TRUMPOWER TMP90-12

TRUMPower
D E SCR IPTION
The TMP90 is a series of AC/DC desktop power
adapters with 60 to 90 watts of continuous output
power. They are enclosed in a 94V-1 rated case with
an IEC 320/C14 or C18 inlet for medical and ITE
applications. All models meet EN55011, EN55022 and
FCC class B emission limits, and are designed for
medical applications.
90 WATTS MEDICAL POWER ADAPTERS
TMP90 SERIES
( LV D )
RoHS
F EATU R E S
‧ Low safety ground leakage current
‧
‧
‧
‧
‧
‧
‧
IEC320/C14 (Class I) or C18 (Class II) Inlet
9 0 t o 2 6 4 VA C w i d e i n p u t r a n g e
Optional output connectors
O v e r v o l ta g e & o v e r c u r r e n t p r o t e c t i o n s
0.98 PFC
100% burn-in
C o m p l i a n t w i t h C E C a n d E n e r g y S ta r e f f i c i e n c y
l e v e l I V r e q u i r e m e n ts ( e x c e p t ≤ 1 2 V o u t p u t m o d e l s )
‧ No load power consumption less than 0.5W
‧ Av e r a g e a c t i v e e f f i c i e n c y ≥ 8 5 %
‧ C o m p l i a n t w i t h R o H S r e q u i r e m e n ts
Saf et y Sta n da rd A ppr o va ls:
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
IN PUT S PE C IF ICAT ION S
I n p u t v o l ta g e :
Input frequency :
Input current :
Earth leakage current :
90 to 264VAC
47 to 63Hz
1.20A (rms) for 115VAC
0.60A (rms) for 230VAC
90µA max. @ 115VAC, 60Hz
150µA max. @ 230VAC, 50Hz
OUTPUT SPECIFICATIONS
Output voltage/current:
See rating chart
Total output power:
See rating chart
Ripple and noise:
2% peak to peak maximum
Over voltage protection:
Set at 112-140% of its nominal
output voltage
Protected to short circuit
conditions
Over current protection:
Temperature coefficient:
±0.04%/ºC maximum
Transient response:
Maximum excursion of 4% or
better on all models, recovering
to 1% of final value within 500µs
after a 25% step load change
E N V IR O N M ENTA L SP E C IF ICATI O N S
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
0 ºC t o + 6 0 ºC
- 4 0 ºC t o + 8 5 ºC
5% to 95% non-condensing
Derate from 100% at +40ºC
linearly to 50% at +60ºC
GENERAL SPECIFICATIONS
Switching frequency: 50 KHz - 110 KHz
Power factor:
0.98 Typical
Efficiency:
85% min. (78% min. for TMP90-05,
TMP90-06, TMP90-09 and TMP90-12)
Hold-up time:
15 msec minimum at 110VAC
Line regulation:
±0.5% maximum at full load
Inrush current:
40 amps @ 115VAC or 60 amps @
230VAC, @ 25ºC cold start
Withstand voltage: 4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
150,000 hours minimum at full load
at 25ºC ambient, calculated per
MIL-HDBK-217F
EMC Performance (IEC60601-1-2: 2001)
E N 5 5 0 11 :
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
E N 6 1 0 0 0 - 4 - 11 :
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Harmonic distortion, Class A and D
Line flicker
ESD, ±8KV air and ±6KV contact
Radiated immunity, 3V/m
Fast transient/burst, ±2KV
Surge, ±1KV diff, ±2KV com.
Conducted immunity, 3Vrms
Magnetic field immunity, 3A/m
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com
MEDICAL GRADE
TMP90 SERIES
OUTPUT VOLTAGE/CURRENT RATING CHART
Output
Product No.
TMP90-05
TMP90-06
TMP90-09
TMP90-12
TMP90-13.5
TMP90-15
TMP90-18
TMP90-19
TMP90-24
TMP90-30
TMP90-48
Vnom.
Imin.
Imax.
Tol.
Maximum
Output Power
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
12.0A
10.0A
7.78A
6.67A
6.3A
5.67A
5.0A
4.74A
3.75A
3.0A
1.87A
5%
5%
5%
5%
5%
5%
5%
5%
5%
5%
5%
60W
60W
70W
80W
85W
85W
90W
90W
90W
90W
90W
5V
6V
9V
12V
13.5V
15V
18V
19V
24V
30V
48V
NOTES:
1. Ripple and noise: Peak-to–peak with 20MHz bandwidth and 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor at rated line
voltage and load ranges.
2. To order a model with IEC320/C18 inlet with class II insulation, add suffix “C2” , e.g. TMP90-12-C2.
MECHANICAL SPECIFICATIONS
NOTES:
1. Dimensions show in inch [mm]
2. Tolerance 0.02 [0.5] maximum
3. Weight: 658 grams approx.
4. Standard output connector is a male 4-pin power DIN. Please contact TRUMPower for other connector options.
5. TMP90-05 and TMP90-06 output connectors are without lock.
PIN CHART
PIN
Product No.
TMP90-05
TMP90-09
TMP90-13.5
TMP90-18
TMP90-24
TMP90-48
TMP90-06
TMP90-12
TMP90-15
TMP90-19
TMP90-30
1
2
3
4
Shell
RETURN
OUTPUT +V
RETURN
OUTPUT +V
AC Ground
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com