200W Medical series 1~4 Output Medical Type MODEL: MPQ-200B OUTPUT FUNCTION TEST NO 1 2 TEST ITEM SPECICATION TEST CONDITION RIPPLE & NOISE V1: 80 mVp-p (Max ) I/P: 230VAC V1: 30 mVp-p (Max ) V2: 120 mVp-p (Max ) 52 mVp-p (Max ) mVp-p (Max ) O/P:FULL LOAD Ta:25℃ V2: V3: 80 V3: 8 mVp-p (Max ) V3: 80 mVp-p (Max ) V4: 8 mVp-p (Max ) OUTPUT VOLTAGE CH1: 4.75 V~ 5.5 V ADJUST RANGE RESULT VERDICT I/P: 230 VAC 4.62 V~ 5.94 V/ 230 VAC I/P: 115 VAC 4.62 V~ 5.94 V/ 115 VAC P O/P:MIN LOAD Ta:25℃ 3 OUTPUT VOLTAGE V1: 2 %~ -2 % (Max) I/P: 115 VAC / 264 VAC V1: 0.2 %~ -0.2 % TOLERANCE V2: 8 %~ -8 % (Max) O/P:FULL/ MIN Ta:25℃ V2: 5.6 %~ -5.6 % V3: 0.15 %~ -0.15 % V4: 0.05 %~ -0.05 % V3: 5 %~ -5 % (Max) LOAD V4: 5 %~ -5 % (Max) 4 LINE REGULATION V1: 0.5 %~ -0.5 % (Max) I/P: 115 VAC ~ 264 VAC V1: 0 %~ 0 % V2: 1 %~ -1 % (Max) O/P:FULL LOAD Ta:25℃ V2: 0 %~ 0 % V3: 0.15 %~ -0.15 % V4: 0.05 %~ -0.05 % V3: 0.5 %~ -0.5 % (Max) V4: 0.5 %~ -0.5 % (Max) 5 LOAD REGULATION V1: 1 %~ -1 % (Max) I/P: 230 VAC V1: 0 %~ 0 % V2: 5 %~ -5 % (Max) O/P:FULL ~MIN LOAD Ta:25℃ V2: 1.8 %~ -0.5 % V3: 0.4 %~ -0.4 % V3: 1 %~ -1 % (Max) V4: 1 %~ -1 % (Max) 6 CROSS REGULATION V4: 0.05 %~ -0.05 % V1: 1 %~ -1 % (Max) I/P: 230 VAC V1: 0 %~ 0 % V2: 5 %~ -5 % (Max) O/P: Testing O/P 60%LOAD V2: 3.1 %~ -3.1 % V3: 0.13 %~ -0.13 % V4: 0 %~ 0 % V3: 1 %~ -1 % (Max) Other O/P 40%LOAD V4: 1 %~ -1 % (Max) 7 SET UP TIME Change Ta:25℃ P P P P 230VAC: 1000 ms (Max) I/P: 230 VAC 230VAC/ 698 ms 115 VAC: 3000 ms (Max) I/P: 115 VAC 115VAC/ 1884 ms O/P:FULL LOAD Ta:25℃ 8 RISE TIME 230VAC: 20 ms (Max) I/P: 230 VAC 230VAC/ 16 ms 115VAC: 20 ms (Max) I/P: 115 VAC 115VAC/ 14 ms O/P:FULL LOAD Ta:25℃ 9 HOLD UP TIME 230VAC: 16 ms(TYP) I/P: 230 VAC 230VAC/ 18 ms 115VAC: 16 ms (TYP) I/P: 115 VAC 115VAC/ 18.3 ms <5 % O/P:FULL LOAD Ta:25℃ 10 OVER/UNDERSHOOT TEST I/P: 230 VAC < +5% TEST: O/P:FULL LOAD Ta:25℃ 11 DYNAMIC LOAD V1: 1000 mVp-p I/P: 230 VAC P P P P P 80 mVp-p O/P:FULL /Min LOAD P 90%DUTY/1KHZ Ta:25℃ Model: MPQ-200B Test Report 1 / 6 200W Medical series 1~4 Output Medical Type INPUT FUNCTION TEST NO 1 TEST ITEM SPECICATION TEST CONDITION INPUT VOLTAGE RANGE 90VAC~264 VAC) I/P:TESTING RESULT VERDICT 75 V~264V O/P:FULL LOAD Ta:25℃ I/P: TEST: OK LOW-LINE-3V= 87 V P HIGH-LINE+15%=300 V O/P:FULL/MIN LOAD ON: 30 Sec . OFF: 30 Sec 10MIN ( AC POWER