tid report - M.S. Kennedy Corp.

Total Dose Radiation Test Report
MSK 5978RH
RAD Hard Positive Adjustable Voltage Regulator
Sept 24, 2010 (TID, First Test)
May 11, 2012 (TID, Second Test, LOT H0923840.4 WF#3)
Jan 29, 2014 (TID, Third Test, LOT H0923840.4 WF#5)
B. Horton
P. Dinneen
M.S. Kennedy Corporation
Liverpool, NY
1
I.
Introduction:
The total dose radiation test plan for the MSK 5978RH was developed to qualify the devices as RAD
Hard to 300 KRADS(Si). The testing was performed beyond 300 KRADS(Si) to show trends in device
performance as a function of total dose. The test does not classify maximum radiation tolerance of the
device, but simply offers designers insight to the critical parameter-shifts up to the specified total dose
level.
MIL-STD-883 Method 1019.7 and ASTM F1892-06 were used as guidelines in the development and
implementation of the total dose test plan for the MSK 5978RH.
II.
Radiation Source:
Total dose was performed at the University of Massachusetts, Lowell, using a cobalt 60 radiation
source. The dose rate was determined to be 85 rads(Si)/sec. The total dose schedule can be found in
Table I.
III. Test Setup:
All test samples were subjected to Group A Electrical Test in accordance with the device data sheet. In
addition, all devices received 240 hours minimum of burn-in per MIL-STD-883 Method 1015 and were
fully screened IAW MIL-PRF-38535 Class V. For test platform verification, one control device was
tested at 25°C. Ten devices were then tested at 25°C, prior to irradiation, and were found to be within
acceptable test limits.
The devices were vertically aligned with the radiation source and enclosed in a lead/aluminum container
during irradiation. Five devices were kept under bias during irradiation. An operating voltage of +26
Volts was used for the bias condition. Five devices had all leads grounded during irradiation for the
unbiased condition.
After each irradiation, the device leads were shorted together and the devices were transported to the
MSK automatic electrical test platform. Testing was performed in accordance with the MSK device data
sheet. Testing was performed on irradiated devices, as well as the control device, at each total dose
level. Electrical tests were completed within one hour of irradiation. Devices were subjected to
subsequent radiation doses within two hours of removal from the radiation field.
IV. Data:
All performance curves are averaged from the test results of the biased and unbiased devices
respectively. If required, full test data can be obtained by contacting M.S. Kennedy Corporation.
V. Summary:
Based on the test data recorded during radiation testing and statistical analysis, the MSK5978RH
qualified as a 300 Krad(Si) radiation hardened device. Set Control Pin Current and Load Regulation
exhibited the most significant shift due to irradiation, however all performance curves stayed within
specification up to 450 Krad(Si) TID.
2
MSK 5978 RH Biased/Unbiased Dose Rate
Schedule
Dosimetry Equipment
Bruker Biospin # 0162
Irradiation Date
1/29/14
Exposure
Length
(min:sec)
Incremental
Dose
rads(Si)
10:06
10:06
10:06
51,510
51,510
51,510
30:18
30:18
154,530
154,530
Cumulative
Dose rads(Si)
51,510
103,020
154,530
309,060
463,590
Biased S/N – 1930, 1931, 1932, 1933, 1936
Unbiased S/N – 1937, 1938, 1939, 1943, 1944
Table 1
Dose Time, Incremental Dose and Total Cumulative Dose
3
4
5
6
7
8
Total Dose Radiation Test Report
MSK 5978RH
RAD Hard Positive Adjustable Voltage Regulator
Sept 24, 2010 (TID, First Test)
May 11, 2012 (TID, Second Test, H09238404.4 WF#3)
B. Horton
K. Conroy
R. Wakeman
M.S. Kennedy Corporation
Liverpool, NY
9
I.
Introduction:
The total dose radiation test plan for the MSK 5978RH was developed to qualify the devices as RAD
Hard to 300 KRADS(Si). The testing was performed beyond 300 KRADS(Si) to show trends in device
performance as a function of total dose. The test does not classify maximum radiation tolerance of the
device, but simply offers designers insight to the critical parameter-shifts up to the specified total dose
level.
MIL-STD-883 Method 1019.7 and ASTM F1892-06 were used as guidelines in the development and
implementation of the total dose test plan for the MSK 5978RH.
II.
Radiation Source:
Total dose was performed at the University of Massachusetts, Lowell, using a cobalt 60 radiation
source. The dose rate was determined to be 109 Rads(Si)/sec. The total dose schedule can be found
in Table I.
III. Test Setup:
All test samples were subjected to Group A Electrical Test in accordance with the device data sheet. In
addition, all devices received 240 hours minimum of burn-in per MIL-STD-883 Method 1015 and were
fully screened IAW MIL-PRF-38535 Class V. For test platform verification, one control device was
tested at 25°C. Ten devices were then tested at 25°C, prior to irradiation, and were found to be within
acceptable test limits.
