NI 5761R User Guide and Specifications The NI 5761 is a 250 MS/s analog input adapter module designed to work in conjunction with your NI FlexRIO™ FPGA module. This document contains signal information and specifications for the NI 5761R, which is comprised of an NI FlexRIO FPGA module and the NI 5761. This document also contains tutorial sections that demonstrate how to acquire data using a LabVIEW FPGA example VI and how to create and run your own LabVIEW project with the NI 5761R. Contents Front Panel and Connector Pinouts ..................................................................................................... 2 Block Diagram..................................................................................................................................... 4 NI 5761 Component-Level Intellectual Property (CLIP) .................................................................... 5 Cables................................................................................................................................................... 6 Clocking............................................................................................................................................... 7 Using Your NI 5761R with a LabVIEW FPGA Example VI.............................................................. 7 Creating a LabVIEW Project and Run a VI on an FPGA Target ........................................................ 9 How to Use Your NI FlexRIO Documentation Set ............................................................................. 12 Specifications....................................................................................................................................... 13 Where to Go for Support ..................................................................................................................... 24 Note Before configuring your NI 5761R, you must install the appropriate software and hardware. Refer to the NI FlexRIO FPGA Module Installation Guide and Specifications for installation instructions. Figure 1 shows an example of a properly connected NI FlexRIO device. NI FlexRIO Adapter Module + NI FlexRIO FPGA Module = NI FlexRIO Device Figure 1. NI FlexRIO Device Front Panel and Connector Pinouts Front Panel Table 1 shows the front panel connector and signal descriptions for the NI 5761. Refer to the Specifications section of this document for additional signal information. Caution To avoid permanent damage to the NI 5761, disconnect all signals connected to the NI 5761 before powering down the module, and only connect signals after the adapter module has been powered on by the NI FlexRIO FPGA module. Table 1. NI 5761 Front Panel Connectors Device Front Panel NI 5761 Connector Signal Description AUX I/O Refer to Table 2 for the signal names and descriptions. D<0..3> LEDs for custom configuration. CLK IN Provides the NI 5761 with an external Sample clock or Reference clock. AUX I/O TRIG Trigger input channel. CH 0 Analog input channels <0..3>. CH 1 CH 2 D0 D1 D2 D3 CH 3 CLK IN TRIG CH 0 CH 1 CH 2 CH 3 250 MS/s Analog Input NI 5761R User Guide and Specifications 2 ni.com AUX I/O Connector Table 2 shows the pins assignments for the AUX I/O connector on the NI 5761. Table 2. NI 5761 AUX I/O Connector Pin Assignments Micro-D Connector Pin Signal Signal Description 1 AUXIO0 General-purpose digital input or output channels. 2 AUXIO1 3 AUXIO2 4 AUXIO3 5 AUXIO4 6 AUXIO5 7 AUXIO6 8 AUXIO7 9 GND 10 GND 11 GND 12 GND 13 GND 14 GND 15 GND 1 9 2 10 3 11 4 12 5 13 6 Ground. 14 7 15 8 Caution Connections that exceed any of the maximum ratings of any connector on the NI 5761R can damage the device and the chassis. NI is not liable for any damage resulting from such signal connections. For the maximum input and output ratings for each signal, refer to the Specifications section of this document. © National Instruments Corporation 3 NI 5761R User Guide and Specifications Block Diagram Figure 2 shows the NI 5761 block diagram and signal flow to and from the NI 5761 component-level intellectual property (CLIP) by way of the adapter module and the corresponding NI 5761 CLIP in LabVIEW FPGA. NI 5761 Adapter Module AUX I/O 8 LabVIEW FPGA CLIP 8 FET Buffer 8 8 D<0..3> (LEDs) 4 Buffer PFI Input PFI Output PFI Write Enable PFI Connector Enable LED <0..3> Trigger Input Internal Reference Clock CLK IN Switch Switch Clock Buffer SPI Engine Interfacing with: Clock Synthesizer TRIG FET Buffer AD9512 ADS62P49 Switches Analog FE Switch Sample Clock AD9512 CH 0 IO Module Clock 0 (n = 1, Single