ON/OFF NO DAMAGE ) 2 3 INPUT FREQUENCY RANGE POWER FACTOR 47HZ ~440HZ I/P: 90 VAC ~ 264 VAC NO DAMAGE OSC O/P:FULL~MIN LOAD Ta:25℃ TEST: OK 0.95/ 230 VAC(TYP) I/P: 230 VAC PF= 0.97 /230 VAC. 0.98/1150 VAC(TYP) I/P: 115 VAC PF= 0.99 /115 VAC P P O/P:FULL LOAD Ta:25℃ 4 EFFICIENCY 78% (TYP) I/P: 230 VAC 79.2 % O/P:FULL LOAD Ta:25℃ 5 INPUT CURRENT P 230V/ 1.6 A (TYP) I/P: 230 VAC I= 1.14 A/ 230 VAC 115V/ 3.5 A (TYP) I/P: 115 VAC I= 2.33 A/ 115 VAC O/P:FULL LOAD Ta:25℃ 6 INRUSH CURRENT 230V/ 40 A(TYP) I/P: 230 VAC I= 29.5 A/ 230 VAC 115V/ 25 A (TYP) I/P: 115 VAC I= 18 A/ 115 VAC O/P:FULL LOAD Ta:25℃ COLD START 7 LEAKAGE CURRENT a. For earth leakage current b. For patient leakage current a. <180 uA / 264 VAC I/P: 264 VAC a. 67.9 uA b. <100 uA / 264 VAC O/P:Min LOAD Ta:25℃ b. 58 uA Model: MPQ-200B P P P Test Report 2 / 6 200W Medical series 1~4 Output Medical Type PROTECTION FUNCTION TEST NO 1 TEST ITEM OVER LOAD PROTECTION SPECICATION 120 %~ 160 % TEST CONDITION I/P: 230 VAC I/P: 115 VAC O/P:TESTING Ta:25℃ 2 OVER VOLTAGE PROTECTION CH1: 5.75 V~ 6.75 V I/P: 230 VAC I/P: 115 VAC 3 OVER TEMPERATURE ROTECTION SPEC: TSW1: 95 + 5℃ O.T.P. RESULT VERDICT 135 %/ 230 VAC 135 %/ 115 VAC P Hiccup Mode 5.95 V/ 230 VAC 5.95 V/ 115 VAC P O/P:MIN LOAD Ta:25℃ Shunt down Re- power ON I/P: 230 VAC O.T.P. Active Shut down o/p voltage,recovers O/P:FULL LOAD NO DAMAGE automatically after P temperature goes down 4 SHORT PROTECTION SHORT EVERY OUTPUT I/P: 264 VAC NO DAMAGE 1 HOUR NO DAMAGE O/P:FULL LOAD Ta:25℃ Hiccup Mode RESULT P CONTROL FUNCTION TEST NO 1 2 TEST ITEM SPECICATION TEST CONDITION REMOTE CONTROL Rc+ / Rc- I/P: 230 VAC 0V~ 0.8V POWER ON 4V~ 10V POWER OFF O/P:FULL LOAD Ta:25℃ DELAY 10ms ~ 500ms I/P: 230 VAC 296 ms/ 230 VAC I/P: 115 VAC 296 ms/ 115 VAC POWER GOOD SIGNAL 0V ~ 2.5 V 2.6 V ~ 10 V VERDICT POWER ON POWER OFF P P O/P:FULL LOAD Ta:25℃ 3 POWER FAIL SIGNAL > 1ms I/P: 230 VAC 5 ms/ 230 VAC I/P: 115 VAC 5 ms/ 115 VAC P O/P:FULL LOAD Ta:25℃ 4 REMOTE SENSE >0.25V I/P: 230 VAC >0.25 O/P:FULL LOAD Ta:25℃ P Model: MPQ-200B Test Report 3 / 6 200W Medical series 1~4 Output Medical Type ENVIRONMENT TEST NO 1 TEST ITEM SPECICATION TEMPERATURE RISE TEST MODEL:MPQ-200B WITH FAN TEST CONDITION RESULT VERDICT 1. ROOM AMBIENT BURN-IN:2HRS I/P: 230VAC O/P: FULL LOAD Ta= 28.8 ℃ 2. HIGH AMBIENT BURN-IN:2HRS