Model 3632 - Honeywell Test and Measurement Sensors

Model 3632
Compression Donut Load Cell
DESCRIPTION
Model 3632 Compression Donut Load Cell features a smooth
available for compression only loads in capacities of 25K, 50K
thru-hole design for use in applications which require the load
and 100K lbs. For optimum performance, this load cell must be
structure to pass through the cell. The hollow load cells are
mounted between load surfaces which are flat and parallel.
FEATURES
•
25000 lb to 100000 lb range
•
Thru-hole design
•
Compact column construction
•
mV/V output
•
0.15 % linearity
Model 3632
PERFORMANCE SPECIFICATIONS
Characteristic
Measure
Load range
25000, 50000, 100000 lb
Non-linearity
±0.15 % of rated output
Hysteresis
±0.15 % of rated output
Repeatability
±0.15 % of rated output
Output @ rated capacity
2 mV/V (nominal)
Operation
Compression
Resolution
Infinite
Standard calibration
Tension (+) and compression (-)
1
TYPICAL SYSTEM DIAGRAM
Customer Supplied
Chart Recorder
Alarm Panel
Data Acquisition
Computer
PLC
In-Line Amplifiers
(Used with unamplified units only)
Output
Amplifier
Universal In-Line Amplifiers
+/-5 VDC
UV
+/-10 VDC
UV-10
0 +/-5 VDC
UBP
4-20mA (3 wire)
U3W
4-20mA (2 wire)
U2W
DIN rail Mount In-Line Amplifiers
0 +/- 5 VDC (3 wire)
DV-05
4-20mA (3 wire)
DA-05
0 +/- 10 VDC (3 wire)
DV10
Power Supply
Honeywell
ENVIRONMENTAL SPECIFICATIONS
In-Line Amplifier
Output
Input
Serial Number
Characteristic
Measure
Temperature, operating
-54 °C to 93 °C [-65 °F to 200 °F]
Temperature, compensated
21 °C to 77 °C [70 °F to 170 °F]
Temperature effect, zero
±0.002 % of rated output/°F
Temperature effect, output
±0.002 % of reading/°F
Display Units
SC500
SC2001
GM
7541
SC1000
SC3004
NK
7558
SC2000
HH
7561-PSD
MODEL 3632
ELECTRICAL SPECIFICATIONS
Characteristic
Measure
Strain gage type
Foil
Excitation (maximum)
20 Vdc or Vac RMS
Insulation resistance
> 5000 mOhm @ 50 Vdc
Bridge resistance, 25K lb,
50K lb
480 ohm
Bridge resistance, 100K lb
240 ohm
Number of bridges
1
Zero balance
±1.0 %
Electrical termination
PT02E-10-6P
MECHANICAL SPECIFICATIONS
Characteristic
Measure
Static overload capacity
150 % of nominal capacity
Material
Carbon steel
Natural frequency
See table
WIRING CODES
Cable
Red
(+) excitation
Black
(-) excitation
Green
(+) output
White
(-) output
2 Honeywell • Sensing and Control
2003.1 lb.
L1 L2 L3 L4
OR
VALUE
CLEAR
UP
DOWN
SETUP
CHAN
TARE
ENTER
EXIT
Mating Connectors & Cables
023-LW181-023-LF Mating Connector
7205-76-XX* Mating connector & 6 conductor cable
(unamplified unit with sense leads but not shunt cal)
7205-75-XX** Mating connector & 4 conductor cable
(unamplified unit without sense leads but not shunt cal)
7205-83-XX* Mating connector & 6 conductor cable
(for connection to instrument 7541)
7205-82-XX** Mating connector & 4 conductor cable
(for connection to instrument 7541)
* XX represents length in feet
** XX represents length in feet
100ft maximum
20ft maximum
Compression Donut Load Cell
NATURAL FREQUENCY
Nominal load limit
capacity Fz
Static extraneous load limits
lb
N
Static overShear FX or
load capacity FY (lb)
(% of nominal
capacity)
Bending MX
or MY (lb-in)
Torque MZ
(lb-in)
Deflection at
nominal load
limit (in)
Ringing frequency (Hz)
25K
125K
150
2750
18K
3K
0.005
5000
50K
200K
150
5500
36K
13K
0.005
7000
100K
500K
150
13K
91K
27K
0.005
10000
MOUNTING DIMENSIONS
Model
A cm [in]
B cm [in]
C cm [in]
D cm [in]
E cm [in]
F cm [in]
G cm [in]
H cm [in]
3632
10,16 [4]
5,08 [2]
7,62 [3]
5,08 [2]
2,54 [1]
8,26 [3.25]
0,15 [0.06]
4,78 [1.88]
A DIA.
B DIA.
C
RECEP.
PT02E-10-6P
E
D
F
G
H
Honeywell • Sensing and Control 3
Model 3632
Compression Donut Load Cell
NOTES
1.
This unit calibrated to Imperial (non-Metric) units.
Warranty. Honeywell warrants goods of its manufacture as
being free of defective materials and faulty workmanship.
Honeywell’s standard product warranty applies unless agreed
to otherwise by Honeywell in writing; please refer to your
order acknowledgement or consult your local sales office for
specific warranty details. If warranted goods are returned to
Honeywell during the period of coverage, Honeywell will repair
or replace, at its option, without charge those items it finds
defective. The foregoing is buyer’s sole remedy and is in lieu
of all warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
While we provide application assistance personally, through our
literature and the Honeywell web site, it is up to the customer to
determine the suitability of the product in the application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this printing.
However, we assume no responsibility for its use.
For more information about Sensing and Control products, visit
www.honeywell.com/sensing or call +1-815-235-6847
Email inquiries to [email protected]
WARNING
PERSONAL INJURY
•DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARNING
MISUSE OF DOCUMENTATION
•The information presented in this catalogue is for
reference only. DO NOT USE this document as product
installation information.
•Complete installation, operation and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
Sensing and Control
Automation and Control Solutions
Honeywell
1985 Douglas Drive North
Golden Valley, MN 55422 USA
+1-815-235-6847
www.honeywell.com/sensing
008651-1-EN IL50 GLO May 2008