1/3 VOFM-5 Series Switching Power Supply Rev. 04-2007 Features • • • • • Universal input 85~264 Vac Designed for medical application Industry standard foot-print Efficiency up to 77% Safety approved to UL/cUL 60601, TUV, CE • Conducted EMI meets EN55022 class B Model VOFM-5-3.3 VOFM-5-5 VOFM-5-9 VOFM-5-12 VOFM-5-15 VOFM-5-18 VOFM-5-24 Voltage (V) 3.3 5.0 9.0 12.0 15.0 18.0 24.0 Current (A) 1.25 1.0 0.55 0.42 0.33 0.28 0.23 Line Reg. 1 (%) 0.5 0.5 0.5 0.5 0.5 0.5 0.5 Load Reg. 2 (%) 1.0 1.0 1.0 1.0 1.0 1.0 1.0 Ripple & Noise 3 (mVp-p max) 50 50 90 120 150 180 240 Efficiency (%) 68 72 74 75 75 76 77 Notes: 1. Measured from High Line to Low Line at Full load. 2. Measured from Full load to 10% load at 110 VAC. 3. Use a 0.1uF ceramic cap. and a 10uF Electrolytic cap. on the output to measure Ripple & Noise at 20MHz BW. VOFM-5 series Derating Curve V-INFINITY LLC, 9615 SW Allen Blvd., Suite 103, Beaverton, OR 97005 1-866-372-1258 www.v-infinity.com 2/3 VOFM-5 Series Switching Power Supply Rev. 04-2007 Input Parameter Conditions/Description Input Frequency Input Voltage Output power derated from 85-90 VAC Input Current Inrush Current Measured at 85VAC, Full load Peak measured at 264 VAC at full load, cold start Min Nom 47 85 Max Units 63 264 Hz VAC 200 40 mA A Output Parameter Efficiency Hold up time Turn on Delay Conditions/Description Measured at 115VAC, 60Hz or 230VAC, 50Hz At 115 VAC, full load Full load Min Nom Max See Table above 8 1 Units mS S Protection Circuit Parameter Short Circuit Over load Output Over-voltage Conditions/Description Auto-recover once short is removed. less than 180% of fulll load 130% to 150% output rating. General and Safety Parameter Operating temp. Storage temp. Operating Humid. Storage Humid. Leakage current Safety EMC Standards Isolation Voltage (HI-POT) Conditions/Description non-condensing Min 0 -20 0 0 Nom Max 70 85 93 93 <100 Units ºC ºC % % µA Approved to UL/cUL 60601-1 (E302945), TUV EN60601-1, CE EN55011, EN55022, EN55024, EN61204-3, EN61000-3-2/3, EN61000-6-1/3, EN60601-1-2 Applied for 1 minute. Primary to secondary: 6565 VDC Primary to transformer core: 1.5K VAC Insulation Resistance Per EN60601-1 RoHS Yes Warranty Standard Warranty Length 2 years V-INFINITY LLC, 9615 SW Allen Blvd., Suite 103, Beaverton, OR 97005 1-866-372-1258 www.v-infinity.com 3/3 VOFM-5 Series Switching Power Supply Rev. 04-2007 Mechanical Parameter Dimensions Conditions/Description 2.17 x 1.38 x 0.87 Cooling method Free air convection Min Nom Max Units inches Outline drawings V-INFINITY LLC, 9615 SW Allen Blvd., Suite 103, Beaverton, OR 97005 1-866-372-1258 www.v-infinity.com