I/P: 230VAC O/P: FULL LOAD Ta= 50.7 ℃ NO 2 OVER LOAD BURN-IN TEST Position ROOM AMBIENT HIGH AMBIENT Ta= 28.8℃ Ta= 50.7 ℃ 52.5℃ 74.5℃ P/N 1 LF2 LF-104 2 D23 31DF6 3A/600V 38.2℃ 60.1℃ IRFP450 14A/500V 45.7℃ 68.0℃ 3 Q20 4 T1 COIL TF-1171 45.3℃ 66.3℃ 5 D100 SBL3040PT 30A/40V LT 80.3℃ 100.6℃ 6 D200 S20LC20U 20A/200V SHI 57.9℃ 79.1℃ 7 D300 F10SC6 10A/60V SHI 57.9℃ 78.8℃ 8 RG70 R05PI 3A/35V KEC 58.6℃ 79.3℃ 9 BD1 D3SB60 4A/600V SHI 48.4℃ 71.8℃ 10 C5 150U/400V RUB 105℃ 40.6℃ 60.2℃ 84.8℃ 11 L2 TR-570 66.3℃ 12 Q1 IRFP460A 20A/500V 63.6℃ 84.4℃ 13 U1 ML4800CP 45.2℃ 66.7℃ 14 L100 TR-523 69.3℃ 91.7℃ 15 C110 2200U/6.3V RUB 105℃ 47.0℃ 69.9℃ NO DAMAGE I/P: 230 VAC 1 HOUR ( MIN ) O/P: V1@20A P TEST:OK V2@9A P [email protected] [email protected] Ta:25℃ 3 LOW TEMPERATURE TURN ON AFTER 2 HOUR TURN ON TEST I/P: 230 VAC O/P: 80 ﹪LOAD TEST:OK P Ta= -20 ℃ 4 AFTER 12 HOURS I/P: 272 VAC HIGH TEMPERATURE IN CHAMBER ON CONTROL 50 ℃ O/P:FULL LOAD Ta= 50 ℃ NO DAMAGE + 0.04 %(0~50℃) I/P: 230 VAC HIGH VOLTAGE TURN ON TEST 5 TEMPERATURE VIBRATION TEST HUMIDITY= P 95 %R.H + 0.02 %(0~50℃) P O/P:FULL LOAD COEFFICIENT 6 TEST:OK HIGH HUMIDITY 1 Carton & 1 Set Operating TEST:OK (1) Waveform: Sine Wave (2) Frequency:10~500Hz (3) Sweep Time:10min/sweep cycle P (4) Acceleration:2G (5) Test Time:1 hour in each axis (X.Y.Z) (6) Ta:25℃ Model: MPQ-200B Test Report 4 / 6 200W Medical series 1~4 Output Medical Type SAFETY TEST NO 1 2 3 4 TEST ITEM SPECICATION TEST CONDITION RESULT WITHSTAND VOLTAGE I/P-O/P: 4 KVAC/min I/P-O/P: 4.2 KVAC/min I/P-O/P: 2.05 mA I/P-FG: 1.5 KVAC/min I/P-FG: 1.8 KVAC/min I/P-FG: 0.82 mA O/P-FG: 1.5 KVAC/min O/P-FG: 1.8 KVAC/min Ta:25℃ O/P-FG: 0.13 mA I/P-O/P:500VDC>100MΩ I/P-O/P: 500 VDC I/P-O/P: 2 GΩ I/P-FG: 500VDC>100MΩ I/P-FG: 500 VDC O/P-FG:500VDC>100MΩ O/P-FG: 500 Ta:25℃ VDC I/P-FG: 30 GΩ O/P-FG: 30 GΩ ISOLATION RESISTANCE GROUNDING CONTINUITY FG(PE) TO CHASSIS APPROVAL OR TRACE < 100 mΩ TUV: Certificate NO :TA50069957 VERDICT P NO DAMAGE P NO DAMAGE 40 A / 2min Ta:25℃ 17 mΩ P P UL: File NO :E227340 E.M.C TEST NO 1 2 3 4 5 TEST ITEM SPECICATION TEST CONDITION RESULT HARMONIC EN61000-3-2 I/P: 230 VAC/50HZ PASS CLASS A O/P:FULL LOAD Ta:25℃ EN55022 I/P: 230 VAC (50HZ) PASS CLASS B O/P:FULL/50% LOAD Ta:25℃ Test by certified Lab CONDUCTION RADIATION E.S.D E.F.T 6 SURGE P EN55022 I/P: 230 