Test Report: GSM36B48 36W AC-DC Single Output Medical Adaptor DESIGN VERIFY TEST Output Function Test Input Function Test Protection Function Test Component Stress Test SAFETY & E.M.C. TEST Safety Test E.M.C. Test RELIABILITY TEST ENVIRONMENT TEST 36W AC-DC Single Output Medical Adaptor GSM36B series DESIGN VERIFY TEST OUTPUT FUNCTION TEST NO 1 TEST ITEM SPECIFICATION TEST CONDITION RIPPLE & NOISE V1:240 I/P: 230VAC mVp-p (Max ) RESULT VERDICT V1: 80 mVp-p (Max ) I/P: 115VAC P O/P:FULL LOAD Ta:25℃ 2 OUTPUT VOLTAGE V1:2 %~ -2 % (Max) I/P:85 VAC / 264 VAC O/P:FULL/ MIN TOLERANCE V1: 1 %~ -1 % LOAD P Ta:25℃ 3 LINE REGULATION V1:1 %~ -1 % (Max) I/P:85 VAC ~ 264 VAC V1: 0 %~ -0 % O/P:FULL LOAD P Ta:25℃ 4 LOAD REGULATION V1:2 %~ -2 % (Max) I/P: 230VAC V1: 1 %~ -1 % I/P: 115VAC P O/P:FULL ~MIN LOAD Ta:25℃ 5 SET UP TIME 230VAC: 500 ms (Max) I/P: 230VAC 230VAC/ 180 ms 115VAC: 1000 ms(Max) I/P: 115 VAC 115VAC/ 390 ms 230VAC/ 20 ms 115VAC/ 20 ms P O/P:FULL LOAD Ta:25℃ 6 RISE TIME 230VAC: 30 ms (Max) I/P: 230VAC 115VAC: 30 ms (Max) I/P: 115 VAC P O/P:FULL LOAD Ta:25℃ 7 HOLD UP TIME 230VAC: 16 ms (TYP) I/P: 230VAC 230VAC/ 80 ms 115VAC: 16 ms (TYP) I/P: 115 VAC 115VAC/ 40 ms P O/P:FULL LOAD Ta:25℃ 8 OVER/UNDERSHOOT < +5% TEST I/P: 230VAC TEST: <5 % I/P: 115VAC TEST: <5 % O/P:FULL LOAD P Ta:25℃ 9 DYNAMIC LOAD V1: 4800 mVp-p I/P: 230VAC (1)400 mVp-p /230V I/P: 115VAC (2)362 mVp-p /230V (1).O/P:FULL /Min LOAD 90%DUTY/ (3)356 mVp-p /230V 1KHZ (2).O/P:FULL /Min LOAD 90%DUTY/ (4)336 mVp-p /230V 3KHZ (3).O/P:FULL /Min LOAD 90%DUTY/ (1)428 mVp-p /115V (2)374 mVp-p /115V 5KHZ (4).O/P:FULL /Min LOAD 50%DUTY/ (3)350 mVp-p /115V (4)370 mVp-p /115V P 120HZ Ta:25℃ Model: GSM36B48 Test Report 2 / 7 36W AC-DC Single Output Medical Adaptor GSM36B series INPUT FUNCTION TEST NO 1 TEST ITEM SPECIFICATION INPUT VOLTAGE RANGE TEST CONDITION 85 VAC~264 VAC RESULT I/P:TESTING VERDICT 66.2 V~264V O/P:FULL LOAD Ta:25℃ I/P: TEST:OK LOW-LINE-3V= 82 V P HIGH-LINE+15%=300 V O/P:FULL/MIN LOAD ON: 30 Sec . OFF: 30 Sec 10MIN ( POWER ON/OFF NO DAMAGE ) 2 INPUT FREQUENCY RANGE I/P:85 47HZ ~63 HZ VAC ~ 264 VAC TEST:OK P O/P:FULL~MIN LOAD NO DAMAGE OSC Ta:25℃ 3 EFFICIENCY 88 % (TYP) I/P: 230VAC 89.102 % /230V I/P: 115VAC 88.819 % /115V O/P:FULL LOAD P Ta:25℃ 4 INPUT CURRENT 230V/ 0.45 A (TYP) I/P: 230VAC I= 0.339 A/ 230 VAC 115V/ 0.9 A (TYP) I/P: 115 VAC I= 0.594 A/ 115 VAC O/P:FULL LOAD P Ta:25℃ 5 INRUSH CURRENT 230V/ 55 LEAKAGE CURRENT NO LOAD CONSUMPTION I= 43 A/ 230 VAC I= 26 A/ 115 VAC A(TYP) COLD START O/P:FULL LOAD < uA / 264 VAC I/P: 264 VAC L-FG: 44 uA O/P:Min LOAD N-FG: 44 