eldrs report - M.S. Kennedy Corp.

ELDRS Radiation Test Report
MSK106RH
Radiation Hardened High Power Op Amp
August 25, 2008
Revised February 7, 2014
B. Erwin
F. Freytag
M.S. Kennedy Corporation
Liverpool, NY
1
I.
Introduction:
The ELDRS radiation test for the MSK106RH was developed to qualify the device to 50 Krad(Si).
MIL-STD-883 Method 1019 Condition C and ASTM F1892-06 were used as guidelines in the
development and implementation of the ELDRS test plan for the MSK106RH.
II.
Radiation Source:
ELDRS test was performed at the University of Massachusetts, Lowell, using a cobalt 60 radiation
source. Dosimetry was performed prior to device irradiation and the dose rate was determined to be
0.01 rads(Si)/sec. The ELDRS dose schedule can be found in Table I.
III. Test Setup:
All test samples were subjected to Group A Electrical Test in accordance with the device data sheet.
In addition, all devices received 320 hours of burn-in per MIL-STD-883 Method 1015 and were
electrically tested prior to irradiation. For test platform verification, one control device was tested at
25°C.
The devices were vertically aligned with the radiation source in the University of Massachusetts
ELDRS facility, chamber #2. Four devices were kept under bias during irradiation. Maximum
recommended operating voltages of ± 18 Volts were used for bias. Four devices had all leads
grounded during irradiation for the unbiased condition.
After each irradiation, the device leads were shorted together then the devices were packaged for
shipment, and shipped overnight to the MSK facility. Electrical testing to the MSK device data sheet
was performed within 72 hours from the removal of the radiation source. Testing was performed on
irradiated devices, as well as the control device, at each total dose level. Devices were then
repackaged for shipment and returned to the UMass facility overnight for subsequent dose level.
Devices were returned to the irradiation field within 120 hours of removal from the radiation source
per MIL-STD-883 Method 1019 Condition C.
IV. Data:
All performance curves are averaged from the test results of the biased and unbiased devices
respectively.
V. Summary:
The devices passed all post irradiation specifications up to 50 Krads(Si). The unbiased devices
exhibited a more significant shift than the biased devices throughout testing.
Devices exhibited a slight quiescent current decrease.
Input bias current increased significantly, but stayed within pre irradiation limits to 50 Krads(Si).
Positive and negative slew rate decreased as testing progressed to 50 Krads(Si). All devices passed
pre irradiation test limits to 50 Krads(Si) except serial number 0220, which had a final slew rate of
1.19V/uS. The device was still within the post irradiation limits.
Transition times showed a slight increase as well. However, all devices except 0220 stayed within
pre irradiation limits up to 50 Krads(Si). Serial number 0020 increased to 1.02 uS, which is still within
post irradiation limits.
2
MSK106RH
Biased/Unbiased Dose Rate Schedule
Dosimetry Equipment
Bruker Biospin # 0141
Irradiation Date
6/03/08 – 8/07/08
Exposure
Length
(hr:min:sec)
Incremental
Dose
rads(Si)
Cumulative
Dose
rads(Si)
332:04:00
264:58:00
335:38:00
475:34:00
1.24E4
9.84E3
1.24E4
1.75E4
12,380
22,220
34,620
52,090
Biased S/N – 0186, 0187, 0197, 0196
Unbiased S/N – 0199, 0203, 0219, 0220
Table 1
Dose Time, Incremental Dose and Total Cumulative Dose
3
Positive Quiescent Current vs. Total Dose
1.50
Positive Quiescent Current (mA)
1.45
1.40
1.35
Avg. Bias
Avg. Gnd
1.30
1.25
1.20
1.15
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Negative Quiescent Current vs. Total Dose
1.50
Negative Quiescent Current (mA)
1.45
1.40
1.35
Avg. Bias
Avg. Gnd
1.30
1.25
1.20
1.15
0
12.4
22.2
34.6
Total Dose in Krads(Si)
4
52
Input Offset Voltage vs. Total Dose
0.50
0.40
Input Offset Voltage (mV)
0.30
0.20
0.10
Avg. Bias
0.00
Avg. Gnd
-0.10
