tid report - M.S. Kennedy Corp.

Total Dose Radiation Test Report
MSK0002RH
Radiation Tolerant High Speed, Buffer Amp
March 29, 2007 (First Test)
April 7, 2009 (Second Test)
February 25, 2011 (Third Test)
June 1, 2012 (Fourth Test)
NPN Wafer Lot: J1978W#PD-10
June 13, 2014 (5th Test,
PNP Wafer Lot: J1951W#SQ-7)
B. Horton
K. Conroy
M.S. Kennedy Corporation
Page 1 of 42
I.
Introduction:
The total dose radiation test plan for the MSK0002RH was developed to qualify the device as
radiation tolerant up to 100 Krad(Si). The testing was performed beyond 100 Krad(Si) to show trends in
device performance as a function of total dose. The test does not classify maximum radiation tolerance
of the hybrid, but simply offers designers insight to the critical parameter-shifts up
to the specified total dose level.
MIL-STD-883 Method 1019.7 and ASTM F1892-06 were used as guidelines in the development and
implementation of the total dose test plan for the MSK0002RH.
II.
Radiation Source:
Total dose was performed at the University of Massachusetts, Lowell, using a cobalt 60 radiation
source. Dosimetry was performed prior to device irradiation and the dose rate was determined to be
120 rads(Si)/sec. The total dose schedule can be found in Table I.
III. Test Setup:
All test samples were subjected to Group A Electrical Test in accordance with the device data sheet. In
addition, all devices received 320 hours of burn-in per MIL-STD-883 Method 1015 and were electrically
tested prior to irradiation. For test platform verification, one control device was tested at 25°C.
The devices were vertically aligned with the radiation source and enclosed in a Pb/Al container during
irradiation to minimize dose enhancement effects. Four devices were kept under bias during irradiation.
Four devices had all leads grounded during irradiation for the unbiased condition.
After each irradiation, the device leads were shorted together and were transported to the MSK
automatic electrical test platform and tested IAW MSK device data sheet. Testing was performed on
irradiated devices, at each total dose level. The control device was tested at the 0K rad(Si) and 150K
rad(Si) test points. Electrical tests were completed within one hour of irradiation.
IV. Data:
All performance curves are averaged from the test results of the biased and unbiased devices
respectively.
V. Summary:
Based on post irradiation performance, the MSK0002RH qualifies as 100K rad hardened. Input current
and output offset voltage both increases with increasing irradiation. Vout2, which is output voltage swing
at maximum load, decreased with increased irradiation. However all parameters stayed well with post
irradiation specifications.
Page 2 of 42
MSK0002RH Biased/Unbiased Dose Rate Schedule
Dosimetry Equipment
Bruker Biospin # 0371
Irradiation Date
6/13/2014
Exposure Length
(min:sec)
Incremental
Dose
rads(Si)
Cumulative Dose
rads(Si)
7:09
7:09
7:09
51,480
51,480
51,480
51,480
102,960
154,440
Biased S/N – K632, K650, K651, K655
Unbiased S/N – K673, K680, K681, K688
Table 1
Dose Time, Incremental Dose and Total Cumulative Dose
Page 3 of 42
Page 4 of 42
Page 5 of 42
Page 6 of 42
Page 7 of 42
Page 8 of 42
Total Dose Radiation Test Report
MSK 0002RH
Radiation Tolerant High Speed, Buffer Amp
March 29, 2007 (First Test)
April 7, 2009 (Second Test)
February 25, 2011 (Third Test)
June 1, 2012 (Fourth Test)
B. Horton
C. Salce
M.S. Kennedy Corporation
Liverpool, NY
Page 9 of 42
I.
Introduction:
The total dose radiation test plan for the MSK 0002RH was developed to qualify the device as
radiation tolerant up to 100 Krad(Si). The testing was performed beyond 100 Krad(Si) to show trends in
device performance as a function of total dose. The test does not classify maximum radiation tolerance
of the hybrid, but simply offers designers insight to the critical parameter-shifts up
to the specified total dose level.
MIL-STD-883 Method 1019.7 and ASTM F1892-06 were used as guidelines in the development and
implementation of the total dose test plan for the MSK 0002RH.
II.
Radiation Source:
Total dose was performed at the University of Massachusetts, Lowell, using a cobalt 60 radiation
source. Dosimetry was performed prior to device irradiation and the dose rate was determined to be
107 rads(Si)/sec. The total dose schedule can be found in Table I.