uA I/P:115VAC < 0.061 W I/P:230VAC < 0.08 W 30 < 50 0.1 P Ta:25℃ Ta:25℃ TOUCH CURRENT 7 I/P: 230VAC I/P: 115 VAC 115V/ 6 A (TYP) W O/P:NO LOAD P P Ta:25℃ Model: GSM36B48 Test Report 3 / 7 36W AC-DC Single Output Medical Adaptor GSM36B series PROTECTION FUNCTION TEST NO 1 TEST ITEM SPECIFICATION OVER LOAD PROTECTION 105 % ~ 170 % TEST CONDITION I/P: 230VAC I/P: 115 VAC O/P:TESTING RESULT VERDICT 156%/ 230 VAC 142 %/ 115 VAC P Hiccup Mode Ta:25℃ 2 OVER VOLTAGE PROTECTION CH1:50.4 V ~64.8 V I/P: 230VAC I/P: 115 VAC O/P:MIN LOAD 64 V/ 230 VAC 63 V/ 115 VAC P Shut down Re- power ON Ta:25℃ 3 SHORT PROTECTION SHORT EVERY OUTPUT I/P: 264 VAC NO DAMAGE 1 HOUR NO DAMAGE O/P:FULL LOAD Hiccup Mode P Ta:25℃ COMPONENT STRESS TEST NO 1 TEST ITEM SPECIFICATION Power Transistor Q1 Rated: 10A/600V ( D to S) or (C to E) Peak Voltage TEST CONDITION RESULT I/P:High-Line +3V = 267 V (1) 474 V O/P: (1)Full Load Turn on (2) 480 V (3) 470 V I/P:High-Line +3V = 267 V (1) 352 V O/P: (1)Full Load Turn on (2) 346 V (3) 342 V I/P:High-Line +3V = 267 V (1) 376 V O/P: (1)Full Load Turn on /Off (2) 384 V (2) Min load Turn on /Off (3) 380 V I/P:High-Line +3V = 267 V (1) 14.4 V O/P: (1)Full Load Turn on /Off (2) 14 V (2) Min load Turn on /Off (3) 14.4 V (2) Output Short VERDICT P (3)Full load continue Ta:25℃ 2 Diode Peak Voltage D100 Rated: 6A/400V (2)Output Short P (3)Full load continue Ta:25℃ 3 Input Capacitor Voltage C5 Rated:56u/400V P (3)Full Load /Min load Change Ta:25℃ 4 Control IC Voltage Test U1 Rated: 10V~27V P (3)Full Load /Min load Change Ta:25℃ Model: GSM36B48 Test Report 4 / 7 36W AC-DC Single Output Medical Adaptor GSM36B series SAFETY & E.M.C. TEST SAFETY TEST NO 1 2 TEST ITEM SPECIFICATION TEST CONDITION RESULT WITHSTAND VOLTAGE I/P-O/P: 4 KVAC/min I/P-O/P: 4.4 KVAC/min I/P-O/P: Ta:25℃ NO DAMAGE ISOLATION RESISTANCE I/P-O/P:500VDC>100MΩ I/P-O/P: 500 VDC Ta:25℃/70%RH VERDICT 0.63 mA 30 GΩ I/P-O/P: P P NO DAMAGE E.M.C TEST NO 1 2 TEST ITEM SPECIFICATION TEST CONDITION RESULT HARMONIC EN61000-3-2 I/P:240VAC/230VAC/220VACVAC/60HZ PASS CLASS A O/P:100%,75%,50%,25%LOAD Ta:25℃ EN55011 I/P: 230VAC CLASS B O/P:FULL/50% LOAD CONDUCTION (50HZ) VERDICT P PASS P Test by certified Lab Ta:25℃ 3 RADIATION EN55011 I/P: 230VAC (50HZ) PASS CLASS B O/P:FULL LOAD Test by certified Lab P Ta:25℃ 4 5 6 E.S.D E.F.T SURGE EN61000-4-2 I/P: 230VAC MEDICAL AIR:8KV / Contact:6KV O/P:FULL LOAD /50HZ EN61000-4-4 I/P: 230VAC MEDICAL INPUT: 2KV O/P:FULL LOAD EN61000-4-5 I/P: 230VAC /50HZ MEDICAL L-N :1KV O/P:FULL LOAD CRITERIA A P Ta:25℃ /50HZ CRITERIA A P Ta:25℃ CRITERIA A P Ta:25℃ L,N-PE:2KV 7 Test by certified Lab & Test Report Prepare Model: GSM36B48 Test Report 5 / 7 36W AC-DC Single Output Medical Adaptor GSM36B series RELIABILITY TEST ENVIRONMENT TEST NO 1 TEST ITEM SPECIFICATION TEMPERATURE RISE TEST MODEL:GSM36B12 TEST CONDITION RESULT VERDICT 1. ROOM AMBIENT BURN-IN: 2 HRS I/P: 230VAC O/P: FULL LOAD Ta= 27.1℃ 2. HIGH AMBIENT BURN-IN: 2 HRS I/P: 230VAC O/P: FULL LOAD Ta= 46.1 ℃ NO 2 OVER LOAD BURN-IN TEST Position ROOM AMBIENT HIGH AMBIENT Ta= 27.1℃ Ta= 46.1℃ 67.7℃ 80.6℃ 1 LF1 2 BD1 79.0℃ 92.0℃ 3 Q1 96.1℃ 110.6℃ 4 D1 90.5℃ 104.7℃ 5 C5 69.2℃ 82.3℃ 6 C40 72.2℃ 86.1℃ 7 U1 71.9℃ 85.6℃ 8 T1 85.5℃ 99.0℃ 9 C105 72.7℃ 87.3℃ 10 D100 83.8℃ 97.9℃ 11 C110 63.5℃ 76.7℃ NO DAMAGE I/P: 230VAC 1 HOUR ( MIN ) O/P: 155 I/P: 115VAC P TEST:OK P ﹪LOAD Ta:25℃ 3 LOW TEMPERATURE TURN ON AFTER 2 HOUR I/P: 264VAC/100VAC TURN ON TEST Ta= 4 5 AFTER 12 HOURS I/P: 272 VAC HIGH TEMPERATURE O/P:FULL LOAD HIGH VOLTAGE IN CHAMBER ON CONTROL 40 ℃ TURN ON TEST NO DAMAGE HUMIDITY= + 0.03 %/℃(0~40℃) STORAGE TEMPERATURE TEST TEST:OK P Ta= 40 ℃ 95 %R.H I/P: 230VAC + 0.023 %/℃(0~40℃) O/P:FULL LOAD COEFFICIENT 6 P -30 ℃ HIGH HUMIDITY TEMPERATURE TEST:OK O/P: 100 ﹪LOAD 1. Thermal shock Temperature: -45℃~ +90℃ P OK 2. Temperature change rate :25℃ / MIN 3. Dwell time low and high temperature : 30 MIN/EACH P 4. Total test cycle:5 CYCLE 5. Input/Output condition: STATIC 7 THERMAL SHOCK TEST 1. Thermal shock Temperature: -30℃~ +45℃ OK 2. Temperature change rate :25℃ / MIN 3. Dwell time low and high temperature : 30 MIN/EACH P 4. Total test cycle:10 CYCLE 5. Input/Output condition: 230VAC/Full Load AC ON/OFF TEST turn on 58sec;turn off 2sec Model: GSM36B48 Test Report 6 / 7 36W AC-DC Single Output Medical Adaptor 8 VIBRATION TEST GSM36B series TEST:OK 1 Carton & 1 Set (1) Waveform: Sine Wave (2) Frequency:10~500Hz (3) Sweep Time:12min/sweep cycle P (4) Acceleration:2G (5) Test Time:60min in each axis (X.Y.Z) (6) Ta:25℃ 9 CAPACITOR LIFE CYCLE 10 11 MTBF DMTBF/Accelerated Life Test SUPPOSE C110 (1) I/P: 230VAC IS THE MOST CRITICAL COMPONENT O/P:FULL LOAD Ta= 25 ℃ LIFE TIME (1) 99386HRS (2) I/P: 230VAC O/P:FULL LOAD Ta= 40 ℃ LIFE TIME (2) 47723HRS (3) I/P: 230VAC O/P:75% LOAD Ta= 40 ℃ LIFE TIME (3) 76944HRS (4) I/P: 230VAC O/P:50% LOAD Ta= 40 ℃ MIL-HDBK-217F NOTICE S2 PARTS COUNT TOTAL FAILURE RATE:657.4 KHRS LIFE TIME (4) 151689HRS P P Demonstration Mean Time Between Failure (Expected Life): Above 30,000 hours @ TA 40℃ P SAMPLE TESTER REVIEW APPROVAL PRODUCT SAMPLE DANIEL GAO SANFORD SU VINCENT TSENG 13.11.25 A50-F031 Model: GSM36B48 Test Report 7 / 7