-0.20
-0.30
-0.40
-0.50
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Power Supply Rejection Ratio vs. Total Dose
Power Supply Rejection Ratio (dB)
140.00
130.00
120.00
Avg. Bias
110.00
Avg. Gnd
100.00
90.00
80.00
0
12.4
22.2
34.6
Total Dose in Krads(Si)
5
52
Positive Input Bias Current vs. Total Dose
300.00
Positive Input Bias Current (nA)
250.00
200.00
Avg. Bias
150.00
Avg. Gnd
100.00
50.00
0.00
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Negative Input Bias Current vs. Total Dose
Negative Input Bias Current (nA)
300.00
250.00
200.00
Avg. Bias
150.00
Avg. Gnd
100.00
50.00
0.00
0
12.4
22.2
34.6
Total Dose in Krads(Si)
6
52
Input Offset Current vs. Total Dose
50.00
45.00
Input Offset Current (nA)
40.00
35.00
30.00
Avg. Bias
25.00
Avg. Gnd
20.00
15.00
10.00
5.00
0.00
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Positive Output Swing (RL = 100 Ω) vs. Total Dose
15.00
Positive Output Swing (V)
14.50
14.00
Avg. Bias
13.50
Avg. Gnd
13.00
12.50
12.00
0
12.4
22.2
34.6
Total Dose in Krads(Si)
7
52
Negative Output Swing (RL = 100 Ω) vs. Total Dose
15.00
Negative Output Swing (V)
14.50
14.00
Avg. Bias
13.50
Avg. Gnd
13.00
12.50
12.00
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Positive Output Swing (RL =10 Ω) vs. Total Dose
15.00
Positive Output Swing (V)
14.50
14.00
Avg. Bias
13.50
Avg. Gnd
13.00
12.50
12.00
0
12.4
22.2
34.6
Total Dose in Krads(Si)
8
52
Negative Output Swing (RL = 10 Ω) vs. Total Dose
15.00
Negative Output Swing (V)
14.50
14.00
Avg. Bias
13.50
Avg. Gnd
13.00
12.50
12.00
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Positive Slew Rate vs. Total Dose
1.80
1.60
Positive Slew Rate (V/µS)
1.40
1.20
1.00
Avg. Bias
Avg. Gnd
0.80
0.60
0.40
0.20
0.00
0
12.4
22.2
34.6
Total Dose in Krads(Si)
9
52
Negative Slew Rate vs. Total Dose
1.80
1.60
Negative Slew Rate (V/µS)
1.40
1.20
1.00
Avg. Bias
Avg. Gnd
0.80
0.60
0.40
0.20
0.00
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Positive Short Circuit Current (Rsc = 0.5 Ω) vs. Total Dose
1.60
Positive Short Circuit Current (A)
1.50
1.40
1.30
Avg. Bias
1.20
Avg. Gnd
1.10
1.00
0.90
0.80
0
12.4
22.2
34.6
Total Dose in Krads(Si)
10
52
Negative Short Circuit Current (Rsc = 0.5 Ω) vs. Total Dose
1.60
Negative Short Circuit Current (A)
1.50
1.40
1.30
Avg. Bias
1.20
Avg. Gnd
1.10
1.00
0.90
0.80
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Positive Short Circuit Current (Rsc = 5 Ω) vs. Total Dose
200.00
Positive Short Circuit Current (mA)
190.00
180.00
170.00
160.00
Avg. Bias
150.00
Avg. Gnd
140.00
130.00
120.00
110.00
100.00
0
12.4
22.2
34.6
Total Dose in Krads(Si)
11
52
Negative Short Circuit Current (Rsc = 5 Ω) vs. Total Dose
200.00
Negative Short Circuit Current (mA)
190.00
180.00
170.00
160.00
Avg. Bias
150.00
Avg. Gnd
140.00
130.00
120.00
110.00
100.00
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Voltage Gain vs. Total Dose
120.00
Voltage Gain (dB)
115.00
110.00
Avg. Bias
105.00
Avg. Gnd
100.00
95.00
90.00
0
12.4
22.2
34.6
Total Dose in Krads(Si)
12
52
Common Mode Rejection Ratio vs. Total Dose
100.00
Common Mode Rejection Ratio (dB)
98.00
96.00
94.00
92.00
Avg. Bias
90.00
Avg. Gnd
88.00
86.00
84.00
82.00
80.00
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Positive Transition Time vs. Total Dose
Positive Transition Time (µS)
1.00
0.80
0.60
Avg. Bias
Avg. Gnd
0.40
0.20
0.00
0
12.4
22.2
34.6
Total Dose in Krads(Si)
13
52
Negative Transition Time vs. Total Dose
Negative Transition Time (µS)
1.00
0.80
0.60
Avg. Bias
Avg. Gnd
0.40
0.20
0.00
0
12.4
22.2
34.6
52
Total Dose in Krads(Si)
Rising Edge Overshoot vs. Total Dose
20.00
18.00
Rising Edge Overshoot (%)
16.00
14.00
12.00
Avg. Bias
10.00
Avg. Gnd
8.00
6.00
4.00
2.00
0.00
0
12.4
22.2
34.6
Total Dose in Krads(Si)
14
52
Falling Edge Overshoot vs. Total Dose
20.00
18.00
Falling Edge Overshoot (%)
16.00
14.00
12.00
Avg. Bias
10.00
Avg. Gnd
8.00
6.00
4.00
2.00
0.00
0
12.4
22.2
34.6
Total Dose in Krads(Si)
15
52