Sample CLIP n = 2, Multiple Sample CLIP) Multiple Sample CLIP Analog Front End ADS62P49 CH 1 1/n SPI Read SPI Write SPI Address SPI Write Data SPI Read Data SPI Device Select SPI Idle Initialization Done Reinitialize Configuration Error Sample Clock Select Sample Clock Commit Synthesizer Locked Data AI 0 Data N Clock AI 0 Data N–1 AI 1 Data N Analog Front End AI 0 AI 1 AI 1 Data N–1 AI 2 Data N ADC Interface AI 2 AI 2 Data N–1 AI 3 Data N CH 2 Single Sample CLIP AI 3 AI 3 Data N–1 Analog Front End Data ADS62P49 CH 3 Clock Analog Front End IOModSyncClock Figure 2. NI 5761 Connector Signals and NI 5761 CLIP Signal Block Diagram NI 5761R User Guide and Specifications 4 ni.com NI 5761 Component-Level Intellectual Property (CLIP) The LabVIEW FPGA Module includes a feature for HDL IP integration called CLIP. NI FlexRIO devices support two types of CLIP: user-defined and socketed. • User-defined CLIP allows users to insert HDL IP into an FPGA target, enabling VHDL code to communicate directly with an FPGA VI. • Socketed CLIP provides the same IP integration functionality of the user-defined CLIP, but also allows the CLIP to communicate directly with circuitry external to the FPGA. Adapter module socketed CLIP allows your IP to communicate directly with both the FPGA VI and the external adapter module connector interface. Figure 3 shows the relationship between an FPGA VI and CLIP. NI FlexRIO FPGA Module FPGA Adapter Module CLIP Socket LabVIEW FPGA VI User-Defined CLIP Socketed CLIP Socketed CLIP Socketed CLIP Fixed I/O DRAM 1 CLIP Socket Fixed I/O DRAM 0 CLIP Socket DRAM0 DRAM1 Fixed I/O Adapter Module External I/O Connector User-Defined CLIP Figure 3. CLIP Relationship © National Instruments Corporation 5 NI 5761R User Guide and Specifications The NI 5761 ships with socketed CLIP items that are used to add module I/O to the LabVIEW project. The NI 5761 ships with the following CLIP items: • NI 5761 Multiple Sample CLIP—Generates two samples per clock cycle at a clock rate that is half the sample rate. The default sample rate is 250 MHz, which sets the default clock rate for this CLIP at 125 MHz. You can set a lower sample rate by using an external Sample clock. This CLIP provides access to four analog input channels, eight PFI lines, and an input clock selector that can be configured to use one of the following settings: – Internal Sample clock – Internal Sample clock locked to an external Reference clock through the CLK IN connector – External Sample clock through the CLK IN connector – Internal Sample clock locked to an external Reference clock through IoModSyncClock – External Sample clock through IoModSyncClock This CLIP also contains an engine to program the CLK chip and ADCs, either through predetermined settings for an easier instrument setup or through a raw SPI address and data signals for a more advanced setup. The NI 5761 Multiple Sample CLIP is the default CLIP. • NI 5761 Single Sample CLIP—Generates one sample per clock cycle at a default sample rate of 250 MHz. You can set a lower sample rate by using an external Sample clock. This CLIP provides access to four analog input channels, eight PFI lines, and an input clock selector that can be configured to use one of the following settings: – Internal Sample clock – Internal Sample clock locked to an external Reference clock through the CLK IN connector – External Sample clock through the CLK IN connector – Internal Sample clock locked to an external Reference clock through IoModSyncClock – External Sample clock through IoModSyncClock This CLIP also contains an engine to program the CLK chip and ADCs, either through predetermined settings for an easier instrument setup or through a raw SPI address and data signals for a more advanced setup. Refer to the NI FlexRIO Help in the LabVIEW Help for more information about NI FlexRIO CLIP items, configuring the NI 5761 with a socketed CLIP, and a list of available socketed CLIP and provided signals. Cables Use the NI CMD-to-CMD cable (NI part number 763194-01) or the NI CMD Single-ended to Pigtail cable (NI part number 763191-01) to connect signals to your AUX I/O connector. Use any 50 Ω SMA cable to connect signals to the other connectors on the front panel of your NI 5761. For more information about connecting I/O signals on your device, refer to the Specifications section of this document. NI 5761R User Guide and Specifications 6 ni.com Clocking The NI 5761 clocks control the sample rate and other timing functions on the device. Table 3 contains information about the possible NI 5761 clock resources. Table 