III. Test Setup:
All test samples were subjected to Group A Electrical Test in accordance with the device data sheet. In
addition, all devices received 320 hours of burn-in per MIL-STD-883 Method 1015 and were electrically
tested prior to irradiation. For test platform verification, one control device was tested at 25°C.
The devices were vertically aligned with the radiation source and enclosed in a Pb/Al container during
irradiation to minimize dose enhancement effects. Four devices were kept under bias during irradiation.
Four devices had all leads grounded during irradiation for the unbiased condition.
After each irradiation, the device leads were shorted together and were transported to the MSK
automatic electrical test platform and tested IAW MSK device data sheet. Testing was performed on
irradiated devices, as well as the control device, at each total dose level. Electrical tests were
completed within one hour of irradiation.
IV. Data:
All performance curves are averaged from the test results of the biased and unbiased devices
respectively.
V. Summary:
Based on post irradiation performance, the MSK0002RH qualifies as 100KRad hardened. Input current
and output offset voltage both increases with increasing irradiation. Vout2, which is output voltage swing
at maximum load, decreased with increased irradiation. However all parameters stayed well with post
irradiation specifications.
Page 10 of 42
MSK0002RH Biased/Unbiased Dose Rate Schedule
Dosimetry Equipment
Bruker Biospin # 0162
Irradiation Date
6/1/2012
Exposure Length
(min:sec)
Incremental
Dose
rads(Si)
Cumulative Dose
rads(Si)
0:08:01
0:08:01
0:08:01
51,467
51,467
51,467
51,467
102,934
154,401
Biased S/N – F875, F876, F877, F878
Unbiased S/N – F879, F880, F881, F882
Table 1
Dose Time, Incremental Dose and Total Cumulative Dose
Page 11 of 42
Page 12 of 42
Page 13 of 42
Page 14 of 42
Page 15 of 42
Page 16 of 42
Total Dose Radiation Test Report
MSK 0002RH
Radiation Tolerant High Speed, Buffer Amp
March 29, 2007 (First Test)
April 7, 2009 (Second Test)
February 25, 2011 (Third Test)
B. Horton
R. Wakeman
M.S. Kennedy Corporation
Liverpool, NY
Page 17 of 42
I.
Introduction:
The total dose radiation test plan for the MSK 0002RH was developed to qualify the device as
radiation tolerant up to 100 Krad(Si). The testing was performed beyond 100 Krad(Si) to show trends in
device performance as a function of total dose. The test does not classify maximum radiation tolerance
of the hybrid, but simply offers designers insight to the critical parameter-shifts up
to the specified total dose level.
MIL-STD-883 Method 1019.7 and ASTM F1892-06 were used as guidelines in the development and
implementation of the total dose test plan for the MSK 0002RH.
II.
Radiation Source:
Total dose was performed at the University of Massachusetts, Lowell, using a cobalt 60 radiation
source. Dosimetry was performed prior to device irradiation and the dose rate was determined to be
125 rads(Si)/sec. The total dose schedule can be found in Table I.
III. Test Setup:
All test samples were subjected to Group A Electrical Test in accordance with the device data sheet. In
addition, all devices received 320 hours of burn-in per MIL-STD-883 Method 1015 and were electrically
tested prior to irradiation. For test platform verification, one control device was tested at 25°C.
The devices were vertically aligned with the radiation source and enclosed in a Pb/Al container during
irradiation to minimize dose enhancement effects. Four devices were kept under bias during irradiation.
Four devices had all leads grounded during irradiation for the unbiased condition.
After each irradiation, the device leads were shorted together and were transported to the MSK
automatic electrical test platform and tested IAW MSK device data sheet. Testing was performed on
irradiated devices, as well as the control device, at each total dose level. Electrical tests were
completed within one hour of irradiation.
IV. Data:
All performance curves are averaged from the test results of the biased and unbiased devices
respectively.
V. Summary:
The devices performed very well with respect to TID tolerance. Input current and output offset voltage
both increases with increasing irradiation. Vout2, which is output voltage swing at maximum load,
decreased with increased irradiation. The devices also exhibited a slight voltage gain decrease. This
was most pronounced from 0Krad(Si) to 100Krad(Si) with less change from occurring from 100Krad(Si)
to 300Krad(Si).