VAC (50HZ) PASS CLASS B O/P:FULL LOAD Ta:25℃ Test by certified Lab EN61000-4-2 I/P: 230 VAC/50HZ CRITERIA A INDUSTRY AIR:8KV / Contact:4KV O/P:FULL LOAD Ta:25℃ EN61000-4-4 I/P: 230 VAC/50HZ INDUSTRY O/P:FULL LOAD Ta:25℃ INPUT: 2KV IEC61000-4-5 I/P: 230 VAC/50HZ INDUSTRY O/P:FULL LOAD Ta:25℃ L-N :2KV VERDICT P P P CRITERIA A P CRITERIA A P L,N-PE:4KV 7 Test by certified Lab & Test Report Prepare M.T.B.F & LIFE CYCLE CALCULATION NO 1 TEST ITEM SPECICATION CAPACITOR SUPPOSE C110 IS THE MOST CRITICAL COMPONENT I/P: 230VAC O/P:FULL LOAD Ta= 25 ℃ LIFE TIME= 437900 HRS LIFE CYCLE I/P: 230VAC 2 MTBF TEST CONDITION O/P:FULL LOAD Ta= 50 ℃ RESULT VERDICT P LIFE TIME= 72238 HRS MIL-HDBK-217F NOTICES2 PARTS COUNT P TOTAL FAILURE RATE: 262.1k HRS Model: MPQ-200B Test Report 5 / 6 200W Medical series 1~4 Output Medical Type COMPONENT STRESS TEST NO 1 TEST ITEM SPECICATION TEST CONDITION RESULT Power Transistor Q20 Rated I/P:High-Line +3V = 267 V (1) 382 V ( D to S) or (C to E) Peak Voltage IRFP450 : 500 V 14A O/P: (1)Full Load Turn on (2) 390 V (3) 390 V I/P:High-Line +3V = 267 V (1) 21.2 V O/P: (1) Full Load Turn on (2) 18.9 V (3) 21.2 V (1) 47.8 V (2) 42.6 V (3) 47.8 V (1) 26 V (2) 24.9 V (3) 50 V D300 Rated (1) 66.4 V FMX-12SL : 200 V 10 A (2) 50 V (3) 70.4 V I/P:High-Line +3V = 267 V (1) 386 V O/P: (1)Full Load Turn on /Off (2) 386 V (2) Min load Turn on /Off (3) 396 V (3)Full Load /Min load Change (4) 390 V (2) Full Load VERDICT P (3)Output Short Ta:25℃ 2 Diode Peak Voltage D100 Rated SBL3040PT : 40 V 30 A (2) Full Load (3)Output Short Ta:25℃ D200 Rated S20LC20U : 200 V 20 A D300 Rated SF10SC6 : 60 V 3 Input Capacitor Voltage 10 A C 5 Rated : 150u / 400V/ 105℃ P P (4) Burn in 1hour Ta:25℃ 4 Control IC Voltage Test U1 Rated I/P:High-Line +3V = 267 V (1) 15 V ML4800 : 18 V O/P: (1)Full Load Turn on /Off (2) 15.2 V (2) Min load Turn on /Off (3) 15.6 V P (3)Full Load /Min load Change Ta:25℃ 5 Power Transistor Q1 Rated I/P:High-Line +3V = 267 V (1) 388 V ( D to S) or (C to E) Peak Voltage IRFP460A : 500 V 20A O/P: (1)Full Load Turn on (2) 406 V (3) 386 V (2) Full Load P (3)Output Short Ta:25℃ DATE SAMPLE TEST RESULT TESTER APPROVAL 2005/5/17 RD SAMPLE PASS VINCENT TSENG MAX LIN 2005/9/3 PRODUCT SAMPLE W0507B20 PASS VINCENT TSENG MAX LIN 2003/12/12 Model: MPQ-200B A50-F023 Test Report 6 / 6