3. NI 5761 Clock Sources Clock Frequency Sample Clock 175 MHz to 250 MHz Reference Clock 10 MHz Source Options • Internal VCO locked to a Reference clock • External through the CLK IN connector • External through IoModSyncClock • Internal • External through the CLK IN connector • External through IoModSyncClock Using Your NI 5761R with a LabVIEW FPGA Example VI The NI FlexRIO Adapter Module Support software includes a variety of example projects to help get you started creating your LabVIEW FPGA program. This section explains how to use an existing LabVIEW FPGA example project to generate and acquire samples with the NI 5761R. This example requires at least one SMA cable for connecting signals to your NI 5761R. Note Examples available for your device are dependent on the device-specific minimum software requirements. For more information about software requirements for your device, visit ni.com/ info and enter rdsoftwareversion as the Info Code to determine which minimum software versions you need for your device. Each NI 5761R example project includes the following components: • • A LabVIEW FPGA VI that can be compiled and run embedded in FPGA hardware A VI that runs in LabVIEW for Windows and interacts with the LabVIEW FPGA VI Note In NI application software, NI FlexRIO adapter modules are referred to as IO Modules. Complete the following steps to run an example that acquires a waveform on CH 0 of the NI 5761. 1. Connect one end of an SMA cable to CH 0 on the front panel of the NI 5761 and the other end of the cable to your device under test (DUT). 2. Launch LabVIEW. 3. In the Getting Started window, click Find Examples to display the NI Example Finder. 4. In the NI Example Finder window, select Hardware Input and Output»FlexRIO»IO Modules» NI 5761. 5. Select NI 5761 - Getting Started.lvproj. 6. In the Project Explorer window, open NI 5761 - Getting Started (Host).vi under My Computer. The host VI opens. This VI uses the NI 7952R as the FPGA target by default. If you are using an NI FlexRIO FPGA module other than the NI 7952R, complete the following steps to change to an FPGA VI that supports your target. a. Select Window»Show Block Diagram to open the VI block diagram. b. On the block diagram, right-click the Open FPGA VI Reference (PXI-7952R) function and select Configure Open FPGA VI Reference. © National Instruments Corporation 7 NI 5761R User Guide and Specifications 7. c. In the Configure Open FPGA VI Reference dialog box, click the Browse Project button in the Open VI section. d. In the Select VI dialog box that opens, expand the tree view for your device, select the VI under your device and click OK. e. Click OK in the Configure Open FPGA VI Reference dialog box. f. Save the VI. On the front panel, in the RIO Resource box, select an NI 5761R resource that corresponds with the target configured in step 6. 8. Select AI 0 in the AI Channel box. 9. Set the Trigger Level (V) and the Record Size controls to the desired value. 10. In the Trigger Type box, you can select either Software Trigger or Data Edge. If you select Software Trigger, the VI acquires data every time you click the Software Trigger button on the front panel of the VI. If you select Data Edge, the VI acquires data every time an edge occurs. 11. Click the Run button to run the VI. 12. Click the Software Trigger button if you selected Software Trigger in the Trigger Type box. The VI acquires data and displays the captured waveform on the Acquired Waveform graph as shown in Figure 4. 13. Click STOP to stop the VI. 14. Close the VI. Figure 4. NI 5761 - Getting Started (Host) VI Front Panel NI 5761R User Guide and Specifications 8 ni.com Creating a LabVIEW Project and Run a VI on an FPGA Target This section explains how to setup your target and create an FPGA VI and host VI for data communication. For more detailed information about acquiring data on your NI 5761R, refer to the example VIs available on your NI FlexRIO Adapter Module Support software. Creating a Project 1. Launch LabVIEW, or if LabVIEW is already running, select File»New. 2. In the New dialog box, select Project»Empty Project. Click OK. The new project opens in the Project Explorer window. 3. Save the project as 5761SampleAcq.lvproj. Creating an FPGA Target VI 1. In the Project Explorer window, right-click My Computer and select New»Targets and Devices. 2. In the Add Targets and Devices on My Computer dialog box, select the Existing Target or Device option button and expand the FPGA Target. The target is displayed. 