Page 18 of 42
MSK0002RH Biased/Unbiased Dose Rate Schedule
Dosimetry Equipment
Bruker Biospin # 0162
Irradiation Date
02/25/2011
Exposure
Length
(min:sec)
Incremental Dose
Cumulative
rads(Si)
Dose rads(Si)
6:52
51,500
51,500
6:52
51,500
103,000
6:52
51,500
154,500
6:52
51,500
206,000
13:44
103,000
309,000
Biased S/N – D202, D204, D205, D206
Unbiased S/N – D207, D212, D213, D215
Table 1
Dose Time, Incremental Dose and Total Cumulative Dose
Page 19 of 42
Page 20 of 42
Page 21 of 42
Page 22 of 42
Page 23 of 42
Page 24 of 42
Total Dose Radiation Test Report
MSK 0002RH
Radiation Tolerant High Speed, Buffer Amp
March 29, 2007 (First Test)
April 7, 2009 (Second Test)
B. Erwin
R. Wakeman
M.S. Kennedy Corporation
Liverpool, NY
Page 25 of 42
I.
Introduction:
The total dose radiation test plan for the MSK 0002RH was developed to qualify the device as
radiation tolerant up to 100 Krad(Si). The testing was performed beyond 100 Krad(Si) to show trends in
device performance as a function of total dose. The test does not classify maximum radiation tolerance
of the hybrid, but simply offers designers insight to the critical parameter-shifts up
to the specified total dose level.
MIL-STD-883 Method 1019.7 and ASTM F1892-06 were used as guidelines in the development and
implementation of the total dose test plan for the MSK 0002RH.
II.
Radiation Source:
Total dose was performed at the University of Massachusetts, Lowell, using a cobalt 60 radiation
source. Dosimetry was performed prior to device irradiation and the dose rate was determined to be
180 rads(Si)/sec. The total dose schedule can be found in Table I.
III. Test Setup:
All test samples were subjected to Group A Electrical Test in accordance with the device data sheet. In
addition, all devices received 320 hours of burn-in per MIL-STD-883 Method 1015 and were electrically
tested prior to irradiation. For test platform verification, one control device was tested at 25°C.
The devices were vertically aligned with the radiation source and enclosed in a Pb/Al container during
irradiation to minimize dose enhancement effects. Four devices were kept under bias during irradiation.
Four devices had all leads grounded during irradiation for the unbiased condition.
After each irradiation, the device leads were shorted together and were transported to the MSK
automatic electrical test platform and tested IAW MSK device data sheet. Testing was performed on
irradiated devices, as well as the control device, at each total dose level. Electrical tests were
completed within one hour of irradiation.
IV. Data:
All performance curves are averaged from the test results of the biased and unbiased devices
respectively.
V. Summary:
The devices performed very well with respect to TID tolerance. Some shift in the output offset current
was seen from 0Krad(Si) to 300Krad(Si)..
Negative Vout2, which is output voltage swing at maximum load, decreased with increased irradiation.
However, the devices stayed within post irradiation limits throughout all testing.
The devices also exhibited a slight voltage gain decrease. This was most pronounced from 0 Krad(Si)
to 50Krad(Si).
