Model 1815 - Honeywell Test and Measurement Sensors

Model 1815
Enhanced Accuracy Rotary
Torque Sensor
DESCRIPTION
A line of enhanced accuracy rotary transformer torque sensors
Carrier frequency excitation provides increased signal/noise
often ideal for test installations running long-term durability test-
immunity. Extended speed range allows shaft rotation up to
ing. Non-contact nature of the signal transfer virtually eliminates
25000 RPM. Capacity range from 50 in-lb to 10000 in-lb for shaft
the need for maintenance. High overload protection with high
mounting configurations.
signal output helps prevent potential damage to the sensor.
FEATURES
•
50 in-lb to 10000 in-lb capacity
•
0.05 % non-linearity and hysteresis
•
Spline drive/flange mount
•
Extended speed range, up to 25000 RPM
•
Minimal maintenance due to “bearings only” contact
Safety considerations: It would be unsafe to operate Honeywell torque
sensors and load cells beyond static overload or ultimate extraneous
load limits as defined in the glossary of terms or, when applicable, higher
than maximum speed. When in doubt, consult factory. Honeywell is not
responsible for any property damage or personal injury which may result
because of the misapplication of the transducer.
Model 7927 shunt cal reference box included with each purchase of
1800 Series.
Flange housing mount with pads to match Army-Navy mountings standard. Spline drive.
Model 1815
PERFORMANCE SPECIFICATIONS
TYPICAL SYSTEM DIAGRAM
Characteristic
Measure
Mating Connectors & Cables
Torque range
50, 100, 200, 500, 1K, 2K, 5K, 10K lb-in
Non-linearity
±0.05 % of rated output
Hysteresis
±0.05 % of rated output
Repeatability
±0.03 % of rated output
Output @ rated capacity
2 mV/V (nominal)
064-LW13621 Mating connector
064-LW13656 Mating connector, speed pickup
064-LW25469 Mating connector, zero velocity speed pickup
7200-81-XX* Mating connector & 6 conductor cable
(unamplified unit with sense leads but not shunt cal)
7204-00-XX* Speed sensor cable, pigtail leads at instrument
7204-16-XX* Speed sensor cable to instrument 7541
7220-47-XX* Zero velocity speed sensor cable, pigtail leads at instrument
7220-119-XX* Zero velocity speed sensor cable to instrument 7541
* XX represents length in feet 100ft maximum
ENVIRONMENTAL SPECIFICATIONS
Characteristic
Measure
Temperature, operating
-29 °C to 93 °C [-20 °F to 200 °F]
Temperature, compensated
21 °C to 77 °C [70 °F to 170 °F]
Temperature effect, zero
±0.002 %/of rated output °F
Temperature effect, output
±0.002 %/of reading °F
Customer Supplied
Chart Recorder
Alarm Panel
Data Acquisition
Computer
PLC
Lebow Products
MENU
VIEW
TEST
Display Units
7541-111 (Single Channel Display)
7541-112 (Dual Channel Display)
7541-115 (For Torque and Optional
Speed Display)
MODEL 7541
TARE
HOLD
ESC
RESET
Model 7927-108 for
connection to 7541
ENTER
RUN
ELECTRICAL SPECIFICATIONS
Characteristic
Measure
Excitation
3.28 kHz optimum @ 10 Vac RMS
Insulation resistance
> 5000 mOhm @ 50 Vdc
Bridge resistance
350 ohm (nominal)
Number of bridges
1
Zero balance
±1.0 % of rated output
2 Honeywell • Sensing and Control
Optional
24368-2 Speed Sensor
22236 Zero Velocity Pickup
CAL
Model 1815
Enhanced Accuracy Rotary Torque Sensor
MECHANICAL SPECIFICATIONS
Model
Capacity lb-in
Capacity N-m
Max speed RPM
Protected for overloads to Nm [lb-in]
1815A-50
50
5
25000
15 [150]
1815A-100
100
10
25000
30 [300]
1815A-200
200
20
25000
60 [600]
1815A-500
500
55
25000
165 [1500]
1815A-1K
1000
115
25000
165 [1500]
1815K-50
50
5
15000
15 [150]
1815K-100
100
10
15000
30 [300]
1815K-200
200
20
15000
60 [600]
1815K-500
500
55
15000
165 [1500]
1815K-1K
1000
115
15000
330 [3000]
1815K-2K
2000
225
15000
675 [6000]
1815K-5K
5000
565
15000
1695 [15000]
1815K-10K
10000
1130
15000
1695 [15000]
Internal and external spline data
Model
Pressure angle
Pitch dia. cm [in]
Pitch
Number of teeth
1815A
30°
2,03 [0.80]
20/40
16
1815K
30°
3,05 [1.20]
20/40
24
MOUNTING DIMENSIONS
Model
A
B
C
cm [in] cm [in] cm [in]
D
cm [in]
E
F
G
H
cm [in] cm [in] cm [in] cm [in]
J
cm [in]
K
L
cm [in] cm [in]
N
P
cm [in] cm [in]
1815A
23,55
[9.27]
2,67
[1.05]
1,60
[0.63]
2,84
[1.12]
20,96
[8.25]
0,36
[0.14]
10,47
[4.12]
10,48
[4.13]
2,62
[1.03]
1,03
[0.41]
12,70
[5.00]
15,24
[6.00]
3/8-24
1815K
25,25
[9.94]
4,29
[1.69]
1,60
[0.63]
4,01
[1.58]
20,96
[8.25]
0,36
[0.14]
10,47
[4.12]
10,48
[4.13]
2,62
[1.03]
1,03
[0.41]
12,70
[5.00]
15,24
[6.00]
3/8-24
MS-3102E-L4S-5P
OPTIONAL
SPEED SENSOR
C
H
G
N
D
P
J
A
E
F
B
FLANGE & SHAFT PER AND 10262
K
6 HOLES
FLANGE & SHAFT PER AND 20002
For reference only
Honeywell • Sensing and Control 3
Model 1815 Enhanced Accuracy Rotary Torque Sensor
Warranty. Honeywell warrants goods of its manufacture as
being free of defective materials and faulty workmanship.
Honeywell’s standard product warranty applies unless agreed
to otherwise by Honeywell in writing; please refer to your
order acknowledgement or consult your local sales office for
specific warranty details. If warranted goods are returned to
Honeywell during the period of coverage, Honeywell will repair
or replace, at its option, without charge those items it finds
defective. The foregoing is buyer’s sole remedy and is in lieu
of all warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
While we provide application assistance personally, through our
literature and the Honeywell web site, it is up to the customer to
determine the suitability of the product in the application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this printing.
However, we assume no responsibility for its use.
For more information about Sensing and Control products, visit
www.honeywell.com/sensing or call +1-815-235-6847
Email inquiries to [email protected]
WARNING
PERSONAL INJURY
•DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARNING
MISUSE OF DOCUMENTATION
•The information presented in this catalogue is for
reference only. DO NOT USE this document as product
installation information.
•Complete installation, operation and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
Sensing and Control
Automation and Control Solutions
Honeywell
1985 Douglas Drive North
Golden Valley, MN 55422 USA
+1-815-235-6847
www.honeywell.com/sensing
008781-1-EN IL50 GLO June 2008