PDRx PDRx

PDRx
Main features
„„
Fully welded version
„„
Robust stainless steel housing
„„
High overpressure resistance
„„
Excellent long term stability
Applications
„„Hydraulic
Measuring range
0 … 60 bar up to 0 … 1600 bar
Long term stability
≤ ± 0.1% FS / Year
Accuracy (20 °C)
≤ ± 0.5% FS, 0.3% FS
(Linearity, hysteresis,
repeatabillity, error of span)
Technical specifications
Measuring principle Resistive thin film
Measuring ranges
0 … 60 bar up to 0 … 1600 bar
Type of pressure
Relative
Accuracy (20 °C)
≤ ± 0.5% FS, 0.3% FS
(Linearity, hysteresis,
repeatabillity, error of span)
Error of zero point
4...20 mA
0...10 V
≤ ± 1% FS
≤ ± 0.2% FS
Zero thermal drift
≤ ± 0.3% FS/10 K
Span thermal drift
≤ ± 0.3% FS/10 K
Long term stability
≤ ± 0.1% FS / Year
Response time
(10 … 90%)
≤ 1 ms
Process connections See page 3
Environment
-40 ... + 85°C
-25 ... + 85°C
-25 ... + 85°C
IP67
Vibration
IEC60068-2-6
1.5 mm p-p (10 – 57 Hz), 10 g (58 Hz – 2 KHz)
10 cycles within 2.5 h per axis
Shock
IEC60068-2-27
50 g/11 ms 100 g/6 ms
10 x Imp. per Axis and direction
Bump
IEC60068-2-27
100 g/2 ms
4000 x Imp. per Axis and direction
Random
IEC60068-2-64
0.1 g2/Hz (20 Hz – 1 KHz)
30 min per axis (>10 g RMS)
Electrical specification
Output signal /
Power supply
Load impedance
Current ouput
Voltage ouput
4 … 20 mA / 10 … 33 VDC
0…10 V / 14 … 33 VDC
RΩ = (Usupply-10 V) / 20 mA
> 5 KΩ
Insulation resistance >100 MΩ at 250 VDC
Electrical connections See page 3
Material
Process connection SS 1.4301 AISI 304
Housing
SS 1.4301 AISI 304
Diaphragm
SS 1.4542 AISI 630
Sealing
NBR
Approvals
CE conformity
EMC directive 2004/108/CE in accordance with
EN61000-6-2, EN 61000-6-3
2013-04-17
Temperature
Storage
Medium
Ambient
Protection rating
Design and specifications subject to change without notice
Main characteristics
www.baumer.com
Data sheet A11.41
Page 1 / 4
PDRx
Measuring ranges and overpressure safety
Pressure in bar
Pressure range
0 ... 60
Over
pressure
120
200
320
500
800
1200
2000
3200
Burst
pressure
480
800
1280
2000
3200
4000
4000
4000
0 ... 160
0 ... 250
0 ... 400
0 ... 600
0 ... 1000
0 ... 1600
2013-04-17
Design and specifications subject to change without notice
0 ... 100
www.baumer.com
Data sheet A11.41
Page 2 / 4
PDRx
Electrical connections
Signal at 4...20 mA / Signal at 0...10V
Signal at 4...20 mA / Signal at 0...10 V
Connector
DIN 43650
M12
4-pins
+VS / +VS
3
IOUT / GND
n.c. / +VOUT
4
1
3
2
n.c. / +VOUT
IOUT / GND
+VS / +VS
1
2
n.c. / n.c.
earth
earth
Dimensions (mm)
Cable
output
Field housing
(Red) / +VS / +VS
M12
4 pins IP67
(Blue) / IOUT / GND
(White) / n.c. / +VOUT
Not included in delivery
clamping range: 8...10 mm
n.c. / +VOUT
4
3
2
1
IOUT / GND
+VS / +VS
(Shield) / earth
48
(Shield) / earth
ø 18,8
Code S14
Code E101
Code E002
Code E102
ø
G 1/4
5
G 1/4
ø 18,9
ø
26,5
35,5
M12 x 1,5
M12 x 1,5
ø 18,9
Design and specifications subject to change without notice
5
Code E107
2
12
21
3
16
23
12
3
23
16
21
2
Code E001
ø 4,6
M14 x 1,5
ø 18,9
G1/2" DIN 3852 E
(HEX 19)
M12x1.5 EN837
(HEX 19)
M12x1.5 DIN 3852 E
(HEX 19)
M14x1.5
cone 60°
(HEX 19)
2013-04-17
G1/4" EN 837
(HEX 19)
www.baumer.com
Data sheet A11.41
Page 3 / 4
PDRx
Ordering details PDRx
PDR
Model
Pressure transmitter
xxxx
xxx
x
xx
B
D
P ≤ 1000 bar
P ≤ 600 bar
P ≤ 1000 bar
P ≤ 600 bar
P ≥ 1000 bar
E001
E002
E101
E102
E107
Output connection
M12, 4 pins
S14
Accuracy
0.5% FS
0.3% FS
C
B
Pressure range and unit in bar
0...60
0...100
0...160
0...250
0...400
0...600
0...1000
0...1600
360
410
416
425
440
460
510
516
In combination with process connection code E001/E002/E101/E102
DE
2013-04-17
Design and specifications subject to change without notice
Damping element
Without damping element
With damping element
xxx
PDR
Output signal
4 ... 20 mA
0 ... 10 V
Process connection
G¼ EN 837
G¼ DIN 3852 with sealing
M12x1.5 EN 837
M12x1.5 DIN 3852 with sealing
M14x1.5 cone 60°
x
www.baumer.com
Data sheet A11.41