Page 26 of 42
Dosimetry
Irradiation Date
Equipment
Bruker Biospin #0141
4/07/2009
Exposure
Length
(min:sec)
4:46
4:46
4:46
4:46
9:33
Incremental
Dose
rads(Si)
51,480
51,480
51,480
51,480
103,140
Cumulative
Dose
rads(Si)
51,480
102,960
154,440
205,920
309,060
Biased Devices: 621, 622, 623, 624
Unbiased Devices: 625, 641, 651, 652
Table 1
Dose Time, Incremental Dose and Total Cumulative Dose
Page 27 of 42
Quiescent Current (mA)
MSK0002RH
Positive Quiescent Current
vs. Total Dose
6.8
6.6
6.4
Avg gnd
Avg bias
Control
6.2
6.0
5.8
5.6
0K
50K
100K
150K
200K
300K
Total dose (rad(Si))
MSK0002RH
Negative Quiescent Current vs. Total Dose
Quiescent Current (mA)
-5.6
-5.8
-6.0
Avg gnd
Avg bias
Control
-6.2
-6.4
-6.6
-6.8
0K
50K
100K
150K
200K
300K
Total dose (rad(Si))
Page 28 of 42
MSK0002RH
Input Current vs. Total Dose
40.0
Input Current (uA)
35.0
30.0
25.0
Avg gnd
Avg bias
Control
20.0
15.0
10.0
5.0
0.0
0K
50K
100K
150K
200K
300K
Total dose (rad(Si))
MSK0002RH
Output Offset Voltage vs. Total Dose
Output Offset V (mV)
24.0
19.0
14.0
Avg gnd
Avg bias
Control
9.0
4.0
-1.0
-6.0
0K
50K
100K
150K
200K
300K
Total dose (rad(Si))
Page 29 of 42
MSK0002RH
Positive V out Swing 1 vs. Total Dose
V out Swing (V pk)
13.0
12.5
12.0
Avg gnd
Avg bias
Control
11.5
11.0
10.5
10.0
0K
50K
100K
150K
200K
300K
Total dose (rad(Si))
MSK0002RH
Negative V out Swing 1 vs. Total Dose
V out Swing (V pk)
-9.0
-9.5
-10.0
Avg gnd
Avg bias
Control
-10.5
-11.0
-11.5
-12.0
0K
50K
100K
150K
200K
300K
Total dose (rad(Si))
Page 30 of 42
MSK0002RH
Positive V out Swing 2 vs. Total Dose
11.60
V out Swing (V pk)
11.50
11.40
11.30
Avg gnd
Avg bias
Control
11.20
11.10
11.00
10.90
10.80
0K
50K
100K
150K
200K
300K
Total dose (rad(Si))
Page 31 of 42
MSK0002RH
Negative V out Swing 2 vs. Total Dose
-8.00
V out Swing (V pk)
-8.50
-9.00
-9.50
Avg gnd
Avg bias
Control
-10.00
-10.50
-11.00
-11.50
-12.00
0K
50K
100K
150K
200K
300K
Total dose (rad(Si))
Gain (V/V)
MSK0002RH
Voltage Gain vs. Total Dose
1.00
0.99
0.98
0.97
0.96
0.95
0.94
0.93
0.92
0.91
0.90
Avg gnd
Avg bias
Control
0K
50K
100K
150K
200K
300K
Total dose (rad(Si))
Page 32 of 42
MSK0002RH
Rise Time vs. Total Dose
9.0
Rise Time (nS)
8.5
8.0
Avg gnd
Avg bias
Control
7.5
7.0
6.5
6.0
0K
50K
100K
150K
200K
300K
Total dose (rad(Si))
Page 33 of 42
Total Dose Radiation Test Report
MSK 0002RH
Radiation Tolerant High Speed, Buffer Amp
March 29, 2007
J. Douglas
B. Erwin
P. Musil
M.S. Kennedy Corporation
Liverpool, NY
Page 34 of 42
I.
Introduction:
The total dose radiation test plan for the MSK 0002RH was developed to qualify the device as
radiation tolerant up to 100 Krad(Si). The testing was performed beyond 100 Krad(Si) to show trends
in device performance as a function of total dose. The test does not classify maximum radiation
tolerance of the hybrid, but simply offers designers insight to the critical parameter-shifts up
to the specified total dose level.
MIL-STD-883 Method 1019.7 and ASTM F1892-06 were used as guidelines in the development and
implementation of the total dose test plan for the MSK 0002RH.
II.
Radiation Source:
Total dose was performed at the University of Massachusetts, Lowell, using a cobalt 60 radiation
source. Dosimetry was performed prior to device irradiation and the dose rate was determined to be
108 rads(Si)/sec. The total dose schedule can be found in Table I.
III. Test Setup:
All test samples were subjected to Group A Electrical Test in accordance with the device data sheet.
In addition, all devices received 320 hours of burn-in per MIL-STD-883 Method 1015 and were
electrically tested prior to irradiation. For test platform verification, one control device was tested at
25°C.
The devices were vertically aligned with the radiation source and enclosed in a Pb/Al container during
irradiation to minimize dose enhancement effects. Four devices were kept under bias during
irradiation. Four devices had all leads grounded during irradiation for the unbiased condition.
After each irradiation, the device leads were shorted together and were transported to the MSK
automatic electrical test platform and tested IAW MSK device data sheet. Testing was performed on
irradiated devices, as well as the control device, at each total dose level. Electrical tests were
completed within one hour of irradiation.
