolerance 0.02 [0.5] maximum 3. Weight: 0.80Kgs (1.76Lbs.) approx. 4. Standard output connector is a male 8-pin DIN plug. Please contact TRUMPower for other options. 5. To order a model with on/off switch, add suffix “S” to the model number, e.g. TMP130-24-S. PIN CHART PIN 1 2 3 4 5 6 7 8 TMP130-05 TMP130-07 TMP130-10 RETURN -SENSE OUTPUT +V RETURN OUTPUT +V RETURN OUTPUT +V +SENSE TMP130-12 TMP130-15 TMP130-18 TMP130-24 TMP130-30 TMP130-36 TMP130-48 RETURN RETURN OUTPUT +V RETURN OUTPUT +V RETURN OUTPUT +V OUTPUT +V Product No. Tumbler Technologies + TRUMPower 3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com