Model 1205B Rotary Torque/RPM/HP Measurement System DESCRIPTION Model 1205B is a wireless telemetry system capable of monitoring the parameters of torque, RPM, and horsepower utilizing a 1226A collar assembly, 2120B transmitter, stationary loop antenna, 1211D power seperation unit, and 1642A receiver unit. FEATURES • dc to 1000 Hz frequency response • Three analog outputs and optional isolated output channel • 10 kHz sampling rate • 175 microstrain (0.5 mV/V) transmitter sensitivity • Bench-mounted signal conditioner and readout Model 1205B PERFORMANCE SPECIFICATIONS ELECTRICAL SPECIFICATIONS Characteristic Measure Characteristic Measure Frequency response dc to 1000 Hz Induction power source 160 kHz ±6 % Sensitivity (transmitter) 175 microstrain (0.5 mV/V) 4,6 m [15 ft] standard 1 Number of channels 1 to 3 Cable length (IPS to receiver) Sampling rate 10 kHz 152,4 m [500 ft] Receiver readout display type 3.5 digit 7 segment LED Max. cable length (IPS to receiver) Power supply voltage (receiver) 115/230 Vac user-selectable Power consumption (receiver) 100 VA max. RF operating frequency 10.7 MHz2 RF modulation type FSK Analog output (receiver) ±10 Vdc @ 5 mA max. Output impedance (receiver) < 1 Ohm Electrical termination (receiver output) BNC female Mating connector (receiver output) BNC male Electrical termination (transmitter) 9 pins, 0,64 mm [0.025 in] diameter ENVIRONMENTAL SPECIFICATIONS Characteristic Measure Temperature, operating (transmitter) -55 °C to 125 °C [-67 °F to 257 °F] Temperature, operating (receiver/IPS/PSU) 10 °C to 50 °C [50 °F to 122 °F] Temperature, compensated (transmitter) 0 °C to 75 °C [32 °F to 167 °F] Temperature, compensated (receiver) 10 °C to 50 °C [50 °F to 122 °F] Static G force (rotating components) 15000 G RPM limits (rotating components) See graph - depends on shaft diameter Vibration (rotating components) 100 G to 1 kHz Collar weight Varies with shaft diameter Electronics weight (not including cables) Receiver and PSU 12,2 kg [27 lb] 2 Honeywell • Sensing and Control 1. 4,6 m [15 ft] is the standard cable length, other lengths are available. Consult factory. 2. Consult factory for other frequencies. Rotary Torque/RPM/HP Measurement System OPTION CODES RPM limits graph Range Code Many range/option combinations are available in our quick-ship and fast-track manufacture programs. Please see http://sensing.honeywell.com/ TMsensor-ship for updated listings. Compensated transmitter temperature 62s. 100 °C [212 °F] Display (receiver) 62c. Remote display Vibration and shock (PSU) 62j. Vibration/shock rated PSU RF operating frequency 10.7 MHz1 Analog output (receiver) ±10 Vdc @ 1 mA 62e. 1 channel 4 mA to 20 mA isolated output 1. 4,6 m [15 ft] is the standard cable length, other lengths are available. Consult factory. TYPICAL SYSTEM DIAGRAM MECHANICAL SPECIFICATIONS Characteristic Measure Dimensions (collar cross section) 31,8 mm x 31,8 mm [1.25 in x 1.25 in] typical (152,4 mm [6 in] OD min.) Dimensions (receiver) (L x W x H) 431,8 mm x 457,2 mm x 133,3 mm [17 in x 18 in x 5.25 in] Dimensions (IPS) (L x W x H) 152,4 mm x 101,6 mm x 50,8 mm [6 in x 4 in x 2 in] Shaft diameter limits (collar) 50,8 mm to 1066,8 mm [2 in to 42 in] 3 Collar material Type G-11 glass epoxy 3. 50,8 mm to 203,2 mm [2 in to 8 in] is standard shaft diameter, consult factory for other shaft diameters. Honeywell • Sensing and Control 3 Model 1205B Rotary Torque/RPM/HP Measurement System Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages. While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application. Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use. For more information about Sensing and Control products, visit www.honeywell.com/sensing or call +1-815-235-6847 Email inquiries to [email protected] WARNING PERSONAL INJURY •DO NOT USE these products as safety or emergency stop devices or in any other application where failure of the product could result in personal injury. Failure to comply with these instructions could result in death or serious injury. WARNING MISUSE OF DOCUMENTATION •The information presented in this catalogue is for reference only. DO NOT USE this document as product installation information. •Complete installation, operation and maintenance information is provided in the instructions supplied with each product. Failure to comply with these instructions could result in death or serious injury. Sensing and Control Automation and Control Solutions Honeywell 1985 Douglas Drive North Golden Valley, MN 55422 USA +1-815-235-6847 www.honeywell.com/sensing 008790-1-EN IL50 GLO June 2008 Copyright © 2008 Honeywell International Inc. All rights reserved.