3. Select your device and click OK. The target and target properties are loaded into the project tree. 4. In the Project Explorer window, expand FPGA Target (RIOx, PXI-79xxR). 5. Right-click the FPGA target and select New»FPGA Base Clock. 6. In the Resource pull-down menu, select IO Module Clock 0. 7. Enter 125 MHz in the Compile for single frequency control. Click OK. 8. Right-click the FPGA target and select New»FPGA Base Clock again. 9. In the Resource pull-down menu, select 200 MHz Clock. Click OK. 10. Right-click IO Module and select Properties. In the General category, you can see the available CLIP for the NI 5761 in the Component Level IP pane. If the category information is dimmed, select the Enable IO Module checkbox. 11. Select NI 5761 Multi Sample CLIP. 12. In the Clock Selections category, select 200 MHz Clock from the pull-down menu for Clk200. Leave Clk40 configured as the Top-Level Clock. This step is necessary to compile the FPGA Target VI correctly. Click OK. Note Configuring these clocks is required for proper CLIP operation. Refer to the NI 5761 CLIP topics in the NI FlexRIO Help for more information about configuring your clocks. 13. In the Project Explorer window, right-click the FPGA target and select New»VI. A blank VI opens. 14. Select Window»Show Block Diagram to open the VI block diagram. 15. In the Project Explorer window, expand the IO Module (NI 5761 : NI 5761) tree view. 16. Select AI 0 Data N and AI 0 Data N-1 and drag them onto the block diagram. 17. Add a Timed Loop around the two nodes. 18. Wire an indicator to the output terminals of the IO Module\AI 0 Data N and IO Module\AI 0 Data N-1 nodes. 19. Wire an FPGA Clock Constant to the input node of the Timed Loop. Set this constant to IO Module Clock 0. © National Instruments Corporation 9 NI 5761R User Guide and Specifications Your block diagram should now resemble the block diagram in Figure 5. Figure 5. 5761SampleAcq (FPGA).vi Block Diagram Tip Click the Clean Up Diagram button on the toolbar to cleanly organize the VI block diagrams. 20. Save the VI as 5761SampleAcq (FPGA).vi. 21. Close the VI. 22. In the Project Explorer window under My Computer, expand the tree view for your device, right-click 5761SampleAcq (FPGA).vi and select Compile to compile the files for your target. The Generating Intermediate Files window opens and displays the compilation progress. The LabVIEW FPGA Compile Server window opens and runs. The compilation takes several minutes. 23. When the compilation completes, click the Stop Server button. 24. Click Close in the Successful Compile Report window. 25. Save and close the VI. 26. Save the project. Creating a Host VI 1. In the Project Explorer window, right-click My Computer and select New»VI. A blank VI opens. Select Window»Show Block Diagram to open the VI block diagram. 2. Add the Open FPGA VI Reference function, located on the FPGA Interface palette, to the block diagram. 3. Right-click the Open FPGA VI Reference function, labeled No Target, and select Configure Open FPGA VI Reference. 4. In the Configure Open FPGA VI Reference dialog box, select the VI option button. 5. In the Select VI window that opens, select 5761SampleAcq (FPGA).vi under your device, and click OK. 6. Click OK in the Configure Open FPGA VI Reference dialog box. The new target name appears under the Open FPGA VI Reference function in the block diagram. 7. Add a While Loop to the block diagram. 8. Right-click the stop condition inside the While Loop and select Create Control to create a STOP control on the VI front panel. 9. Add the Read/Write Control function, located on the FPGA Interface palette, inside the While Loop. 10. Wire the FPGA VI Reference Out indicator of the Open FPGA VI Reference function to the FPGA VI Reference In control on the Read/Write Control function. NI 5761R User Guide and Specifications 10 ni.com 11. Wire the error out indicator of the Open FPGA VI Reference function to the error in control of the Read/Write Control function. 12. Configure the Read/Write Control function by clicking the terminal section labeled Unselected, and selecting IO Module/AI 0 Data N. 13. Add the IO Module/AI 0 Data N-1 I/O item to the Read/Write Control function by clicking the lower edge of the control node with the Positioning tool and dragging the edge down. 14. Wire an indicator to the output terminal of the IO Module\AI 0 Data N node. 15. Wire an Indicator from the output terminal of the IO Module\AI 0 Data N-1 node. 16. Add the Close FPGA VI Reference function, located on the FPGA Interface palette, to the right of the While Loop on the block diagram. 