The devices were vertically aligned with the radiation source and enclosed in a lead/aluminum container
during irradiation. Five devices were kept under bias during irradiation. An operating voltage of +26
Volts was used for the bias condition. Five devices had all leads grounded during irradiation for the
unbiased condition.
After each irradiation, the device leads were shorted together and the devices were transported to the
MSK automatic electrical test platform. Testing was performed in accordance with the MSK device data
sheet. Testing was performed on irradiated devices, as well as the control device, at each total dose
level. Electrical tests were completed within one hour of irradiation. Devices were subjected to
subsequent radiation doses within two hours of removal from the radiation field.
IV. Data:
All performance curves are averaged from the test results of the biased and unbiased devices,
respectively. If required, full test data can be obtained by contacting M.S. Kennedy Corporation.
V. Summary:
Based on the test data recorded during radiation testing and statistical analysis, the MSK5978RH
qualified as a 300 Krad(Si) radiation hardened device. Set Control Pin Current and Load Regulation
exhibited the most significant shift due to irradiation, however all performance curves stayed within
specification up to 450 Krad(Si) TID.
10
MSK 5978 RH Biased/Unbiased Dose Rate
Schedule
Dosimetry Equipment
Bruker Biospin # 0162
Irradiation Date
5/11/12
Exposure
Length
(min:sec)
Incremental
Dose
rads(Si)
Cumulative
Dose rads(Si)
08:00
08:00
08:00
23:50
23:50
51,840
51,840
51,840
154,440
154,440
51,840
51,480
103,680
155,520
309,960
464,400
51,480
08:00
Biased S/N – 0487, 0488, 0489, 0490, 0491
Unbiased S/N – 0492, 0493, 0494, 0495, 0496
Table 1
Dose Time, Incremental Dose and Total Cumulative Dose
11
12
13
14
15
16
Total Dose Radiation Test Report
MSK 5978RH
RAD Hard Positive Adjustable Voltage Regulator
Sept 24, 2010
C. Salce
F. Freytag
M.S. Kennedy Corporation
Liverpool, NY
17
I.
Introduction:
The total dose radiation test plan for the MSK 5978RH was developed to qualify the devices as RAD
Hard to 300 KRADS(Si). The testing was performed beyond 300 KRADS(Si) to show trends in device
performance as a function of total dose. The test does not classify maximum radiation tolerance of
the device, but simply offers designers insight to the critical parameter-shifts up to the specified total
dose level.
MIL-STD-883 Method 1019.7 and ASTM F1892-06 were used as guidelines in the development and
implementation of the total dose test plan for the MSK 5978RH.
II.
Radiation Source:
Total dose was performed at the University of Massachusetts, Lowell, using a cobalt 60 radiation
source. The dose rate was determined to be 132 Rads(Si)/sec. The total dose schedule can be
found in Table I.
III. Test Setup:
All test samples were subjected to Group A Electrical Test in accordance with the device data sheet.
In addition, all devices received 320 hours of burn-in per MIL-STD-883 Method 1015 and were fully
screened IAW MIL-PRF-38534 Class K. For test platform verification, one control device was tested
at 25°C. Ten devices were then tested at 25°C, prior to irradiation, and were found to be within
acceptable test limits.
The devices were vertically aligned with the radiation source and enclosed in a lead/aluminum
container during irradiation. Five devices were kept under bias during irradiation. An operating
voltage of +26 Volts was used for the bias condition. Five devices had all leads grounded during
irradiation for the unbiased condition.
After each irradiation, the device leads were shorted together and the devices were transported to the
MSK automatic electrical test platform. Testing was performed in accordance with the MSK device
data sheet. Testing was performed on irradiated devices, as well as the control device, at each total
dose level. Electrical tests were completed within one hour of irradiation. Devices were subjected to
subsequent radiation doses within two hours of removal from the radiation field.
IV. Data:
All performance curves are averaged from the test results of the biased and unbiased devices,
respectively. If required, full test data can be obtained by contacting M.S. Kennedy Corporation.
V. Summary:
Based on the test data recorded during radiation testing and statistical analysis, the MSK5978RH
qualified as a 300 Krad(Si) radiation hardened device. Set Pin Current and Line Regulation exhibited
the most significant shift due to irradiation, however all performance curves stayed within
specification up to 450 Krad(Si) TID.
18
MSK 5978 RH Biased/Unbiased Dose Rate
Schedule
Dosimetry Equipment
Bruker Biospin # 0162
Irradiation Date
9/24/10
Exposure
Length
(min:sec)
Incremental
Dose
rads(Si)
Cumulative
Dose rads(Si)
6:30
6:30
6:30
6:30
13:00
19:31
51,480
51,480
51,480
51,480
102,960
154,572
51,480
102,960
154,440
205,920
308,880
463,452
Biased S/N – 0037, 0038, 0039, 0042, 0046
Unbiased S/N – 0049, 0053, 0055, 0059, 0060
Table 1
Dose Time, Incremental Dose and Total Cumulative Dose
19
20
21
22
23
24