Model 3632 Compression Donut Load Cell DESCRIPTION Model 3632 Compression Donut Load Cell features a smooth available for compression only loads in capacities of 25K, 50K thru-hole design for use in applications which require the load and 100K lbs. For optimum performance, this load cell must be structure to pass through the cell. The hollow load cells are mounted between load surfaces which are flat and parallel. FEATURES • 25000 lb to 100000 lb range • Thru-hole design • Compact column construction • mV/V output • 0.15 % linearity Model 3632 PERFORMANCE SPECIFICATIONS Characteristic Measure Load range 25000, 50000, 100000 lb Non-linearity ±0.15 % of rated output Hysteresis ±0.15 % of rated output Repeatability ±0.15 % of rated output Output @ rated capacity 2 mV/V (nominal) Operation Compression Resolution Infinite Standard calibration Tension (+) and compression (-) 1 TYPICAL SYSTEM DIAGRAM Customer Supplied Chart Recorder Alarm Panel Data Acquisition Computer PLC In-Line Amplifiers (Used with unamplified units only) Output Amplifier Universal In-Line Amplifiers +/-5 VDC UV +/-10 VDC UV-10 0 +/-5 VDC UBP 4-20mA (3 wire) U3W 4-20mA (2 wire) U2W DIN rail Mount In-Line Amplifiers 0 +/- 5 VDC (3 wire) DV-05 4-20mA (3 wire) DA-05 0 +/- 10 VDC (3 wire) DV10 Power Supply Honeywell ENVIRONMENTAL SPECIFICATIONS In-Line Amplifier Output Input Serial Number Characteristic Measure Temperature, operating -54 °C to 93 °C [-65 °F to 200 °F] Temperature, compensated 21 °C to 77 °C [70 °F to 170 °F] Temperature effect, zero ±0.002 % of rated output/°F Temperature effect, output ±0.002 % of reading/°F Display Units SC500 SC2001 GM 7541 SC1000 SC3004 NK 7558 SC2000 HH 7561-PSD MODEL 3632 ELECTRICAL SPECIFICATIONS Characteristic Measure Strain gage type Foil Excitation (maximum) 20 Vdc or Vac RMS Insulation resistance > 5000 mOhm @ 50 Vdc Bridge resistance, 25K lb, 50K lb 480 ohm Bridge resistance, 100K lb 240 ohm Number of bridges 1 Zero balance ±1.0 % Electrical termination PT02E-10-6P MECHANICAL SPECIFICATIONS Characteristic Measure Static overload capacity 150 % of nominal capacity Material Carbon steel Natural frequency See table WIRING CODES Cable Red (+) excitation Black (-) excitation Green (+) output White (-) output 2 Honeywell • Sensing and Control 2003.1 lb. L1 L2 L3 L4 OR VALUE CLEAR UP DOWN SETUP CHAN TARE ENTER EXIT Mating Connectors & Cables 023-LW181-023-LF Mating Connector 7205-76-XX* Mating connector & 6 conductor cable (unamplified unit with sense leads but not shunt cal) 7205-75-XX** Mating connector & 4 conductor cable (unamplified unit without sense leads but not shunt cal) 7205-83-XX* Mating connector & 6 conductor cable (for connection to instrument 7541) 7205-82-XX** Mating connector & 4 conductor cable (for connection to instrument 7541) * XX represents length in feet ** XX represents length in feet 100ft maximum 20ft maximum Compression Donut Load Cell NATURAL FREQUENCY Nominal load limit capacity Fz Static extraneous load limits lb N Static overShear FX or load capacity FY (lb) (% of nominal capacity) Bending MX or MY (lb-in) Torque MZ (lb-in) Deflection at nominal load limit (in) Ringing frequency (Hz) 25K 125K 150 2750 18K 3K 0.005 5000 50K 200K 150 5500 36K 13K 0.005 7000 100K 500K 150 13K 91K 27K 0.005 10000 MOUNTING DIMENSIONS Model A cm [in] B cm [in] C cm [in] D cm [in] E cm [in] F cm [in] G cm [in] H cm [in] 3632 10,16 [4] 5,08 [2] 7,62 [3] 5,08 [2] 2,54 [1] 8,26 [3.25] 0,15 [0.06] 4,78 [1.88] A DIA. B DIA. C RECEP. PT02E-10-6P E D F G H Honeywell • Sensing and Control 3 Model 3632 Compression Donut Load Cell NOTES 1. This unit calibrated to Imperial (non-Metric) units. Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages. While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application. Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use. For more information about Sensing and Control products, visit www.honeywell.com/sensing or call +1-815-235-6847 Email inquiries to [email protected] WARNING PERSONAL INJURY •DO NOT USE these products as safety or emergency stop devices or in any other application where failure of the product could result in personal injury. Failure to comply with these instructions could result in death or serious injury. WARNING MISUSE OF DOCUMENTATION •The information presented in this catalogue is for reference only. DO NOT USE this document as product installation information. •Complete installation, operation and maintenance information is provided in the instructions supplied with each product. Failure to comply with these instructions could result in death or serious injury. Sensing and Control Automation and Control Solutions Honeywell 1985 Douglas Drive North Golden Valley, MN 55422 USA +1-815-235-6847 www.honeywell.com/sensing 008651-1-EN IL50 GLO May 2008 Copyright © 2008 Honeywell International Inc. All rights reserved.