17. Wire the FPGA VI Reference Out indicator of the Read/Write Control function to the FPGA VI Reference In control of the Close FPGA VI Reference function. 18. Wire the error out indicator of the Read/Write Control function to the error in control of the Close FPGA VI Reference function. Your block diagram should now resemble the block diagram in Figure 6. Figure 6. 5761SampleAcq(Host).vi Block Diagram 19. Save the VI as 5761SampleAcq(Host).vi. Running the Host VI 1. Connect one end of an SMA cable to CH 0 on the front panel of the NI 5761 and the other end of the cable to your DUT. 2. Open the front panel of 5761SampleAcq(Host).vi. 3. Click the Run button to run the VI. 4. The VI acquires data from the DUT on AI 0 Data N and AI 0 Data N-1. 5. Click the STOP button on the front panel and close the VI. © National Instruments Corporation 11 NI 5761R User Guide and Specifications How to Use Your NI FlexRIO Documentation Set Refer to the Figure 7 and Table 4 for information about how to use your NI FlexRIO documentation set. INSTALL Hardware and Software NI FlexRIO FPGA Module Installation Guide and Specifications CONNECT Signals and Learn About Your Adapter Module NI FlexRIO Adapter Module User Guide and Specifications PROGRAM Your NI FlexRIO System in LabVIEW FPGA Module Are You New to LabVIEW FPGA Module? LEARN About LabVIEW FPGA Module Yes No No LabVIEW FPGA Module Help NI FlexRIO Help LabVIEW Examples Figure 7. How to Use Your NI FlexRIO Documentation Set Table 4. NI FlexRIO Documentation Locations and Descriptions Document Location Description NI FlexRIO FPGA Module Installation Guide and Specifications* Available in your FPGA module hardware kit and from the Start Menu. Contains installation instructions for your NI FlexRIO system and specifications for your FPGA module. NI Adapter Module User Guide and Specifications* Available in your adapter module hardware kit and from the Start Menu. Contains signal information, examples, and specifications for your adapter module. LabVIEW FPGA Module Help* Embedded in LabVIEW Help. Contains information about the basic functionality of LabVIEW FPGA Module. NI FlexRIO Help* Embedded in LabVIEW FPGA Module Help. Contains FPGA module, adapter module, and CLIP configuration information. LabVIEW Examples Available in LabVIEW Example Finder. Contains examples of how to run FPGA VIs and Host VIs on your device. Other Useful Information on ni.com ni.com/ipnet Contains LabVIEW FPGA functions and intellectual property to share. ni.com/flexrio Contains product information and data sheets for NI FlexRIO devices. * These documents are also available at ni.com/manuals. NI 5761R User Guide and Specifications 12 ni.com Specifications This section lists the specifications of the NI FlexRIO adapter module (NI 5761). Pair these specifications with the specifications listed in the NI FlexRIO FPGA Module Installation Guide and Specifications. For more information about safety and electromagnetic compatibility refer to the Read Me First: Safety and Electromagnetic Compatibility document included in your hardware kit or available at ni.com/manuals. Caution To avoid permanent damage to the NI 5761, disconnect all signals connected to the NI 5761 before powering down the module, and only connect signals after the module has been powered on by the NI FlexRIO FPGA module. Note All numeric specifications are typical unless otherwise noted. All graphs illustrate the performance of a representative module. Typical values describe useful product performance that are not covered by warranty. Typical values cover the expected performance of units over ambient temperature ranges of 23 ±5 °C with a 90% confidence level, based on measurements taken during development or production. Analog Input (AI CH 0 through AI CH 3) General Characteristics Number of channels ...............................................Four, single-ended, simultaneously sampled Connector...............................................................SMA Input impedance.....................................................50 Ω, per connector Sample rate Internal Sample clock ....................................250 MHz External Sample clock ...................................175 MHz to 250 MHz Digital data range...................................................±8,191 ADC part number...................................................ADS62P491; 14-bit resolution, dual ADC AC-Coupled Specifications Input range (normal operating conditions) ............2.07 Vpk-pk Absolute maximum input.......................................