IV. Data:
All performance curves are averaged from the test results of the biased and unbiased devices
respectively.
V. Summary:
The devices performed very well with respect to TID tolerance. Some negative shift in the input offset
current was seen from 0Krad(Si) to 100Krad(Si). The devices stabilized and input offset current
stayed fairly constant with a slightly negative shift from 100Krad(Si) up to 300Krad(Si).
Vout2, which is output voltage swing at maximum load, decreased with increased irradiation. The
decrease occurred between 0Krad(Si) and 100Krad(Si). This parameter also stabilized at higher
radiation doses.
The devices also exhibited a slight voltage gain decrease. Again, this was most pronounced from
50Krad(Si) to 100Krad(Si) with less change from occurring from 100Krad(Si) to 300Krad(Si).
Page 35 of 42
Irradiation Date
3/29/2007
Dosimetry Equipment
Bruker Biospin #0141
Exposure Incremental
Length
Dose
(min:sec)
rads(Si)
7:57
51,516
7:57
51,516
7:57
51,516
7:57
51,516
7:57
51,516
7:57
51,516
Cumulative
Dose
rads(Si)
51,516
103,032
154,548
206,064
257,580
309,096
Table 1
Dose Time, Incremental Dose and Total Cumulative Dose
Page 36 of 42
Positive IQ vs. Total Dose
7.00
6.80
Quiescent Current (mA)
6.60
6.40
6.20
Avg Bias
6.00
Avg Gnd
Control
5.80
5.60
5.40
5.20
5.00
0
50
100
150
200
250
300
Total Dose In Krad(Si)
Negative IQ vs. Total Dose
-5.00
-5.20
Quiescent Current (mA)
-5.40
-5.60
-5.80
Avg Bias
-6.00
Avg Gnd
Control
-6.20
-6.40
-6.60
-6.80
-7.00
0
50
100
150
200
250
300
Total Dose In Krad(Si)
Page 37 of 42
Offset Current vs. Total Dose
0.00
-2.00
Offset Current (uA)
-4.00
-6.00
Avg Bias
Avg Gnd
Control
-8.00
-10.00
-12.00
-14.00
0
50
100
150
200
250
300
Total Dose In Krad(Si)
Offset Voltage vs. Total Dose
0.00
-2.00
Offset Voltage (mV)
-4.00
-6.00
Avg Bias
Avg Gnd
Control
-8.00
-10.00
-12.00
-14.00
0
50
100
150
200
250
300
Total Dose In Krad(Si)
Page 38 of 42
Vout 1 Positive vs. Total Dose
15.00
14.50
14.00
Vout (Volts-peak)
13.50
13.00
Avg Bias
12.50
Avg Gnd
Control
12.00
11.50
11.00
10.50
10.00
0
50
100
150
200
250
300
Total Dose In Krad(Si)
Page 39 of 42
Vout 1 Negative vs. Total Dose
-10.00
-10.50
-11.00
Vout (Volts-peak)
-11.50
-12.00
Avg Bias
-12.50
Avg Gnd
Control
-13.00
-13.50
-14.00
-14.50
-15.00
0
50
100
150
200
250
300
Total Dose In Krad(Si)
Vout 2 Positive vs. Total Dose
15.00
14.00
Vout (Volts-peak)
13.00
12.00
Avg Bias
Avg Gnd
Control
11.00
10.00
9.00
8.00
0
50
100
150
200
250
300
Total Dose In Krad(Si)
Page 40 of 42
Vout 2 Negative vs. Total Dose
-8.00
-9.00
Vout (Volts-peak)
-10.00
Avg Bias
-11.00
Avg Gnd
Control
-12.00
-13.00
-14.00
-15.00
0
50
100
150
200
250
300
Total Dose In Krad(Si)
Rise Time vs. Total Dose
8.00
7.50
Rise Time (nS)
7.00
6.50
Avg Bias
6.00
Avg Gnd
Control
5.50
5.00
4.50
4.00
0
50
100
150
200
250
300
Total Dose In Krad(Si)
Page 41 of 42
Gain vs. Total Dose
1.00
0.99
0.98
Gain (V/V)
0.97
0.96
Avg Bias
0.95
Avg Gnd
Control
0.94
0.93
0.92
0.91
0.90
0
50
100
150
200
250
300
Total Dose In Krad(Si)
Page 42 of 42