±10 V DC, 5 Vpk-pk AC Bandwidth (–3 dB).................................................0.1 MHz to 500 MHz Bandwidth (–1 dB).................................................1 MHz to 250 MHz 1 For additional information on the ADS62P49, refer to the Texas Instruments device data sheet at www.ti.com. © National Instruments Corporation 13 NI 5761R User Guide and Specifications Table 5 lists the AC-coupled spectral performance measurements. All values are measured with a 250 MHz external Sample clock. Table 5. AC-Coupled Spectral Performance 20.17 MHz 70.17 MHz* 123.17 MHz* SNR 72.5 dB 71.4 dB 70.5 dB SINAD 72.3 dB 71.2 dB 70.3 dB SFDR 88 dB 84 dB 80 dB Measurement * These measurements were extrapolated from a –4 dBFS plot. Channel to channel isolation 1 MHz ............................................................>90 dB 100.1 MHz .....................................................90 dB 501 MHz ........................................................80 dB Gain (dB) AC-Coupled Measurements 5 4 3 2 1 0 –1 –2 –3 –4 –5 –6 –7 –8 –9 –10 –11 –12 –13 –14 –15 80k 100k 1M 10M 100M 700M Frequency (Hz) Figure 8. AC-Coupled Bandwidth (Passband) NI 5761R User Guide and Specifications 14 ni.com Amplitude (dBFS) 0 –5 –10 –15 –20 –25 –30 –35 –40 –45 –50 –55 –60 –65 –70 –75 –80 –85 –90 –95 –100 –105 –110 –115 0 20 40 60 80 100 120 Frequency (MHz) Amplitude (dBFS) Figure 9. AC-Coupled Spectral Measurements: 19.5 MHz and 20.5 MHz, –13 dBFS, 8,192 Point FFT, 10 RMS Average 0 –5 –10 –15 –20 –25 –30 –35 –40 –45 –50 –55 –60 –65 –70 –75 –80 –85 –90 –95 –100 –105 –110 –115 0 10 20 30 40 50 60 70 80 90 100 110 120 Frequency (MHz) Figure 10. AC-Coupled Spectral Measurements: 20.1 MHz, –1 dBFS, 8,192 Point FFT, 10 RMS Average © National Instruments Corporation 15 NI 5761R User Guide and Specifications Amplitude (dBFS) 0 –5 –10 –15 –20 –25 –30 –35 –40 –45 –50 –55 –60 –65 –70 –75 –80 –85 –90 –95 –100 –105 –110 –115 0 20 40 60 80 100 120 Frequency (MHz) Figure 11. AC-Coupled Spectral Measurements: Terminated, 8,192 Point FFT, 10 RMS Average DC-Coupled Specifications Input range (normal operating conditions) ............1.23 Vpk-pk Absolute maximum input.......................................±4.5 V DC Bandwidth (–3 dB).................................................DC to 500 MHz Bandwidth (–1 dB).................................................DC to 250 MHz Table 6 lists the DC-coupled spectral performance measurements. All values are measured with a 250 MHz external Sample clock. Table 6. DC-Coupled Spectral Performance Measurement 20.1 MHz 70.1 MHz 122.1 MHz SNR 65.7 dB 64.3 dB 63 dB SINAD 65.2 dB 61.8 dB 56.6 dB SFDR 76 dB 65 dB 58 dB Channel to channel isolation 1 MHz ............................................................>90 dB 100.1 MHz .....................................................80 dB 501 MHz ........................................................70 dB NI 5761R User Guide and Specifications 16 ni.com To ensure proper device operation, DC-coupled channels on the NI 5761 must see 50 Ω at DC looking back at their source. If you are connecting a to source with an output impedance less than 50 Ω, you can add series resistance close to the source to properly bias the NI 5761 input terminals. You can also use the NI 5761 input channel bias DACs to remove DC offset present in the system. For more information about programming the bias DACs, refer to the NI 5761 CLIP topics in the NI FlexRIO Help. Note Gain (dB) DC-Coupled Measurements 5 4 3 2 1 0 –1 –2 –3 –4 –5 –6 –7 –8 –9 –10 –11 –12 –13 –14 –15 0 10 100 1k 10k 100k 1M 10M 100M 700M Frequency (Hz) Figure 12. DC-Coupled Bandwidth (Passband) © National Instruments Corporation 17 NI 5761R User Guide and Specifications Amplitude (dBFS) 0 –5 –10 –15 –20 –25 –30 –35 –40 –45 –50 –55 –60 –65 –70 –75 –80 –85 –90 –95 –100 –105 –110 –115 0 25 50 75 100 125 Frequency (MHz) Amplitude (dBFS) Figure 13. DC-Coupled Spectral Measurements: 19.5 MHz and 20.5 MHz, –13 dBFS, 8,192 Point FFT, 10 RMS Average 0 –5 –10 –15 –20 –25 –30 –35 –40 –45 –50 –55 –60 –65 –70 –75 –80 –85 –90 –95 –100 –105 –110 –115 0 10 20 30 40 50 60 70 80 90 100 110 120 Frequency (MHz) Figure 14. DC-Coupled Spectral Measurements: 20.1 MHz, –1 dBFS, 8,192 Point FFT, 10 RMS Average NI 5761R User Guide and Specifications 18 ni.com Amplitude (dBFS) 0 –5 –10 –15 –20 –25 –30 –35 –40 –45 –50 –55 –60 –65 –70 –75 –80 –85 –90 –95 –100 –105 –110 –115 0 20 40 60 80 100 120 Frequency (MHz) Figure 15. DC-Coupled Spectral Measurements: Terminated, 8,192 Point FFT, 10 RMS Average Analog Input Phase Noise –70.0 –80.0 Phase Noise (dBc/Hz) –90.0 –100.0 –110.0 –120.0 –130.0 –140.0 –150.0 –160.0 0 1k 10k 100k 500k Frequency Offset from 10 MHz Carrier (Hz) Figure 16. Analog Input Phase Noise © National Instruments Corporation 19 NI 5761R User Guide and Specifications Internal Sample Clock General Characteristics Oscillator type........................................................Fixed frequency synthesizer Frequency...............................................................250 MHz Reference spurs......................................................<70 dBc Phase noise 10 kHz offset..................................................–100 dBc/Hz 100 kHz offset................................................–120 dBc/Hz Clock distribution part number ..............................AD95121; clock distribution Reference clock sources.........................................Internal, External through the CLK IN connector, or IoModSyncClk2 Internal reference type ...........................................TCXO Internal reference stability .....................................±1 ppm Internal reference frequency ..................................10 MHz CLK IN General Characteristics Number of channels ...............................................1, single-ended Connector...............................................................SMA Input impedance.....................................................50 Ω Input coupling ........................................................AC External Sample Clock Input voltage range ................................................0.63 Vpk-pk to 2.5 Vpk-pk Input frequency range ............................................175 MHz to 250 MHz Absolute maximum input.......................................±10 V DC, 3.1 Vpk-pk AC External Reference Clock Input voltage range ................................................1.4 Vpk-pk to 4.4 Vpk-pk Input frequency range ............................................10 MHz Absolute maximum input.......................................±10 V DC, 5 Vpk-pk AC External Trigger General Characteristics Number of channels ...............................................1, single-ended Connector...............................................................SMA Input impedance.....................................................50 Ω Input coupling ........................................................DC 1 2 For additional information about the AD9512, refer to the Analog Devices device data sheet at www.analog.com. IoModSyncClk is available only on the NI PXIe-796xR FPGA modules. NI 5761R User Guide and Specifications 20 ni.com Input levels Voltage Level Minimum Maximum VIL 0V 0.7 V VIH 1.7 V 5.5 V Absolute maximum input.......................................–0.5 V to 7 V PFI<0..7> General Characteristics Number of channels ...............................................8 bidirectional Connector type.......................................................Micro–D Interface standard...................................................2.5 V LVCMOS Interface logic Voltage Level Minimum Maximum VIL 0V 0.7 V VIH 1.7 V 5.5 V VOL 0V 0.4 V VOH 1.9 V 2.5 V Zin ...................................................................17.5 kΩ Zout .................................................................50 Ω Iout...................................................................2 mA Maximum toggle frequency...................................500 kHz Absolute maximum input.......................................–0.5 V to 7 V EEPROM Map Byte Address Size (Bytes) Field Name 0x0 2 Vendor ID 0x2 2 Product ID 0x4 4 Serial Number 0x8 116 Reserved 0x7C 132 User Space Caution Only write to User Space. Writing to any other offset may cause the NI 5761 to stop functioning. © National Instruments Corporation 21 NI 5761R User Guide and Specifications Power Total power, typical operation................................5.3 W Physical Dimensions ............................................................12.9 × 2.0 × 12.1 cm (5.1 × 0.8 × 4.7 in.) Weight ....................................................................312 g (11 oz) Front panel connectors...........................................6 SMA and one Micro-D connector Environmental The NI 5761 is intended for indoor use only. Operating environment1 .........................................0 °C to 55 °C, tested in accordance with IEC-60068-2-1 and IEC-60068-2-2. Relative humidity range .........................................10% to 90%, noncondensing, tested in accordance with IEC-60068-2-56. Altitude ..................................................................2,000 m at 25 °C ambient temperature. Pollution Degree ....................................................2 Storage environment Ambient temperature range ...........................–20 °C to 70 °C, tested in accordance with IEC-60068-2-1 and IEC-60068-2-2. Relative humidity range .................................5% to 95%, noncondensing, tested in accordance with IEC-60068-2-56. Note Clean the device with a soft, non-metallic brush. Make sure that the device is completely dry and free from contaminants before returning it to service. Safety This product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: • IEC 61010-1, EN 61010-1 • UL 61010-1, CSA 61010-1 Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. 1 For PXI/PXI Express chassis configurations that group NI FlexRIO adapter modules in three or more contiguous slots, National Instruments recommends limiting the ambient operating temperature to less than 50 °C. NI 5761R User Guide and Specifications 22 ni.com Electromagnetic Compatibility This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity • EN 55011 (CISPR 11): Group 1, Class A emissions • AS/NZS CISPR 11: Group 1, Class A emissions • FCC 47 CFR Part 15B: Class A emissions • ICES-001: Class A emissions Note For EMC declarations and certifications, refer to the Online Product Certification section of this document. Caution For EMC compliance, operate this device with shielded cables and accessories. CE Compliance This product meets the essential requirements of applicable European Directives as follows: • 2006/95/EC; Low-Voltage Directive (safety) • 2004/108/EC; Electromagnetic Compatibility Directive (EMC) Online Product Certification To obtain product certifications and the Declaration of Conformity for this product, visit ni.com/ certification, search by model number or product line, and click the appropriate link in the Certification column. Environmental Management National Instruments is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial not only to the environment but also to NI customers. For additional environmental information, refer to the NI and the Environment Web page at ni.com/ environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. Waste Electrical and Electronic Equipment (WEEE) EU Customers At the end of the product life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste and Electronic Equipment, visit ni.com/environment/weee. ⬉ᄤֵᙃѻક∵ᶧࠊㅵ⧚ࡲ⊩ ˄Ё RoHS˅ Ёᅶ᠋ National Instruments ヺড়Ё⬉ᄤֵᙃѻકЁ䰤ࠊՓ⫼ᶤѯ᳝ᆇ⠽䋼ᣛҸ (RoHS)DŽ ݇Ѣ National Instruments Ё RoHS ড়㾘ᗻֵᙃˈ䇋ⱏᔩ ni.com/environment/rohs_chinaDŽ (For information about China RoHS compliance, go to ni.com/environment/rohs_china.) © National Instruments Corporation 23 NI 5761R User Guide and Specifications Where to Go for Support The National Instruments Web site is your complete resource for technical support. At ni.com/support you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers. National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. National Instruments also has offices located around the world to help address your support needs. For telephone support in the United States, create your service request at ni.com/support and follow the calling instructions or dial 512 795 8248. For telephone support outside the United States, contact your local branch office: Australia 1800 300 800, Austria 43 662 457990-0, Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599, Canada 800 433 3488, China 86 21 5050 9800, Czech Republic 420 224 235 774, Denmark 45 45 76 26 00, Finland 358 (0) 9 725 72511, France 01 57 66 24 24, Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737, Italy 39 02 41309277, Japan 0120-527196, Korea 82 02 3451 3400, Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710, Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466, New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60, Poland 48 22 328 90 10, Portugal 351 210 311 210, Russia 7 495 783 6851, Singapore 1800 226 5886, Slovenia 386 3 425 42 00, South Africa 27 0 11 805 8197, Spain 34 91 640 0085, Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151, Taiwan 886 02 2377 2222, Thailand 662 278 6777, Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545 LabVIEW, National Instruments, NI, ni.com, the National Instruments corporate logo, and the Eagle logo are trademarks of National Instruments Corporation. Refer to the Trademark Information at ni.com/trademarks for other National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. © 2010 National Instruments Corporation. All rights reserved. 375509A-01 May10